Survey							
                            
		                
		                * Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Renatron® II Service Maintenance Seminar Class Schedule DAY ONE 8:30 AM- Seminar start: Introduction 10:00 AM- Break 12:00 PM- Lunch 2:30 PM - Break 5:00 PM- End of first day Class Schedule DAY TWO 8:30 AM- Seminar start 10:00 AM- Break 12:00 PM- Lunch 2:30 PM - Break 3:30 PM- End of Seminar Cautions  The Renatron® 100 is intended only for use with Renalin® 100 Cold Sterilant  Used only for Hemodialyzers that have been labeled multiple use.  Never run the machine without the uptake tube assembly  All hoses and lines for the Renatron® are cut to length and should not be altered. Indications The Renatron® Dialyzer Reprocessing System is intended for in vitro, rinsing, cleaning testing, and delivery of proportioned Renalin® solution to hollow fiber dialyzers that have been validated to be reprocessed with the Renatron® II System and Renalin® sterilant. It is also indicated for record keeping of these functions. Specifications Electrical Leakage Current  100 micro amps  Selectable voltage 100, 120, 230, 240 VAC  The supply line voltage must not fluctuate more than 10% of the nominal voltage.  Fuses 100-120vac, one 4A  Fuses 220-240vac, two T2A Environmental Limits  Operation Ambient Temperature Range 41-104° F (5-40° C) Water Requirements  Water must be filtered to particle size no larger than 1 micron.  Inlet Water Must Meet or Exceed AAMI Guidelines Water Temperature  Must be 59 -75°F (15-24° C)  A maximum temperature of 86° F is allowable if the dialyzer is stored immediately after reprocessing in an area where the temperature is between 59-75° F. (Must have letter) Feed Line Flow  1.725 liters (.46 gallons) per minute average  Maximum water pressure is 55 PSI static  Minimum water pressure is 20 PSI dynamic read in step 4.  Optimum pressure of 30-35 PSI in Step 4. Peak Flow Rate  6 liters (1.6 gallons) per minute  15-35 seconds in duration  Occurs at approximately 3 minute intervals  3 times per cycle Drain Requirements  Must be able to accept 6 liters (1.6 Gallons) per minute per Renatron®. Materials of Construction  The drain should be constructed of PVC/CPVC or other Non-corrosive Material Renatron II Station Components Renatron® II Station Components  Renatron® II Dialyzer Reprocessing Station  Has a hydraulic side and electronic side.  1 - 6 ft water supply hose with back flow preventor.  1 - 6 ft drain hose.  1 - 6 ft Renalin® hose.  1 - 6 ft RS-232 communication cable Renatron® II Station Components  Pressure gauge block/regulator  70 cc cal cell  Drip tray-single or dual  Renalin®100 Uptake tube assembly Renatron® II Station Components Computer Top of the line computer system Windows XP Monitor  VGA Monitor- CRT or Flat Screen Renatron® II Station Components  Renalog RM Data Management System  Dedicated label printer  Dedicated Report Printer  Interface Module  Multiplexing module that can connect up to 6 Renatrons®.  Renalog RM – 12 Renatrons per computer Overview of Dialyzer Reprocessing End of Treatment After treatment is completed, the dialyzer is brought to the reuse room. The time interval between the end of the treatment and reprocessing the dialyzer should be kept within the time limits prescribed by the physician  The suggested maximum time interval is two hours Dialyser Cleaning Many clinics will clean or rinse a dialyzer before reprocessing. Blood side flush RUF  Renaclear® Reprocessing  Attaching a dialyzer Connect: Venous blood side Venous dialysate side Arterial dialysate side Arterial blood side End of Reprocessing Cycle  Make sure reset is pressed before the dialyzer is disconnected.  Disconnect the dialyzer in the reverse order that it was connected. Cap each port as soon as it is disconnected from the Renatron®.  Aseptic technique must be followed while removing the reprocessed dialyzer from the Renatron® station to prevent contamination of the dialyzer blood pathway Dialyzer Storage  Storage Area  Well ventilated  Temp. range 59-75°F  Out of direct Sun light  Dialyzer Storage  Minimum storage time is 11 hours  Maximum storage time ?  