MYOCARDIAL PERFUSION STUDY
(Tc-99m-Sestamibi / Tc-99m Tetrofosmin / Tl-201 Chloride)
2.7.4
Radiology Associates of Clearwater
Overview
The Myocardial Perfusion Study demonstrates the distribution of blood flow and
perfusion to the myocardium at stress and rest (1).
The preferred study will be a same day exam performed with Tc-99m MIBI.
Indications
Detection of coronary artery disease (2).
Emergency room evaluation of chest pain (3).
Documentation of myocardial perfusion abnormalities before and after
interventional therapy (4).
Detection of hibernating myocardium in conjunction with thallium-201 or
F-18-fluorodeoxyglucose (5-7).
*Exams ordered for indications which are not listed above need to be discussed
with the Nuclear Medicine Physician.
Examination Time: One Day Tc-MIBI/MIBI or 201-Tl/MIBI Study
Rest study (8):
1. Initially: 15 minutes for injection.
2. 45 minutes later: 15 minutes for image acquisition (MIBI)
5-10minutes later: 15 minutes for imaging acquisition (201-Tl).
Stress study (Immediately after rest imaging)
1. Initially: 20 minutes for stress and injection.
2. 45 minutes later: 15 minutes for stress acquisition.
Patient Instructions
Instruct patient to schedule treadmill/medication induced stress test through the
Nuclear Medicine Department. If patient weight is >300 pounds or BMI is >= 35%
then the exam will be schedules as a two day Tc-MIBI study.
Obtain patient's weight, as treadmill limit is 350 lb., and pharmacologic stress
calculations must be based on weight.
Instruct outpatients to bring/wear comfortable shoes and clothing appropriate for
exercise.
For optimal results the patient should discontinue all cardioactive medications
before the study (9). See protocol 2.6 for list of medications:
1. Beta-blockers, e.g. propranolol, for at least 24 hours.
2. Long acting nitrates for at least 4 hours, nitroglycerin for at least 1 hour.
3. Calcium channel blockers.
Dietary restrictions
-No caffeine for 24-36 hours prior to pharmacologic stress with adenosine (10).
-Patients scheduled for stress before 11:00am should be NPO after midnight.
-Patients scheduled for stress test after 11:00am may have a light breakfast
(cereal, toast) to be completed no later than 4 hours prior to testing.
-Determine if the patient is an insulin dependent diabetic. These patients should be
scheduled at the beginning of the day whenever possible. Usually breakfast and
insulin are held until completion of the test. Patients should be instructed to contact
their referring physician for specific instructions.
Lab / Imaging correlation:
For inpatients initially admitted to rule out myocardial infarction, all ordered cardiac
enzyme levels must be obtained.
Obtain all previous myocardial perfusion scans and MUGA studies when available.
Patient Preparation:
Before proceeding, check with the patient to ascertain that all instructions
regarding medications, oral intake, and clothing have been followed.
Explain the test to the patient, including the stress consent form(s). This will
require signatures of the patient, witness, and supervising physician.
A team member is to insert a #20 or 22 gauge IV line in a vein as close to the
antecubital as possible. The brachial(medial) vein is preferred.
Outpatients are to remove clothing from waist up, and wear proper protective
gowns
Equipment & Energy Windows
Gamma camera: Rotating gamma camera for SPECT, preferably a dual head
system with the heads at 90 degrees (13). A camera with hardware and software
attenuation correction may be used.
Collimator: Low energy, high resolution, parallel hole (14).
Energy window: 20% window centered at 140 keV. (99m-Tc-MIBI)
20% window centered at 70 keV and 167 keV. (Tl-201)
Computer with SPECT capability (15).
Radiopharmaceutical and Dose
Dual Isotope – Same Day
Rest: 2.5 mCi (92.5 MBq), Tl-201-Chloride IV.
Stress: 22 mCi (925 MBq) Tc-labeled agents (Tetrofosmin, Sestamibi), IV.
Tc-labeled agent (Myoview, Cardiolite) – Same Day
Rest: 8mCi (296 MBq) Tc-labeled agents (Tetrofosmin, Sestamibi), IV.
