Download From 3-D Positron Emission Tomography to 3

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Backscatter X-ray wikipedia , lookup

Image-guided radiation therapy wikipedia , lookup

Nuclear medicine wikipedia , lookup

Medical imaging wikipedia , lookup

Positron emission tomography wikipedia , lookup

Transcript
doi:10.1016/j.mibio.2004.06.003
REVIEW ARTICLE
Molecular Imaging and Biology
Vol. 6, No. 5, 275–290. 2004
Copyright ! 2004 Elsevier Inc.
Printed in the USA. All rights reserved.
1536-1632/04 $–see front matter
From 3-D Positron
Emission Tomography to 3-D
Positron Emission Tomography/
Computed Tomography: What Did We Learn?
David W. Townsend, PhD
Departments of Medicine and Radiology, University of Tennessee Medical Center, Knoxville, TN
The recent introduction of combined positron emission tomography (PET)/computed
tomography (CT) scanners is having a far-reaching effect on the field of medical imaging
by bringing functional imaging to the forefront in radiology, oncology and other specialties.
The PET/CT scanner is an evolution in technology combining two well-developed imaging
modalities: anatomical imaging with CT and functional imaging with PET. The first prototype
PET/CT scanner was a consequence of a succession of steps that, in chronological order,
included the development of the High Density Avalanche Chamber (HIDAC) PET camera,
3-D PET methodology and the rotating partial-ring tomograph (PRT). The successful
completion of each step was a prerequisite to progress to the next phase, and the lessons
learned could then be applied to subsequent initiatives. This review will map the
milestones from 3-D PET to 3-D PET/CT and assess the role each step played in the
development of PET instrumentation over the past two decades. ! 2004 Elsevier Inc. All
rights reserved.
Key Words: PET instrumentation; 3-D PET; Rotating tomograph; PET/CT scanners.
Introduction
T
he past 20 years in particular have seen dramatic
advances in the performance of imaging instrumentation for positron emission tomography
(PET). From the low resolution, low sensitivity, single
slice designs of the early 1980s to the high resolution,
multi-slice, scanners of today, key imaging parameters
have in most cases improved by at least an order of
magnitude. Current high performance clinical PET
scanners comprise more than 20,000 individual detector
elements, with an axial coverage of 16 cm (or greater), a
4.5 ns coincidence time window, and 20% (or better)
energy resolution. With such specifications, modern
PET scanners attain a measured spatial resolution of
around 2 mm for brain research studies and 4 mm for
clinical whole-body studies, a 3-D scatter fraction of 25%
to 30%, and a peak noise equivalent count rate (NECR)
for 3-D whole-body imaging that approaches 100 kcps.
Address correspondence to: David W. Townsend, PhD, Department
of Medicine, University of Tennessee Medical Center, 1924 Alcoa
Highway, Knoxville, Tennessee 37920-6999. E-mail: dtownsend@
mc.utmck.edu
Financial support for the PET/CT development was provided by NCI
Grant CA 65856
Also of note is a six-decade increase in active coincidence lines, from a few thousand in the early 1980s to
over 109 today. This impressive progress is due to developments in detector construction, new scintillators,
better scanner designs, improved reconstruction algorithms, integration of application-specific electronics,
and, of course, the vast increase in computer power, all
of which have been achieved without a corresponding
order-of-magnitude increase in cost.
In the late 1980s, an important advance occurred with
the introduction of 3-D PET methodology for brain
imaging with a multi-ring scanner1–3. Acquisition of PET
data in 3-D makes optimal use of the emitted radiation,
improving sensitivity compared to 2-D acquisition by a
factor of five even after accounting for the increase in
scatter and randoms3. However, for whole-body imaging, the successful implementation of 3-D methodology has had to await the appearance of faster scintillators,
accurate scatter correction models and improved statistically-based reconstruction algorithms. The recent
availability of scintillators such as gadolinium oxyorthosilicate (GSO) and lutetium oxyorthosilicate (LSO) with
short decay times, accurately modelled scatter distributions, and attenuation-weighted ordered-subset EM reconstruction algorithms have all helped to bring 3-D
275
276 Molecular Imaging and Biology, Volume 6, Number 5
whole-body imaging into the clinical arena. More recently, the importance of routinely imaging function
in conjunction with high resolution anatomy has been
recognized and within the past three years, the combined PET/CT scanner has been introduced into clinical practice4–6. The device acquires accurately-aligned
anatomical (CT) and functional (PET) images in the
same scanner during a single imaging session, overcoming many of the disadvantages of retrospective software
fusion7. The combined PET/CT scanner thus provides
physicians with a powerful tool to diagnose and stage
disease, monitor the effects of treatment, and potentially design better, patient-specific therapies.
This paper maps the milestones from the early PET
scanner designs and development of 3-D methodology to
the advanced combined PET/CT scanners of today.
The topics will include 3-D reconstruction algorithms8,9
developed for the High Density Avalanche Chamber
(HIDAC) positron camera10,11, the application of 3-D
techniques to multi-ring tomographs1–3, the first multiring scanners with retractable septa12, and the development of the first partial-ring rotating tomograph
(PRT)13,14. The PRT design in turn helped to launch the
development of the PET/CT scanner4, the combination
of anatomical and functional imaging in the same
scanner. From the acquisition of the first images in
1998, there are now more than 350 combined PET/CT
devices installed worldwide and they represent up to
80% of new PET scanner sales.
The past two decades of progress in PET instrumentation have not been achieved without considerable effort,
dedication and investment, both intellectual and financial. While mapping the milestones of instrumentation advances it is worth surmising whether such
progress, impressive though it undoubtedly is, has enabled PET to really achieve its full promise and potential. The lessons thus learned may help guide future
directions and identify opportunities that are worthy of
further attention.
The HIDAC Camera and 3-D PET
Since the very early development of nuclear medicine
instrumentation, scintillators such as sodium iodide
(NaI) have formed the basis for the detector systems.
Thallium-activated NaI is an ideal scintillator to convert
the 140 keV photons from the decay of technetium
(99mTc) into light of a wavelength matched to a photomultiplier tube (PMT). The combination of scintillator
and PMT, such as found in the conventional gamma
camera, has dominated nuclear medicine instrumentation since the 1950s. The challenge presented by positron tomography is the detection of the higher energy,
511 keV photons from the annihilation of positrons in
tissue. Early PET scanner designs, however, were based
on NaI (Tl) even though a typical thickness (3/8”)
used in a gamma camera resulted in a very low efficiency
at 511 keV. To improve sensitivity, an increased depth
of scintillator was adopted in some early PET scanners
based on individual crystal arrays15. An alternative approach that was explored in the mid 1970s comprised a
pair of rotating gamma cameras16 in either a standard
configuration or with thicker, 1” crystals for better sensitivity. Curiously, the rotating dual-head gamma camera
experienced a short-lived revival for PET imaging in the
mid 1990s.
A real breakthrough for PET came with the introduction of bismuth germanate (BGO)17 as an alternative
scintillator to sodium iodide. BGO has only 15% of
the light output of sodium iodide but is much denser
and has a greater stopping power and photofraction.
The first PET scanner designs based on BGO appeared
in the late 1970s, and the scintillator eventually became
the most widely-used detector material for PET, a situation that has continued for more than 20 years. Even
so, PET scanner designs based on sodium iodide
continued to be developed until recently and at one
point a significant fraction of PET scanners in clinical
operation were sodium iodide based18,19.
The late 1970s and early 1980s was a period of considerable innovation in PET instrumentation with a
number of different scanner designs under development. While the majority were scintillator-based, some
groups explored a completely different technology,
such as the HIDAC10,11, a detector that originated from
the field of high energy particle physics as the brainchild
of Dr. Alan Jeavons, a physicist working at the European
Center for Nuclear Research (CERN) in Geneva, Switzerland. The HIDAC detector is a multi-wire proportional chamber (MWPC) that incorporates novel lead
converters to improve sensitivity for the detection of
511 keV photons. The standard MWPC was originally
developed to detect high energy charged particles (! 1
GeV) with high spatial resolution and such devices are
relatively insensitive to neutral gamma rays. The addition of perforated lead converters within the chamber
significantly increases the intrinsic sensitivity due to the
photoelectric conversion in the lead of the incoming
gamma. The photoelectron is ejected from the lead into
one of the perforations (holes) and, being charged,
ionises the gas within the chamber allowing the event
to be detected and accurately localized. Groups at the
Lawrence Berkeley Laboratory20, Queens University in
Kingston, Ontario21, University of Pisa in Italy22, Massachusetts Institute of Technology (MIT) in Boston23 and
the Royal Marsden Hospital in London, England24 also
explored the use of MWPC-based designs for PET, each
with a different approach to increasing sensitivity; the
drilled lead converter was unique to the HIDAC detector. Figure 1A shows the first HIDAC camera built for
From 3-D PET to 3-D PET/CT / Townsend
277
Figure 1. The development of
the HIDAC positron camera
for medical imaging. (A) Dr. A.