Temperature differences may cause pressure buildup in dialyzer Dialyzer Testing before Treatment The dialyzer must be tested for Renalin® 100 content after proper storage time- Perrassay Test Strip The dialyzer is rinsed on the dialysis machine and a Residual Test Strip is used to verify that the Renalin® 100 content has been rinsed to acceptable levels Recirculate until treatment is started to prevent rebound ® RENATRON II SYSTEM OPERATION Power Up Sequence  Power on sequence      Printers Monitor Computer Interface module Renatron® II  All at once if using the power strip Renalog® RM Start Up  Log in  User ID  Password  Mode  Change Time- Activation key needed  Start Up Screens        Renalin Verifications Service Rinse Sanitize Renatrons Calibration Water Verifications- 20 psi to 55 psi Water Cultures Dialyzer Verifications- Precleaned, Expired, and not Processed Renatron® II 100 Startup Renalog® RM Startup Renalin® 100 is connected The Renatron® II 100 is sanitized The Renatron® II 100 is calibrated Static and dynamic water pressure readings Renatron® II 100 Startup Calibration verification Scan barcodes- Authentication label, Renatron, and Dialyzer The Renatron® II 100 runs to step 57 Calibration information is stored in the computer and a calibration report can be generated. Modes of Operation Program Mode  Low flux dialyzers (00) kuf <8. Program time = 8 minutes.  High efficiency dialyzers (CH) kuf 8-15. Program time= 10 minutes.  High flux dialyzers (HF) kuf 15+. Program time = 10 minutes. Automatic Reprocessing  Scan Authentication label, dialyzer barcode label, and Renatron® barcode.  Minimum Volume  Program Mode  When reset is pressed in step 57, labels are printed and the database is updated. Semi-Automatic  Click on F6 to display Patient list  Select patient’s dialyzer  Select type of reprocessing  Preclean  Reprocess without a use  Reprocess after use  Sets the Renatron® II 100 to the proper settings for that dialyzer Manual Reprocessing With Computer Process dialyzer on Renatron® II 100 Set Hold to Set volume to dialyzer minimum volume setting  Press mute and reset button to select program setting  At end of cycle, enter information into computer Preprocess Reprocess without use Reprocess after use Renatron number Container number Current Volume Did the dialyzer fail? Manual Reprocessing with Renatron® II 100 only Process dialyzer on Renatron® II 100 Set Hold to Set volume to dialyzer minimum volume  Press mute and reset button to select program setting  At end of cycle, hand write label Preclean Mode Pre-clean Mode Pre-cleaning a dialyzer will help breakdown residual blood in the dialyzer. Pre-clean program works for all types of dialyzers. Program time  4 minutes. Recommended contact time for pre-cleaned dialyzers is 10 minutes to two hours. Caution  Pressure buildup may occur in dialyzers left on the Renatron®  Take care when removing the dialyzer  Pressure buildup may cause fluid to spray Barcode labels User Authentication Label Information Label Dialyzer Barcode Label Pre-Cleaned Label Calibration Cell Label Failed Label Sanitize Program Daily Sanitize  The Renatron needs to be sanitized once a day.  The machine must be connected in sanitize configuration. “SEL” must be in the top display and any of the programs “00”, “CH” or “HF” in the bottom display before “Start Sanitize” is pressed  The machine will display “sanitize complete” and a 3 second tone will sound at the end of step 83.  Turn off the Renatron and the water supply to the Renatron. Minimum Sanitize Time The Renatron must sit for a minimum of 6 hours before it can be used. Maximum Sanitize Time The Renatron® can sit for a maximum of seven days before it needs to be resanitized. 10 Minute Sanitize Cycle  Press start process at the end of step 83, after a 10 minute hold in step 84, the machine will run to step 97, and signal sanitize complete. All fluid will be removed from the machine.  