Stress: 22mCi (814 MBq)Tc-labeled agents (Tetrofosmin, Sestamibi), IV.
Two Day Study
Rest: 22mCi (814 MBq) Tc-labeled agents (Tetrofosmin, Sestamibi), IV.
Stress: 22mCi (814 MBq)Tc-labeled agents (Tetrofosmin, Sestamibi), IV.
Patient Positioning & Imaging Field
Patient position: Supine on the SPECT imaging table (20).
The left arm is placed above the patient's head. (If the patient is unable to keep the
left arm above the head, planar imaging may need to be substituted for SPECT
imaging.
Imaging field: Lower chest.
Acquisition Protocol - One Day Study (14,17)
General:
1. The rest study is performed first followed immediately by the stress study. (14,
17, 21-24). Attenuation correction (25,26) will be used at the discretion of the
nuclear medicine physician.
Rest study:
1. An intravenous line is started and the radiopharmaceutical is injected at rest at
time zero.
2. After 45 minutes the patient is positioned supine for SPECT imaging (Tc-MIBI).
After 5-10 minutes the patient is positioned supine for SPECT imaging (201Tl).
3. SPECT acquisition parameters:
a)
180° collection arc beginning at 45° RAO and ending at 45°
LPO (28).
b)
Orbit: Circular
c)
Projections: 32 images (over 180°) for 60sec each.
d)
e)
Image matrix: 64 x 64 matrix.
Magnification: 1.33
Stress study:
1. EKG leads are placed and gating will be performed at 8 frames per cardiac cycle
(14).
2. The patient undergoes exercise (or pharmacologic) stress.
3. The radiopharmaceutical is injected 1-2 minutes before the end of exercise.
4. The patient is instructed to drink a can of ensure or they may go to the
cafeteria to eat.
5. After a 45 minutes wait the patient is positioned supine for SPECT acquisition.
6. SPECT acquisition parameters (same as rest study - see above + Gating).
Viability Assesment (Tl-201)
If a rest thallium defect is present, a 24 hour redistribution acquisition may be
obtained to differentiate hibernating from infarcted myocardium (34). SPECT
acquisition parameters (same as rest study - see above).
Data Processing (14)
SPECT images are processed for display on the reading stations.
The reconstruction process in general terms is:
1. Correct the planar acquisition images for decay from the start of image
acquisition.
2. Correct the 64 planar images for uniformity (camera non-uniformity) using a high
count, e.g. 30 million count, cobalt-57 flood acquisition.
3. Check the images for patient motion and apply a motion correction algorithm if
indicated.
4. Indicate the superior and inferior limits of the heart so that computer time is not
expended in reconstructing tomograms outside of the heart.
5. Specify the filters to be used in the reconstruction process; the filters for rest and
stress reconstruction may be somewhat different because the injected doses are
quite different.
6. Specify the pixel thickness of the tomogram (usually 1 or 2 pixels).
7. The computer then constructs tomograms through the heart that are transaxial to
the long axis of the body using filtered backprojection. (These initial tomograms
will be oblique to the long and short axes of the left ventricle.)
8. In order to obtain images in standardized anatomic orientations, indicate the long
axis of the left ventricle; the initial tomograms are then reoriented to give
transverse, sagittal, and coronal tomograms of the left ventricle relative to the
long axis of the left ventricle. An automated program has been reported (29).
9. A 1x dataset will also be saved for use with Emory Toolbox.
OPTIONAL (Emory Cardiac Toolbox): The transaxial tomograms of the left ventricle
are then quantitatively analyzed and compared to normal ranges for perfusion at
stress and change from stress to redistribution, e.g. bullseye display and analysis
(7):
1. This analysis usually requires the technologist to indicate the center and outer
limits of the left ventricle in each transaxial tomogram.
The following are routinely saved and submitted for interpretation:
1. Tomograms of myocardial perfusion for both stress and rest in the transaxial,
sagittal, and coronal planes.
2. Gated images will also be included for evaluation of LV EF.
3. The EKG report with interpretation and exercise data.
Optional Protocols
Same day (Tc-labeled) protocol
1. Inject 8 mCi (296 MBq) of Tc-99m-Sestamibi or Tc-99m Tetrofosmin at rest.
2. After a 45-60 minute delay acquire a resting SPECT study (Use rest acquisition
protocol).
