Biopotential Amplifier
Speaker: Sun Shih-Yu
3/20, 2006
Outline




Requirements
A standard ECG
Problems frequently encountered
Amplifiers for various biopotential
signals
Requirements





Large input impedance; small output
impedance
Frequency response
High gain
Protection
Differential amplifier


High CMRR (common mode rejection ratio)
Quick calibration
Problems






Frequency distortion
Saturation or cutoff distortion
Ground loop
Open lead wires
Artifact from large electric transients
Interference
Voltage and freq. ranges for
common biopotential signals
Large electric transient




Defibrillation
Motion of the electrodes
Built-up static electric charge
Older equipment: different offset
voltage from one lead to another
Interference



Electric power system
Magnetic induction
EM interference


Shunting a small capacitor (200pF)
EMG interference
Interference from electric
power systems
v A  vB  id1Z1  id 2 Z 2
id 1  id 2
v A  vB  id1 (Z1  Z 2 )
vA  vB  (6nA)(20k)  120V
Interference from electric
power systems (cont’d)
vcm  idb Z G
vcm  (0.2A)(50k)  10mV
v A  vB  vcm (
Z in
Z in

)
Z in  Z1 Z in  Z 2
v A  vB  vcm (
Z 2  Z1
)
Z in
vA  vB  (10mV )(20k / 5M)  40V
Voltage and freq. ranges for
common biopotential signals
Interference observable!
Interference from magnetic
induction



Shielding
Keep away from magnetic-field regions
Reduce the effective area of the single
turn coil
Amplifiers for various
biopotential signals



EMG amplifier
Amplifiers for intracellular electrodes
EEG amplifier
Amplifiers for various
biopotential signals

different spectrum and amplitude
constraints
EMG amplifier



Amplitude depends on the electrode
used and signal
Frequency spectrum wider than ECG
Less motion interference due to higher
frequency band
Amplifiers for intracellular
electrodes



measure the potential across the cell
membrane
Frequency response must be wide
Amplitude in the order of 50 to 100mV;
gain needs not be high
Amplifiers for intracellular
electrodes (cont’d)


Even large input impedance due to
large source one
Geometry results in a relatively large
shunting capacitance

Use positive feedback to produce negative
capacitance
Compensating positive
feedback
Compensating positive
feedback (cont’d)
1
vi 
Cf
 i dt  A v
i
1
vi 
(1  Av )C f
 vi 
v i
 i dt
i
1
ii dt

Ceq
 Ceq  (1  Av )C f
However……
• gain is frequency dependent
• may be unstable because of positive feedback
• tends to be noisy
EEG amplifier




Low level of signal; Higher gain
Small electrodes; higher input
impedance
Higher CMRR
Low noise amp