1
Florida Heart CPR*
Safer Needle Devices: Protecting Health Care Workers
1 hour
Prepared by the Occupational Safety and Health Administration
Directorate of Technical Support
Office of Occupational Health Nursing
Objectives:
Upon completion of this course the student will be familiar with the epidemiology of
Needlestick Injuries, understand the occupational risks of Needlestick Injuries,
understand the definition and efficacy of safer needle devices as well as how to
evaluate and select safer needle devices and OSHA's Position on safer needle devices.
Acronyms and Abbreviations
CDC
CSHO
FDA
HBV
HCV
HIV
IM
IV
MMWR
NCID
OPIM
OSHA
Centers for Disease Control and Prevention
Compliance Safety and Health Officer
Food and Drug Administration
Hepatitis B Virus
Hepatitis C Virus
Human Immunodeficiency Virus
Intramuscular
Intravenous
Morbidity and Mortality Weekly Report
National Center for Infectious Diseases
Other potentially infectious materials
Occupational Safety and Health Administration
Purpose of this Course
Q. What is the purpose of this course?
A. The purpose of this packet is to:



Update OSHA staff's knowledge of current statistical data on exposure risk,
prevalence, and incidence of human immunodeficiency virus (HIV), hepatitis B
virus (HBV), and hepatitis C virus (HCV) in health care workers.
Provide technical assistance to OSHA staff in understanding the role of safer
needle devices in preventing needlestick injuries.
Give guidance on the process of selecting safer devices.
Florida Heart CPR*
Safer Needle Devices
2
Q. What are some questions that may be asked about safer needle devices and
needlestick prevention programs?
A. Although OSHA does not review, approve, license, or endorse products, OSHA staff
may receive questions about safer needle devices and therefore have an excellent
opportunity to educate the public about the role of these devices in preventing
needlestick injuries. Questions that may be asked include:






What are the current risks to health care workers who sustain a needlestick
injury?
What is the definition of a safer needle device?
Why do we need safer needle devices?
Does OSHA require us to use these devices?
What features make a safer needle device an effective engineering control?
How can a health care facility develop a program to implement safer needle
devices?
This guide will provide answers to these questions.
Epidemiology of Needlestick Injuries
Q. What is the epidemiology of needlestick injuries?
A. Five primary activities are associated with the majority of needlestick injuries. They
are:





Disposing of needles, including collection and disposal of materials used during
patient care procedures
Administering injections
Drawing blood
Recapping needles (not allowed under the Bloodborne Pathogens Standard)
Handling trash and dirty linens (termed "downstream injuries", these usually
affect the housekeeping department (Chiarello, 1992).
Since 1992, the International Health Care Worker Safety Center has gathered data on
needlestick injuries from 63 cooperating hospitals around the country. The results of the
data analysis provide us with a useful picture of the pattern of needlestick injuries
leading to occupational exposure to bloodborne pathogens.



Participating hospitals reported an overall rate for sharp-object injuries of 27.0
per 100 occupied beds per year.
Most exposure incidents occurred in patient rooms (37%), the operating room
(16%), the emergency department (7%) or the critical care units (6%). (Figure 1)
Nurses report the most frequent exposures (49.7%), while physicians rank
second at 12.6%. Nursing assistants and housekeeper/laundry workers account
for 5.3% and 5.1% of exposures respectively. (Table 1)
Florida Heart CPR*
Safer Needle Devices
3

Hollow-bore needles are the cause of injury in 68.5% of cases. Hollow-bore
needles (the type of needle used for giving injections or drawing blood) also are
implicated as the devices most often associated with the transmission of
bloodborne pathogen infections, because the blood remaining inside the bore of
the needle after use contains a larger volume of virus than the relatively small
amount of blood remaining on the outside of a solid core needle (e.g., a suture
needle). The amount of blood the health care worker is exposed to during the
exposure event is one of the risk factors assessed to determine the type of postexposure prophylaxis needed. (Figure 2)
Figure 1
Locations Where Percutaneous Injuries and Mucocutaneous
Blood Exposures Occurred,
U.S. EPINet, 1995
63 hospitals, 3,552 total incidents
(Source: Ippolito, 1997)
Safer Needle Devices:
Protecting Health Care Workers
Occupational Risk of Needlestick Injuries
Q. How serious is the occupational risk of needlestick injuries to health
care workers?
A. One of the most critical control components of health care worker protection
against bloodborne pathogens must be the reduction of sharps-related
incidents. The statistics cited below provide a picture of the seriousness of the
Florida Heart CPR*
Safer Needle Devices
4
problem.




