Survey
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Radiation physics University hospital Linköping Report Radfys-02-02 Image quality – Conventional X-ray Jonas Nilsson, Michael Sandborg Lab Image Quality and Dose in Projection Radiography Before the lab Read this tutorial Image Quality – Conventional X-ray in advance so you have an idea of what the lab is about. During the lab Go to the reception at Röntgenkliniken at the University hospital on level 11 in the main building. Your lab supervisor will guide you to the viewing stations at the radiology clinic. Join one or two other students to form a group and do the lab together. Get the correct login to the PACS from the supervisor. The lab consists of seven cases where one imaging or phantom parameter has been altered in each case; for example, the tube kilovoltage, phantom thickness or focal spot size. Go through the seven cases one after the other and assess image quality and patient dose (dose-area product, DAP). Make sure you understand why image quality and patient dose has changed before you go to the next case. After the lab Write a short report together with your fellow students and summarise for each case 1-7 how image quality and patient dose is altered and why. Submit your lab report in lisam. Lab: Image quality – Conventional X-ray INTRODUCTION You are going to study a number of image pairs during this session. Each pair contains X-ray images produced while varying one or more X-ray system parameters, e.g. tube load or X-ray tube voltage etc.. The images depict objects (also called test phantoms) that make it possible to evaluate the image quality. You are supposed to work together during this session as a group (all participants must be active), discussing and explaining why the images vary in quality. This discussion is going to be based on the theoretical knowledge gained during lectures, PBL group meetings, literature etc. You are going to discuss and compare the images in terms of image resolution, contrast and dose to the patient. Since the images are digital, you should use the possibilities (region of interest definitions, zoom possibilities etc.) offered to you by the PACS software. At the end of this lab you will be able to describe how the X-ray system parameters affect image quality and patient dose. AIM: This lab session aims at increasing your knowledge of how image quality and dose to the patient are affected when some system parameters are altered. The parameters are: tube kilovoltage (kV), tube current and exposure time (mAs), phantom or object-thickness, focus spot size, x-ray beam size, degree of magnification (air gap length) and grid usage. Since we want to concentrate on the effect of each parameter, as few parameters as possible are going to get altered at a time. A dedicated image quality phantom has been used in order to illustrate the different physical effects of these parameter alterations. At the end of the laboratory session, more human like phantoms will be shown in order to demonstrate the clinical effects in a more realistic fashion. DEFINITIONS: We use three concepts in order to define image quality: Contrast: The difference in grey level between the object detail and the surrounding background. The greater the contrast of an object the better the object stands out in the image. Sharpness: An imaging system’s ability to image a sharp edge. In practice the concept of spatial resolution is used meaning the system’s ability to resolve small details in the image that are adjacent to each other or placed close together. Spatial resolution is quantified in terms of line pairs per mm (lp/mm). Noise: Depending on the number of photons contributing to the image formation, the image may be perceived as being noisy to some degrees. A larger number of photons yields less noise and possibly a better image quality. This type of noise is usually referred to as quantum noise. There are other types of noise as well, e.g. electronic noise and detector noise. Normally, quantum noise is the dominating noise factor, hence ‘quantum-limited’ imaging. Dose-area product DAP is a concept that often occurs when discussing the patient dose. The DAP is the product of the dose (in air) and the area of the x-ray beam in the same plane; typically, just beneath the collimator below the x-ray tube. Hence the DAP depends on both dose and beam area. (Dose in air at charge particle equilibrium is called air kerma.) Scatter-to-primary ratio (S/P). The relationship between the amount of scattered radiation (S) and primary radiation (P) at the image detector is often