Download Quantal and visual efficiency of fluorescence in the lens of

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Human eye wikipedia , lookup

Contact lens wikipedia , lookup

Cataract wikipedia , lookup

Corrective lens wikipedia , lookup

Eyeglass prescription wikipedia , lookup

Transcript
Quantal and Visual Efficiency of Fluorescence
in the Lens of the Human Eye
Thomas]. T. P. van den Berg
Purpose. To document quantitatively the fluorescence in the human lens relevant to its interference with visual function. To explain quantitatively the experimental findings relative to loss of
visual function. To study the relation between the fluorescence and the light transmission of
the lens.
Methods. Three normal lenses, from 22-, 28-, and 69-year-old donors, were used. Fluorescent
light was induced by a 4-mm diameter pencil beam of 380, 400, or 420 nm. It was measured as
a function of the deflection angle from —10 to 150 degrees for different wavelengths.
Results. The shapes of the emission spectra were comparable to those reported in the literature. Total quantal efficiency of fluorescence was between 5% for 69 years and 380 nm excitation and 0.4% for 22 years and 420 nm excitation. The forward intensity was less than the
backward intensity because of secondary absorption.
Conclusions. Fluorescence of the lens causes light with wavelengths of 420 nm and lower to be
much more visually effective. A marked homogeneous veil is added to the point spread function. The total increase in luminous efficiency was a factor 3 to 6 at 400 nm, and a factor 70 to
150 at 380 nm. For other visual effects (glare) the increase can be larger. Invest Ophthalmol
VisSci. 1993;34:3566-3573.
JL luorophores in the human lens transform (barely
visually effective) ultraviolet (UV) light into visually
more effective light. This is because the emitted wavelengths are longer than the wavelength of the exciting
light (Stoke's rule). This effect has long been realized.1"3 Fluorescence may be assumed to add a uniform component to the retinal point spread function
(PSF), which is strongly dependent on wavelength. It
adds to the uniform component already present 4 due
to translucency of the eye wall, 56 and other yet unidentified causes.
The human lens may contain many fluorophores
(see references 7 and 8 for reviews). The chief fluorophore is tryptophan (a basic constituent of protein)
with excitation maximum at 290 nm and emission
maximum at 340 nm. 9 The young lens has a second
major component with excitation maximum at 360 nm
and emission at 440 nm. 1011 When induced by longer
wavelengths, the emission also shifts to longer wave-
From The Netherlands Ophthalmic Research Institute and the Laboratory of
Medical Physics and Informatics, University of Amsterdam.
Submitted for publication: November 30, 1992; accepted February 10, 1993.
Proprietary interest category: N.
Reprint requests: T. J. T. P. van den Berg, Laboratory of Medical Physics and
Informatics, Meibergdreef 15, 1105 AZ Amsterdam, The Netherlands.
3566
lengths.12 For young lenses the excitation limit for
measurable fluorescence was 500 nm.13 With aging,
fluorescent intensity increases11 and fluorophores
emerge that emit at even longer wavelengths.1011-14
For older lenses the excitation limit was 650 nm.13
A typical UV source is a so-called black light (UVemitting low-pressure mercury lamp). A black light not
only causes fluorescent effects on clothes and other
objects, but alsofillsthe entire visual field with nondescript light. This light might be assumed to originate
from the fluorescence of the lenses of our eyes. Because fluorescence is essentially isotropic, the light
seems to have no origin. Also on other occasions fluorescence interferes with vision.3 Night driving may become important in the future. The automobile industry is considering the use of UV headlights to make
traffic objects visible by their fluorescence. Because
the luminous efficiency function is very low for UV
(phakic eyes), UV headlights supposedly give no glare
problems from oncoming traffic. But smallfields,eg, 1
degree radius,15 were used to derive the luminous efficiency curve. Within 1 degree only 0.015% of isotropic
radiation would be collected. Because fluorescence is
isotropic, the luminous efficiency curve could seriously underestimate the visual effect of UV.
Investigative Ophthalmology & Visual Science, December 1993, Vol. 34, No. 13
Copyright © Association for Research in Vision and Ophthalmology
Downloaded From: http://iovs.arvojournals.org/pdfaccess.ashx?url=/data/journals/iovs/933172/ on 06/17/2017
Efficiency of Lens Fluorescence
3567
In a recent article, Zuclich et al12 experimentally
verified in monkeys that fluorescence of the lens interferes with visual function. They projected a 413 nm
laser beam onto the lens at 45 degrees to the visual axis
while the foveal visual evoked potential was recorded.