3. At 2-3 hours after the initial injection, stress the patient and inject 22 mCi (814
MBq) of Tc-99m-Sestamibi or Tc-99m Tetrofosmin.
(Patient is asked to drink a can of ensure after stress).
4. 30-45 minutes later acquire a stress Tc-99m-Sestamibi or Tc-99m Tetrofosmin
SPECT study.
Two Day (Tc-labeled) Protocol
1. Inject 22 mCi (814 MBq) of Tc-99m-Sestamibi or Tc-99m Tetrofosmin at rest.
2. After a 60 minute delay acquire a resting SPECT study (Use rest acquisition
protocol).
3. The following day, stress the patient and inject 22 mCi (814 MBq) of
Tc-99m-Sestamibi or Tc-99m Tetrofosmin. (Patient is asked to drink a can of
ensure after stress).
4. 45 minutes later acquire a stress Tc-99m-Sestamibi or Tc-99m Tetrofosmin
SPECT study (Use stress acquisition protocol).
Adenosine / Lexiscan Stress Test (35,36):
1. A physician experienced in managing cardiac emergencies must be immediately
available in case of complications.
2. Mechanism of action: Adenosine/LexiScan causes a direct increase in the
extravascular concentration of adenosine promoting coronary vasodilatation.
3. Contraindications:
a) severe asthma or bronchospasm.
b) unstable angina.
c) recent myocardial infarction, e.g. less than 48 hours.
d) sick sinus syndrome, and 2nd and 3rd degree AV block unless the patient has
a functioning cardiac pacemaker.
e) hypotension, e.g. resting systolic pressure < 80 mm Hg
4. The patient should be NPO for 4-6 hours prior to the study (adenosine may
cause nausea and vomiting).
5. The patient should be off all interfering medications such as Xanthine derivatives
(aminophylline) and caffeine containing products for 24 hours prior to the test (See
protocol 2.6 for complete list).
6. Drug administration:
Adenosine: Infuse 140ug/kg per minute intravenously for 6 minutes.
The clinic computer can be used to calculate the volume of adenosine to be
drawn based on patients weight. The computer will then provide the parameters
needed for the Graseby pump.
7. Blood pressure, Pulse oximetry and electrocardiogram will be monitored
throughout the study.
8. Timing of radiopharmaceutical injection:
Inject Tc-99m-Myoview 3 minutes after the start of the adenosine infusion.
9. Acquire images as described above beginning 45minutes following the end of
the drug infusion.
10. Treatment of severe side effects (39):
Termination of infusion. Aminophylline may also be given (50-75mg IV).
Lexiscan Stress Test:
1. A physician experienced in managing cardiac emergencies must be immediately
available in case of complications.
2. Mechanism of action: Lexiscan is a selective A2A adenosine receptor agonist.
3. Contraindications:
a) severe asthma or bronchospasm.
b) unstable angina.
c) recent myocardial infarction, e.g. less than 48 hours.
d) sick sinus syndrome, and 2nd and 3rd degree AV block unless the patient has
a functioning cardiac pacemaker.
e) hypotension, e.g. resting systolic pressure < 80 mm Hg
4. The patient should be NPO for 4-6 hours prior to the study (Lexiscan may cause
nausea and vomiting).
5. The patient should be off all interfering medications such as Xanthine derivatives
(aminophylline) and caffeine containing products for 24 hours prior to the test (See
protocol 2.6 for complete list).
6. Drug administration:
Lexiscan: Infuse 0.4mg / 5ml intravenously over 20-30 seconds.
7. Blood pressure, Pulse oximetry and electrocardiogram will be monitored
throughout the study.
8. Timing of radiopharmaceutical injection:
Inject Tc-99m-Myoview immediately after the Lexiscan injection.
9. Acquire images as described above beginning 45minutes following the end of
the drug injection.
10. Treatment of severe side effects (39): Aminophylline may be given (150mg IV).
Dobutamine Stress Test (40-42):
1. A physician experienced in managing cardiac emergencies must be immediately
available in case of complications.
