Survey
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Neurological Patient and Technical Services Department November 2005 Page 1 of 27 1-800-328-0810 Compatibility Guidelines for Neurostimulation Products Contents: Device Compatibility Bone Growth Stimulators: 2 Defibrillation/Cardioversion: 2 Dental Drills and Ultrasonic Probes: 2 Diathermy (Shortwave, Microwave, Therapeutic Ultrasound): 3 Effects on Other Implanted Devices: 4 Electrocautery: 4, 5 Electrolysis: 6 EMI/EMC: 6, 7 High-output Ultrasonics/Lithotripsy: 7 Home or Occupational Environment (electromagnetic field devices): 7, 8 Household Items: 9, 10 Laser Procedures: 10 MRI: 10 Other Medical Procedures (CT Scan, Diagnostic Ultrasound, Diagnostic Xray, MEG, PET): 11 Patient Activities (Unexpected Changes in Stimulation, Driving, Operating Power Tools): 11, 12 Psychotherapeutic Procedures: 12 Radiation Therapy: 12 RF or Microwave Ablation: 12 Scuba Diving or Hyperbaric Chambers: 13 Skydiving, Skiing, or Hiking in the Mountains: 13 TENS: 13 Theft Detectors: 13, 14 Therapeutic Magnets: 14, 15 Programmer/Programming Compatibility Patient Control Devices May Affect Other Implanted Devices: 15 Patient Magnet/Control Magnet: 15 Programmer Interaction with Cochlear Implant: 15 Programmer Interaction with Other Implanted Devices/Inadvertent Programming: 15, 16 Telemetry Disruption from EMI: 16 Appendices Table of Potential Effects of EMI from Devices or Procedures: 17 Appendix for MRI and Activa Therapy (Deep Brain Stimulation): 18-22 Appendix for MRI and neurostimulation therapy for chronic pain: 23-27 Neurological Patient and Technical Services Department SCS/PNS 7425, 7427, 7427V, 7479, 7479B, 37711 November 2005 Page 2 of 27 DBS I nterStim 7424, 7426, 7428 3023 Enterra 7425G, 3116 Bone growth stimulators – Keep external magnetic field bone growth stimulator coils 45 cm (18 in.) away from the neurostimulation system. When using either an implantable or external bone growth stimulator, ensure that both the bone stimulator and neurostimulator are working as intended. Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. Precautions Required Bone Growth Stimulators – Keep external magnetic field bone growth stimulator coils 18 in (45 cm) away from the neurostimulator / lead system. When using either an implantable or external bone growth stimulator, ensure that both the bone stimulator and neurostimulator are working as intended. Defibrillation or cardioversion – When a patient is in ventricular or atrial fibrillation, the first consideration is patient survival. External defibrillation or cardioversion can damage a neurostimulation system and cause induced currents in the leadextension portion of the neurostimulation system that can injure the patient. Minimize the current flowing through the neurostimulator system by following these guidelines: • Position defibrillation paddles as far from the neurostimulator as possible. • Position defibrillation paddles perpendicular to the neurostimulation system. • Use the lowest clinically appropriate energy output (watt seconds). After defibrillation, confirm the neurostimulation system is functioning as intended. External Defibrillators – If a patient requires external defibrillation, the first consideration should be patient survival. Safety for use of external defibrillatory discharges on patients with neurostimulation systems has not been established. External defibrillation may damage a neurostimulator. If external defibrillation is necessary, follow these precautions to minimize current flowing through the neurostimulator and lead system: • Position defibrillation paddles as far from the neurostimulator as possible. • Position defibrillation paddles perpendicular to the implanted neurostimulator-lead system. • Use the lowest clinically appropriate energy output (watt seconds). • Confirm neurostimulation system function following any external defibrillation. External Defibrillators –Safety for use of external defibrillatory discharges on patients with neurostimulation systems has not been established. External defibrillation may damage or cause reprogramming of a neurostimulator. If external defibrillation is necessary, follow these precautions to minimize current flowing through the neurostimulator and lead system: • Position defibrillation paddles as far from the neurostimulator as possible. • Position defibrillation paddles perpendicular to the neurostimulatorlead system. • Use the lowest clinically appropriate energy output (watt seconds). • Confirm the neurostimulation system function following any external defibrillation. Defibrillation / Cardioversion – When a patient is in ventricular or atrial fibrillation, the first consideration should be patient survival. External defibrillation or cardioversion can damage a neurostimulation system. It is recommended not to use defibrillation or card ioversion paddles near the neurostimulator. When external defibrillation or cardioversion is necessary, minimize the current flowing through the neurostimulator and lead system as follows: • Position paddles as far from the neurostimulator as possible. • Position paddles perpendicular to the neurostimulation system. • Use the lowest clinically appropriate energy output (watt seconds). • Neurostimulation system function should be confirmed after external defibrillation. Defibrillation or cardioversion may also cause induced currents in the lead portion of the neurostimulation system that could be hazardous or cause further injury. Dental drills and ultrasonic probes – Turn OFF the neurostimulator. Keep the drill or probe 15 cm (6 in) away from the neurostimulator. Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. Dental Drills and Ultrasonic Probes – Turn OFF the neurostimulator. Keep the drill or probe 6 in (15 cm) away from the neurostimulator. Neurological Patient and Technical Services Department DBS SCS/PNS 7425, 7427, 7427V, 7479, 7479B, 37711 November 2005 Page 3 of 27 I nterStim 3023 7424, 7426, 7428 Enterra 7425G, 3116 Contraindication Contraindication Contraindication Contraindication Diathermy – Do not use shortwave diathermy, microwave diathermy or therapeutic ultrasound diathermy (all now referred to as diathermy) on patients implanted with a neurostimulation system. Energy from diathermy can be transferred through the implanted system and can cause tissue damage at the location of the implanted electrodes, resulting in severe injury or death. Diathermy can also damage the neurostimulation system components, resulting in loss of therapy and requiring additional surgery for system explantation and replacement. Advise your patient to inform all their healthcare professionals that they should not be exposed to diathermy treatment. Injury to the patient or damage to the device can occur during diathermy treatment when: • the neurostimulation system is turned ON or OFF. • diathermy is used anywhere on the body—not just at the location of the neurostimulation system. • diathermy delivers heat or no heat. any component of the neurostimulation system (lead, extension, neurostimulator) remains in the body. Implantation of an Activa Brain Stimulation System is contraindicated for: • Patients exposed to diathermy. Do not use shortwave diathermy, microwave diathermy or therapeutic ultrasound diathermy (all now referred to as diathermy) on patients implanted with a neurostimulation system. Energy from diathermy can be transferred through the implanted system and can cause tissue damage at the location of the implanted electrodes, resulting in severe injury or death. Diathermy is further prohibited because it can also damage the neurostimulation system components resulting in loss of therapy, requiring additional surgery for system explantation and replacement. Injury or damage can occur during diathermy treatment whether the neurostimulation system is turned “on” or “off.” Advise your patients to inform all their healthcare professionals that they should not be exposed to diathermy treatment. Diathermy –Do not use shortwave diathermy, microwave diathermy or therapeutic ultrasound diathermy (all now referred to as diathermy) on patients implanted with a neurostimulation system. Energy from diathermy can be transferred through the implanted system and can cause tissue damage at the location of the implanted electrodes, resulting in severe injury or death. Diathermy is further prohibited because it can also damage the neurostimulation system components resulting in loss of therapy, requiring additional surgery for system explantation and replacement. Injury or damage can occur during diathermy treatment whether the neurostimulation system is turned “on” or “off.” Advise your patients to inform all their healthcare professionals that they should not be exposed to diathermy treatment. Diathermy – Do not use shortwave diathermy, microwave diathermy or therapeutic ultrasound diathermy (all now referred to as diathermy) on patients implanted with a neurostimulation system. Energy from diathermy can be transferred through the implanted system and can cause tissue damage at the location of the implanted electrodes, resulting in severe injury or death. Diathermy is further prohibited because it can also damage the neurostimulation system components resulting in loss of therapy, requiring additional surgery for system explantation and replacement. Injury or damage can occur during diathermy treatment whether the neurostimulation system is turned “on” or “off.” Advise your patients to inform all their healthcare professionals that they should not be exposed to diathermy treatment. Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. Neurological Patient and Technical Services Department SCS/PNS 7425, 7427, 7427V, 7479, 7479B, 37711 Neurostimulator interaction with implanted cardiac devices – When a patient’s medical condition requires both a neurostimulator and an implanted cardiac device (eg, pacemaker, defibrillator), physicians involved with both devices (eg, neurologist, neurosurgeon, cardiologist, cardiac surgeon) should discuss the possible interactions between the devices before surgery. To minimize or prevent the effects described below, implant the devices on opposite sides of the body and follow any additional instructions. • Defibrillation therapy from an implanted defibrillator may damage the neurostimulator. • The electrical pulses from the neurostimulation system may interact with the sensing operation from cardiac devices and could result in an inappropriate response of the cardiac devices. To minimize or prevent the cardiac device from sensing the neurostimulator output, program the neurostimulator to a bipolar configuration and to a minimum rate of 60 Hertz. Program the cardiac device to bipolar sensing. DBS 7424, 7426, 7428 Effects on Other Medical Devices – The Activa System may affect the operation of other implanted devices, such as cardiac pacemakers and implantable defibrillators. Possible effects include sensing problems and inappropriate device responses. If the patient requires concurrent implantable pacemaker and/or defibrillator therapy, careful programming of each system may be necessary to optimize the patient’s benefit from each device. Cal Tech Services for further information. November 2005 Page 4 of 27 I nterStim Enterra 3023 7425G, 3116 Effects on Other Implanted Devices The neurostimulation system may affect the operation of other implanted devices, such as cardiac pacemakers and implantable cardioverter defibrillators. Physical proximity may cause sensing problems and inappropriate device responses. If the patient requires concurrent implantable pacemaker and/or defibrillator therapy, evaluation of any potential interference problems and careful programming of each system may be necessary to optimize the patient’s benefit from each device. Interaction with other Implantable Devices – When another implantable device (e.g., pacemakers, defibrillators, or cochlear implants) is required, the physicians involved in both therapies should discuss the possible interaction between the devices. Electrical impulses from the neurostimulation system may affect the sensing operation and cause inappropriate device response of other implanted devices. Careful programming of each system may optimize the benefit from each device. Follow these suggested guidelines: • The neurostimulator should be placed on the opposite side of the body from the other implanted device. • The neurostimulator should be reprogrammed to bipolar stimulation • Each system should be checked to ensure that it is working as intended. Cal Tech Services for further information. Cal Tech Services for further information. Cal Tech Services for further in formation. See also “Programmer interaction with other active implanted devices”. Electrocautery – If electrocautery is used near an implantable device or contacts a device or insertion-needle, the following effects may occur: • The tissue surrounding the insertionneedle (during placement of a percutaneous lead) may be damaged. Electrocautery – Electrocautery can damage the lead, the extension, or both. It can also cause temporary suppression of neurostimulator output and/or reprogramming of the neurostimulator. If use of electrocautery is necessary, the current path (ground plate) should be kept as far away from the neurostimulator, Electrocautery –Electrocautery can cause temporary suppression of neurostimulator output and/or reprogramming of the neurostimulator. For more information, refer to “Cautery Protection” and “Power ON Reset” below. No electrosurgical tip should ever be used in the vicinity of an implanted system. If Caution: Do not use an electrosurgical tip in close proximity (closer than 2 inches) to the neurostimulator or lead. Electrocautery – Electrocautery can damage the lead or neurostimulator. It can also cause temporary suppression of neurostimulator output and/or reprogram Neurological Patient and Technical Services Department SCS/PNS 7425, 7427, 7427V, 7479, 7479B, 37711 • The insulation on the lead or extension may be damaged, resulting in component failure or induced currents into the patient that may damage tissue or stimulate or shock the patient. • The neurostimulator may be damaged, output may be temporarily suppressed or increased, or stimulation may stop because parameters were changed to poweron-reset settings (eg, output OFF, amplitude 0.0V). When electrocautery is necessary, follow these precautions: • Before using electrocautery, turn OFF the neurostimulator. • Disconnect any cable connecting the lead or extension to a screener or external neurostimulator. • Use only bipolar cautery. • If unipolar cautery is necessary: • use only a low-voltage mode. • use the lowest possible power setting. • keep the current path (ground plate) as far from the neurostimulator, extension, and lead as possible. • do not use full-length operating room table grounding pads. After using electrocautery, confirm that the neurostimulator is functioning as intended. DBS November 2005 Page 5 of 27 I nterStim Enterra 7424, 7426, 7428 3023 7425G, 3116 extension, and lead as possible, and use of bipolar electrocautery is recommended. use of electrocautery is necessary: • Turn off the neurostimulator before performing electrocautery. • Avoid using spray coagulation. If spray coagulation is necessary, keep the power setting at less than 50 watts. • Keep the current path (ground plate) as far away from the neurostimulator as possible. • Bipolar cautery is recommended. Refer to SCS/PNS guidelines, except for first bulet. Cautery Protection Cautery protection is incorporated into the InterStim Model 3023 Neurostimulator. It enables the neurostimulator to return to normal function following electrosurgery. It is recommended that the electrical current path between the cautery tip electrode and the indifferent electrode be kept as far away from the neurostimulator/lead system as possible. The use of bipolar cautery is recommended. Caution: Do not use an electrosurgical tip in close proximity (that is, closer than 2 inches) to the neurostimulator, extension (if used), or lead. Power ON Reset As a safety feature, if the neurostimulator is subjected to extreme electromagnetic stress (such as from cautery or defibrillation procedures), its parameters will automatically reset to the original shipping values except for amplitude resolution set at fine. Once the neurostimulator is removed from the interference source, the device must be reprogrammed to the desired settings. In addition, the serial number is lost and cannot be reprogrammed. Refer to SCS/PNS guidelines. the neurostimulator to Power ON Reset parameters (output off, amplitude = 0V), which requires neurostimulator reprogramming. Electrocautery may also cause induced currents in the lead portion of the neurostimulation system that could be hazardous or cause further injury. Follow these precautions when using electrocautery: • Turn off the neurostimulator before performing electrocautery. • Only bipolar cautery is recommended. • If unipolar cautery is necessary: – Do not use high voltage modes. – Keep the power setting as low as possible. – Keep the current path (ground plate) as far away from the neurostimulator and lead as possible. • Confirm the neurostimulator function after electrocauterization. Refer to SCS/PNS guidelines, except for first bulet. Neurological Patient and Technical Services Department SCS/PNS 7425, 7427, 7427V, 7479, 7479B, 37711 November 2005 Page 6 of 27 DBS I nterStim 7424, 7426, 7428 3023 Enterra 7425G, 3116 Electrolysis – Turn OFF the neurostimulator. Keep the electrolysis wand at least 15 cm (6 in) away from the neurostimulator. Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. Electrolysis – Turn OFF the neurostimulator, and keep the electrolysis wand at least 6 in (15 cm) away from the neurostimulator. Electromagnetic interference (EMI) – Electromagnetic interference is a field of energy generated by equipment found in the home, work, medical or public environments that is strong enough to interfere with neurostimulator function. Neurostimulators include features that provide protection from electromagnetic interference. Most electrical devices and magnets encountered in a normal day are unlikely to affect the operation of a neurostimulator. However, sources of strong electromagnetic interference can result in the following: • Serious patient injury or death, resulting from heating of the implanted components of the neurostimulation system and damage to surrounding tissue. • System damage, resulting in a loss of or change in symptom control and requiring surgical replacement. • Operational changes to the neurostimulator, causing it to turn ON or OFF (particularly in neurostimulators enabled for magnet use), or to reset to power-on-reset (POR) settings, resulting in loss of stimulation, return of symptoms, and in the case of POR, requiring reprogramming by a clinician. • Unexpected changes in stimulation, causing a momentary increase in stimulation or intermittent stimulation, which some patients have described as a jolting or shocking sensation. Although the unexpected change in stimulation Electromagnetic Interference (EMI) Electromagnetic interference is a field (electrical, magnetic or a combination of both) that is generated by various medical or environmental devices. These medical and environmental (home, occupational, and other) devices may generate enough interference to change the parameters of a neurostimulator; turn a neurostimulator off and on, or cause a neurostimulator to surge, shock, or jolt the patient. In addition, it is possible for the extension, lead, or both to “pick up” electromagnetic interference and deliver an excess voltage, which can in turn deliver an excessive amount of heat to the brain. Refer to the sections that follow for guidelines on the interaction of electromagnetic interference and an implanted Activa System. Electromagnetic Interference – Patients should exercise reasonable caution in avoidance of devices which generate a strong electric or magnetic field. Close proximity to high levels of electromagnetic interference (EMI) may cause a neurostimulator to unexpectedly cease to function or cause sensitive patients to experience a momentary increase in their perceived level of stimulation. Also, severe EMI can permanently erase the neurostimulator serial number, causing “????” to be displayed in place of the serial number. Electromagnetic interference (EMI) Electromagnetic interference is a field of energy (electric, magnetic, or a combination of both) generated by equipment found in the home, work, medical or public environments that is strong enough to interfere with neurostimulator function. Neurostimulators include features that provide protection from electromagnetic interference. Most electrical devices and magnets encountered in a normal day are unlikely to affect the operation of a neurostimulator. However, strong sources of electromagnetic interference can result in the following: • Serious injury or death, resulting from heating of the implanted neurostimulation system components, which can damage surrounding tissue. • System damage, requiring surgical replacement or result in a loss of or change in symptom control. • Operational changes to the neurostimulator, causing it to switch ON or OFF (particularly in neurostimulators enabled for magnet use), or to reset to default factory settings which may result in loss of stimulation, return of symptoms, and require reprogramming by the physician. Hospital or Medical Environment Kinetra neurostimulators have been tested for electromagnetic compatibility (EMC) in electrical and magnetic fields that simulate common sources of EMI found in a hospital environment. Severe electromagnetic fields, such as those that exist near electrocautery equipment, can reprogram a neurostimulator, temporarily suppress the output of a neurostimulator, and cause damage to the implanted system components. Telemetry usually continues to function but may indicate that the neurostimulator output is On when, in fact, no stimulation is being delivered because the amplitude is set to 0.0 V. Kinetra neurostimulators have a protective circuit