Survey
* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project
Test stand and SRF cavity test plan at the FREIA Laboratory Han Li On behalf of FREIA team FREIA Laboratory, Uppsala University 13th of June 2017 What and Whom? Facility for Research Instrumentation and Accelerator Development State-of-the-art Equipment cryogenics - liquid helium - liquid nitrogen control room - equipment controls - data acquisition Competent and motivated staff collaboration with physics (IFA), engineering (Teknikum), TSL and Ångström workshop Funded by KAWS, Government, Uppsala Univ. Han Li 13th June 2017 vertical cryostat RF power sources 3 bunkers with test stands horizontal cryostat 2 Responsibility for ESS Accelerator FREIA has developed the prototype spoke RF source and validated the prototype spoke cavity and, will validate spoke cryomodule, prototype high-beta elliptical cavity with nominal RF power. 1) Validation of high power RF amplifier (HPA) 2) Validation of dressed spoke cavity 3) Validation of cavity package Prototype cavity package (in test cryostat) Requires the availability of one high power RF amplifier tested during phase 1 4) Validation of elliptical cavity package 5) Validation of spoke prototype cryomodule 6) Validation of prototype modulator and klystron 7) Acceptance testing of series spoke cryomodules Han Li 13th June 2017 3 The Test Stand Four main subsystems needed RF Power Station Courtesy of P. Duthil Cryogenics Superconducting cavity Cryostat Han Li 13th June 2017 LLRF 4 Cryogenics Helium liquefaction 150 l/h at 4.5K (LN2 pre-cooling) 2000 l LHe dewar/buffer, 3+1 outlets 100 m3 gasbag + recovery system cryostats connected in closed loop Liquid nitrogen 20 m3 LN2 tank Han Li 13th June 2017 5 HNOSS Horizontal Cryostat HNOSS: Horizontal Nugget for Operation of Superconducting Systems Main Vacuum Vessel Valve box (on top of main vessel) Total volume ca. 7 m3 Han Li 13th June 2017 LN2 and LHe Cold gas heater for return flow Distributes cryogens to HNOSS and CM Cryogenic transfer lines Distribute cryogens 4K and 2K pots, JT-valve, heat exchanger 5K supercritical helium Magnetic shield (room temperature) Thermal shield (LN2 temperature) Interconnection box (ICB) 3240 x ø1200mm inner volume “beam” axis at 1600mm Magnetic shield (room temperature) Thermal shield (LN2 temperature) re-heating from 2K to 300K Subatmospheric pumps with cooling capacity of 90W at 1.8K Control system 6 RF Distribution Main Hall Two 400kW RF Power Stations Main Bulk Outside Bunker Coaxial & Waveguide Lines Patch-Panels for Rearrangement Han Li 13th June 2017 Bunker 7 352MHz RF Stations Dual Tetrodes Per Station 400kW Peak at 14 Hz, 3.5ms Crowbar & Series Switches Two Stations Commissioned Station Efficiency Peak ~43% Operating Point Han Li 13th June 2017 8 Circulators Thermal Detuning Thermal Oscillations Han Li 13th June 2017 9 704 MHz RF station Prototype modulator, lend by ESS Ampegon, 330 kW (for 2 klystrons) Commissioning with two dummy loads, then attach klystron Prototype klystron, lend by ESS Toshiba type, 1.2 MW output power RF distribution, lend by ESS 704 MHz waveguides, circulator and loads Han Li 13th June 2017 10 Solid State Amplifier Single ended RF Power Amplifier (BLF188XR) tested in pulsed mode at 14 Hz and 3.5 ms Individually: delivering up to 1500 W • • • • amplifier slightly modified on RO3003 for series production Max efficiency: 67% (pulsed) Gain: 19 dB Independent measurements at NXP Two combined: Delivering up to 1700 W • • Max efficiency: 50% (1 dB IL in the output combiner) Working on improving values 100 kW radial power combiner output coupler connected to 3-1/8 inch coaxial line 30 mm 500 mm 7/16 inch input connector 1 kW level hot-S parameter measurements Han Li 13th June 2017 11 The Self-excited Loop Test Stand (I) FREIA developed a test stand based on SEL for demonstrating the performance of superconducting cavities both at low and high power, CW or pulse mode . FREIA SEL block diagram for high power test FREIA SEL block diagram for low power test Han Li 13th June 2017 12 The Self-excited Loop Test Stand (II) Combination with a DB RF station which can output up to 400 kW peak power. Developed digital phase shifter and gain-controller. Introduce interlock system for safety consideration. Introduce RF switch in order to manage a pulse operation mode. Developed SEL control and data acquisition system in LabView. Interlock and RF switch RF station SEL loop installed into a cabinet FPGA