Jeavons makes an adjustment to
the first HIDAC camera used for
medical imaging in 1980, (B) the
first bone scan of a mouse injected with 5 µCi of 18F-fluoride,
(C) the current design of the
quad-HIDAC small animal scanner, and (D) a bone scan of a 27
g mouse injected with 320 µCi
of 18F-fluoride and scanned for
20 minutes, one hour post-injection (Data courtesy of Universitätsklinikum Münster)
medical imaging around 1980 and Figure 1B is the first
18
F-fluoride bone scan of a mouse imaged with the
HIDAC; only 5 µCi of 18F were injected into the mouse.
Even with the lead converters, a HIDAC detector at
that time was, at best, only 10% sensitive to 511 keV
photons compared to an efficiency in excess of 80% for
BGO. The advantage of the HIDAC approach is the
potential to achieve high spatial resolution in a wire
chamber with closely-spaced wires. An intrinsic resolution of 1.5 mm was measured with a radioactive line
source in air. A PET camera comprising a pair of 20
cm × 20 cm HIDAC modules mounted on a rotating
support was assembled and the performance evaluated
with some limited clinical imaging25 at Geneva University Hospital, Geneva, Switzerland. Owing to the low
intrinsic sensitivity of the HIDAC modules, it was essential to operate the camera in 3-D acquisition mode to
make maximum use of the available photon flux. The
development of fully 3-D image reconstruction algorithms was, therefore, an important aspect of this work.
Since the BGO scanners in the early 1980s consisted of
only one or two rings of small crystals with an axial
coverage of 1 cm, acquisition and reconstruction were
intrinsically 2-D. The 3-D acquisition demands of the
HIDAC camera therefore provided an early impetus to
the development of fully 3-D reconstruction algorithms.
A first step was to derive an appropriate filter to be
used in 3-D filtered backprojection to reconstruct data
acquired at a number of discrete positions of the HIDAC
camera9. Independently, Colsher, working on 3-D filtered backprojection for a dual head rotating gamma
camera for PET, derived the general reconstruction
filter for continuously rotating area detectors26. The
result was also applicable to a continuously rotating
HIDAC camera (where continuous in this context
means a large number of small steps), with the condition
that the system point response function is spatially
invariant and appears the same for all points within
the imaging field-of-view (FOV). To satisfy this condition
for the HIDAC camera, the acceptance angle for annihilation events was limited to a value less than the maximum, thus defining a smaller imaging volume within
which the point response function is spatially invariant27. The algorithm was implemented as 3-D backprojection followed by a 3-D filtering step, unlike the
mathematically-equivalent but more conventional implementation where the 2-D projections are first filtered
and then backprojected in 3-D28.
In the early 1980s, a key advantage of the rotating
HIDAC camera was that the reconstructed spatial resolution at the center of the FOV was isotropic and around
3.5 mm, considerably better than the 8 mm to 10 mm
for scintillator-based scanners at that time; the corresponding volume resolution was 27 ml for the HIDAC
compared with around 1000 ml for the scintillator-based
scanner. However, one lesson learned from this device is
that very high spatial resolution does not compensate
for low efficiency, except when imaging small animals
or small organs such as the thyroid. A coincidence time
window (2τ) of 30 ns and the lack of effective energy
resolution results in high background levels due to randoms and scatter that, combined with a low 511 keV
sensitivity, is a challenging environment in which to
278 Molecular Imaging and Biology, Volume 6, Number 5
image patients. Some limited clinical imaging was nevertheless attempted successfully, including 18F-fluoride
for bone, 124I for thyroid, 68Ga-colloid for liver, and
55
Co-bleomycin for brain tumors. While the original
HIDAC camera design never achieved prominence as
a clinical scanner, improvements in the technology have
established the quad-HIDAC camera as a serious contender for animal imaging (Figure 1C). The most recent
version achieves sub-millimetre resolution at sensitivities
adequate for imaging small animals; the image of a 27
g mouse scanned on a quad-HIDAC after injection of
320 µCi of 18F-fluoride is shown in Figure 1D. The quantitative imaging potential of the quad-HIDAC has also
recently been demonstrated29.
The HIDAC-related efforts in 3-D reconstruction and
the concept of a dual detector rotating scanner were
seminal to subsequent work on 3-D reconstruction for
multi-ring scanners, the low-cost partial ring rotating
scanner and, ultimately, the combined PET/CT. The
HIDAC approach highlighted the necessity, in high
quality clinical imaging, of achieving both good resolution and high sensitivity while minimizing the acquisition of randoms and scatter. Nevertheless, despite the
low sensitivity and high background levels, the HIDAC
camera demonstrated the feasibility of fully 3-D imaging
with an area detector rotating scanner, including both
acquisition and reconstruction.
Multi-ring PET Scanners and 3-D Imaging
With the exception of the dual-head rotating gamma
camera16,30 and the HIDAC camera11, most PET scanner
designs in the 1980s were single or dual rings of crystals
covering an axial extent of 1 cm to 2 cm31. Such a small
axial coverage was especially a limitation for PET studies
in neuroscience research where ideally the whole brain
should be imaged at the same time rather than with
multiple bed positions at different times. The architectural and financial difficulties of increased axial coverage were largely solved by a breakthrough invention of
Drs. Ron Nutt and Mike Casey at CTI in Knoxville,
Tennessee – the block detector in which a 5 cm × 5 cm
block of scintillator (BGO) is bonded to four 1″ photomultiplier tubes (PMTs)32. The first blocks were cut into
8 × 4 smaller crystals and the four PMTs localized the
incident photon to one of the 32 crystals based on
the light sharing between the PMTs. The coupling of
32 crystals to four PMTs significantly reduced cost (by
reducing the number of PMTs relative to a 1-1 coupling)
and a single ring of blocks covered 5 cm axially, subdivided into four rings of crystals. The first multi-ring
scanners based on the block design appeared in the mid1980s, and by 1987 scanners incorporating two rings of
blocks covering 10 cm axially with eight rings of crystals
were starting to appear at a small number of research
institutions33. With that coverage, much of the brain
could be imaged in a single scan.
PET was originally conceived in the 1950s as a 3D imaging modality with the physical collimation of
conventional single-photon emitters replaced by the
electronic collimation of coincidence imaging. However, concerns over the high level of scattered photons
and the perceived absence of an effective fully 3-D reconstruction algorithm resulted in the first multi-ring
PET scanners being equipped with lead annuli, or septa,
between each ring of crystals (Figure 2A). Septa shield
the detectors from out-of-plane scatter, effectively subdividing the 3-D imaging volume into a set of independent
2-D slices analogous to multiple CT sections (Figure 2B).
Each slice was reconstructed with a 2-D algorithm such
as filtered backprojection that had been exhaustively
validated for CT and single photon emission computed
tomography (SPECT). The first multi-ring scanners
therefore imaged a 3-D positron-emitting distribution
as a set of contiguous 2-D sections, a procedure that
continues to this day some two decades later even
though it makes poor use of the available photon flux.
Following the installation of one of the early eight-ring
PET scanners, the ECAT 931/08-1233 at Hammersmith
Hospital, London in 1987, Dr. Terry Jones was among
the first to propose removing the septa and operating the
scanner fully in 3-D. With foresight, the eight-ring ECAT
had actually been designed to acquire all possible coincidence lines-of-response (LORs) and collect a full 3D data set of 8 × 8 sinograms. After physically removing
the septa, the first 3-D data sets were acquired for a multiring BGO PET scanner in early 1988 and reconstructed
using a backproject and filter algorithm34. This algorithm satisfies the condition that the point response
function must be spatially invariant by subdividing the
reconstruction volume into smaller sub-volumes within
each of which invariance is preserved. The filter used
in this algorithm is that derived by Colsher26, the same as
used with the rotating area detectors. However, the
algorithm is different to that for the HIDAC camera
where spatial invariance was achieved by restricting the
acceptance angle to a value less than the maximum
possible, ensuring a central volume of uniform response. This procedure could not be applied to the multiring scanner without restricting the acceptance angle
to a small value equivalent to the acceptance angle in
2-D (Figure 2B), eliminating all the benefits of 3-D
acquisition.