This cycle is to be used after a 409® procedure, when maintenance is performed on the machine, or when the machine is being returned for maintenance. External Disinfection (Daily)  1 % Renalin® solution is used to wipe down external surfaces of the Renatron® station  Actril® undiluted is also acceptable Formula 409 Cleaning 409® Cleaning  Institutional Formula 409®  Institutional Formula 409® Cleaner Degreaser only  Must say NQF  Do not use any other solutions  Internal fluid pathways must be cleaned periodically ( typically once every two weeks )  This should be performed on Saturday a long dwell time is preferred 409® Cleaning  Institutional Formula 409® is free when Renalin® 100 is ordered in proper quantities Dialyzer Reprocessing Cycle Cleaning Cycle Step 8 to 24  Chemical Draw- Step 9  Flush Blood Circuit Arterial to Venous-Step 14  Reverse Ultrafiltration- Steps 16 and 18  Flush Dialysate Compartment- Step 20  Flush Blood Circuit (Arterial to Venous)-Step 21  Flush Blood Circuit (Venous to Arterial)-Step 25 Test- Step 25 to 40 Blood Compartment Volume Measurement Step 25 to 35  Fill Blood Compartment from venous to arterial- Step 25  Sweep Blood Compartment- Steps 29, 31 and 33 Measure Mass- Step 35  Repeatability ± 5% Test- Step 25 to 40 Pressure Leak Test- Steps 36 to 39 Draw vacuum- Steps 36 and 38 Measure bleed off rate- Step 39 Dialysate compartment is at -250 mmHg or less  Blood compartment atmospheric pressure  Leakage rate of .83mmHg per second fails ± 10 % for 00, CH mode  Leakage rate of 1.25mmHg per second fails ± 10% for HF program Disinfect- Steps 41 to 57  Draw Renalin® - Step 45 Verify Renalin® Draw- Steps 46 and 47  Proportion Renalin® solution- Step 48  Flush Blood Compartment- Step 50 Flush Dialysate Compartment- Step 51  Draw Renalin® Solution to Dialysate Compartment- Step 53  Draw Renalin® Solution to Blood Compartment- Step 54 Interface Module/Scanner Operation Interface Module  Yellow power indicator light.  Green On Line light- Indicates that the computer is in communication mode. This light will only be on in the Automatic Reprocessing, Automatic Calibrations, or Renalin Connection screens.  On/Off switch and fuse at power entry module.  Older version of scanner will beep twice when scanning.  The newer version of the scanner will beep once when scanning. Scanner  For best operation of Scan Team 3000 and 3400 models, hold the scanner so it is touching the barcode and press the trigger. Continue to press the trigger while moving the scanner away form the barcode until the scanner beeps.  Scan Team 3800 models, hold the scanner 3-9 inches from the barcode.  Make sure scanner lens is clean.  Clean the scanner with a nonabrasive cloth. Do not use a cleaner that contains alcohol. Renalog® RM Label Printer Label Printer Should be connected to the computer with a Parallel Printer Cable. If printing is not aligned on label, perform the Calibration Procedure. Turn power switch to Off Hold Green Button down Turn Power switch to On Let go of Green Button when the red light flashes Press Green Button to get “out of dump” label Red light means that:  Cover is not completely closed  Ribbon is not advancing  Labels are not advancing Renatron® II 100 Controls and Indicators Description of Controls  On  Off  Hold to Set Set Knob  Start Process  Start Sanitize  Mute  Reset  Display Indicators  Volume Display  Program Step  Clean  Test  Disinfect  Process Complete  Sanitize Complete Error Messages  Volume Fail- Step 35  Pressure Fail- Steps 36, 38, or 39  Add Chemical- Steps 9, 45, 67, or 77  Tank Volume- Many steps throughout the program  Alarm  Interlock Safety Features  Tracks Preclean Dialyzers  Tracks Maximum storage time  Tracks Dialyzers that have not been processed- Preprocessed and Dry Packs  Information stored on hard drive  Precautions for similar name patients  Patient data and program settings are password protected Renatron® Storage  Long term storage  run machine through step 97 to ensure fluid pathways are drained  Box in Renatron® box  Keep in low traffic area Technical Service  Telephone Support 24 hours a day 7 days a week  1-800-328-3324 Press “2” for “Medical Equipment Emergency” to get a live person  Service  Loaners available upon request  Board exchanges for $200  RMA’s must accompany all returns Warranty / Promotions  Warranty  1 year warranty from date of install  90 days for service warranty  90 days for parts orders  Promotions  PM contracts  Service contracts Renalin® 100 Warnings     Corrosive. Use universal precautions. Can cause eye damage and skin irritation. When handling Renalin®100, wear eye protection, rubber gloves and protective clothing.  