As a result the visual evoked potential declined. Assuming 413 nm to be entoptically scattered according to
Vos's white light model,16 scattering could not explain
the decline. Because 413 nm light is not only scattered,
but also gives fluorescence, it was concluded that fluorescence was the cause.
The current study aimed to gain quantitative insight into the visual effect of lens fluorescence. Specifically, the uniform component added to the retinal PSF
was sought. Intact donor eye lenses were used. We
wanted to determine the intensity of fluorescent light
emitted by the lens in the forward direction (the direction of the retina). But in this direction the excitation
light interferes with the measurement. This is why
standard fluorescence measurements are conducted
at right angles. But with intact lenses, light deflected at
right angles travels laterally through 6 mm of the lens
toward the optically less well-defined equator. This
light would not properly represent the fluorescence
emitted toward the retina. The problem was solved by
making a scan around the lens. From these measurements the uniform component of the PSF could be
estimated. Taking the wavelength dependencies into
consideration, the disturbance of visual function can
be understood and quantified from the modified
PSFs.
METHODS
Three lenses were used (lenses 1,2, and 3, from donors aged 28, 69, and 22 years, respectively), forming
part of a larger study17 on light scattering in donor
lenses. Eyes with short postmortem enucleation times
(8, 1, and 6 hours, respectively) and no potential damage to the lens (eg, no trauma to the head) were obtained from the Hoornvliesbank (cornea bank) in Amsterdam. The lenses were carefully extracted and
mounted in a special holder with a free diameter of 8
mm, and placed between, but not in contact with, specially selected cover glasses that were renewed for each
lens. As immersion medium the isotonic solution (Na+
77, K+ 4, Ca2+ 2.2, Cl" 85 and Glucose 139 mmol/1,
index of refraction 1.336) was chosen.17 The mounted
lenses were checked using a slit lamp. They were normal for their age, but postmortem changes (some superficial irregularity, spokelike structures) could not
be excluded. The 69-year-old lens showed sclerosis including early irregularities in the nucleus. Measurements were performed during a few hours (2,4, and 4
hours, respectively) directly after extraction of the
lens. A first scattering measurement was repeated at
the end of the measurements to check that no change
in scattering characteristics had taken place during the
experiment.
Figure 1 is a simplified drawing of the setup (not
to scale). The hot spot of a high-pressure mercury
lamp was focused on diaphragm 1. The light passed
through (excitation) interference filter 1 (half band
width 10 nm). The beam divergence at the site of this
interference filter was approximately 0.2 degrees. This
is small enough for proper functioning of the interference filter, with no need for collimation. For lens 1
excitation wavelength 400 nm was used, for lenses 2
and 3 excitation wavelengths 380, 400, and 420 nm
were used. Diaphragm 1 was focused in front of the
lens (small cross in Fig. 1). The divergence of the beam
was adjusted by means of diaphragm 2, to obtain a
Lens3 &
Diaphragm3
Mercury
Lamp
Lensi
Diaphragmi
Lens2 &
Diaphragm2
Interf. PhotoFilter2 multiplier
Donor
Lens
Interf
Filteri
FIGURE l. Schematic drawing of the setup, not to scale. The combination of lens 3, diaphragm
3, interference filter 2, and photomultiplier could rotate, with the donor lens as the center of
rotation.
Downloaded From: http://iovs.arvojournals.org/pdfaccess.ashx?url=/data/journals/iovs/933172/ on 06/17/2017
3568
Investigative Ophthalmology & Visual Science, December 1993, Vol. 34, No. 13
beam width of 4 mm diameter at the site of the donor
lens. Shifting lens 2 of the setup, the distance between
focus (small cross) and donor lens was adjusted so that
a sharp image (radius about 1 mm) was obtained at 550
mm behind the donor lens. A 9 mm radius diaphragm
(diaphragm 3) was positioned at this point. This diaphragm could rotate with the donor lens as center of
rotation. With the position of the image defined as 0
degrees, the diaphragm was rotated from - 1 0 degrees
to 150 degrees. Light collected by this diaphragm was
fed into a calibrated photomultiplier with a calibrated
interference filter in front. Analysis wavelengths of
400, 420, 440, 460, 480, 500, 520, 540, 562, 602, and
700 nm were used, all with half band width of about 10
nm. Not all these wavelengths were used for all series.
The photomultiplier was used as photon counter.
After each experiment the donor lens was removed
from the holder and the measurement was repeated
for each excitation wavelength used (analysis filter removed), with diaphragm 3 at 0.645 mm radius. In the
absence of the donor lens, no imaging of the focus
(small cross) in diaphragm 3 took place. The beam divergence was approximately 3 degrees radius. This
scan was used to calculate (by integration over solid
angle) the total number of excitation quanta/second
(q/s; Table 1). For 400 nm, log(q/s) was 12.57, 12.24,
and 12.54 for donor lenses 1, 2, and 3, respectively.