In general, a more efficient implementation of filtered backprojection, particularly in 3-D, is to filter the
2-D projection data and then backproject in 3-D. The
requirement that the point response function is spatially
invariant is equivalent to a requirement that all projections are fully measured. For a scanner of limited axial
extent, this is obviously not the case since projections
From 3-D PET to 3-D PET/CT / Townsend
279
Figure 2. (A) In the late 1980s, multi-ring PET scanners were designed with inter-plane lead or tungsten septa between each
detector ring, (B) when extended the septa limited the acceptance of coincidence events to small incident angles subdividing
the imaging volume into a set of 2-D transverse sections, (C) from the early 1990s, multi-ring scanners incorporated retractable
septa that could be removed allowing all coincidence lines to be acquired and thus imaging the volume fully in 3-D, and (D)
the NECR curves for the NEMA NU-2001 phantom acquired for an LSO PET scanner with septa extended (2-D) and septa
retracted (3-D); the advantage of imaging in 3-D at low activity concentrations is evident.
at oblique angles are incompletely measured. A solution
to the problem of incompletely-measured projections,
the reprojection algorithm35 was proposed by Kinahan
and Rogers in 1989 while investigating image reconstruction for a volume PET scanner design. Instead of
restricting the acceptance angle, all angles are allowed
and the partially measured parallel projections are completed by re-projecting the missing projection data
through an initial estimate of the volume image. The
initial estimate is obtained by reconstructing the complete set of 2-D sections and stacking them to form the
volume through which the missing LORs can be re-projected. Volume images reconstructed with the reprojection algorithm typically incorporated more than 90%
of the full 3-D data set.
The results from these preliminary studies with the
septa removed were encouraging, particularly for the
brain, where a factor of 8 increase in count rate was
observed. Although the random and scattered coincidence rates also increased, there was still a net gain
of at least a factor of 2 in signal-to-noise for the 3D acquisition, particularly at low activity concentrations.
In addition, for the brain, the peak signal-to-noise in 3D was achieved at an activity concentration of about
10% that required to achieve a similar signal-to-noise
with 2-D acquisition. The lessons learned from this work
clearly highlighted some of the advantages of 3-D PET
imaging with a multi-ring scanner, at least for brainsized distributions. However, physical removal of the
septa from this first multi-ring scanner was impractical
on a routine basis and a more efficient procedure to offer
both 2-D and 3-D acquisition capability was required.
Scanners with Retractable Septa
The results from these preliminary 3-D studies were
promising and in 1989 CTI PET Systems in Knoxville,
Tennessee embarked on the design and construction
of the first multi-ring PET scanner with retractable septa,
the ECAT 953B. With a patient port of 35 cm, the scanner was designed specifically for imaging the brain.
Push-button control was provided to extend or retract
the septa within two minutes. The ECAT 953B was also
the first 16-ring PET scanner comprising two rings of 5
cm × 5 cm BGO blocks cut into 8 × 8 small crystals. The
first scanner of this design was installed in Hammersmith Hospital London in early 199012. With septa retracted (Figure 2C), many more LORs are active and a
total of 256 (16 × 16) sinograms are acquired.
The capability of this system to acquire data in either
2-D or 3-D offered an excellent environment in which
to assess the advantages and challenges of 3-D PET.
Exhaustive phantom studies were conducted to compare the 2-D and 3-D performance of the ECAT 953B3,36.
Studies with a 20 cm diameter uniform cylinder showed
that the system count rate increased by a factor of 7.8
when the septa were retracted, reducing to a factor of
4.7 after scatter subtraction. Other results confirmed
those found previously during the preliminary studies,
such as a factor of 2 to 3 improvement in NECR at low
activity concentrations despite a factor 3 increase in
scatter fraction. Similar NECR improvement factors of
3 to 5 were found for flow and ligand studies in the brain.
Procedures for 3-D normalization, attenuation and
scatter correction were developed and image reconstruction was based on an implementation of the reprojection algorithm with the Colsher filter35. At that time
280 Molecular Imaging and Biology, Volume 6, Number 5
a drawback to the adoption of 3-D methodology was
that the reconstruction of a 128 × 128 × 31 voxel matrix
took several hours on a standard workstation. Accurate
scatter correction was also a concern for quantitative
studies especially for receptor measurements. Despite
these and other concerns, by the early 1990s, 3-D PET
methodology for the brain had been developed to a
point where it was being applied in practice for activation studies. Accurate quantitation was not essential
and 12 mCi of 15O-water could be injected in 3-D rather
than the 50 mCi required in 2-D. Some brain receptor studies were also performed in 3-D since the high
sensitivity was of benefit for measurements of low receptor populations, but here accurate quantitation is required. Other PET scanner designs with retractable septa
followed the ECAT 953B and retractable septa quickly
became standard on all scanners that were not already intrinsically 3-D. Whole-body designs with retractable septa appeared in the early 1990s for cardiac and
tumor imaging although valid concerns about high
levels of random and scattered events limited the use
of 3-D outside the brain. The lessons learned from this
work firmly established 3-D PET methodology for brain
studies capitalizing on the increased sensitivity essential
for ligand studies and enabling the injected activity
level for activation studies to be significantly reduced.
The latest septa-retractable LSO-based PET scanner
designs show a similar trend in 3-D (Figure 2D) as the
earlier designs. The measurements of NECR in Figure
2D are for the NEMA NU-2001 phantom in the ECAT
ACCEL with septa extended (2-D) and septa retracted
(3-D); the advantages of 3-D methodology at low activity
concentrations can clearly be seen. The success of 3-D
methodology such as demonstrated in Figure 2D suggested a possible design for a low-cost or entry-level PET
scanner. The design, completed in 1990 and commercially available even today, is based on dual rotating
arrays of block detectors, and the first prototype was
the partial ring rotating tomograph (PRT-1)13.
The PRT
PET tomographs based on the rotation of a pair of
opposing 2-D area detectors were thoroughly explored
during the late 1970s and early 1980s11,15,16,30. As described, most of these approaches had low sensitivity
or poor count rate performance and generated little
clinical interest. With few exceptions, by 1988, rotating
designs had been abandoned for PET imaging in favor
of the stationary multi-ring BGO scanners. In the late
1980s, the capital outlay for the imaging equipment was
perceived as a major impediment to the wider acceptance of PET and efforts to limit the financial investment
required to establish a PET center were considered important. For a typical BGO scanner, more than half the
cost lies in the detectors – scintillator and PMTs – and
eliminating a significant number of detectors can have
a major impact on cost. Thus, increasing sensitivity
by acquiring data in 3-D presented an opportunity to
design and build a lower cost scanner by removing over
half the detector blocks from a multi-ring scanner and
rotating the remaining partial rings of detectors to acquire the full set of 3-D projection data (Figure 3A).
This design, with two banks of block detectors, achieves
a clinically-acceptable sensitivity by acquiring data fully
in 3-D. In fact, for a fixed number of detectors, it can
be shown that a higher sensitivity (a greater number of
active LORs) is achieved by arranging the detectors as
two rotating arrays rather than a continuous stationary
ring due to the greater axial coverage of the arrays37.