Hospital-based health care workers experience some 800,000
needlestick injuries each year in the United States (Jagger,1990).
About 2% or approximately 16,000 of these are likely to be
contaminated by HIV (American Hospital Association, 1992).
Needlestick injuries account for up to 80% of all accidental exposures
to blood. (Jagger, 1988).
Based on various studies, researchers have documented that
needlestick injuries are under reported by health care workers and the
number of exposures could potentially be much higher (Hamory, 1983).
Chiarello (1992) cites several studies that found rates of under-reporting
between 40.4% and 53% for nurses and 92% for laboratory personnel.
Physicians under reported needlestick injuries by 70% to 95%.
In well-documented studies, injuries from contaminated needles and
other sharp devices used in the health care setting have been associated
with transmission of bloodborne pathogens to health care workers. In
fact, more than 20 pathogens have been transmitted through sharps
or needlestick injuries (Chiarello, 1992). Of these, HIV, HBV and HCV
pose the greatest risk to the health care worker (Ippolito et al., 1997). The
risk of transmission of HBV and HCV through percutaneous injury is
much higher than for HIV (CDC, 1991).
Understanding the prevalence of these bloodborne diseases in the
patient population gives a better picture of the risk health care workers
face:
o A study of the Johns Hopkins Hospital emergency room
determined the prevalence of the HCV, HBV and HIV in blood
samples from 2523 patients. Of those patients:
 18% were seropositive for HCV
 5% were seropositive for HBV
 6% were seropositive for HIV (Kalen et al, 1997).
o An estimated 1.25 million people in the U.S. are chronically
infected with HBV and 6,000 die each year as a result of HBVrelated liver disease (Moyer & Hodgson, 1996).
o In 1989, the CDC studied the prevalence of HIV in three inner-city
hospitals and three suburban hospitals. In the three inner-city
hospitals, seroprevalence of HIV ranged from 4.1 to 8.9 per 100
patients. The suburban hospitals' seroprevalence rates ranged
from 0.2 to 6.1 per 100 patients. (Marcus et al., 1989)
o In 1997, the National Institutes of Health estimated that nearly 4
million people in the U.S. are infected with Hepatitis C (NIH, 1997).
Of these, 85% develop chronic HCV infection and the potential for
transmission of HCV to others (CDC, 1997).
Occupational Risk of Hepatitis B
Florida Heart CPR*
Safer Needle Devices
5
Q. What occupational risk does Hepatitis B pose to the health care
worker?
A. For more than 50 years, HBV infection, a well-documented and recognized
occupational hazard, has been and continues to be one of the most common
bloodborne pathogens among health care workers. Studies conducted prior
to implementation of recommendations to prevent bloodborne pathogen
transmission (1976-1985) show that health care workers had a prevalence of
HBV infection three to five times higher than the general U.S. population
(Moyer & Hodgson, 1996).