Do not mix Renalin®100 with alkaline substances such as bleach. Indications Renalin® 100 Cold Sterilant is indicated for the in vitro cleaning and sterilizing of hollow fiber dialyzers with the RS8300 and RS8330 models of the Renatron® Dialyzer Reprocessing System labeled for use with Renalin®100. Renalin®100 Cold Sterilant also may be used for disinfecting dialysis equipment (e.g. kidney machines) and supplies (e.g. port caps). Contraindications Renalin® 100 Cold Sterilant is not designed, sold, or intended for use except as indicated. Patients with known hypersensitivity to hydrogen peroxide and/or peroxyacetic acid should not be treated using dialyzers reprocessed with Renalin® 100 Cold Sterilant. Water Requirements  AAMI quality water filtered to particle size no larger than 1 micron  Water temperature should be between 59-75° F  Water Temperature may be up to 86° F if the dialyzers are stored at the proper temperature Renalin®100 Storage  Renalin®100 Storage Upright Never in direct sunlight At temperatures between 32-75° F Do not freeze  Never tamper with the cap  Never replace the vented cap with a nonvented cap Renalin® can be removed from the shipping carton if it is not stored in direct sunlight Renalin® 100  Renalin® 100 replaces Formaldehyde Bleach and/or hydrogen peroxide  Active ingredients Hydrogen peroxide Peroxyacetic acid Acetic acid  Renalin® 100 degrades quickly and will not harm the environment. It breaks down into ; oxygen, water, and acetic acid. Renalin 100®  Renalin® 100 is supplied in three different volumes: Quart bottles 2.5 gallon bottles that contain 2 liters of Renalin® concentrate 1 gallon bottles containing 3 liters of Renalin® 100 Renalin® Nomenclature Renalin® 100 Cold Sterilant is a 100% Renalin® Solution. Proportioned Renalin® - A 3.25% or 3.5% Renalin® solution that is used to sterilize the reprocessed dialyzer. 1% Renalin® solution- A solution made from 1 part Renalin® concentrate and 99 parts AAMI quality water. Renalin® Nomenclature 4% Renalin® solution is used to sanitize the Renatron®  2% Renalin® solution is used during the preclean program Auto Dilutor The On Demand™ Auto Dilution System proportions Renalin® 100 Cold Sterilant and dispenses a continuous supply of 1% Renalin solution. Shelf Life Renalin® 100 has a shelf life of 1 year from date of manufacture when properly stored. Following dilution, Renalin® 100 exhibits a gradual loss of potency at a rate such that 50% of the active ingredients remain after a seven day period Shelf Life When diluting Renalin® Cold Sterilant to a 1% solution (e.g. port cap disinfection) fresh Renalin® Cold Sterilant should be used each time. Do not allow 1% Renalin® solution to sit for longer than 24 hours. Test Strips  Perassay Test Strip Simple go/no go test Indicates a 1% Renalin® solution Green stick for identification Do not need to be refrigerated Test Strips  Residual Test Strip  Color chart for interpreting test results Do not need to be refrigerated Verifies the proportioned Renalin® solution has been properly rinsed from the dialyzer prior to clinical use Renalin® 100 Presence Test (Perassay)  Remove 1 ml of solution from the dialysate port of the dialyzer  Immerse the entire reaction pad of the Perassay strip in the solution  Shake off excess solution  Read the reaction pad after 2-5 seconds. The test strip should turn a blue/gray or blue/black color. This will indicate 500 PPM of peracetic acid 1% Renalin® solution. Renalin® 100 Residual Test Strips  Turn off the blood pump and clamp the arterial and venous blood lines adjacent to the access ends. Separate the blood lines where the access ends are joined and unclamp the venous blood line.  Allow a few drops of saline to drip from the venous line onto the residual test strip.  After 5 seconds, shake the excess liquid from the reaction zone and compare with the color scale on the container.  Per AAMI standards, a reading of less than 3 PPM is acceptable Handling and Storage of Reprocessed Dialyzers  If a dialyzer is foaming after being reprocessed it should be reprocessed again without a use  Store a reprocessed dialyzer in a cool, well ventilated area that is shielded from any direct sunlight. Recommended temperature range is 59-75° F  A reprocessed dialyzer should not be used until a minimum of 11 hours has elapsed  After storage and before rinsing, both headers of the dialyzers must be at least 2/3 full Renalin® Spill  Flush spill with large quantities of water and also consult the handout “What to do if Renalin® spills”  After hours, contact Chemtrec for emergency spill information at 800-424-9300 (202-483-7616 in the Washington D.C. area) Questions