* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Download Section 3 - Haley Alison Colby
Hearing loss wikipedia , lookup
Auditory system wikipedia , lookup
Evolution of mammalian auditory ossicles wikipedia , lookup
Sound localization wikipedia , lookup
Sound from ultrasound wikipedia , lookup
Noise-induced hearing loss wikipedia , lookup
Sensorineural hearing loss wikipedia , lookup
Audiology and hearing health professionals in developed and developing countries wikipedia , lookup
Operating Instructions Manual Permission from Grason-Stadler Page 1 Contents Section 1: Operation Manual Introduction………………………… 3 1.1 – Introduction……………………………………………………………………………………. 1.2 – Audience ……………………………............................................................... 1.3 - Purpose……………………………………………………………………………………………. 1.4 – Audiometer…………………………………………………………………………………….. 3 3 3 Section 2: Terms…………………………………………………………………….. 4 2.1- Terms to be familiar with………………………………………………………………….. 4-5 Section 3: Care and Operations…………………………………………. 5 3.1- Care of the Audiometer……………………………………………………………………. 5 3.2- Mechanical Function Check……………………………………………………………… 5-6 Section 4: LCD Screen Display……………………………………......... 7 4.1- Figure 1.1…………………………………………………………………………….. 7 4.2- Screen Indicators…………………………………………………………………. 7 Section 5: Front Panel……………………………………………………………. 8 5.1- Figure 1.2…………………………………………………………………………….. 8 5.2- Front Panel Control and Indicators………………………………………. 8-9 5.3- Typical Testing Session………………………………………………………… 9 Section 6: Instructions………………………………………………………….……… 6.1- Procedure…….…………………………………………………………………………………… 10 10-12 Section 7: Safety Warnings…………………………………………………………. 13 Section 8: Troubleshooting………………………………………………………… 14 Permission from Grason-Stadler Page 2 Section 1 – Operation Manual Introduction 1.1 INTRODUCTION Hearing screening with a pure tone audiometer conducted by properly trained personnel will lead to early identification of hearing loss. The hearing screening process can identify individuals with outer and middle ear problems, as well as hearing loss caused by damage to the inner ear. The screening program is intended to identify individuals in need of further diagnosis and treatment. 1.2 AUDIENCE This manual is intended for use of performing a hearing screening in clinical and school based settings. 1.3 PURPOSE The purpose of this hearing training manual is to provide the screener with instructional information to conduct pure tone hearing screening in the school or clinic setting. Detailed description of each procedure will be presented in this manual. These guidelines are here to help perform a screening. 1.4 AUDIOMETER An audiometer is a device that produces tones at set frequencies and known amplitudes through headphones worn by the patient. The patient is required to indicate what tones he/she can hear. The pure tone audiometer is the preferred instrument used in school and clinic hearing screening programs. The audiometer has test tones ranging from 250 to 8,000 Hz (Hz refers to frequency, or cycles per second). This is the pitch of the tones that range from very low pitch to high pitch. The tones most often used are 500, 1,000, 2,000, and 4,000 Hz. The intensity (loudness) is signified in decibels (dB). Thirty is considered the level of a soft whisper. Sixty-five is the average intensity for a normal conversation, and eight-five is so loud that individuals must be protected when working under exposure. *Note, the audiometer that will be used in this screening is GSI 61. Permission from Grason-Stadler Page 3 Section 2 – Terms 2.1 Terms to be Familiar With Audiogram – A graph on which a person’s ability to hear sounds at different frequencies are recorded. Decibel((dB) – Expresses the intensity of a sound. Frequency Dial: This dial rotates from 250 to 8000 Hz. It controls the test frequency, or pitch of the tone. Hearing Level Dial or Decibel (dB) Dial: This controls the test intensity, or loudness of the tone. Normally this is a rotary dial, which allows you to vary the tones presented in 5 dB steps. Hearing Level – The amount of hearing loss indicated by audiometry and measured in terms of decibels for any given frequency. Hearing Loss – Varies greatly from person to person. The American Speech Language and Hearing Association (ASHA) classify hearing loss as follows: 0-20 dB – normal 21-25 dB – slight 26-40 dB – mild 41-55 dB – moderate 56-70 dB – moderately severe 71-90 dB – severe 91+ dB – profound Power or On/ Off Switch: This switch needs to be on when testing. Pure Tone – A single frequency produced by an audiometer. A pure tone contains no harmonics or overtones. Permission from Grason-Stadler Page 4 Threshold – The softest (minimum) hearing level at which an individual is able to respond to a tone. Tone Presentation Bar or Stimulus Switch: Pressing this bar (or switch) presents the tone to the person being tested. Section 3 – Care and Operations 3.1 Care of the Audiometer General care: Handle gently Earphones: Clean routinely with cleaning agent, alcohol free wipes. When the earphone cushions need cleaning, remove them from the headset, clean and dry thoroughly before replacing. The audiometer is in need of repair if: 1.) Tone does not sound normal or sound is not produced when tone lever or button is pressed or static is heard. 2.) Earphones do not remain in proper position over ears 3.2 Mechanical Function Check Before using the audiometer, plug it in for ten minutes and then proceed to the following list. 1. Power On- Make sure there is power to the audiometer 2. Earphone Cushions- The cushions should be clean and free from breaks or tears. DO NOT use alcohol wipes on the cushions. Permission from Grason-Stadler Page 5 3. Headband Tension- Put on the earphones. There should be enough tension so that when positioned on the head, the earphones rest snugly on the ears. 4. Tone ON/OFF – Turn the switch to on position and the sound should be on. The sound should be off in (REV) position. 5. Volume Increase/Decrease- Turn the hearing level dial and listen for loudness changes. 