Hepatitis B is much more transmissible than HIV. Health care workers
at greatest risk work in areas where they are directly exposed to blood (
in emergency rooms, clinical laboratories, operating rooms, hemodialysis
units, etc.) (CDC, 1991).
The risk of a health care worker contracting HBV from needlestick injuries
ranges from 6% to 30% (CDC, 1991).
In 1994, 1000 health care workers developed HBV infection, and each
year 100 to 200 health care workers die from this disease (CDC, 1997).
The Centers for Disease Control and Prevention (CDC) estimates that
the annual number of new HBV infections in health care workers has
steadily declined from 12,000 in 1985, due in part to the widespread
adoption of universal precautions, vaccination against hepatitis B, and the
implementation of OSHA's Bloodborne Pathogens Standard. (Moyer &
Hodgson, 1996).
Occupational Risk of Hepatitis C
Q. What occupational risk does Hepatitis C pose to the health care
worker?
A. Hepatitis C virus infection is a major cause of chronic liver disease in the
United States and worldwide. The virus, because of its similarity to HBV,
presents an occupational risk to persons whose work activities involve
handling human blood and body fluids (CDC, 1997).




Needlestick injuries are the most common cause of occupational HCV
exposure (Hibberd, 1995).
In 1995, an estimated 560 to 1,120 cases of HCV infection occurred
among health care workers who were occupationally exposed to blood
(Alter, 1993).
No vaccine is available for hepatitis C and no effective post-exposure
prophylaxis is known at this time (CDC, 1997).
Screening tests for hepatitis C antibodies are commercially available,
but interpretation of the results, especially in a post-exposure situation, is
limited by several factors.
Florida Heart CPR*
Safer Needle Devices
6




A positive result does not distinguish between acute, chronic, or past
infection, and a negative result does not indicate the absence of acute
infection, only the absence of antibodies to HCV.
False positives are common in populations with a low prevalence of HCV.
The tests do not detect HCV antibodies in approximately 5% of people
(CDC, 1997).
As many as 85% of all HCV-infected persons develop chronic
infection. Persons with chronic hepatitis are at increased risk for
cirrhosis and primary hepatocellular carcinoma. Hepatitis C is now the
leading reason for liver transplantation in the United States. (NIH, 1997).
Occupational Risk of HIV
Q. What occupational risk does HIV pose for the health care worker?
A. HIV infection has been reported after occupational exposures to HIV-infected
blood through needlesticks or cuts; splashes in the eyes, nose, or mouth;
and skin contact.




