Product Information
PI
“ SD Series I.I.”
Toshiba X-ray Image Intensifier
E5863SD-P4
Input Field : 150 mm (6″ )
Output Image Diameter: 20 mm
FEATURING:

High Resolution

High Conversion Factor

High contrast ratio

“ SD Series I.I.” E5863SD-P4 is used for medical radiographic diagnosis.

This X-ray Image Intensifier is intended to be connected with IEC ClassⅠequipment.

“ SD Series I.I.” E5863SD-P4 has been perfected by the integration of several aspects
of advanced technology;
The input window is of a thin metal plate, which has good X-ray transparency
characteristics and reduces X-ray scatter.
It uses a thick input phosphor screen made with an extremely fine pillar shaped
structure. This has high X-ray absorption and very high conversion efficiency. In
addition, there is a thin output phosphor screen on a single thick-glass output window
with an anti-reflective coating, offering very high contrast for small details, very low
structure noise and high MTF characteristics.
The E5863SD-P4, employing these manufacturing techniques, provides outstanding
performance in resolution, brightness, contrast and noise.
★The information contained herein is presented only as a guide for the applications of our products.
No responsibility is assumed by TOSHIBA ELECTRON TUBES & DEVICES CO., LTD. (TETD) for any infringements of patents or other rights of
the third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of TETD or
others.
★The information contained herein may be changed without prior notice. It is therefore advisable to contact TETD before proceeding with the
design of equipment incorporating this product.
E5863SD-P4
PERFORMANCE DATA
Specification
Central Resolution
Conversion Factor (Gx)
Contrast Ratio
10 % area
10 mm dia.
DQE (IEC standard)
54 Lp/cm typ.
18 (cd/m2)/ (μ Gy/s) typ.
160 (cd/m2)/ (mR/s) typ.
30:1 typ.
20:1 typ.
65 % typ.
GENERAL DATA
 MECHANICAL
Specification
Overall Length
Maximum Diameter
Nominal Entrance Field Size
Useful Entrance Field Size
Output Image Diameter
Weight (approx.)
Dimensional Outlines
276±5 mm
206±2 mm
150 mm min.
140 mm min.
20±0.5 mm
13 kg
Refer to page 5
 POWER SUPPLY
Specification
DC Power Supply
Input Voltage
Output Voltage
High Voltage Power Supply (Built-in)
Input Voltage
 ENVIRONMENTAL
Specification
Temperature
Humidity
Pressure
AC100 to 240Vac
24±1Vdc
24±1 Vdc
Under delivery and stock
-15 to 45 ℃
10 to 90 %
50 to 106 kPa
-2-
Under operating
5 to 40 ℃
30 to 85 %
70 to 106 kPa
E5863SD-P4
 ABSOLUTE MAXIMUM RATINGS
Specification
Anode Voltage
G3 Electrode Voltage
Potential Difference between Anode and
G3 Electrode
Potential Difference between G3 Electrode and G2 Electrode
G2 Electrode Voltage
G1 Electrode Voltage
Photocathode Current
Continuous
Pulsed
Input X-ray Dose
32 kV
5 kV
30 kV
5 kV
700 V
300 V
0.1 µA
2 µA
8.73×10-3Gy/min [1R/min]
-3-
E5863SD-P4
ELECTRICAL WIRING AND CONNECTING
X-ray image intensifier is operated with 100 to 240Vac input voltage applied to the DC power
supply.
Wiring and connecting are shown as follows.
①: I.I. input cable
②: AC input cable
③: GND cable
-4-
E5863SD-P4
DIMENSIONAL OUTLINE
Unit: mm
-5-
E5863SD-P4
OVERSEAS SUBSIDIARIES AND AFFILIATES
For Sales & Technical Services, please contact the following representative:
・TOSHIBA ELECTRONICS EUROPE GMBH
HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296 107 FAX +49 (211) 5296 402
・TOSHIBA AMERICA ELECTRONIC COMPONENTS, INC.
2150 EAST LAKE COOK ROAD, SUITE 310 BUFFALO GROVE,
ILLINOIS 60089 U.S.A
PHONE +1 (847) 484-2400 FAX +1 (847) 541-7287
・TOSHIBA ELECTRON DEVICES & MATERIALS (SHANGHAI) CO., LTD. (TEMS)
RM1606, SH-PLAZA,
No.336, XIZANG ROAD (MIDDLE), SHANGHAI, 200001, CHINA
PHONE +86 (21) 6361-0077 FAX +86 (21) 6351-5760
(Product Development Engineering Dept.)
1385 SHIMOISHIGAMI, OTAWARA-SHI, TOCHIGI-KEN, 324-8550, JAPAN
PHONE: +81-287-26-6666
FAX: +81-287-26-6060
http://www.toshiba-tetd.co.jp/
● Toshiba Electron Tubes & Devices Co., Ltd. meets the Environmental Management System Standard, SO14001
● Toshiba Electron Tubes & Devices Co., Ltd. meets internationally recognized Standards for Quality Management System ISO9001, ISO13485.
PE-E5863SD-P4 2013