The first design of this type (Figure 3B, top), the PRT1 consisted of two arrays of 48 (transaxial) × 2 (axial)
blocks, each block cut into 8 × 8 crystals13. The blocks
in this prototype were identical to those of the 953B12 and
the PRT-1 contained one third of the number of detectors in the full-ring scanner. The absolute sensitivity
of the prototype was 0.51%, slightly less than the 0.63% of
the ECAT 953B with septa extended38. The data were
acquired in step-and-shoot mode at six positions of
the detectors for a 20 cm transverse FOV and a 10.8 cm
axial FOV. The minimum scan time including rotation
was 60 seconds and the prototype included both static
and dynamic imaging capability. The transmission
source for attenuation correction in 3-D was a 6 mm
thick solid germanium plane source containing 0.9 mCi
of 68Ge; the source was 10.8 cm wide to cover the axial
FOV and it could be attached to the front of one detector array. Scatter correction was based on an iterative
deconvolution approach. Excellent results were obtained for imaging brain (Figure 3B, bottom) and heart
with 2-deoxy-2-[18F]fluoro-D-glucose (FDG), including a
13-frame dynamic acquisition for the brain. The device
was also capable of imaging 82Rb in the heart for myocardial flow studies, an impressive achievement with a rotating scanner considering the 75 seconds half-life of the
tracer13.
A second prototype14, PRT-2, was built in 1993 incorporating dual arrays of ECAT EXACT block detectors
covering 16.2 cm axially (Figure 4A). Each array comprised 10 (transaxial) × 3 (axial) blocks representing
40% of the detectors in the equivalent full-ring scanner,
the ECAT EXACT39. Data acquisition was again in stepand-shoot mode with the number of angular steps
depending on the transaxial imaging FOV required,
typically 18 positions for a 47 cm FOV. Similar to the
PRT-1, the transmission source was a rectangular planar
source covering the 16.2 cm axial FOV; for this prototype, the source was shaped to follow the curvature of the
detectors. The sensitivity of PRT-2 was approximately
double that of PRT-1 due to the additional axial coverage and was actually 25% higher than that of the EXACT
From 3-D PET to 3-D PET/CT / Townsend
281
Figure 3. (A) a more cost effective design of PET scanner
was developed by removing
detectors from a full-ring
scanner (top) and then rotating the dual partial ring arrays
(bottom) to acquire a full 3D data set, (B) the first partial
ring rotating PET scanner
PRT-1 built at the University
of Geneva Hospital in collaboration with CTI PET Systems (top) and (bottom) the
first FDG brain scan acquired
with the prototype in May,
1991.
with septa extended; with septa retracted, the 3-D sensitivity of the EXACT was a factor 3.5 higher than that of
the PRT-2. The scanner was evaluated clinically, showing
promise for imaging FDG in brain and heart (Figure
4C, right), 13N-ammonia for myocardial flow (Figure 4C,
left), and even 15O-water for blood flow activation studies (Figure 4B). Whole-body FDG-PET studies of oncology patients were also acquired in step-and-shoot mode14.
The success of the two prototypes stimulated the development of a commercial version of the partial ring
tomograph, announced in 1995 (Figure 5a). The design
of the advanced rotating tomograph (ART) (CPS Innovations, Knoxville, TN), was based on dual arrays of 11
(transaxial) × 3 (axial) ECAT EXACT blocks corresponding to 46% of the detectors in the ECAT EXACT.
The sensitivity was therefore similar to that of the PRT2. The design eliminated cables by incorporating slipring technology, both mechanical for power and serial
communications and optical for high-speed data transfer, and rotated constantly at 30 rpm. The early ART
Figure 4. (A) a second prototype partial ring tomograph
PRT-2 still requiring the use of
cables, (B) cerebral blood flow
imaged with 15O-water in the
PRT-2; the images are summed
over 12 injections each of 12
mCi of 15O-water and the scan
time is 90 s, and (C) a cardiac
viability study with (left) 8 mCi
of 13N-ammonia to image myocardial blood flow, and (right)
10 mCi of FDG to image
myocardial metabolism. Data
courtesy of the University of
Pittsburgh.
282 Molecular Imaging and Biology, Volume 6, Number 5
designs incorporated dual 68Ge rod transmission
sources mounted at one end of each detector array,
later replaced by collimated 137Cs point sources that
scanned axially during rotation; transmission acquisition was in singles mode40. A number of these low-cost
scanners, the first dedicated PET scanner to be sold for
under $1 million, appeared at clinical installations from
the mid to late 1990s and were used effectively for FDG
imaging of brain, heart and whole-body. The ART was
also used for brain activation studies with 15O-water,
blood flow measurements in peripheral muscle, and
82
Rb for gated cardiac studies. Figure 5B shows the
experimental set-up for brain activation studies with
laser pain stimulation, and summed flow images in a
normal volunteer subjected to pain stimulation are
shown in Figure 5C. The ART played a particularly important role in clinical whole-body imaging at several
institutions41,42.
The second and more significant impact of the PRT1 design was the realization that, with dual arrays of
detectors, there was potentially enough void between
the arrays (Figure 6A, arrowed) to mount an X-ray tube
and detectors and acquire anatomical images in addition to the functional images from PET, the two data
sets being accurately co-registered. Such a concept
(Figure 6B), first envisaged in 1991, would address many
of the difficulties of software registration for the whole
body where patient positioning differences and internal
organ motion are particularly challenging issues.
Although the combined PET/CT concept was proposed
in 1991, it was another seven years before the first prototype appeared in the clinic, as described in detail below.
Whole-body PET Imaging for Oncology
The success of 3-D PET methodology in the brain was
not matched by a comparable situation for the rest of
the body. Poor whole-body image quality was due to
several reasons: randoms and scatter arising from activity outside the imaging FOV could not be effectively
shielded as they can for the brain; the uptake in the
brain is generally greater than for any other organ and
hence in the rest of the body the true coincidence rate
decreases while the scatter and randoms increase; patient movement is more difficult to control than for the
brain; patient body habitus varies considerably and, for
the larger patients, the transmission and emission scans
are highly noisy; and the reprojection algorithm35, a 3D version of filtered backprojection does not perform
optimally in a high noise situation, introducing streak
and other artifacts that interfere with image interpretation (Figure 7A). The situation for whole-body imaging
in 3-D was therefore not favorable despite extensive
effort to produce an accurate scatter correction
Figure 5. (A) a commercial ECAT ART scanner with the front cover removed; the data and power cables required in the two
prototypes are replaced by mechanical and optical slip rings, (B) the ART scanner being used for brain activations studies; the
device to the left of the scanner is a laser for stimulating a response to differing levels of pain, and (C) summed blood flow
images acquired with 15O-water injections during pain stimulation. (Activation studies performed in collaboration with Drs.
Anthony Jones and Stuart Derbyshire, Manchester University, UK.)
From 3-D PET to 3-D PET/CT / Townsend
283
Figure 6. (A) the proposal for
a combined PET/CT scanner arose from the observation of the empty space
(arrowed) between the detector arrays in the PRT-1, and
(B) the initial design envisaged for the combined PET/
CT with the X-ray tube and
detectors mounted in the
gaps between the PET detectors. The final design actually
had the PET detectors
mounted on the rear of the
CT support.
model43,44. At that time, most institutions with retractable septa scanners elected to perform whole-body scanning in 2-D since the septa shielded the detectors from
the high level of randoms and scatter and a fast 2-D
reconstruction algorithm could be used.
Nevertheless, from 1995 onwards an increasing
number of 3-D only, dedicated PET scanners, including the ART scanner, appeared in clinical practice even
though it was not until 1998 that Medicare approved
reimbursement in the US for PET in lung and a few
other cancers. In addition to the ECAT ART, designs
such as the QUEST (UGM Medical, Philadelphia, PA)
based on the PENN-PET18, and later the C-PET19 were
3-D only and used the superior energy resolution of
sodium iodide to limit the 3-D scatter fraction. Nevertheless, for all but the lighter patients, whole-body image
quality in 3-D was poor, further degraded by attenuation
correction due to the magnitude of the correction factors. Many physicians insisted on reviewing the less-noisy
non-corrected images. Even so, for a typical patient, the
NECR averaged over all body positions is about a factor
of 2 higher in 3-D than 2-D45. Owing to the success (and
low capital outlay) of the sodium iodide-based designs,
and with a small contribution from the installed base
of ART scanners, by the late 1990s a significant fraction of
tomographs in clinical operation for oncology were 3D only. However, for the larger patients, the image
quality was far from optimal and whole-body BGO scanners with septa extended were generally superior. The
lesson to be learned is that BGO is not the best scintillator for whole-body imaging in 3-D and filtered backprojection is not the optimal algorithm to reconstruct
the images.