Exposures from needlesticks or cuts cause most infections. The
average risk of HIV infection after a needlestick exposure to HIV-infected
blood is 0.3% or 1 in 300. Even though the risk of seroconversion after
needlestick is relatively rare, injured health care workers may suffer
disabling physical side effects from post-exposure anti-viral medication as
well as severe emotional trauma as they await their test results (CDC,
1991).
The risk after exposure of the eye, nose, or mouth to HIV-infected blood
is estimated to be, on the average, 0.1% or 1 in 1,000 (CDC, 1987).
The risk after exposure of the skin to HIV-infected blood is estimated to
be less then 0.1%. The risk may be higher if the skin is broken or if the
contact involves a large area of skin or is prolonged (CDC, 1987).
As of December 1998, CDC had received reports of 54 documented
cases and 134 "possible" cases (i.e., documentation was unclear) of
occupationally acquired HIV among health care workers in the United
States. Of the 54 documented cases of occupationally acquired HIV
infection, 46 resulted from needle sticks or cuts (CDC HIV/AIDS
Surveillance Report, 1998) (Table 2). The total number of occupationally
acquired HIV infections in health care workers is increasing each year
(Figure 3).
Q. Why are universal precautions and personal protective equipment not
adequate to protect the health care worker against needlestick injuries?
A. Using universal precautions, along with personal protective equipment,
engineering controls and other work practice controls, reduces employee
exposure to bloodborne pathogens. Personal protective equipment provides a
Florida Heart CPR*
Safer Needle Devices
7
barrier to protect skin and mucous membranes from contact with blood and
other potentially infectious material (OPIM), but most personal protective
equipment is easily penetrated by needles. Needlestick injuries are caused
by unsafe needle devices rather than careless use by health care workers.
(Jagger, 1988). Safer needle devices have been shown to significantly reduce
the incidence of accidental needlesticks and exposure to potentially fatal
bloodborne illnesses (CDC, 1997).
Definition of a Safer Needle Device
Q. What is a safer needle device?
A. A safer needle device incorporates engineering controls to prevent
needlestick injuries before, during, or after use through built-in safety features.
The term, "safer needle device," is broad and includes many different types of
devices from those that have a protective shield over the needle to those that do
not use needles at all. The common feature of effective safer needle devices
is that they reduce the risk of needlestick injuries for health care workers.
Safer Needle Devices:
Protecting Health Care Workers
Efficacy of Safer Needle Devices
Q. Can safer needle devices prevent needlestick injuries?
A. All needlestick injuries are not preventable, but research has shown that
almost 83% of injuries from hollow-bore needles can be prevented (Ippolito et al,
1997). Many of these needlesticks can be prevented by using devices that
have needles with safety features or eliminate the use of needles
altogether (e.g., needleless IV connectors, self re-sheathing needles, and
blunted surgical needles).
Most current research is hospital based and studies have indicated that a
significant portion of needlestick injuries occur when manipulating IV lines or
administering IV and IM injections (Jagger, 1988). In 1992 the FDA published a
safety alert warning of the risk of needlestick injuries from the use of hypodermic
needles as a connection between two pieces of IV equipment. This alert was
based on research that demonstrated that secondary IV tubing with connector
needles was associated with the highest risk of needlestick injury. The use of
needleless IV systems or systems with recessed needles to connect adjoining
equipment was strongly encouraged in this alert (See Appendix A).
Florida Heart CPR*
Safer Needle Devices
8
Two new studies indicate that the use of safer needle designs can reduce the
risk of needlestick injuries among health care workers. The National Center for
Infectious Diseases (NCID) Hospital Infections Program conducted the studies
in collaboration with eight hospitals in three U.S. cities. The results appear in the
January 17, 1997 issue of MMWR. (See References)
The first study describes the use of blunt suture needles during gynecologic
surgery, and indicates that blunt suture needles may reduce the likelihood of a
needlestick during surgery by as much as 86%. The second study examined the
use of safer needles for drawing blood and found that safer needles for drawing
blood may reduce needlesticks to health care workers by 27% to 76%. The
investigations also found that the use of safer needles did not lessen the quality
of patient care. Further, the safer needles were generally accepted by health
care workers.
Some devices have not been well-accepted in the clinical setting nor associated
with a significant decrease in injury rate. These results may be explained by lack
of training or lack of support for change in the clinical setting (Chiarello, 1992).
Although all major medical device manufacturers market devices with safety
features, no standard criteria exist for evaluating the safety claims of these
features. Employers implementing needlestick prevention programs should
evaluate the effectiveness of various devices in their specific settings.
Characteristics of Safer Needle Devices
Q. What are the design features of a safer needle device?
A. The Food and Drug Administration (FDA, 1992, 1995) has suggested that a
safety feature designed to protect health care workers should:





Provide a barrier between the hands and the needle after use;
The safety feature should allow or require the worker's hands to remain
behind the needle at all times;
Be an integral part of the device and not an accessory;
Be in effect before disassembly and remain in effect after disposal to
protect downstream workers; and
Be simple and self evident to operate and require little or no training to
use effectively.
Features designed to protect the health care worker should not "have a negative
impact on the delivery of patient care. As an example, there are a few reports of
increased bloodstream infections with the use of needleless IV systems"
(Chiarello, 1997).
Florida Heart CPR*
Safer Needle Devices
9
Q. What are some of the types of safety features used in safer needle
devices?
A. The types of safety features used in safer needle devices can be categorized
according to certain aspects of the safety feature, i.e., whether the feature is
active or passive and whether the engineering control is part of the device
(Chiarello, 1995).