For 380 nm these values were 1.13 log unit lower and
for 420 nm 0.18 log unit higher. To derive log(lumens), the log(q/s) values have to be decreased by
19.86, 18.87, and 17.89 log units for 380, 400, and
420 nm, respectively.15
Unwanted refraction, reflection (limit angle of a
Ringer-air interface is about 48 degrees), and screening effects occurred in the holder. Fluorescence is basically isotropic, so the fluorescence scans should be
TABLE l.
flat. Figure 2 shows the actual shape of two fluorescence scans. They are from lens 1 at excitation wavelength 400 nm and analysis wavelengths 420 (large
peak) and 480 nm (small peak). Although interference
filters with very good side band suppression were
used, also a small part of nonfluorescent light was
transmitted to the photomultiplier. This is seen as the
peak at 0 degrees. It is small for 480 nm. Thus, for 480
nm, an absolute value for the fluorescent intensity in
the zero direction could be estimated by interpolation
of the flat portions around —10 and 10 degrees. For
wavelengths closer to the excitation wavelength (420
nm in the example of Fig. 2), the peak is much larger.
To arrive at a proper estimate in the zero direction,
the 420 nm curve was shifted vertically so that the part
of the forward branch outside the peak coincides with
that of the 480 nm curve. (The peak could have been
reduced if a second set of filters would have been available to improve the side band suppression.) Because
the holder was opaque, at 90 degrees the dark current
from the photomultiplier was measured. Taking into
account the size of the diaphragm, the spectral transmission characteristics of the respective filters, the
spectral quantal efficiency curve of the photomultiplier, the vertical shift, and the number of counts/s at
480 nm, for each analysis wavelength the number of
quanta/s X nm due to fluorescence emitted by the
donor lens in the forward half space (in the direction
of the retina) was derived. For this derivation, fluorescent emission was assumed to be independent of angle, implying that emission in the forward half space
was calculated from forward (0 degrees) quanta/
s X steradian by multiplication with 27r. But, at larger
angles, fluorescent light travels a larger distance
through the lens. This will give extra losses. Conversely, in vivo, the light travels through vitreous hu-
The Data Set Derived from the Scans
Forward Fluorescent Emission log(q/s X nm)
Forward/Backward Ratio in Fluorescent Emission,
relative to 480 nm (log units)
Lens 1 Lens 2 Lens 2 Lens 2 Lens 3 Lens 3 Lens 3 Lens 1 Lens 2 Lens 2 Lens 2 Lens 3 Lens 3 Lens 3
Excitation
400
380
400
420
400
420
420
380
400
420
400
380
380
wavelength 400
Log q/s
12.57 12.24 12.43 11.12 12.54 12.70 11.41
Wavelength
(nm)
400
5.20
6.42
-1.53
-0.36
420
7.59
7.08
6.64
7.57
7.04 -0.14 -0.53
-0.64 -0.21
-0.26
440
8.08
7.81
7.82
7.23
8.05
7.98
7.44 -0.05 -0.22 -0.11 -0.31 -0.10
0.00 -0.06
460
8.12
7.96
7.96
7.38
8.08
8.03
7.56 -0.02 -0.11 -0.08 -0.21 -0.03
0.03
0.00
480
8.25
8.20
8.31
7.55
8.20
8.21
7.63
0.00
0.00
0.00
0.00
0.00
0.00
0.00
500
8.15
8.23
8.34
8.14
7.51 -0.03
0.09
0.02
0.00 -0.05
520
8.15
8.36
8.10 ' 8.18
0.00
0.04
-0.07
0.06
540
8.00
8.12
8.32
8.10
0.01
0.16
0.08
0.00
562
7.85
0.00
602
7.52
7.68
0.00
0.18
700
6.31
Downloaded From: http://iovs.arvojournals.org/pdfaccess.ashx?url=/data/journals/iovs/933172/ on 06/17/2017
3569
Efficiency of Lens Fluorescence
25
50
75
100
measurement angle (degree)
125
150
175
of the donor lens, it follows that the ratio between the
forward and backward branch is T1-2f. For example, if
the ratio between the forward and backward branch is
0.2, and the total transmission of the donor lens is
0.01 (this could be the case at 400 nm), then (1 - 2f) X
log(0.01) = log(0.2), therefore f = 0.33. If the thickness of the donor lens is 4 mm, the effective (mean)
depth of emission would be 1.3 mm. Note that the
actual data tabulated are the vertical distances between the backward branches, as in Figure 2, with the
forward branches shifted on top of each other. So, the
actual data are relative to 480 nm. They equal logfT1"
2f
(wavelength)/T1-2f(480 nm)). (Throughout this article Briggsian (base 10) logarithms are used.)