The situation began to change with the introduction
of new fast scintillators for PET. For the first time in over
two decades, an alternative to BGO appeared in the
form of lutetium oxyorthosilicate (LSO), a scintillator
with five times the light output and seven times faster
than BGO. Discovered around 199046, LSO first appeared in commercial PET scanners some ten years
Figure 7. A coronal section
from an FDG-PET whole-body
scan acquired in 3-D mode
with septa retracted and reconstructed using (A) 3-D filtered
back-projection algorithm with
reprojection, (B) Fourier rebinning (FORE) and attenuationweighted OSEM (AWOSEM).
The noise streaks that are a
characteristic feature of filtered
backprojection are eliminated
by the statistical approach (reconstructions courtesy of Dr
David Brasse).
284 Molecular Imaging and Biology, Volume 6, Number 5
later. LSO offers better positioning accuracy, shorter
coincidence time window, better energy resolution, and
reduced detector dead time due to the superior physical
characteristics. The scintillator was introduced by CTI,
Inc (Knoxville, Tennessee) and was incorporated into
the ECAT ACCEL, the first LSO-based clinical scanner47. Compared to the ECAT EXACT39, the BGOequivalent of the ACCEL, the impact of LSO on performance includes higher spatial resolution, reduced scatter and randoms rates and better count rate
performance. The NECR curves for the ACCEL in 2-D
and 3-D are shown in Figure 2D. At about the same
time, Philips Medical introduced the ALLEGRO, a gadolinium oxyorthosilicate (GSO) based scanner48. First
described in 198349, GSO is less sensitive than LSO, has
about 50% longer decay time, and only 36% of the
light output; however, it outperforms BGO in all aspects
except stopping power.
In addition to these improvements in detector technology and hardware, significant software advances have
taken place over the past decade. The considerable effort
invested in scatter correction has converged to an accurate image-based scatter correction model50, and with
improvements in computing power and algorithm development, a statistically-based reconstruction algorithm
has emerged as a successor to filtered backprojection. In
order to reduce the computational burden, the 3-D acquired data set is rebinned into an equivalent 2-D set51
and then reconstructed using the attenuation-weighted
ordered-subset EM (AWOSEM) algorithm52. Currently,
Fourier rebinning (FORE)51 is performed to reduce the
3-D sinogram set to 2-D for reconstruction by AWOSEM
in which a coincidence LOR is assigned a weight related
to the attenuation factor of the line – LORs with large
factors are assigned the smallest weight. The combination of FORE with AWOSEM has been implemented
efficiently and a reconstructed image can be produced within a couple of minutes. The same data set
as that used for Figure 7A has been reconstructed with
FORE " AWOSEM; the significantly improved image
quality is shown in Figure 7B. A fast implementation of
3-D AWOSEM is under evaluation and initial results
show improved clinical image quality compared to the
2-D rebinning approach.
The most obvious lesson to be learned from this
decade of focussed effort is that it is a challenge to
obtain high quality 3-D whole-body images. The combination of BGO, filtered backprojection reconstruction
and a simple scatter correction model failed to produce
whole-body images of sufficiently high quality that
were diagnostically useful and acceptable to physicians,
especially in large patients. Fortunately, these lessons
have been well-learned and the situation is now very
different with 3-D whole-body imaging becoming an
established methodology in many institutions.
The Combined PET/CT Scanner
The significance of anatomical landmarks in functional
images has long been recognized, at least since the days
of hand-drawn neck outlines on the early nuclear medicine thyroid scans. With the advent of digital imaging,
a more rigorous approach to combining anatomy and
function became possible by using software to align the
two image sets. Nevertheless, most image alignment is
still performed visually by physicians with CT and PET
images presented together on adjacent displays. Despite
receiving little attention clinically, over the past decade
or so software registration algorithms for aligning
image sets acquired by two different modalities have
evolved from simple matching procedures to complex
non-linear warping techniques7,53–55. Motivated by the
belief that the combination of anatomical and functional images is beneficial and complementary to both
modalities, sophisticated software tools have been developed to perform registration. Outside the brain, however, despite the sophistication, software registration
has had only limited success56 and is still not widely used
clinically. Difficulties include allowing for inconsistent
patient positioning between the different scanners, and
uncontrolled internal organ movement. While reasonably accurate localized alignment is possible for specific
regions such as the thorax, co-registration of the whole
body remains problematic and validation of the techniques remains a challenge. Nevertheless, limited studies
that have been made using software-based registration
have highlighted, despite practical difficulties, the intrinsic
advantages of combining anatomical and metabolic
information56.
The general concept of fusing anatomy and function
in the same device is not new. Historically, one of the
first dual modality systems reported in the literature was
a combination of CT and SPECT with work in the late
1980s by Hasegawa et al.57 resulting in a device that used
the same detector material, high purity germanium,
for both modalities. However, to avoid compromising
either modality by the choice of a common detector
material, Lang, Hasegawa and coworkers later developed a device based on two separate scanners – a CT and
a SPECT scanner58. In patenting the concept Hasegawa
noted that it could equally well apply to CT and PET59,
although to progress from concept to realization of a
functioning PET/CT scanner required funding, adequate engineering facilities, and a motivated team of
people. While the PET/CT concept emerged independently in 1991 from a proposal by Townsend and Nutt
after observing the void within the PRT-1, the completed
prototype did not appear in the clinic until 1998.
Funded in part by a grant from the National Cancer
Institute in 1995, the final prototype design4 was a considerable departure from the initial concept.
From 3-D PET to 3-D PET/CT / Townsend
285
Figure 8. A schematic of the original PET/CT prototype design combining a Siemens Somatom AR.SP spiral CT scanner with
an ECAT ART PET scanner. Acquisition and reconstruction were kept separate and the CT images were imported into the
PET environment to generate CT-based attenuation correction factors and for the fused image display on the PET console.
The prototype design (Figure 8) incorporated a Siemens Somatom AR. SP spiral CT scanner with an ART
scanner. The high packing density of components on
the CT rotating assembly precluded the mounting of
the ART detectors as originally envisaged (Figure 6B).
Instead, a second annular support was attached to the
rear of the CT assembly, set back from the slip rings,
and the PET detectors mounted on the additional support. The complete assembly rotated at 30 rpm and to
avoid any cross interaction the imaging components
were operated consecutively and not concurrently. The
data acquisition systems and image reconstruction computers were not integrated and the CT and PET images
were acquired and processed separately (Figure 8). The
CT images were imported into the PET environment
and used to generate the PET attenuation correction
factors to be applied to the PET emission data. The CT,
PET and fused images were displayed on the PET console for review and interpretation. The device was
installed at the University of Pittsburgh PET Facility in
April 1998 for clinical evaluation. During the subsequent three-year evaluation program, more than 300
cancer patients were scanned on the prototype60–63.
Some important lessons were learned from the six-year
PET/CT development program, covering scientific, engineering and clinical aspects. The scientific and engineering experience influenced the design of the first
commercial PET/CT scanners, while the initial clinical
experience clarified some of the strengths and challenges
of the approach, identified the application areas where
PET/CT could bring the greatest benefits, and defined
protocols appropriate for these applications.
Following the promising clinical results, the major
vendors moved rapidly to bring a commercial design to
the marketplace. A common feature of the designs that
eventually emerged is the low level of hardware integration – all essentially comprise a CT scanner and a PET
scanner in tandem64. The integration of the two modalities is more focused on the computer systems in order
to present a unified operation for acquisition and image
reconstruction with the underlying computational complexity transparent to the user. Interestingly, of the four
major distributors, three opted for 3-D-only PET scanners and eliminated the septa and two of them opted not
to provide standard PET transmission sources and base
the attenuation correction on the CT images. It is less
than three years since the first commercial PET/CT
scanner was installed and the technology is still evolving,
and will doubtless continue to evolve in the future. CT
in particular is experiencing a period of rapid change
as multi-slice CT scanners, in only three years, have progressed from 2 to 16 slices, with 32 and 64 slices recently
announced. There is now a realistic prospect of 2-D
area detectors and cone-beam CT. The appropriate configurations for PET/CT have still to be established and
may well depend upon the application area targeted: oncology, cardiology or neurology. High performance CT
may not be required for imaging cancer, and PET/
286 Molecular Imaging and Biology, Volume 6, Number 5
Figure 9. (A) the latest high-performance 16-slice LSO PET/CT scanner at the University of Tennessee Medical Center in Knoxville,
Tennessee. The unique design of the patient couch eliminates any vertical deflection due to patient weight as the bed moves
into the scanner, and (B) a typical PET/CT scan acquired with this system; the patient is a 155 lb, 50 year-old male undergoing
treatment for colon cancer with bladder metastases. The PET scan (left) was acquired 148 minutes post-injection of 10 mCi
of FDG for four minutes per bed position, and the scan demonstrated a focal area of elevated uptake in the right lower
quadrant; the PET/CT scan (right) accurately localizes the uptake anatomically within the abdomen.
