Survey
* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project
CARDIOVASCULAR INSTRUMENTATION Techniques and Significance of Threshold Measurement for Cardiac pacing* Relationship t o Output Circuit of Cardiac Pacemakers S . Serge Barold, M .B., F.C.C.P., '* and lames A. Winner, B.S.E.E. 1. Because it determines the Yeserven of the pacemaker's output circuit, threshold measurement remains the single most important factor that will ultimately ensure successful pacing. In this review, we describe the various techniques and significance of threshold measnrement 'in relation to the property of pacemaker output circuits. There are two basic types of pacing circuits: (1) those with coastant current, as in most external (temporary) pulse generators, and (2) those with constant voltage, as in many implantable pulse generators. Current-limited pulse generators have features of both constant-current and constant-voltage circuitry. The current threshold is helpful in determining the integrity of the electrode-tissue interface and reflectsthe density of the current, which is the prime factor responsible for successful stirnulation. Voltage thresholds are useful for information on lead position and integrity, especially when voltage and current are measured simultaneously. Impedance (calculated from voltage and current during stimulation) can be helpful in the diagnosis of lead fractures, insulation breaks, and position problems. Threshold and impedance are entirely unrelated factors, each providing specific and different information about a pacing 'system. Threshold may also be measured in t e m s of charge and energy and in relation to the width of the pulse (at a constant impulse amplitude). The concept of safety margins is important when measuring long-term threshold at the time of replacement of a pulse generator. We have analyzed this problem and have attempted to make acceptable recommendations in the absence of clear information in the literature about this subject. roper evaluation of a pacemaker's function rep quires some knowledge of the basic design characteristics of the electronic circuits that deliver The threshold for cardiac pacing may be defined as the smallest amount of electrical energy that produces consistent capture outside the refractory period of the heart3 This threshold may be measured by gradually decreasing the output of the generator to the point where capture no longer occurs consistently or by increasing the output from a low value to the point where consistent capture occurs. The small difference in threshold obtained by these two techniques is almost always clinically unimportant. The threshold may vary according to the site of stimulation, the type of electrode and its age, the duration of the stimulus, the polarity, and the wave shape. It may also be influenced by administration of drugs, changes in electrolyte levels, the state of the myocardium, and the type of measuring equipment u t i l i ~ e d . ~ the pacemaker's impulse. Unfortunately, confusing terminology has made an already dBcult technical Because threshold measubject even more -cult. surement actually determines "reserve" of the output circuit, it remains the single most important measurement that will ultimately ensure successful pacing, and it is essential for the determination of optimal electrode position, lead integrity, and the design and use of low-output pulse generators to increase longevity.'" In this review, we dezcribe the various techniques and significance of threshold measurement in relation to the properties of pacemaker output circuits. This information is essential for trouble-shooting problems &th pacemakers. 'From the Division of Cardiology, Department of Medicine, the Genesee Hospital and the University of Rochester School of Medicine and Dentistry, Rochester, NY. "Chief of Cardiology, the Genesee Hospital, and Associate Professor of Medicine. +Engineer, Medtronic, Inc. Rkptint requests: Dr. Barold, 224 A l e d r Street, Rochester, New York 14607 Lead impedance in this discussion refers to impedance when the electrode stimulates the heart, in contrast to source impedance, which is related to the'sensing property of the electrode. The two are not the' same, and the complex subject of source 760 BAROLD, WINNER Downloaded From: http://publications.chestnet.org/pdfaccess.ashx?url=/data/journals/chest/20987/ on 05/03/2017 CHEST, 70: 6, DECEMBER, 1976 impedance is beyond the scope of this discussion. The in uiuo impedance (resistance) imposed upon a pulse generator consists of the sum of ( 1) the resistance of the pacing lead (equivalent to the in uitro value), and ( 2 ) the impedance of the body tissue between the electrodes, which is nonlinear and rises during the pulse. The impedance of the body tissue can be divided into the following two parts: ( 1 ) the pure tissue resistance and ( 2 ) the nonlinear polarization effect, caused by oppositely charged particles that accumulate at the electrodeheart interface during the pacemaker pulse. This alignment of charges acts like a capacitor. The effect is essentially zero during the extremely rapid initial part of the pacemaker's stimulus and increases to the trailing edge, after which most of the accumulated charges dissipate by diffusion, usually within 200 msec. Polarization increases as the surface area of the electrode decreases. Polarization also depends on the mode of stimulation (unipolar or bipolar), the maturity of the lead, the tissue chemistry, the material of the electrode, and the amplitude and duration of the output-pulse ~idth.~-~ VOLTAGE @ 10mA MEDTRONIC 5880A LEADING EDGE ONLY RESISTANCE The output of a temporary constant-current pulse generator is independent of the resistance (impedance) to the limits of the power supply, which is generally about 15 volts ( Fig 1) . Consequently, if the resistance of the lead increases, the voltage automatically increases (within the range of power supply) to keep the current constant according to Ohm's law (V = IR, where V is voltage, I is current, and R is resistance or impedance). The smaller the required current, the greater the range of resistances over which current remains truly constant. Figure 1 shows that when the dial for current output is set at 1 ma, one commercially available external pulse generator ( Medtronic 588014) is almost a perfect source of constant current over a wide range of resistances. Because of design, difficulties in maintaining constant current may arise when the load exceeds 800 to 1,000 ohms (Fig 1 ) . Thus, the curves for current (Fig 1) at various dial settings remain flat in the anticipated clinical range (300 to 1,000 ohms) but begin to bend down (ie, amplitude decreases) as the limits of the power supply are approached. The absolute limits of the power supply are approximately 15 v for this particular unit (Medtronic 5880A). Constant-current circuitry would thus appear to be advantageous during external ( temporary ) pacing because of its unique compensating feature allowing constant current in the face of a possible CHEST, 70: 6, DECEMBER, 1976 / VOLTAGE / - FIGURE1. Current and voltage output curves (leading edge only) of constant-current pulse generator ( Medtronic 5880A ) at various resistive loads. Corresponding current and voltage curves are shown for current settings of 1, 2, 5 and 10 ma ( K = 1,000 ohms). Resistance is shown in log scale. In this and subsequent figures, measurements were made on bench, and standard technique of voltage drop across a 10-ohm resistor was used to measure current on oscilloscope. Pulse retains its square wave form under almost all circumstances. increase in the resistance of the lead; however, this is done at the cost of an increased drain on current and a consequent reduced longevity of the power sources. Nevertheless, longevity is of little practical importance during temporary pacing, as power sources may be readily replaced. With constant-voltage circuitry, the output of voltage will remain constant (within limits), regardless of the load imposed on the pulse generator. Only special pacemakers, used mostly for research, have true constant-voltage output pulses; however, most implantable pulse generators are termed constant voltagea because the leading edge of the voltage pulse remains constant, regardless of the load (Fig 2 ) . Historically, the outputs of pulse generators have been given in terms of current (milliamps), and it is commonly stated that the pulse generator has an output of 10 ma. This 10 ma is specified with an arbitrary load of 500 ohms only. The cardiac load and the resistance of the lead THRESHOLD MEASUREMENT FOR CARDIAC PACING 761 Downloaded From: http://publications.chestnet.org/pdfaccess.ashx?url=/data/journals/chest/20987/ on 05/03/2017 MEDTRONlC XYTRON SOH) EDGE VOLTAGE TRAILING EDGE won CURRENT (I( IU 101 RESISTANCE --) FIGURE2. Voltage and current output curves of a constantvoltage implantable pulse generator ( Medtronic Xytron 5950) at various resistive loads. Both leading and trailing edges are shown. Note that only leading edge of voltage remains constant with all resistances over 200 ohms, but current falls as resistance increases. Resistance is shown in log scale. usually differ from this arbitrary value, and the current output will, therefore, adjust itself according to Ohm's law and decrease in order to maintain constant voltage as the resistance increases. When applied to the leading edge, the concept of constant-voltage output should not be confused with changes in voltage that occur during the pacemaker pulse after the constant leading edge. The voltage of the pulse wave form droops from leading to trailing edge as a function of the resistance and the size of the pulse generator's output capacitor (Fig 2). Any constant-voltage course may cease to remain constant when the resistance is very low be cause the battery then becomes unable to supply all of the current required to maintain a steady voltage (Fig 2), but such low resistances are not seen clinically. The threshold has historically been evaluated with temporary pulse generators with constantcurrent characteristics, yet many implanted pulse generators are of the constant-voltage type. At the time of replacement of a pulse generator, the threshold measured with an external constant-current device could be 1 ma, and a decision could, therefore, be made to simply implant a new pulse generator based only on this information that seemingly indicates a good threshold. If the impedance is 7,000 ohms because of a partial fracture, the fact that the compensation feature has increased the voltage output of the constant-current external pulse generator to 7.0 v would remain unknown because the only visible control is the current setting ( V/I = R; 7 v/ 1 ma = 7,000 ohms). This 7-v stimulating voltage is higher than that available from conventional implantable (constant-voltage) pulse generators. Therefore, the implantable pulse generator would not capture the heart when connected to the lead. This point becomes clearer by considering yet another example. Assume a current threshold of 3 ma and a lead impedance of 5,000 ohms. A current setting of 5 ma on one constantcurrent pulse generator (Medtronic 5880A) will, therefore, deliver the necessary3 ma to capture the heart, so that pacing will occur at a seemingly acceptable long-term threshold of 5 ma according to the settings on the pulse generator (Fig 1 ) ; however, pacing will not occur when a constant-voltage (5-v) generator is connected to the lead, as this generator can only deliver 1 ma, which is below the required current threshold at this particular lead impedance ( I = V/R = 515,000 = 1 ma). These two examples illustrate that with respect to the delivery of current, the constant-voltage pulse generator simply cannot compensate for a higher Iead impedance, whereas adequate compensation may occur with a constant-current device.O The true incidence of this type of problem is unknown, but these situations may be easily avoided by testing thresholds with a constant-voltage device that matches or closely matches the output characteristics of the implantable unit.O The constantvoltage threshold should be measured routinely whenever a pulse generator is replaced or a problem with pacing is suspected. Interestingly, studies by Fontaine and associates4*10have found that voltage thresholds are more consistent and reproducible than constant-current values for the routine measurement of threshold. This design aims at increasing the life of an implanted pacemaker by controlling the maximal current flow in the electrode circuit. "Current-limited" circuitry has features of both constant-current and constant-voltage circuitry. The "current-limited" source appears like constant current with low re- 762 BAROLD, WINNER Downloaded From: http://publications.chestnet.org/pdfaccess.ashx?url=/data/journals/chest/20987/ on 05/03/2017 CHEST, 70: 6, DECEMBER, 1976 Energy Threshold CORDIS OMNl STANICOR HIGH CURRENT SETTING The energy threshold is usually expressed in microjoules and gives the total picture of energy requirement for depolarization; it is an indirectmeasurement calculated from the product of voltage, current, and pulse duration at thre~hold.