Han Li 13th June 2017 FREIA Labview SEL control system 13 LLRF system from Lund University Hardware Chassis: 12 slot µTCA, 1 kW power supply MCH: NAT ADC and FPGA: Struck 8300-L2 Vector modulator: DS8VM1 Timing: MRF timing receiver Useful features for the tests at FREIA Possibility to generate arbitrary RF pulses (no feedback) RF signals monitoring with possibility to generate an interlock signal Setting amplitude limit for the output RF signal Software Direct sampling (undersampling) of the RF signal Data acquisition and control algorithms done in the FPGA High level control system: EPICS User Interface: Control System Studio Machine Protection and Safety Systems Machine Protection Fast interlock: Compact Rio, NI cRIO-9024 Cuts off RF signal. Fast digital 150 ns , some 10 – 20 µs 16 analogue channels (100 kHz) programmable thresholds. 300000 data points, post mortem EPICS operator interface for set-up of RF interlocks. Slow interlock, S7-300 PLC system Currently about 100 i/o channels NI cRIO-9024 Fast interlock EPICS interface Radiation protection bunker Bunker with clearing system Door switch trigger PLC and hardware interlock to RF stations Radiation area monitors: Rotem Medismarts Easily integrated with EPICS FREIA bunker status display Dose rates Clearing status Warnings messages Graphs of dose rate Oxygen deficiency detector Tested Cavities Hélène Three spoke cavities , Hélène, Germaine and Romea from IPN Orsay, have been tested at low temperatures at FREIA. Parameter Helene (SSR) Frequency [MHz] 360.4 Germaine and Romea (DSR) 352.2 Beta_optimum 0.2 0.5 r/Q [Ohm] 117 427 Epk/Eacc 6.6 4.3 Bpk/Eacc [mT/MV/m] 13.4 89 6.8 130 Han Li 13th June 2017 Romea Germaine G [Ohm] 16 Typical Measurements Low power: RF behaviour on warm cavity before installation in HNOSS, Loaded Q-factor, eigen and external Q, Q0 = f(E) curve, Lorentz detuning and microphonics, Field emission onset and multipacting barriers, Sensitivity to helium pressure fluctuations, Residual resistance of cavity suface, Cryogenic heat loads. High power: RF behaviour during cool down, Coupler conditioning and cavity package conditioning, Achieve nominal gradient and nominal Q0, Cryogenic heat loads, Loaded Q-factor, eigen and external Q, Q0 = f(E) curve, Dynamic Lorentz detuning and mechanical modes, Field emission onset and multipacting barriers, Sensitivity to helium pressure fluctuations, Tuning sensitiviy. Han Li 13th June 2017 17 Coupler conditioning (I) The warm and first cold RF processing was done using IPN Orsay’s system, Followed by the new FREIA conditioning system to verify its performance, All processes used a traditional signal generator driven loop, The warm RF processing procedure took about 40 hours, while the cold processing took roughly 10 hours, During warm conditioning, lots of outgassing occurred through the forward power region of 40-50 kW at short pulses, 120 kW forward power was reached with 2.86 ms pulse duration at 14 Hz. Forward power (W) Time Han Li 13th June 2017 One phase example of FPC warm conditioning 18 Coupler conditioning (II) An automatic conditioning system, which consists of an acquisition system, a control system based on LabView software and feedback was developed at FREIA. Key paremeters are primarily set. Parameter value The main devices for the RF conditioning process are: 1 Loop control time (s) Signal Generator Power Meter Vacuum Gauge Controller (VGC) Cold Cathode Gauges (CCG) Arc Detector Electron Detector Fast RF Interlock Switch Vacuum Pumping Cart Han Li 13th June 2017 Pulse repeat rate (Hz) Vacuum upper limit (mbar) 14 1e-6 Vacuum lower limit (mbar) Initial pulse length (µs) pulse length step 5e-7 20 20 µs, 50 µs, 100µs, 200 µs, 500 µs, 1 ms, 1.5 ms, 2 ms, 2.5 ms, 2.86 ms 19 Cool down The cavity package in HNOSS cooled down from room temperature to 4K within 30 minutes, From 4K to 2K it took an extra 20 minutes, Cavity frequency was checking during cooldown. 200K to 4K 4K to 2K Temp. 300 K 4K 2K Resonance frequency 351.8939± 0.0001MHz 352.4004 ± 0.0001MHz 352.3725 ± 0.0001MHz 4.48 K/min 3.25 K/min Han Li 13th June 2017 4.10 K/min 20 Cavity Package conditioning Cavity package conditioning was done by FREIA pulse SEL, 2.86 ms pulse with 14Hz repetition rate was used, Cavity package conditioning took about 30 hours, Three major multipacting regions have been found Forward power from 22 to 30 kW, peak gradient from 4.5 to 4.8 MV/m. Forward power form 35 to 48kW, peak gradient from 5.2 to 5.7 MV/m. Forward power form 67 to 76 kW, peak gradient from 7 to 7.5 MV/m. Profile during MP Han Li 13th June 2017 The profile of cavity voltage changes. High radiation. Accompany with vacuum spikes. Coupler temperature increase. 