2. Mechanism of action: Dobutamine increases myocardial contraction by direct
stimulation of the heart's beta-1 receptors.
3. Contraindications
a) severe aortic stenosis.
b) unstable angina.
c) recent myocardial infarction, e.g. less than 48 hours.
d) history of tachyarrhythmias.
e) hypertension, e.g. resting systolic pressure > 200 mm Hg.
f) poor left ventricular function
4. Withhold beta blockers for 48-72 hours.
5. The patient should be NPO for 4-6 hours prior to the study.
6. Monitor the blood pressure, pulse oximetry and electrocardiogram every minute
during administration of the drug and for 6 minutes afterwards.
7. Dobutamine administration and radiopharmaceutical injection:
a) infuse dobutamine at 5 µg/kg/min for 3 minutes followed by stepped
increases to 10, 20, 30, and 40 µg/kg/min for each successive 3 min.
b) inject radiopharmaceutical 1 minute following initiation of the maximum dose
over a 1 minute time period.
c) continue dobutamine infusion for 2 minutes after end of injection of
radiopharmaceutical.
d) Atropine 0.5-1.0 mg may be used to increase the HR.
8. Acquire stress images as described above beginning 45 minutes following
injection of the radiopharmaceutical.
9. Emergency medications: Atropine, Lopressor, Verapamil and Proventil must be
available to treat severe side effects.
Principle Radiation Emission Data - Tc-99m (51)
Physical half-life = 6.01 hours.
Radiation
Gamma-2
Mean % per disintegration
89.07
Mean energy (keV)
140.5
Dosimetry - Tc-99m-Tetrofosmin (at stress) (27)
Organ
Gallbladder wall
Upper large intestine
Urinary bladder wall
Lower large intestine
rads/25 mCi
3.08
1.88
1.45
1.43
mGy/925 MBq
30.8
18.8
14.5
14.3
Small intestine
Kidneys
Ovaries
Thyroid
Red marrow
Testes
Liver
1.13
0.98
0.73
0.40
0.38
0.33
0.30
11.3
9.8
7.3
4.0
3.8
3.3
3.0
Dosimetry - Tc-99m-Sestamibi (at stress) (19)
Organ
Upper large intestine
Lower large intestine
Gallbladder wall
Small intestine
Kidneys
Urinary bladder wall
Ovaries
Thyroid
Red marrow
Whole body
Liver
Testes
rads/25 mCi
3.88
2.68
2.41
2.32
1.39
1.29
1.02
0.68
0.60
0.38
0.36
0.24
mGy/925 MBq
38.8
26.8
24.1
23.2
13.9
12.9
10.2
6.8
6.0
3.8
3.6
2.4
Principle Radiation Emission Data - Tl-201
Physical half-life = 3.05 days.
Radiation
Gamma-4
Gamma-6
ce-K, gamma-8
K alpha-1 x-ray
K alpha-2 x-ray
K beta-1 x-ray
Mean % per disintegration
2.7
10.0
15.4
46.2
27.2
10.5
Mean energy (keV)
135.3
167.4
84.3
70.8
68.9
80.3
Dosimetry - Tl-201 as Thallous Chloride
Organ
rads/3 mCi
Kidneys
Thyroid
Liver
Heart wall
Testes
Ovaries
4.8
2.6
2.2
2.0
2.0
1.9
mGy/111
MBq
48.0
26.0
22.0
20.0
20.0
19.0
Gastrointestinal tract
stomach wall
small intestine
upper colon
lower colon
Total body
1.7
1.5
1.0
0.8
0.8
17.0
15.0
10.0
8.0
8.0
References
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
Borges-Neto S, Watson JE, Miller lMJ: Tc-99m sestamibi cardiac SPECT
imaging during coronary artery occlusion in humans: Comparison with
dipyridamole stress studies. Radiology 198:751-754, 1996.
Zanco P, Zampiero A, Favero A, et al: Myocardial technetium-99m
sestamibi single-photon emission tomography as a prognostic tool in
coronary artery disease: Multivariate analysis in a long-term prospective
study. Eur J Nucl Med 22:1023-1028, 1995.