FIGURE 2. Fluorescent intensity from an isolated human
lens, measured as a function of deflection angle. Excitation
wavelength 400 nm. Emission wavelengths 480 nm (lower
curve) and 420 nm (upper curve). The 420 nm curve was RESULTS
shifted vertically to coincide around 50 degrees with the 480
nm curve. The peak at zero degrees is an artefact (see text).
In Table 1 the data set derived from the scans is shown.
The backward branch is relatively higher for 420 nm beAn estimate for total fluorescent light that in vivo
cause (re)absorption is higher for 420 as compared to 480
would reach the retina is given as explained in the
nm (see text).
mor instead of air. Because of the refraction effects
(ratio of the index of refraction of vitreous humor
versus air is 1.336) involved, fluorescent emission is
underestimated by 1.336 squared. No corrections
were made for these opposing effects. The logarithms
are given in Table 1. These data are plotted on a linear
scale, relative to the peak values, in Figure 4.
It should be realized that the excitation light is not
absorbed homogeneously throughout the donor lens,
even if chromophores/fluorophores were homogeneously distributed. The (short wavelength) excitation
light weakens with penetration in the lens. It is absorbed more at the anterior side of the lens as compared to the posterior side. At the site of absorption,
the emitted light is isotropic. However, the emitted
light subsequently travels through part of the donor
lens and will be partly reabsorbed. The reabsorption
loss will depend on the distance traveled. So, the light
emitted in the forward direction is reabsorbed more as
compared to light emitted in the backward direction.
This is why the backward branches in Figure 2 do not
coincide. This phenomenon can be used to derive the
(effective) depth of emission. For this purpose, the
vertical distance between the backward branches (with
the forward branch on top of the 480 nm forward
branch, as in Fig. 2) was measured. In Figure 2 this
distance was —0.14 log units. These values are given in
Table 1. The fraction of the depth (more precisely: the
fraction of the pigment) at which emission takes place
is called f. Backward emitted light must travel through
the fraction f. Forward emitted light must travel
through the fraction 1-f. If T is the total transmission
Methods section. The log ratio between forward and
backward fluorescent emission (relative to 480 nm) is
given. Negative values indicate backward emission to
be relatively stronger. In Figure 3 absolute results are
shown (not relative to 480 nm) for the ratio of forward
fluorescent emission to backward fluorescent emission, as a function of wavelength. As explained in
Methods these data can be interpreted as T1"2^ Lens 1
(28 years of age) is at the top, lens 2 (69 years of age) is
in the middle, and lens 3 (22 years of age) is at the
bottom. The lines drawn compare with literature data
on lens transmission. Wyszecki and Stiles15 reviewed
the literature on light losses in the eye. They used the
spectral transmission of all eye media to represent the
spectral transmission of the lens. They tabulated the
data of Van Norren and Vos as best estimate for the
young human eye lens.18 The lines drawn in Figure 3
are derived from their tabulation of these data15 (let us
call this "standard" transmission Ts; at 400 nm log(Ts)
= —1.45; at the long wavelength side log(Ts) = 0). Lens
transmission was measured separately17 for wavelengths 400, 500, 602, and 700 nm. We call the transmission of the present donor lenses T{ (i = 1, 2, 3).
With log(Tj) shifted to 0 for the mean of the 602 and
700 nm results, logCTj) for 400 nm was -1.35, -2.76
and —1.28. The differences signify differences in lens
pigmentation among persons. One might assume the
ratio log(T;(400 nm))/log(Ts(400 nm)) (= 0.93, 1.90
and 0.88) to be equal to the ratios between the amount
of chromophore present in the individual lens as compared to the literature18 mean. This ratio was used to
derive T; for the other wavelengths: log(Tj(wavelength)) = log(Ti(400 nm))/log(Ts(400 nm)) X log(Ts
(wavelength)). Using this one-component multiplica-
Downloaded From: http://iovs.arvojournals.org/pdfaccess.ashx?url=/data/journals/iovs/933172/ on 06/17/2017
3570
Investigative Ophthalmology 8c Visual Science, December 1993, Vol. 34, No. 13
380 and 420 nm data are multiplied by these constants
(and shifted vertically). The correspondence with the
400 nm data is rather precise. The continuous curves
resulted from a fit as above. For lens 2 the continuous
curve is 0.99 X log(TJ (f = 0.24, f1/e = 0.16), for lens 3
the continuous curve is 0.20 X log(Ts) (f = 0.39,
fVe = 0-34).