CT protocols are still under development, addressing
issues such as respiration65–67 and the use of CT oral
contrast media68.
One of the latest PET/CT designs (Figure 9A) comprises a Sensation 16 CT scanner (Siemens Medical Solutions, Forchheim, Germany) with a high resolution,
LSO-based PET scanner (CPS Innovations, Knoxville,
TN). This combination is now distributed by Siemens
as the biograph 16. The spiral CT with 5 mm slice spacing
takes less than 25 s to scan from base of brain to upper
thigh, and the combination of fast scintillator and high
count rate electronics ensures a PET imaging time that
can be as short as six to seven minutes for small patients.
This is a significant improvement over the 60-minute
imaging times that were the norm just a few years ago.
Other commercially-available designs include an open
system where the CT and GSO-based PET scanner can be
moved apart (Gemini, Philips Medical), and a BGObased PET scanner with 6 mm detectors and a 4, 8 or
16-slice CT (Discovery ST, GE Medical Systems).
An illustration of the extent of progress in PET instrumentation is demonstrated by the quality of the scans
in Figure 9B; the images show a coronal section of a
patient scanned recently on the 16-slice LSO PET/CT
at the University of Tennessee, Knoxville. The PET scan
identified a focal area of increased uptake in the right
lower quadrant. To date, many thousands of cancer
patients have been imaged in PET/CT scanners installed at more than 400 institutions worldwide. The
impact of this evolution in imaging technology has been
profound within radiology and nuclear medicine especially, and sales this year are predicted to exceed $1
billion. Many lessons have already been learnt from
combined PET/CT imaging, the most important ones
being the added confidence in reading PET scans with
the co-registered CT anatomy routinely available, the
ability to accurately localize focal tracer uptake, and
the potential to distinguish between normal accumulation of tracer and pathology. Many more lessons are to
be learned as the technology matures and becomes
more widely available.
Discussion
In the past two decades, it is evident that there have
been dramatic advances in imaging instrumentation for
PET. Numerous scientists and engineers have contributed to these advances at universities and corporate
institutions worldwide. Much of the pioneering work has
been funded through research grants and with financial
support from national agencies and private foundations.
The unrelenting quest for higher sensitivity, better spatial resolution, improved signal-to-noise and lower cost
has resulted in PET scanners that achieve 2 mm spatial
resolution throughout the brain, can acquire data at
system count rates up to 4 × 107 singles per second, and
attain a peak NECR of up to 100 kcps. To meet these
levels of performance, the number of active LORs has
increased from a few thousands in 1980 to more than
109 today. Fortunately, the progress in computational
and data storage capacity has kept pace with demand and
the resources available today largely satisfy clinical needs.
Until reimbursement was approved for a limited
number of oncology studies in 1998, PET was primarily
a research tool for neuroscience. Consequently, in the
decades preceding reimbursement the main driving
From 3-D PET to 3-D PET/CT / Townsend
force for instrumentation development was brain imaging, specifically for studies of psychiatric and neurodegenerative disorders. These studies demanded high
spatial resolution to resolve small brain structures,
high sensitivity to measure low receptor populations, dynamic scanning capability to follow rapid tracer kinetics,
and accurate quantitation of both specific and nonspecific ligand binding. As the emphasis shifted to
oncology in the latter part of the 1990s, the demands on
the instrumentation also changed. For clinical oncology
imaging, of greater importance are short scan times,
high throughput capability, ease of operation, high reliability, fast image reconstruction and applicationspecific image display and analysis. Obviously, there
are some demands that apply equally to clinical and
research applications such as high sensitivity and low
scatter and randoms rates. Research activities generally
require access to the raw data, to scanner operating
procedures, detector setup, data acquisition settings,
reconstruction parameters and image analysis tools.
Clinical operation has a preference for standard protocols with fixed parameter settings to reduce the opportunity for operator error, fast reconstruction times for
rapid physician feedback, and efficient display tools
for reading and interpreting images. The lessons
learned in one domain obviously impact progress in the
other domain.
From the early beginnings with HIDAC and other
fully 3-D PET scanners, the man-years of effort that has
been devoted to 3-D methodology, first in the brain and
now for the whole-body, is a major investment. Since
PET is intrinsically 3-D, in these days of heightened
sensitivity to radiation exposure, it is an obligation to
make maximum use of the available photon flux from
the patient while ensuring diagnostic image quality. To
limit the potentially high skin exposure from CT, considerable effort has been devoted to making efficient use
of the X-ray dose. Although radiation exposure from a
positron-emitting tracer like FDG is much lower than
the CT skin dose, there is nevertheless an increasing
trend to reduce the injected dose to levels that will
almost obligate the use of 3-D acquisition. Fortunately,
the achievements for the past 15 years have finally established the role of 3-D methodology for both brain and
whole-body.
The addition of anatomical imaging capability to PET
scanners has had a profound effect, particularly on the
clinical operation by essentially eliminating the time for
the transmission scan, improving image quality and increasing patient throughput. Initially, from the performance of the original prototype where five minutes was
required for CT and 45 minutes for PET, the PET/CT
approach was often criticized for making inefficient use
of the CT scanner. As a consequence of the dramatic
advances in both CT and PET technology, the high
287
performance PET/CT scanners of today require 30 seconds for CT and 10 minutes or so for PET. From the
viewpoint of the patient, the actual scan time is now a
small fraction of the time required for the complete
study, including patient preparation, FDG uptake, and
positioning in the scanner. Much has been learned
and much has been achieved towards improving imaging technology for PET.
From its inception, one of the great strengths of positron tomography was the richness of the compounds
that could be labelled with short-lived positron-emitting
biomolecules such as 15O, 13N and 11C. The first images
of the human brain with FDG, were acquired in 1976
at a time when the HIDAC camera was imaging the first
mouse with 18F-fluoride. Now, 28 years later, FDG is still
the only radiopharmaceutical widely used and reimbursed for clinical imaging, with 82Rb on a much smaller
scale for myocardial flow imaging. Whole-body imaging
with FDG can hardly be said to challenge the performance of current instrumentation. There are, of course,
a much wider range of PET radiopharmaceuticals available in the research domain, primarily for brain studies.
It is important for the future of PET that the current
situation evolves so that other tracers such as 31[18F]fluoro-31-deoxythmidine (FLT), 11C-acetate, 18F-choline
and maybe a hypoxia imaging agent such as 18F-miso
or 64Cu-ATSM become available clinically. There are
many aspects of tumor physiology more representative of
cancer than elevated glucose metabolism that could provide an earlier indication of a primary or recurrent
malignancy at a time when effective treatment is still
possible. PET is clinically under-utilized and even with
FDG has not yet achieved the promise of the 1980s
despite these major advances in imaging instrumentation. Progress in tracer development and regulatory approval of PET radiopharmaceuticals is now essential to
match the advances described here that have occurred
in PET instrumentation over the past two decades.
It is a great honor, although with considerable sadness, to
accept this award in memory of my dear friend and colleague
Dr. Peter Valk who passed away December 16, 2003. Peter
himself made numerous contributions to the advances described here, particularly to the evaluation of PET scanner
designs, to 3-D PET methodology, and to clinical applications
of PET in which he was instrumental in obtaining the first
Medicare reimbursement for PET oncology studies in 1998.
He was a great supporter of PET/CT and he will be sadly
missed. It is also a privilege to recognize the many individuals,
both friends and colleagues, who have contributed to this work
over more than two decades. The author is particularly indebted to Dr. Alan Jeavons inventor of the HIDAC camera
This paper reflects the content of the 2003 AMI Distinguished Scientist
Award Lecture presented by Dr. David Townsend at the Annual
Meeting of the Academy of Molecular Imaging in Orlando, Florida
in March 2003.