21 Static Heat Loads 20 mbar • • The cooling of the FPC was checked with Cernox sensors TT303 (inlet ScHe), TT305 (outlet ScHe) and a Pt100 sensor (TT147) connected to the midsection of the FPC. P cavity [mbar] Eacc_peak [MV/m] TT303 [K] TT147 [K] TT305 [K] 20 0 14 39.5 291 20 9 9.8 41.6 278 TT303 (inside HNOSS) TT305 (outside HNOSS) At 20 mbar operating pressure, the static heat loads for Romea with TT147 at ca 41 K amount to ca. 6.6 W FT551 TT147 Identified four (4) regions depending on level FT551 [m3/h] 7.12 6.81 6.54 6.11 Std dev [m3/h] 0.34 0.31 0.33 0.33 LT101 min [%] max [%] 76 80 72 76 69 72 60 69 Note: tests where TT147 was varied between 30K and 90K give an extra heat load of only 1W at 20 mbar. Han Li 13th June 2017 22 Q0 measurement (I) Calorimetrical measurement of Q0 The level in the 2K tank was kept between 60% and 80% Close inlet and outlet valves of the cryomodule Record the pressure as a function of time for three (3) minutes Apply a known amount of resistive heat to the helium again record the pressure a function of time LHe inlet LHe outlet Cavity & 2K tank Heater Han Li 13th June 2017 Pressure curve vs Applied Power 23 Q0 measurement (II) Calorimetrical measurement of Q0 cont. Build the calibration curve: the rate of pressure rise vs. heat Load apply RF to the cavity and the system was left to stabilise only in pressure, record the pressure rise Calculate the dynamic RF load using the calibration curve preliminary data Pm = m*W+c m = 0.0017 c = 0.015 Han Li 13th June 2017 Dissipate power calibration curve Eacc_paek (MV/m) pressure rise slop dynamic RF load (W) 2,004499 2,515368 2,699868 3,02766 3,287095 3,591016 4,008444 0,0152 0,0153 0,0156 0,0158 0,0161 0,0162 0,0187 0,117647 0,176471 0,352941 0,470588 0,647059 0,705882 2,176471 4,317015 5,321938 5,966339 6,301396 6,589388 7,012012 7,360766 7,697375 8,07178 8,31318 8,659001 9,097229 0,0227 0,0359 0,0191 0,0185 0,0195 0,0245 0,0413 0,0432 0,0237 0,0252 0,028 0,0332 4,529412 12,29412 2,411765 2,058824 2,647059 5,588235 15,47059 16,58824 5,117647 6 7,647059 10,70588 24 Double Spoke Cavity Performance (Romea, 2017) Test of cavity package. preliminary data ESS goal : 1.5×109 Han Li 13th June 2017 Low-field Q0 factor of 1.4×109@ 2 K Q factor of 2,8×108@ Eacc_peak=9MV/m Several multipacting regions were found 25 Dynamic Lorentz Force detuning (I) Monitoring and manipulating the complex signal from cavity during the pulse, dynamic Lorentz force detuning at different gradient were studied. preliminary data dVc 1 / 2 j Vc 1 / 2 RL I dt r r 1/2 sin r r1/2 1/2 cos state space equation Han Li 13th June 2017 26 Dynamic Lorentz Force detuning (II) Developed an FPGA-based LabView program for dynamic Lorentz force detuning. Frequency shift at peak accelerating gradient 9MV/m@ 2.86 ms pulse length is about 400Hz. A Loaded Q value of 1.8 × 105 from state space equation caculation is consist with the VNA measurement. preliminary data ∆𝑓 = 400𝐻𝑧 𝑄𝐿 = 1.8 × 105 Han Li 13th June 2017 27 Mechanical Modes Stimulate the cavity by amplitude modulation . By sweeping the modulation frequency up to 800 Hz, the fit of mechanical modes was studied. Slow tuner is in fixed position . preliminary data Han Li 13th June 2017 Simulation from IPNO N° 1&2 3&4 5&6 7 8 to 11 12 Frequency 212 Hz 265 Hz & 275 Hz 285 Hz 313 Hz 315 Hz to 365 Hz 396 Hz 28 Frequency Sensitivity to Pressure By closing both the inlet and outlet of the cryostat, checking the cavity frequency increasing as a function of helium pressure from 20 to 40 mbar. pressure sensitivity = +27.1 Hz/mbar This result is consistent with the 28 kHz frequency shift measured during cool down from 4.2 K (~1030mbar) to 2 K (~20mbar). Han Li 13th June 2017 29 Tuner Sensitivity Measurements Slow tuner is controlled by Lund system. Tuning sensitivity: 150KHz/mm@2K Length is defined by deformation of the cavity beam pipe. Han Li 13th June 2017 30 Test plan & Conclusions Test of Spoke Cryomodule and high-beta Elliptical cavity this year. Conclusions A test stand for high power test based on a self-excited loop was developed at FREIA. The first prototype ESS spoke cavity package has been tested with the tetrode based RF system. System is ready for Spoke Cryomodule. Han Li 13th June 2017 31