Passive safety features remain in effect before, during and after use;
health care workers do not have to activate them. Passive features
enhance the safety design and are more likely to have a greater impact
on prevention.
Active devices require the health care worker to activate the safety
mechanism. Failure to do so leaves the worker unprotected. Proper use
by health care workers is the primary factor in the effectiveness of these
devices.
An integrated safety design means that the safety feature is built in as
an integral part of the device and cannot be removed. This design feature
is preferred.
An accessory safety device is a safety feature that is external to the
device and must be carried to or temporarily or permanently fixed to the
point of use. This design also is dependent on employee compliance and
according to some researchers, is not desirable.
Evaluation and Selection of Safer Needle Devices
Q. How can health care employers evaluate and select safer needle
devices?
A. OSHA's Bloodborne Pathogens Standard requires that "Each employer
having an employee(s) with occupational exposure ... shall establish a written
Exposure Control Plan designed to eliminate or minimize employee exposure"
[29 CFR 1910.1030(c)(1)(ii)(B)]. A variety of safer needle devices is now widely
available. Manufacturers have responded to the need for safer devices and as a
result, a wave of safer medical products have flooded the marketplace. One
thousand U.S. patents for safer medical devices have been issued since 1984
(Ippolito,1997). Employers are faced with the tremendous task of selecting and
evaluating products from the vast array of devices available.
Although OSHA does not require employers to institute the most sophisticated
engineering controls, it does require the employer to evaluate the effectiveness
of existing controls and to review the feasibility of instituting more advanced
engineering controls (CPL 2-2.44C). OSHA staff should encourage an employer
to implement a needlestick prevention program. Research to date has shown
that no single safer needle device will work equally well in every facility so
Florida Heart CPR*
Safer Needle Devices
10
employers must develop their own programs to select the most appropriate
devices. The goal is to choose devices that are:



Clinically effective,
Acceptable to users, and
Most effective in reducing needlestick injuries in that particular
setting (Chiarello, 1995).
To evaluate and select appropriate safer needle devices, health care employers
should review available needlestick injury data including the personnel involved,
the devices used, and the circumstances and frequency of needlestick events.
This information can assist the employer in determining how employees can
maximally benefit from a product change to safer needle devices. Although not
required by OSHA, collection and evaluation of complete needlestick injury
data by hospitals are key to identifying injury patterns and then
implementing an effective abatement plan (Chiarello, 1995).
Q. What are the steps a health care employer should consider in
developing a comprehensive needlestick prevention program and
implementing safer needle devices?
A. Chiarello (1995) describes a framework to aid health care employers in
establishing a comprehensive program to select and evaluate safer medical
devices in a systematic manner. The product choice should ideally be based on:


The needs of the primary users.
The need of the patients who must continue to receive safe, efficient,
and comfortable care. Health care workers are likely to reject products
that they think will interfere with patient care in any way.
Chiarello (1995) suggests that a comprehensive needlestick prevention program
might include the following:





Creating of a multi-disciplinary team.
Defining prevention priorities on the basis of collection and analysis of an
institution's injury data.
Developing design and performance criteria for product selection
according to needs for patient care and health care worker safety.
Planning and implementing an evaluation of products in clinical settings.
Analyzing product performance and cost-effectiveness to choose the
product.
OSHA's Position on Safer Needle Devices
Q. What is OSHA's position on safer needle devices?
Florida Heart CPR*
Safer Needle Devices
11
A. OSHA's position:





Section (d)(2)(i) of the Bloodborne Pathogen Standard requires the use of
engineering and work practice controls to eliminate or minimize
employee exposure.
CPL-2.44C states that "Section (d)(2) shall be cited for failure to use
engineering/work practice controls."
CPL 2-2.44C also states that:
"Most preferable is the use of devices which offer an alternative to
needles being used to perform the procedure. Examples of such devices
include stopcocks (on-off switch), needle-protected systems or
needleless systems which can be used in place of open needles to
connect intravenous lines. Other devices which are integral to the
syringe, such as self-sheathing needles, allow both hands to remain
behind the needle and require very little manipulation to isolate the
needle safely." In addition, "While employers do not automatically have to
institute the most sophisticated engineering controls ( e.g., needleless IV
connectors, self-sheathing needles), it is the employer's responsibility to
evaluate the effectiveness of existing controls and to review the feasibility
of instituting more advanced engineering controls."
Health care workers in settings where safer needle devices have not
been implemented may be at higher risk of sustaining a needlestick
injury.
Therefore, failure to use engineering and work practice controls could
result in a citation.
Q. What are some resources one can use to update their knowledge of
safer needle devices, needlestick prevention programs, and current
statistics about bloodborne pathogens?
A. The face of the bloodborne diseases is changing rapidly as new treatments
are developed and new ways of preventing disease are introduced.
Manufacturers continue to introduce new products with more efficient and less
costly safety features. A variety of resources are available on this subject.
Appendix C contains a partial list of helpful resources as well as the reference
material that is provided on the following pages.
Q. What conclusions can be drawn from this information?
A. Safer needle devices can protect employees from occupational
exposure to blood and other potentially infectious materials. When the first
case of occupationally transmitted HIV through needlestick was reported in
1984, a new awareness about the occupational hazards faced by health care
workers emerged. Employers adopted aggressive prevention strategies in
response to increased awareness of risk to workers from HIV and other
Florida Heart CPR*
Safer Needle Devices
12
bloodborne pathogens, but the health care worker's risk of acquiring a
potentially lethal infection through needlestick injury remains very real. Safer
needle devices can protect health care workers from exposure to lifethreatening diseases by preventing needlestick injuries.
References
Alter, M.J.. "The Detection, Transmission, and Outcome of Hepatitis
C Virus Infection." Infectious Agents and Disease. 2(8): 155-166,
1993.
American Hospital Association. Implementing Safer Needle Devices.
December, 1992.
Centers for Disease Control and Prevention. "Recommendations for
Preventing Transmission of Human Immunodeficiency Virus and
Hepatitis B Virus to Patients During Exposure-Prone Invasive
Procedures." MMWR Recommendations and Reports. 40(RR-8): 19, 1991.
Centers for Disease Control and Prevention. "Evaluation of Safety
Devices for Preventing Percutaneous Injuries Among Health care
Workers During Phlebotomy Procedures - Minneapolis, St. Paul,
New York City, and San Francisco, 1993-1995." MMWR. 46(2): 2129, 1997.
Centers for Disease Control and Prevention. "An Evaluation of Blunt
Suture Needles in Preventing Percutaneous Injuries Among Health
Care Workers During Gynecological Surgical Procedures - New
York City, March 1993-June 1994." MMWR. 46(2): 29-33,1997.
Centers for Disease Control and Prevention. HIV/AIDS Surveillance
Report. Atlanta, GA: Division of HIV/AIDS Prevention, National
Center for HIV, STD, and TB Prevention, Centers for Disease
Control and Prevention. Vol. 8. No. 2. 1996.
Centers for Disease Control and Prevention. HIV/AIDS Surveillance
Report. Atlanta, Ga: Division of HIV/AIDS Prevention, National
Center for HIV, STD, and TB Prevention, Centers for Disease
Control and Prevention. 9 (1), 1997.
Centers for Disease Control and Prevention. "Recommendations for
Prevention of HIV Transmission in Health care Settings." MMWR. 35
( 2S): 1W-18S, 1987.
Centers for Disease Control and Prevention. Draft Guideline for
Florida Heart CPR*
Safer Needle Devices
13
Infection Control in Health Care Personnel, 1997. Washington, D.
C.: GPO, 1997.
Chiarello, L.A. "Selection of Safer Needle Devices: A Conceptual