Weissman IA, Dickinson CZ, Dworkin HJ, et al: Cost-effectiveness of
myocardial perfusion imaging with SPECT in the emergency department
evaluation of patients with unexplained chest pain. Radiology 199:353-357,
1996.
Gibson WS, Christian WF Pellikka PA, et al: Serial tomographic imaging
with technetium-99m-sestamibi for the assessment of infarct-related arterial
patency following reperfusion therapy. J Nucl Med 33:2080-2085, 1992.
Berman DS, Kiat G, Friedman JD, et al: Separate acquisition rest
thallium-201/stress technetium-99m sestamibi dual-isotope myocardial
perfusion single-photon emission computed tomography: A clinical validation
study. J Am Coll Cardiol 22:1455-64, 1993.
Sandler MP, Videlefsky S, Delbeke D, et al: Evaluation of myocardial
ischemia using a rest metabolism/stress perfusion protocol with fluorine-18
deoxyglucose/technetium-99m MIBI and dual-isotope
simultaneous-acquisition single-photon emission computed tomography. J
Am Coll Cardiol 26:870-878, 1995.
Marzullo P, Sambuceti G, Parodi O: The role of sestamibi scintigraphy in the
radioisotopic assessment of myocardial viability. J Nucl Med 33:1925-1930,
1992.
Van Train KF, Garcia EV, Maddahi J, et al: Multicenter trial validation for
quantitative analysis of same-day rest-stress technetium-99m-sestamibi
myocardial tomograms. J Nucl Med 35:609-618, 1994.
Cid E, Verani MS, Mahmarian JJ: Factors affecting the diagnostic accuracy
of quantitative single photon tomography combined with dobutamine stress:
Importance of anti-ischemic medications. J Nucl Med 37:58P, 1996.
Bottcher M, Czernin J, Sun KT, et al: Effect of caffeine on myocardial blood
flow at rest and during pharmacological vasodilation. J Nucl Med
36:2016-2021, 1995.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
Wackers FJT: Exercise myocardial perfusion imaging. J Nucl Med
35:726-729, 1994.
Jang S, Jaszczak RJ, Wang H, et al: Effect of breast size and shape on
defect contrast in myocardial perfusion SPECT imaging. J Nucl Med
37:216P, 1996.
DePuey EG, Melancon S, Masini D: Comparison of single-detector and
90°-angled two-detector cameras for technetium-99m-sestamibi cardiac
SPECT. J Nucl Med Technol 23:158-166, 1995.
Roy L, Van Train K, Bietendorf J, et al: An optimized protocol for detection
of coronary artery disease using technetium-99m-sestamibi. J Nucl Med
Tech 19:63-67, 1991.
Bisi G, Sciagra R, Santoro GM, et al: Comparison of tomographic and
planar imaging for the evaluation of thrombolytic therapy in acute myocardial
infarction using pre- and post-treatment myocardial scintigraphy with
technetium-99m sestamibi. Am Heart J 122:13-22, 1991.
Rigo P, Leclercq B, Itti R, et al: Technetium-99m-tetrofosmin myocardial
imaging: A comparison with thallium-201 and angiography. J Nucl Med
35:587-593, 1994.
Flamen P, Bossuyt A, Franken PR: Technetium-99m-tetrofosmin in
dipyridamole-stress myocardial SPECT imaging: Intraindividual comparison
with technetium-99m-sestamibi. J Nucl Med 36:2009-2015, 1995.
Jain D, Wackers FJT, Mattera J, et al: Biokinetics of
technetium-99m-tetrofosmin: Myocardial perfusion imaging agent:
Implications for a one-day imaging protocol. J Nucl Med 34:1254-1259,
1993.
Wackers FJTh, Berman DS, Maddahi J, et al: Technetium-99m hexakis
2-methoxyisobutyl isonitrile: Human biodistribution, dosimetry, safety, and
preliminary comparison to thallium-201 for myocardial perfusion imaging. J
Nucl Med 30:301-311, 1989.