288 Molecular Imaging and Biology, Volume 6, Number 5
and the first to introduce the author to the challenges of PET
in 1975; to Professor Gabor Herman for teaching the author the
basic mathematics of image reconstruction; to Professor Alfred
Donath who had the foresight to see the promise of clinical
PET from as early as 1980; to Professor Terry Jones for his
enthusiasm, encouragement and motivation for PET, especially 3-D PET and PET/CT; and to Dr. Ron Nutt for his
insight, leadership and dedication to the development of
PET instrumentation for over twenty years, and especially for
his contributions to the work chronicled in this review. The
author offers special thanks: to Dr. Thomas Beyer, who obtained his PhD on the PET/CT project and who made numerous innovative contributions; to Drs. Benno Schorr, Michel
Defrise, Paul Kinahan and Rolf Clackdoyle for their many
original contributions, both theoretical and practical, to the
field of image reconstruction; to Drs. Antoine Geissbuhler,
Terry Spinks, Dale Bailey and Sylke Grootoonk for their numerous contributions to 3-D PET methodology; to Dr. Henri
Tochon-Danguy, Martin Wensveen, Tony Brun, Raymond
Roddy, Larry Byars and Anne Christin who help to design,
build and operate the first rotating PET scanner; and to Drs.
David Brasse and Claude Comtat for their contributions to
the PET/CT development project while at the University of
Pittsburgh; and finally to Drs. Jeffrey Yap, Jonathan Carney
and Nathan Hall from the Departments of Medicine and Radiology, University of Tennessee Graduate School of Medicine,
Knoxville for their friendship and continuing contributions
to the development and application of PET/CT.
The development of the Partial Ring Tomograph was supported by the Swiss government Commission for Scientific
Research (CERS). The PET/CT development project is supported by National Cancer Institute grant CA 65856.
References
1. Townsend, D.W.; Spinks, T.J.; Jones, T.; et al. Three-dimensional reconstruction of PET data from a multi-ring
camera. IEEE Trans. Nucl. Sci. 36:1056–1065; 1989.
2. Cherry, S.R.; Dahlbom, M.; Hoffman, E.J. Three-dimensional positron emission tomography using a conventional
multislice tomograph without septa. J. Comput. Assist.
Tomogr. 15:655–668; 1991.
3. Townsend, D.W.; Geissbuhler, A.; Defrise, M.; et al.
Fully three-dimensional reconstruction for a PET camera with retractable septa. IEEE Trans. Med. Imaging 10:
505–512; 1991.
4. Beyer, T.; Townsend, D.W.; Brun, T.; et al. A combined
PET/CT scanner for clinical oncology. J. Nucl. Med.
41:1369–1379; 2000.
5. Lardinois, D.; Weber, W.; Hany, T.F.; et al. Staging of nonsmall-cell lung cancer with integrated positron emission
tomography and computed tomography. N. Engl. J. Med.
348:2500–2507; 2003.
6. Makhija, S.; Howden, N.; Edwards, R.; et al. Positron emission tomography/computed tomography imaging for the
detection of recurrent ovarian and fallopian tube carcinoma: a retrospective review. Gynecol. Oncol. 85:53–58;
2002.
7. Hawkes, D.J.; Hill, D.L.; Hallpike, L.; et al. Coregistration
of structural and functional images. In: Valk, P.; Bailey,
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
D.L.; Townsend, D.W.; Maisey, M.; eds. Positron Emission
Tomography: Basic science and clinical practice. SpringerVerlag, New York; 2003:181–198.
Schorr, B.; Townsend, D. Filters for three-dimensional
limited-angle tomography. Phys. Med. Biol. 26:305–312;
1981.
Schorr, B.; Townsend, D.; Clack, R. A general method
for three-dimensional filter computation. Phys. Med. Biol.
28:1009–1019; 1983.
Jeavons, A.P.; Hood, K.; Herlin, G.; et al. The high density
avalanche chamber for positron emission tomography.
IEEE Trans. Nucl. Sci. 30:640–645; 1983.
Townsend, D.W.; Frey, P.; Jeavons, A.P.; et al. High density
avalanche chamber (HIDAC) positron camera. J. Nucl.
Med. 28:1554–1562; 1987.
Spinks, T.J.; Jones, T.; Bailey, D.L.; et al. Physical performance of a positron tomograph for brain imaging with
retractable septa. Phys. Med. Biol. 37:1637–1655; 1992.
Townsend, D.; Wensveen, M.; Byars, L.; et al. A rotating
PET using BGO block detectors: design, performance and
applications. J. Nucl. Med. 34:1367–1376; 1993.
Townsend, D.W.; Byars, L.G.; Defrise, M.; et al. Design and
performance of a rotating PET scanner PRT-2. IEEE Medical Imaging Conference Record 2:1058–1062; 1994.
Brownell, G.L.; Burnham, C.A.; Chesler, D.A.; et al. Transverse section imaging of radionuclide distributions in
heart, lung and brain. In: Reconstruction tomography
in diagnostic radiology and nuclear medicine. (Ter-Pogossian, M.M.; Phelps, M.E.; Brownell, G.L.; Eds) Baltimore:
University Park Press; 1977:293–307.
Muehllehner, G.; Atkins, F.; Harper, P.V. Positron camera
with longitudinal and transverse tomographic capabilities.
In: Medical radionuclide imaging, volume 1 STI/PUB/
44C. Vienna: IAEA; 1977:391–407.
Cho, Z.H.; Farukhi, M.R. Bismuth germanate as a potential scintillation detector in positron cameras. J. Nucl.
Med. 18:840–844; 1977.
Karp, J.S.; Muehllehner, G.; Mankoff, D.A.; et al. Continuous-slice PENN-PET: a positron tomograph with volume
imaging capability. J. Nucl. Med. 31:617–627; 1990.
Karp, J.S.; Muehllehner, G.; Geagan, M.J.; et al. Wholebody PET scanner using curve-plate NaI(Tl) detectors. J.
Nucl. Med. 39:50P; 1998[Abstract].
Lim, C.B.; Chu, D.; Kaufman, L.; et al. Initial characterization of a multiwire proportional chamber positron
camera. IEEE Trans. Nucl. Sci. 22:388–394; 1975.
McKee, B.T.A. Towards high-resolution positron emission
tomography for small volumes. In: Positron annihilation. Proceedings of the Sixth International Conference
on Positron Annihilation, University of Texas at Arlington,
955–964; April 1982.
Del Guerra, A.; Lum, G.K.; Perez-Mendez, V.; et al. The
HISPET Project: state of the art. In: Positron annihilation.
Proceedings of the Seventh International Conference
on Positron Annihilation, New Delhi, India: 810–812,
1985.
Lanza, R.C.; Osborne, L.S.; Holman, B.L.; et al. Mesh
chambers for positron emission tomography. IEEE Trans.
Nucl. Sci. NS-33:482–485; 1986.
From 3-D PET to 3-D PET/CT / Townsend
24. Ott, R.J.; Marsden, P.K.; Flower, M.; et al. Clinical PET with
a large area multiwire proportional chamber PET camera.
Nuclear Instruments and Methods in Physics Research.
A269:436–442; 1988.
25. Frey, P.E.; Townsend, D.W.; Flattet, A.; et al. Tomographic Imaging of the human thyroid using 124I. J. Clin.
Endocrinol. Metab. 63:918–927; 1986.
26. Colsher, J.G. Fully three-dimensional positron emission
tomography. Phys. Med. Biol. 25:103–115; 1980.
27. Clack, R.; Townsend, D.; Jeavons, A. Increased sensitivity
and Field of View for a Rotating Positron Camera. Phys.
Med. Biol. 29:1421–1431; 1984.
28. Brookes, R.A.; Di Chiro, G. Principles of computer assisted
Tomography (CAR) in Radiographic and Radioisotopic
Imaging. Phy. Med. Biol. 21:689–732; 1976.
29. Honer, M.; Brühlmeier, M.; Missimer, J.; et al. Dynamic
Imaging of Striatal D2 Receptors in Mice using QuadHIDAC PET. J. Nucl. Med. 45:464–470; 2004.
30. Paans, A.M.J.; Vaalburg, W.; Woldring, M.G. A rotating
double-headed positron camera. J. Nucl. Med. 26:1466–
1471; 1985.