Table 2 gives an overview of these and other derived parameters. Values are given in parentheses for
some of the 380 and 420 nm data. This is because
these data were derived using an inferred, not measured value for Tat these wavelengths. The value for T
was derived using the above one-component multiplicative model for lens transmission. Log(Ts(380 nm)) =
-2.61,log(Ts(420nm)) = -0.59. This gives log(T2(380
-0.4;
nm))
= -4.96, log(T2(420 nm)) = -1.12, log(T3(380
500
550
600
700
400
nm)) = -2.31, log(T3(420 nm)) = -0.52.
(analysis) wavelength (nm)
In Figure 4 the emission spectra for the forward
FIGURE 3. Differences in intensity (due to reabsorption) of
direction
are shown (symbols). The continuous lines
fluorescent light emitted by a human lens in the forward and
are
replots
of Figures 2 and 3 from Zuclich et al12 for
the backward direction. Excitation wavelengths as indicated.
Ages: above, 28 years; middle, 69 years; below, 22 years. Each 360 and 413 nm excitation. At the peaks log(emission
data point followed from a comparison between backward
q/s X nm) — log(excitation q/s) were: lens 1 (400 nm)
branches as depicted in Figure 2.
-4.32, lens 2 (380, 400 and 420 nm) -3.57, -4.02
and -4.07 and lens ^ (380, 400 and 420 nm) -3.79,
—4.34 and —4.51 log(l/nm). The two curves from Zuclich et al12 have more or less the same shape. The
tive model for lens transmission, the formula derived
present data seem to follow the same shape also. For
in the methods section becomes:
each excitation wavelength the 413 nm curve was
shifted horizontally to coincide with the data set.
(l-2f) X (log(Ti(400 nm))/1.45) X log(Ts(wavelength))
These shifted curves were used to arrive at Figure 5
(see below). They were also used to estimate the quanThis formula was fitted to the data of Table 1 (+ a
tal
efficiency of the fluorescence. To arrive at the total
constant representing the 480 nm result) using the
number
of fluorescence quanta generated at the site
least-squares criterion. The constant was simultaof
emission,
a correction had to be made for the abneously fitted. The result is shown in Figure 3. For lens
1 (400 nm excitation only) (l-2f) X 0.93 X log(Ts)
(upper continuous line) was fitted to the data + a constant (symbols). The fit gave f = 0.36 (so, the continuous line is 0.27 X log(Ts)), and -0.02 for the constant.
Compare this value f = 0.36 to the 1/e penetration
depth for 400 nm, that follows from logOTj) = -1.35:
f1/e = 0.32.
With lenses 2 and 3, in addition to 400 nm, measurements were performed with 380 and 420 nm. It
should be realized that the 380 nm excitation light
penetrates less far, whereas the 420 nm excitation
light penetrates farther into the lens. If the analysis
presented above is correct then the log forward/backward ratios for 380 and 420 nm must have the same
wavelength dependence as for 400 nm, apart from a
500
550
600
450
700
400
650
analysis wavelength (nm)
multiplication (and an addition) constant. To check
this, the 380 and 420 nm data were fitted to the 400
FIGURE 4. Data points: relative fluorescence spectra from
nm data, estimating these constants. For 380 nm the
human lenses for emission toward the retina. Excitation
multiplication constant was 1.23 and 1.30 respecwavelengths as indicated. Ages: above, 28 years; middle, 69
tively. For 420 nm it was 0.52 for lens 2. The absorpyears; below, 22 years. Continuous curves are replots from
tion in lens 3 was too weak for the estimation of a 420
Zuclich et al12 for 360 nm excitation (left) and 413 nm excitation (right).
nm multiplication constant (set to 1). In Figure 3 the
Downloaded From: http://iovs.arvojournals.org/pdfaccess.ashx?url=/data/journals/iovs/933172/ on 06/17/2017
Efficiency of Lens Fluorescence
TABLE
3571
2. Main Results
Lens 1
Age (yr)
Excitation wavelength (nm)
Log relative transmission at excitation wavelength
1/e fractional depth of excitation intensity
Mean fractional depth of light absorption
Mean fractional depth of fluorescent emission (f)
Log quantal efficiency of fluorescence
Log visual efficiency of fluorescence
28
400
-1.35
0.32
0.28
0.36
-2.19
0.44
sorption losses occurring between the site of emission
and the posterior pole of the lens. For this purpose the
curves were divided by Tj(1~f). The resulting curves
were integrated over wavelength and multiplied by 2,
to allow for emission toward the backward halfspace.