to prevent Refer to SCS/PNS guidelines. Neurological Patient and Technical Services Department SCS/PNS 7425, 7427, 7427V, 7479, 7479B, 37711 may feel uncomfortable, it does not damage the device or injure the patient directly. In rare cases, as a result of the unexpected change in stimulation, patients have fallen down and been injured. November 2005 Page 7 of 27 DBS I nterStim 7424, 7426, 7428 3023 Enterra 7425G, 3116 damage to the circuitry, but a large burst of electromagnetic energy may temporarily affect the output and/or the ability to reprogram the device. Severe EMI (e.g., electrocautery, defibrillatory discharge, MRI) can also erase the neurostimulator serial number. High-Output Ultrasonics –Use of highoutput ultrasonic devices, such as an electrohydraulic lithotriptor, is not recommended for patients with an implanted neurostimulation system. While there is no danger to the patient, exposure to high-output ultrasonic frequencies may result in damage to the neurostimulator circuitry. If lithotripsy must be used, do not focus the beam near the neurostimulator. High-output ultrasonics or lithotripsy – Use of high-output ultrasonic devices, such as electrohydraulic lithotriptors, is not recommended for patients who have an implanted neurostimulation system. If lithotripsy must be used, do not focus the beam within 15 cm (6 in) of the neurostimulator. Lithotripsy – Use of high output ultrasonic devices, such as an electrohydraulic lithotriptor, is not recommended for patients with an implanted neurostimulation system. While there is no danger to the patient, exposure to high output ultrasonic frequencies may result in damage to the neurostimulator circuitry. If lithotripsy must be used, do not focus the beam near the neurostimulator. Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. Electromagnetic field devices – Patients should exercise care or avoid the following equipment or environments: • Antenna of citizens band (CB) radio or ham radio • Electric arc welding equipment • Electric induction heaters used in industry to bend plastic • Electric steel furnaces • High-power amateur transmitters • High-voltage areas (safe if outside the fenced area) • Linear power amplifiers • Magnetic degaussing equipment • Magnets or other equipment that generates strong magnetic fields • Microwave communication transmitters (safe if outside the fenced area) • Perfusion systems • Resistance welders Occupational Environments – Commercial electrical equipment (arc welders, induction furnaces, resistance welders), communication equipment (microwave transmitters, linear power amplifiers, high-power amateur transmitters), and high voltage power lines may generate enough electromagnetic interference (EMI) to interfere with neurostimulator operation if approached too closely. Occupational Environments – Commercial electrical equipment (arc welders, induction furnaces, resistance welders), communication equipment (microwave transmitters, linear power amplifiers, high-power amateur transmitters), and high-voltage power lines may generate enough EMI to interfere with neurostimulator operation if approached too closely. Home or Job Environment Kinetra neurostimulators should not be affected by normal operation of electrical equipment such as household appliances, electric machine shop tools, microwave ovens, RF transmitting systems, or microwave frequency transmitting systems. A strong magnetic field (electromagnet or permanent magnet) can switch the neurostimulator output Refer to SCS/PNS guidelines. High-Output Ultrasonics / Lithotripsy – Use of high-output ultrasonic devices, such as electrohydraulic lithotriptor, is not recommended for patients with an implanted neurostimulation system. If lithotripsy must be used, do not focus the beam within 6 inches (15 cm) of the neurostimulator. Home, Public, and Occupational Environment Exercise care or avoid the following equipment or environments: • Antenna of citizen band (CB) radio or ham radio • Electric arc welding equipment • Resistance welders • Electric induction heaters used in industry to bend plastic • Electric steel furnaces • High voltage (safe if outside the fenced area) • Television and radio transmitting towers (safe if outside the fenced area) • Microwave communication transmitters (safe if outside the fenced area) • Linear power amplifiers • High power amateur transmitters Neurological Patient and Technical Services Department SCS/PNS 7425, 7427, 7427V, 7479, 7479B, 37711 • Television and radio transmitting towers (safe if outside the fenced area) If patients suspect that equipment is interfering with neurostimulator function, they should do the following: • Move away from the equipment or object. • If possible, turn OFF the equipment or object. • Then, if necessary, use the control magnet or patient programmer to return the neurostimulator to the desired ON or OFF state. • Inform the equipment owner or operator of the occurrence. If the above actions do not resolve the effects of the interference, or the patients suspect that their therapy is not effective after exposure to EMI, they should contact their physician. DBS I nterStim 7424, 7426, 7428 3023 from On to Off or Off to On , but does not change the programmed parameters. In addition, theft detectors and airport/security screening devices can affect the neurostimulator output and patient stimulation. Advise your patients to use care when approaching theft detectors or airport/security screening devices. If they feel unwanted stimulation as they approach the device, suggest that they request assistance to bypass the device. Advise your patients to avoid or to exercise care when approaching the following: • Theft detectors • Airport/security screening devices • Large stereo speakers with magnets • Electric arc welding equipment • Electric steel furnaces • Electric induction heaters (used in industry to bend plastic) • Power lines • Electric substations and power generators If your patient suspects an electrical device or magnet is interfering with the neurostimulator, advise him/her to move away from it, or turn the device off. Then, the patient can use the control magnet or therapy controller to set the neurostimulator back to the desired On or Off state. When switched On, the neurostimulator will resume stimulation at the previously programmed level. 1 Unexpected On/Off switching of the Kinetra Model 7428 Neurostimulator may be avoided by disabling the magnet control circuit with the clinician programmer software. If the magnet control circuit is disabled, patients will require a Model 7436 Therapy Controller to turn their therapy On or Off. 1 November 2005 Page 8 of 27 Enterra 7425G, 3116 • • • Perfusion systems Magnets or other equipment that generates strong magnetic fields Magnetic degaussers Neurological Patient and Technical Services Department SCS/PNS 7425, 7427, 7427V, 7479, 7479B, 37711 Household items – Most household appliances and equipment that are working properly and grounded properly will not interfere with the neurostimulation system. The following equipment is generally safe if patients follow these guidelines: • Freezer, refrigerator, or storm door magnets that hold the door closed: Do not lean against the magnetic strip of the door. • Radio-frequency sources (AM/FM radios, analog and digital cellular telephones, cordless and conventional telephones): Keep these items at least 10 cm (4 in) away from the implanted neurostimulator. • Stereo speakers and radios for the home or car: Do not lift or carry speakers or radios near the neurostimulator. • Sewing machines or salon hair dryer: Keep the neurostimulator away from the motors. • Computer disk drives: Keep the neurostimulator away from the disk drives. • Induction range: Keep the neurostimulator away from the burners while the burners are turned ON. • Power tools: Keep the motor away from the neurostimulator, lead, and extension. DBS November 2005 Page 9 of 27 I nterStim Enterra 7424, 7426, 7428 3023 7425G, 3116 Home Appliances – Home appliances that are in good working order and properly grounded do not usually produce enough electromagnetic interference (EM I) to interfere with neurostimulator operation. However, items with magnets (e.g., stereo speakers, refrigerators, freezers, power tools) may cause the neurostimulator to switch On or Off. Home Appliances –Home appliances that are in good working order and properly grounded do not usually produce enough electromagnetic interference (EM I) to interfere with neurostimulator operation. Home, Public, and Occupational Environment Most household appliances and equipment that are in good working order and properly grounded will not interfere with the neurostimulation system. If patients suspect that equipment is interfering with neurostimulator function, they should do the following: 1. Move away from the equipment or object. 2. If possible, turn off the equipment or object. 3. Inform the equipment owner / operator of the occurrence. If the above actions do not resolve the effects of the interference, or the patients suspect that their therapy is not effective after exposure to EMI, they should contact their physician. Radio Frequency Sources – Analog and digital cellular phones, AM/FM radios, cordless phones, and conventional wired telephones may contain permanent magnets. To prevent undesired turning On or Off of the stimulation, these devices should be kept at least 4 inches away from the implanted neurostimulator. Refer to SCS/PNS guidelines. Physician Instructions to the Patient Advise patients to exercise care when approaching the following: • Large stereo speakers (which contain magnets) • Electric arc welding equipment • Electric steel furnaces • Electric induction heaters used in industry to bend plastic • Power lines • Electric substations and power generators If a patient suspects an electrical device or magnet is interfering with his or her neurostimulator, the patient should move away from or turn the interfering device off. If necessary, the patient can use his or her control magnet or patient programmer to set the neurostimulator back to the desired on or off state. (When switched on, the neurostimulator will resume stimulation at the previously programmed level.) Refer to SCS/PNS guidelines. Generally Safe if precautions are followed: • Freezer, refrigerator, or storm door magnets that hold the door closed: Do not lean against the magnetic strip of the door. • Radio Frequency Sources: Analog and digital cellular phones, AM/FM radios, cordless phones, and conventional wired telephones, etc., should be kept at least 4 in (10 cm) away from the implanted neurostimulator. • Stereo speakers and radios for the home or car: Do not lift or carry them so that they are close to or touching the part of your body where the neurostimulator is located. • Sewing machines or salon hair dryer: Keep the neurostimulator away from the motors. Neurological Patient and Technical Services Department SCS/PNS 7425, 7427, 7427V, 7479, 7479B, 37711 November 2005 Page 10 of 27 DBS I nterStim 7424, 7426, 7428 3023 Enterra 7425G, 3116 • • • Laser