31. Williams, C.W.; Crabtree, M.C.; Burgiss, S.G. Design and
performance characteristics of a positron emission computed axial tomograph, ECAT II. IEEE Trans. Nucl. Sci.
26:619–627; 1979.
32. Casey, M.E.; Nutt, R. A Multicrystal two-dimensional BGO
detector system for positron emission tomography. IEEE
Trans. Nucl. Sci. 33:460–463; 1986.
33. Spinks, T.J.; Jones, T.; Gilardi, M.C.; et al. Physical performance of the latest generation of commercial positron
scanners. IEEE Trans. Nucl. Sci. 35:721–725; 1988.
34. Defrise, M.; Kuijk, S.; Deconinck, F. A new three-dimensional reconstruction method for positron cameras using
plane detectors. Phys. Med. Biol. 33:43–51; 1988.
35. Kinahan, P.E.; Rogers, J.G. Analytic three-dimensional
image reconstruction using all detected events. IEEE
Trans. Nucl. Sci. 36:964–968; 1990.
36. Bailey, D.L.; Jones, T.; Spinks, T.J.; et al. Noise equivalent
count measurements in a neuro-PET scanner with retractable septa. IEEE Trans. Med. Imag. 10:256–260; 1991.
37. Townsend, D.W.; Byars, L.; Defrise, M.; et al. Rotating
positron tomographs revisited. Phys. Med. Biol. 39:401–
410; 1994.
38. Bailey, D.L.; Spinks, T.J.; Jones, T. A method for measuring the absolute sensitivity of positron emission scanners.
Eur. J. Nucl. Med. 18:374–379; 1991.
39. Wienhard, K.; Eriksson, L.; Grootoonk, S.; et al. Performance evaluation of the positron scanner ECAT EXACT.
J. Comput. Assist. Tomogr. 16:804–813; 1992.
40. deKemp, R.A.; Nahmias, C. Attenuation correction in PET
using single photon transmission measurement. Med. Phys.
21:771–778; 1994.
41. Bailey, D.L.; Young, H.; Bloomfield, P.M.; et al. ECAT
ART - a continuosly rotating PET camera: performance
characteristics, inital clinical studies, and installation considerations in a nuclear medicine department. Eur. J.
Nucl. Med. 24:5–15; 1997.
42. Townsend, D.W.; Beyer, T.; Meltzer, C.C.; et al. The ECAT
ART scanner for Positron Tomography: 2. Research and
Clinical Applications. Clin. Posit. Imaging 2:17–30; 1999.
289
43. Watson, C.C.; Newport, D.; Casey, M.E. A single scatter
simulation technique for scatter correction in three-dimensional PET. In: Grangeat, P.; Amans, J.L.; eds.Threedimensional image reconstruction in radiology and nuclear medicine. Dordrecht, The Netherlands: Kluwer,
255–268, 1996.
44. Ollinger, J.M. Model-based scatter correction for fully
3-D PET Phys. Med. Biol. 41:153–176; 1996.
45. Townsend, D.W.; Isoardi, R.A.; Bendriem, B. Volume
Imaging Tomographs. In: Townsend, D.W., Bendriem, B.;
eds. The Theory and Practice of 3-D PET. Dordrecht, The
Netherlands: Kluwer Academic Publishers. 111–132, 1998.
46. Melcher, C.L.; Schweitzer, J.S. Cerium-doped lutetium oxyorthosilicate: a fast, efficient new scintillator. IEEE Trans.
Nucl. Sci. 39:502–505; 1992.
47. Halama, J.R.; Sajdak, R.A.; Henkin, R.E. Performance
measurements utilizing the NEMA NU-2 protocol on
the ALLEGRO GSO 3-D PET scanner for 17F and 82Rb.
J. Nucl. Med. 44:111P; 2003[Abstract].
48. Valk, P.E.; Casey, M.E.; Bruckbauer, T.; et al. Clinical evaluation of an LSO PET system for 3-D whole-body imaging.
J. Nucl. Med. 42:309P; 2001[Abstract].
49. Takagi, K.; Fukazawa, T. Cerium-activated Gd2SiO5 single
crystal scintillator. Applied Physics Letters. 42:43–45; 1983.
50. Watson, C.C. New, faster image-based scatter correction
for 3-D PET. In: IEEE Nuclear Science Symposium and
Medical Imaging Conference. Toronto, Canada, November 1999:1637–1641.
51. Defrise, M.; Kinahan, P.E.; Townsend, D.W.; et al. Exact
and approximate rebinning algorithms for 3-D PET
data. IEEE Trans. Med. Imaging 16:145–158; 1997.
52. Comtat, C.; Kinahan, P.E.; Defrise, M.; et al. Fast reconstruction of 3-D PET data with accurate statistical modeling. IEEE Trans. Nucl. Sci. 45:1083–1089; 1998.
53. Pelizzari, C.A.; Chen, G.T.; Spelbring, D.R.; et al. Accurate three-dimensional registration of CT, PET, and/or
MR images of the brain. J. Comput. Assist. Tomogr. 13:20–
26; 1989.
54. Pietrzyk, U.; Herholz, K.; Heiss, W.D. Three-dimensional
alignment of functional and morphological tomograms.
J. Comput. Assist. Tomogr. 14:51–59; 1990.
55. Woods, R.P.; Mazziotta, J.C.; Cherry, S.R. MRI-PET registration with automated algorithm. J. Comput. Assist.
Tomogr. 17:536–546; 1993.
56. Wahl, R.L.; Quint, L.E.; Cieslak, R.D.; et al. Anatometabolic tumor imaging: fusion of FDG PET with CT or MRI
to localize foci of increased activity. J. Nucl. Med. 34:1190–
1197; 1993.
57. Hasegawa, B.H.; Stebler, B.; Rutt, B.K.; et al. A prototype
high-purity germanium detector system with fast photoncounting circuitry for medical imaging. Med. Phys. 18:
900–909; 1991.
58. Lang, T.F.; Hasegawa, B.H.; Soo Chin, L.; et al. Description of a prototype emission-transmission computed tomography imaging system. J. Nucl. Med. 33:1881–1887;
1992.
59. Hasegawa, B.H.; Gould, R.G.; Lang, T.F.; et al. Emissiontransmission imaging system using single energy and dual
290 Molecular Imaging and Biology, Volume 6, Number 5
60.
61.
62.
63.
energy transmission and radionuclide emission data. US
Patent No. 5376795, 1994.
Charron, M.; Beyer, T.; Bohnen, N.; et al. Image analysis
in patients with cancer studies with a combined PET and
CT scanner. Clin. Nucl. Med. 25:905–910; 2000.
Meltzer, C.C.; Martinelli, M.A.; Beyer, T.; et al. Whole-body
FDG PET imaging in the abdomen: value of combined
PET/CT. J. Nucl. Med. 42:35P; 2001[Abstract].
Meltzer, C.C.; Snyderman, C.H.; Fukui, M.B.; et al. Combined FDG PET/CT imaging in head and neck cancer:
impact on patient management. J. Nucl. Med. 42:36P;
2001[Abstract].
Kluetz, P.G.; Meltzer, C.C.; Villemagne, M.D.; et al. Combined PET/CT imaging in oncology: impact on patient
management. Clin. Posit. Imaging 3:1–8; 2001.
64. Townsend, D.W.; Beyer, T.; Blodgett, T.M. PET/CT Scanners: A Hardware Approach to Image Fusion. Sem. Nucl.
Med. XXXIII(3):193–204; 2003.
65. Goerres, G.W.; Kamel, E.; Heidelberg, T.-N.H.; et al.
PET/CT image co-registration in the thorax: influence of
respiration. Eur. J. Nucl. Med. 29:351–360; 2002.
66. Goerres, G.W.; Burger, C.; Schwitter, M.R.; et al. PET/CT
of the abdomen: optimizing the patient breathing pattern.
Eur. Radiol. 13:734–739; 2003.
67. Beyer, T.; Antoch, G.; Blodgett, T.; et al. Dual modality
PET/CT imaging: the effect of respiratory motion on combined image quality in clinical oncology. Eur. J. Nucl.
Med. 30:588–596; 2003.
68. Carney, J.P.; Beyer, T.; Brasse, D.; et al. Clinical PET/CT
scanning using oral CT contrast agents. J. Nucl. Med.
45:57P; 2002[Abstract].