~l-'~ Due to the capacitive'effects of the electrode-heart interface, the voltage will lag behind the current in time. Theoretically,-this'phasedifference enters into the energy calculation, but it is neglected in practice because the difference is less than 2 percent.12 Because energy = VIT = 12RT (where T is pulse width), a fourfold increase in curkent measured in milliamperes would be equivalent to a 1Bfold increase expressed in microjoules. Strength-DurationCurves (Including Threshold Related to Pulse Width) lmn 11. ~ O U RESISTANCE -+ FIGURE 3. Voltage and current output curves for current-lirnited pulse generator (Cordis OmniStanicor at high current setting). Both leading and trailing edges are shown. Resistance is shown in log scale. sistances, in that the voltage increases as the resistance increases until the voltage supply limit is reached (Fig 3). When the limit is reached, the pulses are more char~cteristicof constant-voltage circuitry since the voltage of the leading edge remains constant (Fig 3). The shape of the impulse resembles that of a constant-voltage pacemaker only when the resistance is high (Fig 3). OTHER WAYS OF MEASURING THRESHOLDS Charge Thresholds Charge thresholds are seldom calculated, but this information could be helpful in the design of lowoutput pulse generators with greater longevity. Charge threshold is usually expressed in microcoulombs and is the product of the current threshold in milliamperes and the duration of the pulse in milliseconds. To calculate longevity, the pulse generator's output charge (and not the charge threshold) must be compared to the pulse generator's battery charge (cell capacity is rated in amphours; the standard mercury-zinc cell wallory RMl] has a capacity of 1 amp-hours). CHEST, 70: 6, DECEMBER, 1976 When the circuitry of the external device measuring threshold matches that of the implanted pulse generator, a useful strength-duration curve may be easily plotted. The voltage or current is plotted on the vertical axis and pulse duration on the horizontal axis12 (Fig 4). The output of the pulse generator may then be superimposed on these curves, both at fuU voltage and after depletion of one cell (eg, mercury-zinc) to assess safety margins. Noninvasive programmability of the duration of the pulse is a relatively new approach to conserve energy and has so far been used with constantvoltage pulse generators.14J5 The shortest duration of an impulse producing consistent capture (at a constant impulse amplitude) represents the threshold of stimulation as a function of pulse duration, with the voltage held constant at the output of the pulse generat~r.~'.'~ Proper adjustment of pulse duration allows long-term pacing with conservation of energy and charge.l5JeThe ability to increase the output energy of an implanted unit in the face of a large, but temporary, change in threshold (eg, anoxia) is an important benefit of noninvasive programming of the width of the pulse. Programmability of the width of the pulse may also be useful in regaining capture and temporarily extending the useful life of an implanted pulse generator with a failing battery if pulse duration is increased. Figure 4 illustrates the value of increasing pulse duration with cell depletion. Some implantable pulse generators automatically adjust the width of their pulse when battery depletion occurs. This feature is called "energy compensation." The increase in the width of the pulse compensates for the drop in voltage that occurs with the depletion of one cell, thereby maintaining the energy delivered and the original safety margins. THRESHOLD MEASUREMENT FOR CARDIAC PACING 763 Downloaded From: http://publications.chestnet.org/pdfaccess.ashx?url=/data/journals/chest/20987/ on 05/03/2017 STRENGTH DURATION CURVE With a constant voltage threshold device incorporating adjustable pulse width, this strength duration curve can be easily created. 