stated as the scatter-to-primary ratio (S/P). This ratio is dependent on how the grid and/or the air gap technique is used in order to minimize the contrast-reducing scattered radiation. It is often stated that contrast, C, is reduced by a factor CDF=(1+S/P)-1, i.e. when S/P is high, CDF gets low and the contrast is reduced. Automatic exposure control AEC is used in all but the last case (case #7). This means that the radiographer selects the tube voltage but the exposure time (s) is terminated when the x-ray system has measured the correct dose (or “photons”) at the AEC-sensors just in front of the image detector. This means that if the photon energy of the beam or phantom thickness is altered the exposure time will change. The tube currant (mA) is depending on the focal spot size. A larger tube current is possible with the large focal spot size (larger filaments), and a smaller tube current is used with the smaller focal spot size. The air gap is the distance between the patient and the image detector. Typically, this distance is small (a few cm), but sometimes the patient couch height is increased by 10 or 20 cm and the patient is positioned closer to the x-ray tube and further away from the image detector. The anti-scatter grid is used in most examination of adults and of larger children with mass >20 kg. The test phantom plate that contains all the low- and high-contrast details, is positioned above 10 cm of Plexiglas (Lucite) and then an additional 10 cm of Plexiglas is positioned on top of the test phantom plate (see illustrations below). The contrast is best assessed by evaluating how many of the ten circular disc (located to the right in the images) that are visible. The sharpness of spatial resolution can be assessed by evaluating which of the parallel lines that are separable from each other; the more line pairs per mm the higher the spatial resolution (top left in image). The noise is best quantified by positioning a circular region of interest in a homogeneous part of the test phantom plate and measure the standard deviation in the pixel values. PROCEDURE: For each pair of images, you are supposed to assess how contrast, sharpness and DAP vary for different values of the system parameters. Useful, program specific advices (mouse pointer on the image under study) 1. In order to zoom continuously, hold Ctrl-button down and press the left mouse button while moving the mouse back or forth in order to zoom in and out respectively. 2. In order to change the grey scale mapping of the image continuously hold the middle button (wheel) of the mouse down while moving the mouse back or forth and/or left or right. Case 1: Dependence on tube kilovoltage kV Image nr Parameter Contrast Sharpness altered (higher/lower) (lp/mm) Dose-area product, DAP (higher/lower) Conclusions Comparison 1 1 50 kV, 338 mAs 2 125 kV, 3,2 mAs Comparison 2 3 70 kV, 35 mAs 4 125 kV, 3,2 mAs Case 2: Dependence on object thickness Image nr Dose-area Parameter Contrast Sharpness product, DAP altered (higher/lower) (lp/mm) (higher/lower) Comparison 3 1 70 kV, 4,1 mAs, 10 cm lucite 2 70 kV, 35 mAs, 20 cm lucite Comparison 4 3 70 kV, 35 mAs, 20 cm lucite 4 70 kV, 104 mAs, 25 cm lucite Conclusions Case 3: Dependence on field size Image nr Dose-area Parameter Contrast Sharpness product, DAP altered (higher/lower) (lp/mm) (higher/lower) Conclusions Comparison 5 1 70 kV, 35 mAs, 20x20 cm2 field 70 kV, 45 mAs, 2 10x10 cm2 field Case 4: Dependence on grid Image nr Dose-area Parameter Contrast Sharpness product, DAP altered (higher/lower) (lp/mm) (higher/lower) Conclusions Comparison 6 1 70 kV, 35 mAs, with grid 2 70 kV, 6,9 mAs, without grid Case 5: Dependence on focal spot size Image nr Parameter altered Comparison 7 1 70 kV, 35 mAs, large focal spot 2 70 kV, 36 mAs, small focal spot Contrast (higher/lower) Sharpness (lp/mm) Dose-area product, DAP (higher/lower) Conclusions Case 6: Dependence on air gap length or geometric magnification Image nr Parameter altered Comparison 8 1 70 kV, 37,7 mAs, 10 cm air gap, large focus 2 70 kV, 39 mAs, 10 cm air gap, small focus Comparison 9 3 70 kV, 36 mAs, 20 cm airgap, large focus 4 70 kV, 37 mAs, 20 cm air gap, small focus Contrast (higher/lower) Sharpness (lp/mm) Dose-area product, DAP (higher/lower) Conclusions Case 7: Dependence on tube load, quantum noise and patient dose Image nr Parameter altered Comparison 10 1 70 kV, 1 mAs 2 70 kV, 2 mAs 3 70 kV, 4 mAs 4 70 kV, 8 mAs 5 70 kV, 16 mAs 6 70 kV, 32 mAs 7 70 kV, 63 mAs 8 70 kV, 125 mAs Contrast Sharpness (higher/lower) (lp/mm) Dose-area product, DAP (higher/lower) Conclusions