Folks R, Garcia E: Supine versus prone Tc-99m sestamibi SPECT in
normals: Effect of correcting for table attenuation. J Nucl Med Tech 19:119,
1991.
Montz R, Perez-Castejon MJ, Jurado JA, et al: Technetium-99m tetrofosmin
rest/stress myocardial SPET with a same-day 2-hour protocol: Comparison
with coronary angiography. A Spanish-Portuguese multicentre clinical trial.
Eur J Nucl Med 23:639-647, 1996.
Heo J, Kegel J, Iskandrian AS, et al: Comparison of same-day protocols
using technetium-99m-sestamibi myocardial imaging. J Nucl Med
33:186-191, 1992.
Taillefer R, Primeau M, Costi P, et al: Technetium-99m-sestamibi
myocardial perfusion imaging in detection of coronary artery disease:
Comparison between initial (1-hour) and delayed (3-hour) postexercise
images. J Nucl Med 32:1961-1965, 1991.
Yong TK, Chambers J, Maisey MN, et al: Technetium-99m tetrofosmin
myocardial perfusion scan: Comparison of 1-day and 2-day protocols. Eur J
Nucl Med 23:320-325, 1996.
25.
26.
27.
28.
29.
30.
31.
32.
33.
34.
35.
36.
37.
38.
39.
40.
Viggiano J, McClellan JM, Alavi A, et al: Improved Tc-99m sestamibi
uniformity with attenuation correction in patients with low likelihood for
coronary artery disease. J Nucl Med 37:148P, 1996.
Matsunari I, Stollfuss J, Schneider-Eicke J, et al: Effect of attenuation
correction on Tc-99m-tetrofosmin SPECT in the detection of viable
myocardial compared to FDG-PET. J Nucl Med 37:180P, 1996.
Tc-99m-Tetrofosmin. Package insert. Medi+Physics, Amersham
Healthcare, Arlington Heights, IL, February, 1996.
Folks R, Van Train K, Wong C, et al: Evaluation of Tc-MIBI SPECT
acquisition parameters: Circular vs. elliptical and 180° vs. 360° orbits. J Nucl
Med 30:795, 1989.
Germano G, Kavanagh PB, Su HT, et al: Automatic reorientation of
three-dimensional, transaxial myocardial perfusion SPECT images. J Nucl
Med 36:1107-1114, 1995.
Klingensmith WC, Stern D, Spitzer VM: Single-composite display of
information from tomographic myocardial perfusion studies. Radiology
141:242-245, 1981.
Berman DS, Kiat HS, Van Train KF, et al: Myocardial perfusion imaging with
technetium-99m-sestamibi: Comparative analysis of available imaging
protocols. J Nucl Med 35:681-688, 1994.
Kiat H, Germano G, Friedman J, et al: Comparative feasibility of separate or
simultaneous rest thallium-201/stress technetium-99m-sestamibi
dual-isotope myocardial perfusion SPECT. J Nucl Med 35:542-548, 1994.
Mahmood S, Gunning M, Bomanji JB, et al: Combined rest
thallium-201/stress technetium-99m-tetrofosmin SPECT: Feasibility and
diagnostic accuracy of a 90-minute protocol. J Nucl Med 36:932-935, 1995.
Kiat H, Biasio Y, Wong FP, et al: Frequency of reversible resting
hypoperfusion in patients undergoing rest Tl-201/stress Tc-sestamibi
separate acquisition dual isotope myocardial perfusion SPECT. J Am Coll
Cardiol 21:222A, 1993.
Iskandrian AS: Adenosine myocardial perfusion imaging. J Nucl Med
35:734-736, 1994.
Cerqueira MD, Verani MS, Schwaiger M, et al: Safety profile of adenosine
stress perfusion imaging: Results from the adenoscan multicenter trial
registry. J Am Coll Cardiol 23:384-389, 1994.
Leppo JA: Dipyridamole myocardial perfusion imaging. J Nucl Med
35:730-733, 1994.
Blust JS, Boyce TM, Moore WH: Pharmacologic cardiac intervention:
Comparison of adenosine, dipyridamole, and dobutamine. J Nucl Med
Technol 20:53-59, 1992.