6. Pitch Change- Change the frequency dial and listen for changes in pitch. 7. Tone Presenter Switch(es)- Press the switch, and the sound should go off when the switch is initiated. 8. Static- No static should be heard. Permission from Grason-Stadler Page 6 Section 4– Screen Display 4.1 Figure 1-2 G1 G2 G3 G4 G5 Section 4.2- Screen Display Indicators G1- Indicates the frequency desired G2- Indicates the hearing level (dB) G3- Indicates that you have changed the stimulus to tone G4- Shows you have changed the transducer to phone G5- Routing will tell you which ear you are testing Permission from Grason-Stadler Page 7 Section 5 – Front Panel 5.1 Figure 1-1 F1 11 F7 F2 F3 F4 F5 F6 5.2 The Front Panel Controls and Indicators (Figure 1-1) F1-This control allows the operator to choose the type of tone presented to the test subject. It can be set on steady, pulsed (press the button 3 times) or FM, which is one steady tone. F2-This button indicates you want to deliver pulsed tones. F3- Indicates a knob for setting the stimulus intensity level. Level is indicated on the center top of the display. Permission from Grason-Stadler Page 8 F4- Present this bar for stimulus presentation. The symbol appears on the left side of the display when the stimulus is being presented. F5-Allows you to talk to the patients using a headphone set. F6-changes the frequency desired F7-This will present a pulse tone. 5.3- Typical Testing Session 5.3.1 Pre-Test Review a. Turn the instrument on. b. Check that the earphones are operating properly. c. Seat subject comfortably in the test area. d. Explain the test procedure. e. Assist subject with the earphones. f. Select the desired type. g. Begin the procedure Permission from Grason-Stadler Page 9 Section 6: Procedures 6.1.1 Instructions Your instructions should be clear and brief, letting the client know what to expect and how to respond. The client needs to understand: They are listening for very soft sounds, they have to respond every time they hear the sound, how to let you know they have heard the sound. One way of instructing the client is to say something like: “You will be hearing a series of beeps, first in the right ear and then in the left. As soon as you hear the sound, no matter how soft it is, let me know by pressing the button. Like this (indicate how to press the button). 6.1.2 Placing the Earphones The earphones must be put on carefully. Place the diaphragm of the earphones directly over the opening of the ear canal and the band sits across the top of the head. If the earphones are not placed directly over the opening of the ear canal, hearing thresholds could appear worse than they are. 6.1.3 What Ear Always ask for better ear. If the client is unaware always test the right ear first. 6.1.4 Client Is Ready Once your client is all set and doesn’t have any further questions, start the testing. 6.1.4.1 Return to the control room and put the headset/microphone on. Permission from Grason-Stadler Page 10 6.1.5 Get Equipment Ready Using Channel 1, set transducer to phone, stimulus to tone, and routing to right ear. To the left of Channel 1, you will push pulse in order to use a pulsed tone. Using the talk forward button, inform the person you are about to begin. 6.1.6 Start Threshold Testing – Using the banana shaped presentation button under channel 1, present a pulsed tone (3-5 beeps) to the right ear at 30 DB AT 1000 Hz. Set the routing selector to route the test tone to the selected ear. Demonstrate the 1000 Hz. Tone at a 30 dB level. Duration of tone should be between one and two seconds (push the button three times). Set HL control to -10 dB, gradually increase the intensity by turning the hl control until a response occurs. Switch the tone off for at least two seconds and present the tone again at the same level. If there is a second positive response, proceed to threshold measurement. If a second response does not occur, repeat this step. NOTE: LOOK AT FIGURE 1-3 FOR WHICH FREQUENCIES TO BE TESTED ON THE NEXT PAGE Permission from Grason-Stadler Page 11 Figure 1-3 Frequency Testing: Right ear Left ear 1. 1000Hz 8. 1000Hz 2. 2000Hz 9. 2000Hz 3. 4000Hz 10. 4000Hz 4. 8000Hz 11. 8000Hz 5. 1000Hz 12. 1000Hz 6. 500Hz 13. 500Hz 7. 250HZ 14. 250HZ 6.1.7 Threshold Determination The level of the first presentation shall be 10 dB below the level at which the subject responded. The tone for a period of one or two seconds. Time between tones can be varied, but not shorter than the duration of the test tone itself. After each failure to respond to a signal, the level is increased by 5 dB until the first response occurs. After the response, the intensity is decreased 10 dB and another series begins. NOTE Descend in 10 dB steps and ascend in 5 dB steps. *The threshold is considered to be the minimum dial setting at which a response has occurred two out of three times Permission from Grason-Stadler Page 12 6.1.8 Testing Procedures Repeat steps (6.1.5 and 6.1.6) for each tone- setting in the following order: 1000, 2000, 3000, 4000, and 8000 Hz. Retest 1000 followed by 500 and 250. If there is a difference of 20 dB or more between two octaves, test the inter-octave responses at 750, 1500, 3000 Hz. 6.1.9 Repeat steps (6.1.5, 6.1.6, and 6.1.7) for the left ear. Section 7: Safety Warnings 1. Do not turn on/ off system while a patient is wearing the headsets or internal earphones. 2. Use only the provided power supply. 3. To ensure patient safety and optimal performance, use only GS1 61 recommended accessories and supplies. Permission from Grason-Stadler Page 13 Section 8: Troubleshooting If the audiometer is no longer working properly, please check the following points: LCD screen does not light up No sound from headphone Is the power switch on? Is the power cord plugged in correctly in the power connector? Is the wall outlet working? Is the headphone cable plugged correctly in socket? Is the head phone defective If your instrument still does not work properly after this short check, please consult the company’s service center. Permission from Grason-Stadler Page 14