Framework for Approaching Product Evaluation." American Journal
of Infection Control. 23( 6):386- 395, 1995.
Chiarello, L.A. Written private communication, to Susan Elliott,
September, 1997.
FDA. "Needlestick and Other Risks From Hypodermic Needles on
Secondary I.V. Administration Sets-Piggyback and Intermittent I.V."
FDA Safety Alert. April 16, 1992.
FDA. Supplementary Guidance on the Content of Premarket
Notification (510K) Submissions for Medical Devices with Sharps
Injury Prevention Features (Draft). Rockville, MD: General Hospital
Devices Branch, Pilot Device Evaluation Division, Office of Device
Evaluation, 1995.
Hamory, B.H. "Under Reporting of Needlestick Injuries in a
University Hospital." American Journal of Infection Control. 11(5):
174-177, 1983.
Hibberd, P.L. "Patients, Needles, and Health Care Workers:
Understanding the Epidemiology, Pathophysiology, and
Transmission of the Human Immunodeficiency Virus, Hepatitis B
and C, and Cytomegalovirus." Journal of Intravenous Nursing. 18 (
6S): S22-S31, 1995.
Ippolito, G., Puro, V., Petrosillo, N., Pugliese, G., Wispelwey, B.,
Tereskerz, P. M., Bentley, M., & Jagger, J. Prevention,
Management, and Chemoprophylaxis of Occupational Exposure to
HIV. Charlottesville, VA: Advances in Exposure Prevention,
International Health Care Worker Safety Center, 1997.
Jagger, J. "Rates of Needlestick Injury Caused by Various Devices
in a University Hospital." N England J Med. 319(5): 284-8, 1988.
Jagger, J. (1990). Preventing HIV transmission in health care
workers with safer needle devices. Sixth International Conference
on AIDS. June 22, 1990: San Francisco, CA.
Kelen, G.D., Green, G.B., Purcell, R.H., Chan, D.W., Qaqish, B.F.,
Sivertson, K.T., Quinn, T.C. "Hepatitis B and Hepatitis C in
Emergency Department Patients." N Engl J Med. 326(21):1399-
Florida Heart CPR*
Safer Needle Devices
14
1404, 1992.
Kroc, K., Pugliese, G. Implementing safer needle devices. Chicago:
Florida Heart CPR*
Safer Needle Devices Assessment
1. One of the most critical control components of health care worker protection
against bloodborne pathogens must be the reduction of:
a. Interaction with patients infected with bloodborne pathogens
b. Activities involving infected blood
c. Sharps-related incidents
d. Clumsy health care providers
2. Based on various studies, researchers have documented that needlestick injuries
are _______by health care workers and the number of exposures could
potentially be much higher.
a. Under reported
b. Over reported
c. Falsely reported
d. Inaccurately reported
3. For more than 50 years, ______, a well-documented and recognized
occupational hazard, has been and continues to be one of the most common
bloodborne pathogens among health care workers.
a. HAV
b. HBV
c. HCV
d. HIV
4. Hepatitis B is much more transmissible than HIV. Health care workers at greatest
risk work in areas where they are directly exposed to blood, for example:
a. In emergency rooms and operating rooms
b. In clinical laboratories
c. In hemodialysis units
d. Any of the above
5. Needlestick injuries are the most common cause of ______ exposure.
a. HAV
b. HIV
c. Occupational HCV
d. None of the above
6. HIV infection has been reported after occupational exposures to HIV-infected
blood through:
Florida Heart CPR*
Safer Needle Devices
15
a.
b.
c.
d.
Skin contact
Splashes in the eyes, nose, or mouth
Needlesticks or cuts
All of the above
7. All needlestick injuries are not preventable, but research has shown that almost
___% of injuries from hollow-bore needles can be prevented.
a. 58
b. 67
c. 83
d. 92
8. The Food and Drug Administration (FDA, 1992, 1995) has suggested that a
safety feature designed to protect health care workers should:
a. Provide a barrier between the needle and the hands after use
b. Be simple and self evident to operate and require little or no training to use
effectively
c. Be in effect before disassembly and remain in effect after disposal to protect
downstream workers
d. All of the above
9. _______means that the safety feature is built in as an integral part of the device
and cannot be removed. This design feature is preferred.
a. Integrated safety design
b. An active device
c. Passive safety features
d. Accessory safety device
10. _______ remain in effect before, during and after use; health care workers do not
have to activate them. This enhances the safety design and is more likely to have
a greater impact on prevention.
a. Integrated safety design
b. An active device
c. Passive safety features
d. Accessory safety device
Florida Heart CPR*
Safer Needle Devices