procedures – Turn OFF the neurostimulator. Keep the laser directed away from the neurostimulation system. Magnetic Resonance Imaging MRI RF transmit body coil – Medtronic recommends that you do not conduct an MRI examination using an RF transmit body coil on a patient with any implanted neurostimulation system component because the interaction of the MRI with the neurostimulation system may lead to serious injury or death. See the section “Risks associated with MRI examination” on page 24. MRI transmit/receive head coil – An MRI examination of the head only (no other part of the body) can be conducted safely using an RF transmit/receive head coil when all instructions in “Appendix B: MRI and neurostimulation therapy for chronic pain” on pages 23-27 are followed. REFER TO MRI APPENDIX FOR SCS/PNS ON PAGE 23. Refer to SCS/PNS guidelines. Contraindication Patients who will be exposed to Magnetic Resonance Imaging (MRI) using a full body transmit radio-frequency (RF) coil, a receive-only head coil, or a head transmit coil that extends over the chest area. Performing MRI with this equipment can cause tissue lesions from component heating, especially at the lead electrodes, resulting in serious and permanent injury including coma, paralysis or death. Warnings Magnetic Resonance Imaging – Do not conduct an MRI examination on a patient with any implanted Activa System component until you read and fully understand all MRI information in the Kinetra or Soletra Manual. Do not conduct an MRI examination at parameters other than those described in this guideline. Failure to follow all warnings and guidelines related to MRI can result in serious and permanent injury including coma, paralysis, or death. REFER TO MRI APPENDIX FOR KINETRA ON PAGE 18. Computer disk drives: If repairing or adding additional components to a computer, keep the neurostimulator away from the disk drives. Induction range: Keep the neurostimulator away from the burners while the burners are turned on. Power tools: Keep the motor away from the neurostimulator and lead. Refer to SCS/PNS guidelines. Laser Procedures – Turn off the neurostimulator, and keep the laser directed away from the neurostimulation system. Magnetic-Resonance Imaging – Patients with an implanted device should not be exposed to the electromagnetic fields produced by magnetic resonance imaging (MRI). Use of MRI may potentially result in system failure or dislodgement, heating, or induced voltages in the neurostimulator and/or lead. An induced voltage through the neurostimulator or lead may cause uncomfortable, “jolting,” or “shocking” levels of stimulation. Clinicians should carefully weigh the decision to use MRI in patients with an implanted neurostimulation system, and note the following: • Magnetic and radio-frequency (RF) fields produced by MRI may change the neurostimulator settings, activate the device, and injure the patient. • Patients treated with MRI should be closely monitored and programmed parameters verified upon cessation of MRI. Magnetic Resonance Imaging (MRI) – Patients with an implanted device should not be exposed to the electromagnetic fields produced by magnetic resonance imaging (MRI). Use of MRI may potentially result in system failure, dislodgement, heating, or induced voltages in the neurostimulator and/or lead. An induced voltage through the neurostimulator or lead may cause uncomfortable, “jolting,” or “shocking” levels of stimulation. Clinicians should carefully weigh the decision to use MRI in patients with an implanted neurostimulation system, and note the following: • Magnetic and radio-frequency (RF) fields produced by MRI may change the neurostimulator settings and injure the patient. • Patients treated with MRI should be closely monitored and programmed parameters verified upon cessation of MRI. Neurological Patient and Technical Services Department SCS/PNS 7425, 7427, 7427V, 7479, 7479B, 37711 Other medical procedures – EMI from the following medical procedures is unlikely to affect the neurostimulation system: • Computerized Axial Tomography (CT or CAT) scans • Diagnostic ultrasound (eg, carotid scan, doppler studies) Note: To minimize potential image distortion, turn OFF the neurostimulator and keep the transducer 15 cm (6 in) away from the neurostimulation system. • Diagnostic x-rays or fluoroscopy Notes: ņ To minimize potential image distortion, turn OFF the neurostimulator. ņ Tight pressure such as used during mammography may damage the neurostimulator or disconnect the neurostimulation system components, which may require surgery to reconnect or replace components. During x-ray procedures that require external compression around implanted components, the x-ray equipment should be adjusted to limit the amount of pressure exerted on the neurostimulator. • Magnetoencephalography (MEG) • Positron Emission Tomography (PET) scans DBS 7424, 7426, 7428 Medical Environment Most routine diagnostic procedures, such as fluoroscopy and x-rays, are not expected to affect system operation. However, because of higher energy levels, sources such as transmitting antennas found on various diagnostic and therapeutic equipment may interfere with the Activa System. Refer to SCS/PNS guidelines. November 2005 Page 1 1 of 27 I nterStim Enterra 3023 7425G, 3116 Medical Environment Effects in Monitoring Devices When using diagnostic monitoring devices such as an electrocardiogram (ECG), Holter Monitor, electroencephalogram (EEG), or implantable heart monitor, pulses from the neurostimulation system may be detected as an electrical signal. When evaluating the diagnostic information, be sure to identify the neurostimulator pulses as intrinsic. Effects from Diagnostic Procedures Most routine diagnostic procedures, such as fluoroscopy and x-rays, are not expected to affect neurostimulation system operation. However, because of higher energy levels, sources such as transmitting antennas may interfere with the system. Effects from Diagnostic Procedures Ultrasound Scanning –Ultrasonic scanning equipment may cause mechanical damage to an implanted neurostimulator or implanted lead if used directly over the neurostimulator or lead implant site. Refer to SCS/PNS guidelines. Diagnostic Ultrasound (e.g., carotid scan, doppler studies) – An implanted neurostimulation system is unlikely to interfere with diagnostic ultrasound. To minimize potential image distortion, turn OFF the neurostimulator and keep the transducer 15 cm (6 in) away from the neurostimulation system. Medical and Hospital Environment Safe from Electromagnetic Interference The following medical procedures are not likely to affect the implanted system: • Computerized Axial Tomography (CT or CAT) Scans • Diagnostic X-rays / Fluoroscopy Note: Tight pressure can affect the system. Refer to “X-rays Requiring Tight Enclosure” on page 46. • Magnetoencephalography (MEG) • Positron Emission Tomography (PET) Scans X-rays Requiring Tight Enclosure – Pressing the neurostimulator too tightly during X-ray procedures that require enclosure of the implant area may damage the neurostimulator or disconnect the neurostimulation system components, which may require surgery to fix the system or replace components. Adjust the X-ray equipment to limit the amount of pressure exerted on the neurostimulator during procedures that require enclosure of the implant area. Refer to SCS/PNS guidelines. Patient activities Unexpected changes in stimulation – Electromagnetic interference, postural changes, and other activities may cause a perceived increase in stimulation, which some patients have described as Patient Activities/Environmental Precautions – Patients should exercise reasonable caution in avoidance of devices which generate a strong electric or magnetic field. Close proximity to high levels of electromagnetic interference Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. Neurological Patient and Technical Services Department SCS/PNS 7425, 7427, 7427V, 7479, 7479B, 37711 DBS 7424, 7426, 7428 November 2005 Page 12 of 27 InterStim 3023 Enterra 7425G, 3116 uncomfortable stimulation (jolting or shocking sensation); therefore, patients should reduce the amplitude to the lowest setting and turn OFF the neurostimulator before engaging in activities that could be unsafe for themselves or others if they received an unexpected jolt or shock (eg, driving, operating power tools). Patients should discuss these activities with their physician. (EM I) may cause a neurostimulator to switch On or Off. The system also may unexpectedly cease to function due to battery depletion or other causes. For these reasons, the patient should be advised about any activities that would be potentially unsafe if their symptoms unexpectedly return. For additional information about devices which generate electromagnetic interference, call Medtronic at 1-800-328-0810. Psychotherapeutic procedures – Safety has not been established for psychotherapeutic procedures using equipment that generates electromagnetic interference (eg, electroconvulsive therapy, transcranial magnetic stimulation) in patients who have an implanted neurostimulation system. Induced electrical currents may cause heating, especially at the lead electrode site, resulting in tissue damage. Psychotherapeutic Procedures – The safety of psychotherapeutic procedures using equipment that generates electromagnetic interference (e.g., electroshock therapy, transcranial magnetic stimulation) has not been established. Radiation therapy – Do not direct high radiation sources such as cobalt 60 or gamma radiation at the neurostimulation system. If radiation therapy is required near the neurostimulation system, place lead shielding over the device to help prevent radiation damage. High Radiation Sources – High radiation sources, such as cobalt 60 or gamma radiation, should not be directed at the neurostimulator. If a patient requires radiation therapy in the vicinity of the neurostimulator, place lead shielding over the device to prevent radiation damage. High Radiation Sources –High-radiation sources, such as cobalt 60 or gamma radiation, should not be directed at the neurostimulator. If a patient requires radiation therapy in the vicinity of the neurostimulator, place lead shielding over the device to prevent radiation damage. Radiation Therapy – Do not direct high radiation sources such as cobalt 60 or gamma radiation at the neurostimulation system. If radiation therapy is required in the vicinity of the neurostimulation system, place lead shielding over the device to prevent radiation damage. Radio-frequency (RF) or microwave ablation – Safety has not been established for RF or microwave ablation in patients who have an implanted neurostimulation system. Induced electrical currents may cause heating, especially at the lead electrode site, resulting in tissue damage. Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. Radiofrequency (RF) / Microwave Ablation – Safety has not been established for radiofrequency (RF) or microwave ablation in patients with an implanted neurostimulation system. Induced electrical currents from these procedures to the neurostimulation system may cause heating, especially at the lead electrode site, resulting in tissue damage. Refer to SCS/PNS guidelines. Psychotherapeutic Procedures – The safety of psychotherapeutic procedures using equipment that generates electromagnetic interference (e.g., electroshock therapy, transcranial magnetic stimulation) has not been established in patients with an implanted neurostimulation system. Refer to SCS/PNS guidelines. Neurological Patient and Technical Services Department SCS/PNS 7425, 7427, 7427V, 7479, 7479B, 37711 November 2005 Page 13 of 27 DBS InterStim Enterra 7424, 7426, 7428 3023 7425G, 3116 Scuba diving or hyperbaric chambers – Patients should not dive below 10 meters (33 feet) of water or enter hyperbaric chambers above 2.0 atmospheres absolute (ATA). Pressures below 10 meters (33 feet) of water (or above 2.0 ATA) could damage the neurostimulation system. Before diving or using a hyperbaric chamber, patients should discuss the effects of high pressure with their physician. Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. Skydiving, skiing, or hiking in the mountains – High altitudes should not affect the neurostimulator; however, the patient should consider the movements involved in any planned activity and take precaution to avoid putting undue stress on the implanted system. Patients should be aware that during skydiving, the sudden jerking that occurs when the parachute opens may cause lead dislodgment or fractures, which may require surgery to repair or replace the lead. Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. Transcutaneous electrical nerve stimulation (TENS) – Do not place TENS electrodes so that the TENS current passes over any part of the neurostimulation system. If patients feel that the TENS may be interfering with the implanted neurostimulator, patients should discontinue using the TENS until they talk with their doctor. Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. TENS (Transcutaneous External Neurostimulator) – Do not place TENS electrodes so that the TENS current passes over any part of the neurostimulator / lead system. If the patient feels that the TENS unit may be interfering with the implantable neurostimulator, the patient should talk with their doctor. Theft detectors and security screening devices – Advise patients to use care when approaching theft detector and security screening devices (such as those found in airports, libraries, and some department stores). When approaching Theft Detectors and Screening Devices Patients should be advised to use care when approaching security arches or gates (such as those found in airports, libraries, and some department stores) because these devices can turn on or turn Theft Detectors and Security Screening Devices – Patient should be advised to use care when approaching theft detector and security screening devices (such as those found in airports, libraries, and some Theft Detectors and Security Screening Devices Patients should be advised to use care when approaching theft detector and security screening devices (such as those found in airports, libraries, and some Neurological Patient and Technical Services Department SCS/PNS 7425, 7427, 7427V, 7479, 7479B, 37711 these devices, patients should do the following: 1. If possible, request to bypass these devices. Patients should show the security personnel their patient identification card for the neurostimulator and request a manual search. Security personnel may use a handheld security wand but patients should ask the security personnel not to hold the security wand near the neurostimulator any longer than is absolutely necessary. Patients may wish to ask for another form of personal search. 2. If patients must pass through the theft detector or security screening device, they should turn OFF their neurostimulator, approach the center of the device and walk through normally. a. If two security gates are present, they should walk through the middle, keeping as far from each gate as possible. b. If one gate is present, they should walk as far from it as possible. Note: Some theft detectors may not be visible. 3. Proceed through the security device. Patients should not linger near or lean on the screening device. 4. After patients pass through the security device, they should turn ON their neurostimulator. Therapeutic magnets (eg, magnetic mattresses, blankets, wrist wraps, elbow wraps) – Keep the magnet at least 25 cm (10 in) away from the neurostimulator. Magnetic fields of 10 gauss or less will generally not affect the neurostimulator. November 2005 Page 14 of 27 I nterStim Enterra 7424, 7426, 7428 3023 7425G, 3116 off their neurostimulator. If an airport security wand is used, they should ask the security personnel to avoid placing the wand over the neurostimulator. When approaching these devices, patients should do the following: 1. If security personnel are present, show them the neurostimulator identification card and request a hand search. 2. If patients must pass through the security device, they should approach the center of the device and walk normally. a. If two security gates are present, they should walk through the middle, keeping as far away as possible from each gate. b. If one gate is present, they should walk as far away as possible from it. Note: Some theft detectors may not be visible. 3. Proceed through the security arch or gate. Do not touch, lean on or linger near the security arch or gate. 4. If patients suspect that their neurostimulator was turned off, they should make sure someone is able to turn on the system again. (This person could be the patient, if his or her medical condition allows it. It could also be a family member or clinician who has been taught how to use the system.) department stores). When approaching these devices, patients should do the following: 1. If possible, request to bypass these devices. The patient should show security personnel their patient identification card for the neurostimulator and request a manual search. Security personnel may use a handheld security wand but the patient should ask the security personnel not to hold the security wand near the neurostimulator any longer than is absolutely necessary. The patient may wish to ask for another form of personal search. 2. If patients must pass through the theft detector or security screening device, they should turn their neurostimulator off, approach the center of the device and walk through normally. a. If two security gates are present, they should walk through the middle, keeping as far away as possible from each gate. b. If one gate is present, they should walk as far away as possible from it. Note: Some theft detectors may not be visible. 3. Proceed through the security device. Do not linger near or lean on the screening device. 4. After patients pass through the security device, they should turn their neurostimulator on again. DBS Refer to SCS/PNS guidelines. Therapeutic Magnets – Therapeutic magnets (for example, those found in bracelets, back braces, shoe inserts and mattress pads) can cause inadvertent on or off activations of the neurostimulator. Therefore, patients should be advised not to use them. Refer to SCS/PNS guidelines. department stores). When approaching these devices, patients should do the following: 1. If possible, request to bypass these devices. The patient should show the security personnel their patient identification card for the neurostimulator and request a manual search. Security personnel may use a handheld security wand but the patient should ask the security personnel not to hold the security wand near the neurostimulator any longer than is absolutely necessary. The patient may wish to ask for another form of personal search. 2. If patients must pass through the theft detector or security screening device, they should approach the center of the device and walk through normally. a. If two security gates are present, they should walk through the middle, keeping as far away as possible from each gate. b. If one gate is present, they should walk as far away as possible from it. Note: Some theft detectors may not be visible. 3. Proceed through the security device. Do not linger near or lean on the screening device. Therapeutic Magnets – If a neurostimulator is enabled for magnet use, therapeutic magnets (e.g., magnetic mattresses, blankets, wrist wraps, elbow wraps, etc.), can inadvertently turn the neurostimulator ON or OFF. Advise patients to keep their therapeutic magnets at least 10 in (25 cm) away from their neurostimulator. Magnetic fields of 10 Neurological Patient and Technical Services Department SCS/PNS 7425, 7427, 7427V, 7479, 7479B, 37711 DBS 7424, 7426, 7428 November 2005 Page 15 of 27 InterStim 3023 Patient control devices may affect other implanted devices – Radio signals (telemetry) from the patient programmer may interfere with the performance of other implantable devices. Enterra 7425G, 3116 gauss or less will generally not affect the neurostimulator. Patient control devices may affect other implanted devices – Do not place the patient control device (ie, patient programmer, control magnet, radiofrequency transmitter) over another active implanted medical device (eg, pacemaker, defibrillator, another neurostimulator). The patient control devices could unintentionally change the operation of the other device. Refer to SCS/PNS guidelines. Patient magnet may damage items – Do not place the patient magnet on or near computer monitors, magnetic storage disks or tapes, televisions, credit cards, or other items affected by strong magnetic fields. If the patient magnet is too close, these items may be damaged. Patient Magnet – The magnet provided to the patient for device activation and deactivation may damage televisions, computer disks, computer monitors, credit cards, and other items affected by strong magnetic fields. Control Magnet –The magnet provided to the patient for device activation and deactivation may damage televisions, computer disks, credit cards, and other items affected by strong magnetic fields. N/A Programmer interaction with a cochlear implant – When the patient has a cochlear implant, minimize or eliminate the potential for unintended audible clicks during telemetry by keeping the external portion of the cochlear system as far from the programming head as possible or by turning OFF the cochlear implant during programming. Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. Programmer interaction with other active implanted devices – When a patient has a neurostimulator and another active implanted device (eg, pacemaker, defibrillator, neurostimulator): • the radio-frequency (RF) signal used to program these devices may reset or reprogram the other device. Inadvertent Programming – If more than one neurostimulator is implanted, then the potential for unintentional programming changes to the other neurostimulator exists. If two neurostimulators are implanted, they must be implanted at least 8 inches apart to minimize interference. Verify final programmed Refer to SCS/PNS guidelines. Inadvertent Programming – To prevent inadvertent programming or interference with other implanted devices, follow these guidelines: • If more than one neurostimulator is implanted, then the potential for unintentional programming changes to the other neurostimulator exists. If Refer to SCS/PNS guidelines. Patient control devices may affect other implanted devices – Do not place the patient control device (ie, patient programmer, control magnet, radiofrequency transmitter) over another active implanted medical device (eg, pacemaker, defibrillator, another neurostimulator). The patient control devices could unintentionally change the operation of the other device. Neurological Patient and Technical Services Department SCS/PNS 7425, 7427, 7427V, 7479, 7479B, 37711 November 2005 Page 16 of 27 DBS I nterStim 7424, 7426, 7428 3023 • the magnet in a cardiac programmer parameters by reviewing both devices at may activate magnetically controlled the conclusion of any programming functions of the neurostimulator. session. To verify that inadvertent programming did not occur, clinicians familiar with each Refer to SCS/PNS guidelines. device should check the programmed parameters of each device before the patient is discharged from the hospital and after each programming session of either device (or as soon as possible after these times). Also, inform patients to contact their physician immediately if they experience symptoms that could be related to either device or to the medical condition treated by either device. Telemetry signal disruption from EMI – Refer to SCS/PNS guidelines. Do not attempt telemetry near equipment that may generate electromagnetic interference (EMI). If EMI disrupts programming, move the programmer away from the likely source of EMI. Examples of sources of EMI are magnetic resonance imaging (MRI), lithotripsy, computer monitors, cellular telephones, x - ray equipment, and other monitoring equipment. Enterra 7425G, 3116 two neurostimulators are implanted, they must be implanted at least 20 cm (8 in) apart to minimize interference. Verify final programmed parameters by reviewing both devices at the conclusion of any programming session. Refer to SCS/PNS guidelines. Refer to SCS/PNS guidelines. Neurological Patient and Technical Services Department Device or procedure November 2005 Page 17 of 27 Table 1. Potential Effects of EMI from Devices or Procedures Device Serious Momentary damage patient increase in injury stimulation Bone growth stimulators Defibrillation/cardioversion 3 Dental drills and ultrasonic probes Diathermy, therapeutic 3 Electrocautery 3 Electrolysis 3 3 3 3 Laser procedures 31 3 3 3 3 3 3 3 Radiation therapy Radio-frequency (RF)/microwave ablation Theft detector Transcutaneous electrical nerve stimulation (TENS) 1. DBS Therapy only. 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 Therapeutic magnets Therapeutic ultrasound Intermittent Stimulation 3 Household items Magnetic Resonance Imaging (MRI) Psychotherapeutic procedures (TMS, ECT) 3 3 3 3 Electromagnetic field devices: (eg, arc welding, power stations) High-output ultrasonics/lithotripsy Device turns OFF/ON 3 3 3 3 3 3 Neurological Patient and Technical Services Department November 2005 Page 18 of 27 FROM KINETRA DEEP BRAIN STIMULATION APPENDIX, MAGNETIC RESONANCE IMAGING The Effects of Magnetic Resonance Imaging (MRI) on Deep Brain Stimulation System (Activa) for Movement Disorders Models: Kinetra®: 7428 SoletraTM: 7426 Itrel® II: 7424 MRI and Activa Therapy Introduction It is important to read this section in its entirety before conducting an MRI examination on a patient with any implanted Activa System component. Contact Medtronic Technical Services at 1-800-328-0810 if you have any questions. Due to the number and variability of parameters that affect MRI compatibility, the safety of patients or continued functioning of Activa Systems exposed to MRI cannot be absolutely ensured. MRI systems generate powerful electromagnetic fields that can produce a number of interactions with implanted components of the Activa neurostimulation system. Some of these interactions, especially heating, are potentially hazardous and can lead to serious injury or death. However, with appropriate control measures, particularly with respect to the selection of MRI parameters and RF coils, it is generally possible to safely perform an MRI head scan on an Activa patient. In addition, Activa System components can affect the MRI image, potentially impacting the diagnostic use of this modality. The following information describes the potential interactions and control measures that should be taken to minimize the risks from these interactions. Contraindication: Implantation of an Activa Brain Stimulation System is contraindicated for patients who will be exposed to Magnetic Resonance Imaging (MRI) using a full body transmit radio-frequency (RF) coil, a receive only head coil, or a head transmit coil that extends over the chest area. Performing MRI with this equipment can cause tissue lesions from component heating, especially at the lead electrodes, resulting in serious and permanent injury including coma, paralysis, or death. Warnings: • Do not conduct an MRI examination on a patient with any implanted Activa System component until you read and fully understand all the information in this section. Failure to follow all warnings and guidelines related to MRI can result in serious and permanent injury including coma, paralysis, or death. • In-vitro testing has shown that exposure of the Activa neurostimulator system to MRI at parameters other than those described in this guideline can induce significant heating at the lead electrodes or at breaks in the lead. Excessive heating may occur even if the lead and/or extension are the only part of the Activa System that is implanted. Excessive heating can result in serious and permanent injury including coma, paralysis, or death. • MRI examinations of patients with an implanted Activa System should only be done if absolutely needed and then only if these guidelines are followed. MRI should not be considered for Activa patients if other potentially safer diagnostic methods such as CT, X-ray, ultrasound, or other methods will provide adequate diagnostic information. • A responsible individual with expert knowledge about MRI, such as an MRI radiologist or MRI physicist, must assure all procedures in this guidelines are followed and that the MRI scan parameters, especially RF specific absorption rate (SAR) and gradient dB/dt parameters, comply Neurological Patient and Technical Services Department • November 2005 Page 19 of 27 with the recommended settings, both for the pre-scan (tuning) and during the actual MRI examination. The responsible individual must verify that parameters entered into the MRI system meet the guidelines in this section. Do not conduct an MRI examination if the patient has any other implants or limiting factors that would prohibit or contraindicate an MRI examination. Cautions: • The neurostimulator, especially those without filtered feedthroughs such as the Itrel II Model 7424, may be reset or potentially damaged when subjected to an MRI examination. If reset, the neurostimulator must be reprogrammed. If damaged, the neurostimulator must be replaced. • MRI images may be severely distorted or image target areas can be completely blocked from view near the implanted Activa System components, especially near the neurostimulator. If the MRI targeted image area is near the neurostimulator, it may be necessary to move the neurostimulator to obtain an image, or use alternate imaging techniques. Do not remove the neurostimulator and leave the lead system implanted as this can result in higher than expected lead heating. • Carefully weigh any decision to perform magnetic resonance imaging (MRI) examinations on patients who require the neurostimulator to control tremor. Image quality during MRI examinations may be reduced, because the tremor may return when the neurostimulator is turned off. • If possible, do not sedate the patient so that the patient can provide feedback of any problems during the examination. • Monitor the patient during the MRI examination. Verify that the patient is feeling normal and is responsive between each individual scan sequence of the MRI examination. Discontinue the MRI immediately if the patient becomes unresponsive to questions or experiences any heating, pain, shocking sensations/uncomfortable stimulation, or unusual sensations. Note: The MRI guidelines provided here may significantly extend the MRI examination time or prevent some types of MRI examinations from being conducted on Activa patients. General Information on MRI An MRI system produces three types of electromagnetic fields that may interact with implanted neurostimulation systems. All three of these fields are necessary to produce an MRI image. Each of these fields can also produce specific but different types of interactions with implanted neurostimulator systems. These fields include: • Static Magnetic Field. This is a steady state non-varying magnetic field that is normally always ON, even when no scan is underway. In a 1.5 Tesla MRI system, the static magnetic field is approximately 30,000 times greater than the magnetic field of the earth. • Gradient Magnetic Field. This is a low-frequency pulsed magnetic field that is only present during a scan. The gradient magnetic field can induce voltages onto the lead system that may result in unintended stimulation or functional interactions with the neurostimulator. • RF Field. This is a pulsed radio frequency (RF) field that is only present during a scan. It can be produced by a variety of transmission RF coils such as a whole body transmit coil or an extremity coil such as a transmit/receive head coil. Only a transmit/receive head coil should be used as the other RF coils can expose more of the lead system to RF energy, thereby increasing the risk of excessive heating and thermal lesions possibly resulting in coma, paralysis, or death. MRI Interactions with Implanted Activa Systems MRI/neurostimulation system interactions are various, and the risk to the patient can range from minimal to severe. These interactions include the following: Heating – The MRI RF field induces voltages onto the lead system that can produce significant heating effects at the lead electrode-tissue interface or at the location of any breaks in the neurostimulator lead system. Component heating from the MRI RF field is the most serious risk from MRI exposure. Failure to follow these MRI recommendations can result in thermal lesions possibly resulting in coma, paralysis, or death. Neurological Patient and Technical Services Department November 2005 Page 20 of 27 Magnetic Field Interactions – Magnetic field interactions such as force and torque effects are produced by the static magnetic field. Any magnetic material will be attracted to the static magnetic field of the MRI. The force and torque effects may produce movement of the neurostimulator that can be uncomfortable to the patient, open a recent incision, or both. Activa System components are designed with minimal