DePuey EG, Rozanski A: Pharmacological and other nonexercise
alternatives to exercise testing to evaluate myocardial perfusion and left
ventricular function with radionuclides. Sem Nucl Med 21:92-102, 1991.
Verani MS: Dobutamine myocardial perfusion imaging. J Nucl Med
35:737-739, 1994.
41.
42.
43.
44.
45.
46.
47.
48.
49.
50.
51.
Marwick T, Willemart B, D'Hondt AM, et al: Selection of the optimal
nonexercise stress for the evaluation of ischemic regional myocardial
dysfunction and Tc-99m-MIBI single photon emission computed
tomography. Circulation 87:345-354, 1993.
Imai K, Arake Y, Saito S, et al: Usefulness and safety of high dose
dobutamine stress thallium scintigraphy. J Nucl Med 33:951-952, 1992.
Gonzalez P, Massardo T, Munoz A, et al: Is the addition of ECG gating to
technetium-99m sestamibi SPECT of value in the assessment of myocardial
viability? An evaluation based on two-dimensional echocardiography
following revascularization. Eur J Nucl Med 23:1315-1322, 1996.
DePuey EG, Rozanski A: Using gated technetium-99m-sestamibi SPECT to
characterize fixed myocardial defects as infarct or artifact. J Nucl Med
36:952-955, 1995.
Chua T, Kiat H, Germano G, et al: Gated technetium-99m sestamibi for
simultaneous assessment of stress myocardial perfusion, postexercise
regional ventricular function and myocardial viability: Correlation with
echocardiography and rest thallium-201 scintigraphy. J Am Coll Cardiol
23:1107-1114, 1994.
Germano G, Kiat H, Kavanagh PB, et al: Automatic quantification of ejection
fraction from gated myocardial perfusion SPECT. J Nucl Med 36:2138-2147,
1995.
Maurea S, Cuocolo A, Soricelli A, et al: Enhanced detection of viable
myocardium by technetium-99m-MIBI imaging after nitrate administration in
chronic coronary artery disease. J Nucl Med 36:1945-1952, 1995.
Bisi G, Sciagra R, Santoro GM, et al: Technetium-99m-sestamibi imaging
with nitrate infusion to detect viable hibernating myocardium and predict
postrevascularization recovery. J Nucl Med 36:1994-2000, 1995.
Sciagra R, Bisi G, Santoro GM, et al: Evaluation of coronary artery disease
using technetium-99m-sestamibi first-pass and perfusion imaging with
dipyridamole infusion. J Nucl Med 35:1254-1264, 1994.
Giubbini R, Campini R, Milan E, et al: Lung uptake of 99mTc-sestamibi:
Correlation with LV ejection fraction. J Nucl Med 33:846, 1992.
43-Tc-99m: In MIRD: Radionuclide Data and Decay Schemes, DA Weber,
KF Eckerman, AT Dillman, JC Ryman, eds, Society of Nuclear Medicine,
New York, 1989, pp 178-179.
Normal Findings
>
Lette J, Caron M. Cerino M, et al: Normal qualitative and quantitative
Tc-99m sestamibi SPECT: Spectrum of intramyocardial distribution during
exercise and at rest. Clin Nucl Med 19:336-343, 1994.
>
Henze E, Lietzenmayer R, Calusen M, et al: Reproducibility of relative blood
flow quantification with Tc-99m sestamibi-SPECT. J Nucl Med 31:821,
1990.
>
He ZX, Verani MS, Mahmarian JJ: Separate normal patient data banks for
thallium-201 and technetium-99m sestamibi are unnecessary when
>
quantifying perfusion defects with single photon tomography. J Nucl Med
37:178P, 1996.
Naruse H, Daher E, Sinusas A, et al: Quantitative comparison of planar and
SPECT normal data files of thallium-201, technetium-99m-sestamibi,
technetium-99m-tetrofosmin and technetium-99m-furifosmin. J Nucl Med
37:1783-1788, 1996.
Note: This procedure complies with the Society of Nuclear Medicine procedure
guideline for myocardial perfusion approved June 15, 2002
JSM
PROTOCOL\2-7-4
Rev. 6/1/2016