Log(total number of emitted quanta thus derived) —
log(number of incident quanta) gave as result: lens 1
(400 nm) -2.19, lens 2 (380, 400 and 420 nm) -1.30,
-1.82, and -1.95, lens 3 (380, 400 and 420 nm)
-1.61, -2.21, and-2.41.
In Figure 5 a measure for the visual effectiveness
of the fluorescence is given. For this purpose the intensities (q/s) were transformed into radiometric units
(Watts) and these were transformed into photometric
units (lumens).15 The last step involves multiplication
by the luminous efficiency function. But this function
is defined for light outside the (standard) eye, whereas
the emission intensity was measured behind the eye
lens. A correction for the transmission Ts of the (standard) eye lens therefore had to be made. Plotted is:
(emission lumens/nm)/Ts (emission wavelength) X
(excitation lumens). These curves are intended to give
by integration the ratio (weighted for photopic vision)
0.2-
450
500
550
600
analysis wavelength (nm)
650
700
FIGURE 5. Fluorescence spectra of human lenses weighted to
represent photopic visual efficiency. The ratios between
emission intensity (per nm) and excitation intensity (both
photopically weighted) are given.
Lens 2
69
400
-2.76
0.16
0.16
0.24
-1.82
0.82
Lens 2
69
420
(-1.12)
(0.39)
(0.31)
(0.36)
-1.95
-0.16
Lens 2
Lens 3
69
22
380
400
(-4.96) -1.28
(0.10)
0.34
(0.10)
0.29
(0.18)
0.39
-1.30
-2.21
0.42
2.17
Lens 3 Lens 3
22
420
(-0.52)
(0.84)
(0.40)
(—)
-2.41
-0.66
22
380
(-2.31)
(0.19)
(0.19)
(0.35)
-1.61
1.83
between the total amount of fluorescent light (totalized over the entire retina) vs the amount of incident
light. The logarithms of these ratios are: lens 1 (400
nm) 0.44, lens 2 (380, 400 and 420 nm) 2.17, 0.82,
and - 0 . 1 6 , lens 3 (380, 400 and 420 nm) 1.83, 0.42,
and - 0 . 6 6 .
DISCUSSION
In Figure 3 the (fractional) transmission spectra
showed good agreement between the different excitation wavelengths. This suggests that the presented
analysis is valid. The validity rests on the implicit assumption that the chromophores responsible for lens
absorption above 400 nm are more or less equally (not
necessarily homogeneously) distributed. The agreement with the curves derived from the literature (valid
for young lenses15) was good for the two young lenses,
but not for the older lens. The transmission of the
older lens increases more gradually. This indicates that
the one-component multiplicative model used to construct log spectral transmission Tor different ages from
the literature18 data was too simple. Indeed, a closer
look at the literature1519 shows older lenses to have the
more gradual shape as found here. This change with
age has already been noted2021 and was used for a
two-component model.21
In the discussion of the f values, it must be realized
that emission does not take place at one site. The
experimentally determined f values represent a
(weighted) mean overall emission site. The f values can
be interpreted as fractional depth in the lens if the
chromophores were distributed homogeneously. However, this is not the case.1114 More precisely, the interpretation of f should be: the fraction of the chromophores passed by the excitation light before arrival at
the emission site. These fractions (0.36, 0.24, and
0.39) were found to be larger than the fractions at
which the penetrating light had decreased to 1/e
(0.32, 0.16, and 0.34). But it should be noted that the
emission does not take place at one site. Briefly, calculation by integration of the mean sites (given as a fraction) for absorption of 400 nm light in the three lenses
Downloaded From: http://iovs.arvojournals.org/pdfaccess.ashx?url=/data/journals/iovs/933172/ on 06/17/2017
3572
Investigative Ophthalmology & Visual Science, December 1993, Vol. 34, No. 13
yielded 0.28, 0.16, and 0.29 (Table 2). These values
are smaller than the experimental values, so it seems
that the fluorescence is not distributed in the same way
as the chromophores. The fluorescence seems to be
concentrated more in the center.
Another indication for dissociation between chromophore and fluorophore can be found in the values
given in Table 2 for quantal efficiency. The maximum
and minimum of these figures —1.30 and -2.41 correspond to 5% and 0.4%, respectively. These values
might be considered rather low for a fluorescent process. But it must be realized what these figures actually
represent: the ratio between the total number of incident quanta (almost equal to absorbed quanta) and the
total number of fluorescent quanta for a whole lens.
So, it is the quantal efficiency for the complete mixture of substances present in the human lens. It might
be that some of these substances are efficient fluorophores, whereas others only contribute to the absorption.