magnetic materials. Induced Stimulation – Gradient magnetic fields may induce voltages onto the lead system that may cause unintended stimulation. The voltage of the induced stimulation pulses is proportional to the time rate of change (dB/dt) of the gradient pulses, the effective loop area created by the neurostimulator lead system, and the location of the lead system with respect to the gradient coils of the MRI. Effects on Neurostimulator Function – The static, gradient, and RF fields of the MRI may affect the neurostimulator operation and programming. The static magnetic field may cause the neurostimulator to turn ON or OFF if the neurostimulator uses a magnetically controlled switch that allows the patient to control stimulation by the application of a handheld magnet. Additionally, the MRI RF, static, and gradient fields may temporarily affect or disable other functions, such as telemetry or stimulation pulses. Parameters will need to be reprogrammed if the MRI causes a POR (Power On Reset) of the neurostimulator. Image Artifacts and Distortion – The neurostimulation system components, particularly the neurostimulator, can cause significant imaging artifacts and/or distortion of the MRI image, particularly if the neurostimulator components contain magnetic material. The neurostimulator can cause the MRI image to be completely blocked from view (i.e., signal loss or signal "void") or severely distorted within several inches of the neurostimulator. MRI Procedure Scope These MRI/neurostimulator exposure guidelines apply to Activa Systems comprising combinations of the following components: • Neurostimulator Models: Itrel II 7424, Soletra 7426, Kinetra 7428 • Lead Extension Models: 7495, 7482 • Lead Models: DBS 3387, 3389 Supervision A responsible individual such as an MRI radiologist or MRI physicist must assure these procedures are followed. If the MRI is operated by an MRI technician, it is strongly recommended the responsible individual verifies that the MRI recommendations are followed. Preparation Do the following prior to performing an MRI examination on an Activa patient: 1. Inform the patient of the risks of undergoing an MRI. 2. Check if the patient has any other implants or conditions that would prohibit or contraindicate an MRI examination. Do not conduct an MRI examination if any are found. 3. Verify that all proposed MRI examination parameters comply with the “MRI Operation Settings” on Table 12. If not, the parameters must be modified to meet these requirements. If this cannot be done, do not perform an MRI. 4. If the patient has implanted leads but does not have an implanted neurostimulator, perform the following steps: a. Wrap the external portion of the leads/percutaneous extensions with insulating material. b. Keep the external portion of the leads/percutaneous extensions out of contact with the patient. c. Keep the external leads/percutaneous extensions straight, with no loops, and running down the center of the head coil. 5. If the patient has an implanted neurostimulator, perform the following steps: a. Review the neurostimulator with a clinician programmer and print out a copy of the programmed parameters for reference. b. Test for possible open circuits by measuring impedance and battery current on all electrodes in unipolar mode (see Table 11). If an open circuit is suspected, obtain an x-ray to identify whether the open circuit is caused by a broken lead wire. If a broken lead wire is found, do not perform an MRI. Neurological Patient and Technical Services Department November 2005 Page 2 1 of 27 Table 11. Measurement Values Indicating Possible Open Circuits Neurostimulator Impedance Battery Current Itrel II Model 7424 >2000 : <10 µA Soletra Model 7426 >2000 : <10 µA Kinetra Model 7428 >4000 : <15 µA Warning: An MRI procedure should not be performed in a patient with an Activa System that has a broken lead wire because higher than normal heating may occur at the break or the lead electrodes which can cause thermal lesions. These lesions may result in coma, paralysis, or death. c. If the Activa System is functioning properly and no broken lead wires are found, program the neurostimulator to the settings provided in Table 12. Table 12. Recommended Neurostimulator Settings for MRI Parameter Stimulation output Setting OFF (all programs) Stimulation mode Bipolar (all programs) Amplitude 0 Volts (all programs) Magnetic (reed) switch Disabled (Kinetra Model 7428 only) Other parameters Do not change MRI Operation Settings Prior to the MRI examination, a responsible individual such as an MRI radiologist or MRI physicist must assure the examination will be conducted according to the following MRI requirements. If standard MRI pulse sequences will be used, they must meet these requirements. If they do not, the pulse parameters must be adjusted so that they comply with these requirements: Warning: In-vitro testing has shown that exposure of the Activa System to MRI under conditions other than described in this guideline can induce excessive heating at the lead electrodes or at breaks in the lead to cause lesions. These lesions may result in coma, paralysis, or death. • • Use only a 1.5 Tesla horizontal bore MRI (do not use open sided or other field strength MRI systems). Use only a transmit/receive head coil. • • Contraindication: Implantation of an Activa Brain Stimulation System is contraindicated for patients who will be exposed to Magnetic Resonance Imaging (MRI) using a full body transmit radio-frequency (RF) coil, a receive-only head coil, or a head transmit coil that extends over the chest area. Performing MRI resulting in serious and permanent injury including coma, paralysis, or death. Enter the correct patient weight into the MRI console to assure the head SAR is estimated correctly. Use MRI examination parameters that limit the applied head SAR to 1/10 (0.1) W/kg or less for all RF pulse sequences. Warnings • Ensure the SAR value is the value for head SAR. Some MRI systems may only display SAR, whole body SAR, or local body SAR. Make sure the value being limited is for head SAR. Excessive heating may occur if the wrong SAR value is used. Neurological Patient and Technical Services Department • November 2005 Page 22 of 27 If MRI parameters must be manually adjusted after the initial automatic MRI prescan, do not make any adjustments that will increase the SAR value. Some MRI machines may not automatically update the displayed SAR value if manual adjustments are made. This may lead to higher than expected temperature increases in the Activa System, particularly at the lead electrodes. • Limit the gradient dB/dt field to 20 Tesla/second or less. Note: The recommendations provided are based on in-vitro testing and should result in a safe MRI examination of a patient with an implanted Medtronic Activa System. However, due to the many variables that affect safety, Medtronic cannot absolutely ensure safety or that the neurostimulator will not be damaged. The user of this information assumes full responsibility for the consequences of conducting an MRI examination on a patient with an implanted Activa System. Prior to the MRI Examination Prior to the scan examination, the responsible individual must verify the MRI examination parameters comply with these guidelines. • Patients with implanted Activa Systems should be informed of the risks of undergoing an MRI. • If possible, do not use sedation so the patient can inform the MRI operator of any heating, discomfort, or other problems. • Instruct the patient to immediately inform the MRI operator if any discomfort, stimulation, shocking, or heating occurs during the examination. During the MRI Examination • • Monitor the patient both visually and audibly. Check the patient between each imaging sequence. Discontinue the MRI examination immediately if the patient is unable to respond to questions or reports any problems. Conduct the examination using only the MRI pulse sequence that the MRI radiologist or physicist has confirmed meets the MRI requirements above. Post MRI Examination Review • Verify that the patient is feeling normal. • Verify that the neurostimulator is functional. • Reprogram the neurostimulator to pre-MRI settings. November 2005 Page 23 of 27 Neurological Patient and Technical Services Department FROM MED TRONIC PAIN THERAPY, INFORMATION FOR PRESCRIBERS, APPENDIX B MRI and neurostimulation therapy for chronic pain Introduction Medtronic recommends that you do not conduct an MRI examination of any part of the body on a patient using a radio-frequency (RF) transmit body coil. If all of the instructions stated in this Appendix B are followed, MRI examinations of the head only using an RF transmit/receive head coil may be safely performed. It is important to read this information in its entirety before conducting an MRI examination on a patient with any implanted component of a Medtronic neurostimulation system for chronic pain. These instructions do not apply to other implantable products or other devices, products, or items. Contact Medtronic at 800328-0810 if you have any questions. Due to the number and variability of parameters that affect MRI compatibility, the safety of patients or continued functioning of neurostimulation systems exposed to MRI cannot be absolutely ensured. MRI systems generate powerful electromagnetic fields that can produce a number of interactions with implanted components of the neurostimulation system. Some of these interactions, especially heating, are potentially hazardous and can lead to serious injury or death. However, when all instructions stated in this Appendix B are followed, MRI examinations of the head only may be safely performed. In addition, neurostimulation system components can affect the MRI image, potentially impacting the diagnostic use of this modality. The following information describes the potential interactions and control measures that should be taken to minimize the risks from these interactions. The instructions in this Appendix B describe how to conduct a head-only MRI examination of a patient with a neurostimulation system implanted for chronic pain, using a transmit/receive head coil of a 1 .5-Tesla horizontal bore MRI. MRI examinations of any other part of the body are not recommended, as these require the use of the MRI RF transmit body coil, which may produce hazardous temperatures at the location of the implanted lead electrodes. Warnings MRI RF transmit body coil – Medtronic recommends that you do not conduct an MRI examination using an RF transmit body coil on a patient with any implanted neurostimulation system component because the interaction of the MRI with the neurostimulation system may lead to serious injury or death. See the section “Risks associated with MRI examination” on page 24. MRI transmit/receive head coil – An MRI examination of the head only (no other part of the body) can be conducted safely using an RF transmit/receive head coil when all instructions in this Appendix B are followed. Limitations • MRI should not be considered for patients with neurostimulation systems if other potentially safer