Download SY1. CT Laser Light Accuracy Laser light accuracy must be

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Photon scanning microscopy wikipedia , lookup

Microscopy wikipedia , lookup

Light wikipedia , lookup

Laser wikipedia , lookup

Anti-reflective coating wikipedia , lookup

Photonic laser thruster wikipedia , lookup

Nonlinear optics wikipedia , lookup

Magnetic circular dichroism wikipedia , lookup

Surface plasmon resonance microscopy wikipedia , lookup

Thomas Young (scientist) wikipedia , lookup

Confocal microscopy wikipedia , lookup

Harold Hopkins (physicist) wikipedia , lookup

Ultraviolet–visible spectroscopy wikipedia , lookup

Ultrafast laser spectroscopy wikipedia , lookup

Retroreflector wikipedia , lookup

Super-resolution microscopy wikipedia , lookup

Optical coherence tomography wikipedia , lookup

Transcript
SY1. CT Laser Light Accuracy
Laser light accuracy must be determined for both axial scan localization lights,
which indicate the location of the radiation beam relative to the external
anatomical structures of the patient, and the sagittal and coronal scan
localization lights, which centre the anatomic structure of interest in the scan field
of view. This can be tested in several ways as shown below:
1. Axial Scan Localization Lights—Axial scan localization light accuracy is
tested either by imaging a thin wire (approximately 1 mm diameter) or by
using a needle to puncture holes in film at the positions of the laser lights
and exposing the film using the smallest available scan width. The
difference between the exposed areas on the film and the locations of the
pin pricks must be less than ± 2 mm.
2. Some manufacturers’ phantoms have fiducial markers which can be
aligned wit the external lasers and then seen in the image using a small
slice width, as in the GE CT phantom shown below. The ACR phantom
(Gammex 464) has beads at the surface which can be used in the same
way.
CT slice through alignment section
of ACR phantom showing surface
steel beads
GE Phantom showing deeper grooves for
alignment test
3. Isocentre Alignment and Sagittal and Coronal Scan Localization Lights—
Place a thin absorber (pencil) centered in the tomographic plane at the
intersection of the sagittal and coronal positioning light fields.
CT slice through pencil
The intersection of the sagittal and coronal scan localization lights must indicate
the centre of the field of view. The midline of the table should be coincident with
the sagittal scan alignment light. The results must be within the manufacturer’s
recommended values and tolerances. A limit of ± 5 mm should be achievable.