Figure 5 and Table 2 indicate that fluorescence is
visually important at excitation wavelengths of 420 nm
and less. For the intact eye, a correction must be made
because of light losses in the cornea. But this correction is small, about 0.1 log unit at 380 nm (manuscript
in preparation). Above 420 nm the luminous efficiency of the excitation light itself dominates in the
visual effect. Below 420 nm, eg, down to 330 nm, the
fluorophore with peak emission at 360 nm (kynurenine metabolites of tryptophan22"24) is probably responsible. Below 330 nm excitation tryptophan fluorescence may become more important, but the resulting
wavelength ranges are of less visual consequence (high
absorption by the eye media, low sensitivity of the visual pigment, low occurrence in ambient light).
The Results section (Table 2) indicates that for the
young eye stimulated with 380 nm light, fluorescence
adds a visually effective stimulus that is larger by 1.83
log unit (a factor 68) than the 380 nm light itself. At
400 nm the addition is larger by 0.43 log unit (a factor
2.7). At 420 nm the addition is smaller by 0.66 log unit
(a factor 0.22). For the older eye the additions were
more than twice as large: for 400 nm 0.82 log unit (a
factor 6.6), for 380 nm 2.17 log unit (a factor 148),
and for 420 nm —0.16 log unit (a factor 0.69). It must
be noted that these figures are based on the luminous
efficiency function. But this function is not valid for
old eyes. The sensitivity of old eyes to short wavelengths is much lower. So, in fact for old eyes the ratio
in luminous effect between the direct light, and the
fluorescence it causes, is much larger. For example, at
380 nm log(Ts) = -2.61 but log(T2) = -4.96. So, for
this old eye the difference in visual effect between 380
nm light and its fluorescence is 2.17 + 4.96 — 2.61 =•
4.52 log units (a factor 33000).
To evaluate the importance of this added light the
following comparison may help. It is based on a
model4 for the PSF, including parameters for age, pigmentation, and pupil diameter: note that the added
light is homogeneously distributed over the retina,
whereas the normal light follows the PSF.416 The PSF
has a high peak and declines with angle up until 90°.
For the 35-year-old, average white person with a 4-mm
pupil, the PSF has a homogeneous component of its
own5 of 0.133 (—0.88 log unit).4 This homogeneous
component dominates the PSF for angles larger than
10°. If the PSF is compared with a homogeneously
distributed light of some other origin (fluorescence),
the two may cross at some point. For homogeneous
light that is larger than the PSF by 2, 1, or 0 log units,
crossing takes place at 0.92°, 2.2°, and 6.4°, respectively. For example, taking the above value for 420 nm
(the addition is 0.66 log unit smaller), at angles larger
than 10° the PSF for 420 nm light is increased by a
factor of 2.7 on account of fluorescence. From the
functional point of view (glare, contrast sensitivity,
etc.)25"28 this is a serious deterioration of the PSF and
hence of vision. The deterioration increases sharply
with decreasing wavelength (and increasing age).
As mentioned in the introduction, Zuclich et al12
experimentally verified that the fluorescence of rhesus
monkey lenses interferes with visual function. They
projected a 1.5 mW 413 nm laser beam at 45 degrees
to the visual axis onto the lens, while the foveal visual
evoked potential was recorded using a 60 cd/m2
screen. The visual evoked potential declined about
20%. The present data form a basis for the interpretation of such findings. According to Vos's PSF model,16
entoptic scattering of the laser beam would correspond to a veiling equivalent luminance of only 0.5
cd/m2. Fluorescence was therefore proposed12 as an
alternative explanation for the decline. From the
above figures for 400 and 420 nm, an interpolated
value for 413 nm of —0.23 log unit follows. The photopic intensity of the laser beam was —2.75 log lumen.
Plus —.23 gives —2.98 log lumen that entered the eye
to be homogeneously distributed over the retina. Assuming a pupillary diameter of 4 mm (surface —4.90
log m2), this gives a veiling equivalent luminance of
2
10(-2.98 + 4.9oy27r = 12 cd/m . This is 20% of the screen
luminance. It should be realized that the correspondence with the 20% mentioned earlier may partly be
fortuitous in view of the uncertainties present, such as
that human data are used here to explain monkey12
results.
Key Words
human lens, fluorescence, aging, glare, straylight
Acknowledgments
The author thanks the staff of the Hoornvliesbank in Am-.
sterdam, J. van Bracht and J. K. IJspeert for data collection,
Downloaded From: http://iovs.arvojournals.org/pdfaccess.ashx?url=/data/journals/iovs/933172/ on 06/17/2017
Efficiency of Lens Fluorescence
3573
13. Yu NT, Bando M, Kuck FR. Fluorescence/Raman intensity ratio for monitoring the pathologic state of human lens. Invest Ophthalmol Vis Sci. 1985;26:97-101.