diagnostic methods such as CT, x-ray, ultrasound, or others will provide adequate diagnostic information. • • These instructions apply only to Medtronic neurostimulation therapies for chronic pain for approved indications. The instructions in this Appendix B apply to all Medtronic fully implantable neurostimulators, leads, and extensions used for chronic pain therapy. Note: The instructions contained in this Appendix B are not applicable to MRI examinations of patients with radiofrequency (RF) neurostimulators. Medtronic recommends physicians not prescribe MRI for a patient who has an implanted Itrel 3 Model 7425 Neurostimulator. The Itrel 3 Neurostimulator is highly susceptible to reset or damage when subjected to an MRI examination. If reset, the neurostimulator must be reprogrammed. If damaged, the neurostimulator must be replaced. The Itrel 3 Neurostimulator has an increased risk of induced electrical current, which may stimulate or shock the patient. Contact Medtronic at 800-328-0810 for information about newer models or any updates. Neurological Patient and Technical Services Department November 2005 Page 24 of 27 • The RF transmit/receive head coil must not cover any implanted system component. • If the patient has any other implants or products that prohibit or contraindicate an MRI examination, follow the instructions from the manufacturer. The instructions in this Appendix B apply only to the Medtronic products listed above. • Do not conduct an MRI examination if the patient’s neurostimulation system has a broken lead wire, because higher than normal heating may occur at the break or lead electrodes. Excessive heating can cause tissue damage and result in severe injury or death. • Physicians should not prescribe MRI for patients undergoing trial neurostimulation and having systems that are not fully implanted. • If the MRI targeted image area is near the neurostimulator, it may be necessary to move the neurostimulator to obtain an image, or use alternate imaging techniques. MRI images may be severely distorted or image target areas can be completely blocked from view near the implanted neurostimulation system components, especially near the neurostimulator. • Do not remove the neurostimulator and leave the lead system implanted as this can result in higher than expected lead heating. Excessive heating can cause tissue damage and result in severe injury or death. Risks associated with MRI examination – Exposing a patient with an implanted neurostimulation system or component to MRI may potentially injure the patient or damage the neurostimulator. The known potential risks are as follows: • Induced electrical currents from the MRI to the neurostimulation system or component may cause heating, especially at the lead-electrode site, resulting in tissue damage. Induced electrical currents may also stimulate or shock the patient. Note: This warning applies even if only a lead or extension is implanted. Factors that increase the risks of heating and patient injury include, but are not limited to, the following: • High MRI specific absorption rate (SAR) RF power levels • Low impedance leads or extensions (Medtronic product names or model numbers designated by a “Z,” an “LZ,” or “low impedance”) • MRI RF transmit/receive coil that is near or extends over the implanted lead system • Implanted lead systems with small surface area electrodes • Short distances between lead electrodes and heat-sensitive tissue • Exposure to gradients exceeding a dB/dt limit of 20 Tesla per second may result in overstimulation or shocking, particularly for unipolar-capable devices. • MRI may permanently damage the neurostimulator, requiring explant or replacement. • MRI may affect the operation of the neurostimulator. The MRI may also reset the parameters to power-on-reset settings, requiring reprogramming with the clinician programmer. • The Itrel 3 Model 7425 Neurostimulator is highly susceptible to reset or damage when subjected to an MRI examination. If reset, the neurostimulator must be reprogrammed. If damaged, the neurostimulator must be replaced. An Itrel 3 Neurostimulator also might exhibit unpredictable behavior if subjected to an MRI examination. The neurostimulator may move within the implant pocket and align itself with the MRI field, which may cause patient discomfort or a recent neurostimulator implant incision to open. Neurological Patient and Technical Services Department November 2005 Page 25 of 27 Cautions Patient interaction during MRI – If possible, do not sedate the patient so that the patient can provide feedback of any problems during the examination. Monitor the patient during the MRI examination. Verify that the patient is feeling normal and is responsive between each individual scan sequence of the MRI examination. Discontinue the MRI immediately if the patient becomes unresponsive to questions or experiences any heating, pain, shocking sensations/uncomfortable stimulation, or unusual sensations. MRI procedure using an RF transmit/receive head coil Supervision If all of the instructions stated in this Appendix B are followed, MRI examinations of the head using an RF transmit/receive head coil may be safely performed. Prior to the MRI examination, an individual with the proper knowledge of MRI equipment such as an MRI radiologist or MRI physicist must ensure the MRI examination will be conducted according to the information outlined in this Appendix B. Note: Due to the additional requirements in these instructions, MRI examination time may be significantly extended. MRI exposure requirements Prior to an MRI examination, determine whether the patient has multiple active medical device implants (such as deep-brain stimulation systems, implantable cardiac defribrillators, and others). The most restrictive MRI exposure requirements must be used if the patient has multiple active medical device implants. Contact the appropriate manufacturers of the devices if you have questions. If the following requirements cannot be met, do not proceed with the MRI examination. • Use only an RF transmit/receive head coil.* • Use only a 1 .5-Tesla horizontal bore MRI (do not use open-sided or other field strength MRI systems). • Enter the correct patient weight into the MRI console to ensure the head SAR is estimated correctly. The MRI scan sequences must meet the following requirements. If they do not, the pulse parameters must be adjusted so that they comply with these requirements. • Use MRI examination parameters that limit the head SAR to 1.5 W/kg or less for all RF pulse sequences. • Limit the gradient dB/dt field to 20 Tesla per second or less. *Important: If you are unsure if your MRI has RF transmit/receive head coil capability or if it displays “head SAR”, check with your MRI manufacturer. Note: The requirements provided are based on in-vitro testing and should result in a safe MRI examination of a patient with an implanted Medtronic neurostimulation system when all instructions in this Appendix B are followed. However, due to the many variables that affect safety, the safety of patients or continued functionality of neurostimulator systems exposed to MRI cannot be absolutely ensured. The user of this information assumes full responsibility for the consequences of conducting an MRI examination on a patient with an implanted neurostimulation system. Neurological Patient and Technical Services Department November 2005 Page 26 of 27 Preparation for the MRI examination Do the following prior to performing an MRI examination on a patient with an implanted neurostimulation component: 1. Inform the patient of all of the risks of undergoing an MRI examination as stated in this Appendix B. 2. If possible, do not use sedation so the patient can inform the MRI operator of any heating, discomfort, or other problems. 3. Instruct the patient to immediately inform the MRI operator if any discomfort, stimulation, shocking, or heating occurs during the examination. 4. Determine if the patient has any other implants or conditions that would prohibit or contraindicate an MRI examination. If you are unclear what implants may be present, perform an x-ray to determine implant type and location. Do not conduct an MRI examination if any conditions or implants that would prohibit or contraindicate an MRI are present. 5. Verify that all proposed MRI examination parameters comply with the “MRI exposure requirements” (refer to page 25). If not, the parameters must be modified to meet these requirements. If parameters cannot be modified, do not perform an MRI. 6. If the patient has implanted leads but does not have an implanted neurostimulator, perform the following steps: a. Wrap the external portion of the leads/percutaneous extensions with insulating material, such as dry gauze. b. Keep the external portion of the leads/percutaneous extensions out of contact with the patient. c. Keep the external leads/percutaneous extensions straight, with no loops, and running down the center of the head coil. 7. If the patient has an implanted neurostimulator, perform the following steps: a. Review the neurostimulator with a clinician programmer and print out a copy of the programmed parameters for reference. b. Test for possible open circuits by measuring impedance on all electrodes. An impedance measurement greater than 4000 ohms for the Synergy Plus+, Synergy Compact+, Synergy Versitrel, Synergy, or Itrel 3 Neurostimulator indicates a possible open circuit. An impedance measurement greater than 3600 ohms for the Restore Neurostimulator indicates a possible open circuit. c. If an open circuit is suspected, obtain an x-ray to identify whether the open circuit is caused by a broken lead wire. If a broken lead wire is found, do not perform an MRI examination. Warning: Do not conduct an MRI examination if the patient’s neurostimulation system has a broken lead wire, because higher than normal heating may occur at the break or lead electrodes. Excessive heating can cause thermal lesions and result in severe injury or death. 8. If the system is functioning properly and no broken lead wires are found, program the neurostimulator to the settings provided in Table 3. Parameters Table 3. Recommended neurostimulator settings for MRI examinations Settings Stimulation output OFF (all programs) Stimulation mode Bipolar (all programs) Amplitude 0 Volts (all programs) Magnetic (reed) switch Disabled (Itrel 3 Model 7425 only) Other parameters Do not change Neurological Patient and Technical Services Department November 2005 Page 27 of 27 During the MRI examination • Monitor the patient both visually and audibly. Check the patient between each imaging sequence. Discontinue the MRI examination immediately if the patient is unable to respond to questions or reports any problems. • Conduct the examination using only the MRI pulse sequence that the MRI radiologist or physicist has confirmed meets the “MRI exposure requirements” outlined in this Appendix B. Post-MRI examination review • Verify that the patient feels normal. • Verify that the neurostimulator is functional. • Reprogram the neurostimulator to pre-MRI settings.