1I I ln~ttal output One Cell Safety Margln pulse width .... :\ wtth decreasIng output voltage I Typtcal chrontc threshold curve 0 I I I I I 05 1.0 1.5 2.0 2.5 Pulse Duration - (milliseconds) FIGURE 4. Typical strengthduration curve for long-term lead. This illustrates importance of pulse width and principle of automatic energy compensation that is feature of some implantable pulse generators ( Medtronic ) . Pulse width increases with battery depletion. Initial output refers to output voltage of implantable pulse generator (5.4 v ) , and "one cell depleted refers to output voltage with depletion of one mercury-zinc cell ( 3.6 v ) . Dashed cume shows long-term threshold. Assume current at 500 ohms. Initial energy delivered by generator = VIT = 5.4 x 10.8 x 0.5 =29 microjoules at 500 ohms. Energy delivered with one cell depleted = VIT = 3.6 X 7.2 X 1.5 = 39 microjoules at 500 ohms. Voltage threshold at initial pulse width = 2.75 v. Voltage threshold with one cell depletion and increased pulse width = 1.25 v. Initial safety margin = Vo/Vt2 = 3.86/1. Safety margin with one cell depleted with increased pulse width = (3.6/1.25)* = 8.29/1. Safety margin with one cell depleted, without energy compensation, by means of increased pulse width = (3.W 2.75)* = 1.71/1. Safety Margins The concept of safety margins is important when measuring long-term threshold at the time of replacement of a pulse generat~r.'~.""~ The threshold in any given patient can vary physiologically during the day13J7J8and can be significantly altered by a number of pharmacologic agents or metabolic abnormalitie~.'~.~~ Much of the original work in this area was done by measuring changes in threshold expressed in microjoules. This stems largely from the work of Preston et al,13who measured threshold as a percentage of available pacemaker energy. They also measured physiologic and pharmacologic changes ( + or -) in threshold as the maximal percentage of change from the baseline energy threshold. Their approach and the important clinical implications become clearer with a specific example. Let us assume a patient with an average threshold of 5 microjoules. A 50-percent decrease in energy threshold gives a value of 2.5 microjoules (2.515 X 100 = 50 percent). A 50-percent increase in energy threshold gives a value of 7.5 microjoules (2.515 X 100 = 50 percent ) . Ordinary physiologic circumstances may increase or decrease the threshold expressed in microjoules, by as much as 50 per~ e n t . ' ~ . ' ~If. ' the ~ patient's threshold is measured at the lowest point (2.5 microjoules ) in its physiologic swing, an adequate safety margin must provide at least 7.5 microjoules to allow for a 50 percent increase in energy requirement above the "baseline" value of 5 microjoules. Safety margins have been traditionally defined as the ratio of output energy of the pulse generator (Eo) to the energy at threshold ( E , ) . In a constant-voltage pulse generator, the output voltage remains constant according to the number and type of cell (5.4 v with four mercuryzinc cells and 6.7 v with five such cells). This allows simple calculation of safety margins by comparing voltage output ( Vo ) to threshold voltage ( V, ), provided measurements are made at the same pulse width ( T ) : Safety margin = Eo = VOIOT = VOIO = VOVO = VOVO = Vf = E. V.1.T V.1. V.V. v: A vtvt - [?I= R According to the previously mentioned considerations in our hypothetical patient, we would recommend (T. A. Preston, personal communication, 1976) an energy safety margin ( & / E t ) of 311, (7.5/2.5), rather than a 211 ratio.'' A 311 energy safety margin may be translated into a voltage ratio of about 1.75 X Vo/V,. The output voltage of most constant-voltage pulse generators is about 5 v or slightly higher. A constant-voltage threshold of 3 v is probably the highest acceptable chronic threshold. The range of 3 to 3.5 v is a gray zone and may be acceptable if the patient is not "pacemaker dependent." Values over 3.5 v are generally unacceptable (G. Fontaine, personal communication, 1976). Firm recommendation about the lower acceptable value for long-term threshold cannot be made in the absence of clearer information in the literature about this problem. In a borderline situation a decision must be made to replace the system, connect a high-output generator, or, at worst, consider the safety margin acceptable. These decisions must be individualized according to the need of the patient and the pacing svstem. No chances should be taken in patients who might be subject to large fluctuations in long-term thresholds because of potentially serious metabolic dis- 764 BAROLD, WINNER Downloaded From: http://publications.chestnet.org/pdfaccess.ashx?url=/data/journals/chest/20987/ on 05/03/2017 CHEST, 70: 6, DECEMBER, 1976 orders or other conditions liable to cause s i g d c a n t changes in threshold, especially if the patient is "pacemaker dependent." C m ~ c fIMPLICATIONS i The current threshold is helpful in determining the integrity of the electrode-tissue interface and reflects the density of current, which is a prime factor responsible for successful stimulation. While current thresholds yield good information on lead position, they are often not helpful in determining lead integrity. Voltage thresholds are useful for information on lead position and lead integrity, especially when voltage and current are measured simultaneously. Energy thresholds may be calculated when mathematically accurate safety margins are required. Impedance is calculated when voltage and current are measured during stimulation of the heart. Voltage and current wave forms may be displayed on an oscilloscope, averaged, and then used in Ohm's law to determine impedance. Average impedance (measured from the ratio of average voltage to average current during the pulse) tends to fall slightly in the first ten days after initial imof the lead and then stabilizes.~.~o the absence of lead fracture, infection, or excessive myocardial scarring,2' long-term values for impedance tend to remain constant and do not differ significantly from the original ~ a l u e s . ~ .Calcula'~,~~ tion of impedance is obviously more accurate when both voltage and current are measured simultaneously. Because polarization is a function of pulse amplitude, the aoerage impedance may vary slightly (but not significantly for clinical purposes) if calculated from measurements obtained at threshold, compared with those obtained at the full output of the implantable pulse generator. We recommend that voltage and current be obtained at the same voltage level as the output of the implantable pulse generator and be measured near the leading edge of the pulse, since more information can be learned about the lead and heart tissue resistance before the polarization effect begins to influence the reading significantly. Currently, it is simpler, more practical, and rob ably more accurate to use devices that measure voltage and current instantaneously and simultaneously very close to the leading edge and to display one or both values digitally (some commercially available threshold analyzers measure voltage and current at an arbitrary time of 90 microseconds into the pulse). Instantaneous display of current when taking the constant voltage threshold may be the only way to make the diagnosis of an intermittent (and rapidly) changing lead impedance due to incomplete fracture. It cannot be overemphasized that threshold and are unrelated each providing and different about a pacing system. Determination of impedance can be helpful in the diagnosis of lead fractures, inTable 1 sulation breaks, and position Presents some guidelines. CONCLUSIONS Proper measurement of stimulation thresholds is important to ensure good function and safety of a pacemaker. The techniques for threshold measurement should be standardized to allow meaningful comparison of published data.4 The external device Table 1-4uideliner for implantation of Permanent Pacemukerr Constant Current Threshold Constant Voltage Threshold Impedance* Increased Increased Normal Displacement ; high threshold Reposition or use high output generator Increased Decreased Decressed Fracture in lead insulation Repair or replace lead Normal (low) or increased Increased Increased Lead fracture Repair or replace lead (consider unipolarization of bipolar system) Decreased Decreased Normal Norma1 situation; if there h malfunction of implanted system, consider faulty pacemaker lead connection or faulty implanted pulse generator Correct connection or replace pulse generator Cause Treatment *In general, impedance ranges from 300 to 1,000 ohms. Higher values may occasionally be seen in normally functioning systems, particularly with ball-tip electrodes. Acceptable ranges of impedance should be obtained from individual manufacturers. CHEST, 70: 6, DECEMBER, 1976 THRESHOLD MEASUREMENT FOR CARDIAC PACING 765 Downloaded From: http://publications.chestnet.org/pdfaccess.ashx?url=/data/journals/chest/20987/ on 05/03/2017 for measuring thresholds should preferably use the same type of output circuitry as the