14.
Jacobs
R, Krohn DL. Fluorescence intensity profile of
References
human lens sections. Invest Ophthalmol Vis Sci.
1981:20:117-120.
1. Hess C. Pathologie und therapie des linsensystems.
In: Graefe-Saemisch Handbuch der gesamten Augenheil- 15. Wyszecki G, Stiles WS. Color Science. 2nd Ed. New
York: John Wiley & Sons; 1982:256-257.
kunde. Teil II. Leipzig: Verlag von Wilhelm Engelmann; 1911.
16. Vos JJ. Disability glare—a state of the art report. Commission International de I'EclairageJournal. 1984;2:392. le Grand Y. Recherches sur la Fluorescence des Milieux
53.
Oculaire. Instituto de biofisica Universidade do Brasil;
17. van den Berg TJTP, IJspeert JK. Light scattering by
1948.
donor lenses. Exp Eye Res. In press.
3. Weale RA. Human lenticular fluorescence and trans18. van Norren D, Vos JJ. Spectral transmission of human
missivity, and their effects on vision. Exp Eye Res.
ocular media. Vision Res. 1974;14:1237-1244.
1985;41:457-473.
19. Said FS, Weale RA. The variation with age of the spec4. IJspeert JK, van den BergTJTP, Spekreijse H. An ocutral transmissivity of the living human crystalline lens.
lar point-spread function model with parameters for
Gerontologica. 1959;3:213-231.
age, pupil size and pigmentation. Vision Res.
20. Tan KEWP. Vision in the Ultraviolet. Thesis, University
1992:33:15-20.
of Utrecht, The Netherlands, 1971.
5. van den Berg TJTP, IJspeert JK, de Waard PWT. De21. Pokorny J, Smith VC, Lutze M. Aging of the human
pendence of intraocular straylight on pigmentation
lens. Applied Optics. 1987:26:1437-1440.
and light transmission through the ocular wall. Vision
Res. 1991:31:1361-1367.
22. van Heiningen R. Fluorescent derivatives of 3-hydroxykynurenine in the lens of man, the baboon, and
6. van den Berg TJTP, IJspeert JK, de Waard PWT,
the grey squirrel. BiochemJ. 197l;123:30-31P.
Meire F. Functional quantification of diaphany. Doc
2 3. van Heiningen R. The glucoside of 3-hydroxykynurenOphthalmol. 1990;75:239-246.
ine and other fluorescent compounds in the human
7. Zigman S. Photobiology of the lens. In: Maisel H, ed.
lens. In: The human lens in relation to cataract. Ciba
The Ocular Lens. New York: Marcel Dekker;
foundation symposium 19, Amsterdam: Elsevier;
1985:301-347.
1973:151-171.
8. Dillon J. Photochemical mechanisms in the lens. In:
24. Bando M, Nakajima A, Satoh K. Spectrophotometric
Maisel H, ed. The Ocular Lens. New York: Marcel
estimation of 3-OH L-kynurenine O-/3-glucoside in
Dekker; 1985:349-366.
the human lens. / Biochem. 1981;89:103-109.
9. Sato K, Ban do M, and Nakajima A. Fluorescence in
25. van den Berg TJTP. On the relation between glare and
human eye lens. Exp Eye Res. 1973; 16:167-172.
straylight. Doc Ophthalmol. 1991;78,177-181.
10. Cooper GF, Robson JG. The yellow colour of the lens
26. van den Berg TJTP. Importance of pathological intraof man and other primates. J Physiol. 1969;203:411ocular light scatter for visual disability. Doc Ophthal-.
417.
mol. 1986:61:327-333.
11. Spector A, Roy D, Stauffer J. Isolation and character27. Elliott DB, Bullimore MA. Assessing the reliability, disization of an age-dependent polypeptide from human
criminative ability and validity of disability glare tests.
lens with non-tryptophan fluorescence. Exp Eye Res.
Invest Ophthalmol Vis Sci. 1993:34:108-119.
1975;21:9-24.
28. Bettelheim FA, Ali S. Light scattering of normal hu12. Zuclich JA, Glickman RD, Menendez AR. In situ meaman lens III. Relationship between forward and backsurements of lens fluorescence and its interference
scatter of whole excised lenses. Exp Eye Res.
with visual function. Invest Ophthalmol Vis Sci.
1985;41:l-9.
1992;33:410-415.
and Drs Felius, Vos, Bettelheim, and the reviewers for valuable comments on the manuscript.
Downloaded From: http://iovs.arvojournals.org/pdfaccess.ashx?url=/data/journals/iovs/933172/ on 06/17/2017