implantable pulse generator. When voltage and current measurements are known, lead impedance can be calculated. This is essential for the differential diagnosis of obvious or impending lead fracture, insulation breaks, and lead dislodgment. The ordinary constant-current external pulse generator has important limitations, and the additional use of a constant-voltage device for determining threshold is essential to insure integrity and optimal performance of the pacing system. ACKNOWLEDGMENT: We are grateful to Larry Shearon, B.S.E.E. (Medtronic, Inc.) for his contribution and enthusiastic support. R E F E R E N m 1 Center S, Tarjan P: The clinical application of low output pacemakers. J Thorac Cardiovasc Surg 64:935, 1972 2 Smyth NPD, Keshishian JM, Baker NR, et al: Physiological rationale for the clinical use of low output pacemakers. Med Ann DC 43:257,1974 3 Furman S, Escher DJW: Principles and Techniques of Cardiac Pacing. New York, Harper and Row, 1970, p 53 4 Fontaine G, Kevorkian M, Bonuet M, et al: Defbition de la mesure du seuil d'entrainement electrique: A propos de I'etude par ardinateur de 2400 mesures ( seuil pl.Bcoce et chronique) avec sondes endocavitahs bipdaires Medtronic 5816 and 5818. Ann Cardid Angeiol20:491, 1971 5 Jarvon D, Briller SA, Schwan HP:Polarization impedance in platinum-iridium and elgiloy cardiac pacemaker e l m trodes. In Engineering in Medicine and Biology: Proceedings of the 19th Annual Conference. 1966, p 9 6 Thalen HJT, Van den Berg JW, Van der Heide JNH, et al: The Artificial Cardiac Pacemaker. Springfield, Ill, Charles C Thomas, 1969, p 189 7 Greatbatch W, Piersna B, Shannon FD, et al: Polarization phenomena relating to physiological electrodes. AM NY Acad Sci 167:722,1969 8 Keller JW Jr: Improving pacemaker electrodes. In T h h n HJT ( e d ): Cardiac Pacing: Proceedings of the Fourth International Symposium. Assen, the Netherlands, Van Gorcum and Co, 1973, p 276 9 Lulu DJ, Buysman JR: A threshold ambiguity problem involving a cardiac pacemaker. Am Surg 41:413, 1975 10 Fontaine G, Kevorkian M, Welti JJ, et al: Comparison between endocardial versus myocardial and unipolar versus bipolar thresholds after long term pacing. In Thalen HJT ( e d ): Cardiac Pacing: Proceedings of the Fourth International Symposium. Assen, the Netherlands, Van Gorcum and Co, 1973, p 304 11 Furman S, Parker B, Escher DJW: Decreasing electrode size and increasing efficiency of cardiac stimulation. J Surg Res 1:105,197 1 12 Davies JG, Sowton E: Electrical threshold of the human heart. Br Heart J 28:231,1966 13 Preston TA, Fletcher RD, Lucchesi BR, et al: Changes in myocardial threshold: Physiologic and pharmacologic factors in patients with implanted pacemakers. Am Heart J 74:235,1967 14 Chardack WM, Gage AA, Federico AJ, et al: Non-invasive, magnetically coupled control of pulse width of implantable pacemakers: Its value in reduction of current drain and facilitation of patient follow-up. In Thalen H P (ed): Cardiac Pacing: Proceedings of the Fourth International Symposium. Assen, the Netherlands, Van Corcum and Co,1973, p 128 15 Furman S, Garvey J, Hurzeler P: Pulse duration variation in pacemaker longevity. J and electrode size as fa& ThoracCardiovasc Surg 69:382,1975 16 Funnan S, Denize A, Escher DJW, et al: Energy consumption for cardiac stimulation as a function of pulse duration. J Surg Res 6:441,1966 17 Siddons H, Sowton E: Cardiac Pacemakers. Springhld, Ill, Charles C Thomas, 1967, p 145 18 Sowton E, Barr I: Physiological changes in threshold. Ann NY Acad Sci 167:679,1969 19 Preston TA, Judge RD: Alteration of pacemaker threshold by drug and physiological factors. Ann NY Acad Sci 167:686,1969 20 Westerholm CJ: Threshold studies in transvenous cardiac pacemaker treatment. Scand J Thorac Cardiovasc Surg, suppl8,1971 21 Parsonnet V, in panel discussion, Indications for long term pacing, surgical techniques and pacemaker teams, elee trodes, energy sources. In Thalen HJT ( e d ) : Cardiac Pacing: Proceedings of the Fourth International Symposium. Assen, the Netherlands, Van Gorcurn and Co, 1973, p 501 22 Tarjan PP: Engineering aspects of implantable cardiac pacemakers. In Samet P (ed) : Cardiac Pacing. New York, Grune and Stratton, Inc, 1973,p 47 786 BAROLD, WINNER Downloaded From: http://publications.chestnet.org/pdfaccess.ashx?url=/data/journals/chest/20987/ on 05/03/2017 CHEST, 70: 6, DECEMBER, 1976