Download Active loads and the RIGOL DP800 and DP1000 Series

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Resistive opto-isolator wikipedia , lookup

Wireless power transfer wikipedia , lookup

Mercury-arc valve wikipedia , lookup

Electrical ballast wikipedia , lookup

Power factor wikipedia , lookup

Electrical substation wikipedia , lookup

Audio power wikipedia , lookup

Power over Ethernet wikipedia , lookup

Opto-isolator wikipedia , lookup

Power inverter wikipedia , lookup

Stray voltage wikipedia , lookup

Voltage regulator wikipedia , lookup

Current source wikipedia , lookup

Pulse-width modulation wikipedia , lookup

Surge protector wikipedia , lookup

TRIAC wikipedia , lookup

Electric power system wikipedia , lookup

Variable-frequency drive wikipedia , lookup

Islanding wikipedia , lookup

Power MOSFET wikipedia , lookup

Amtrak's 25 Hz traction power system wikipedia , lookup

Three-phase electric power wikipedia , lookup

Rectifier wikipedia , lookup

Electrification wikipedia , lookup

History of electric power transmission wikipedia , lookup

Power engineering wikipedia , lookup

Power electronics wikipedia , lookup

Voltage optimisation wikipedia , lookup

Buck converter wikipedia , lookup

Alternating current wikipedia , lookup

Power supply wikipedia , lookup

Switched-mode power supply wikipedia , lookup

AC adapter wikipedia , lookup

Mains electricity wikipedia , lookup

Transcript
Active loads and the RIGOL DP800 and DP1000 Series
1. Introduction
The RIGOL DP800 series and DP1000 series are programmable linear DC power supplies. They
can only provide power for a pure load that does not have the ability to output a current.
Active loads, such as those that can provide power (batteries, solar cells, etc..), should not be used
with the DP800 series or DP1000 series power supply. Active loads can lead to instability in the
power supply control loop and may damage the powered device.
Connecting the power supply to active loads is not recommended.
+
Load
Power Supply
Figure 1 Improper use of DP800 and DP1000 Series Power Supply
2. Detailed Technical Description
The RIGOL DP800 series and DP1000 series programmable linear DC power supply can only work
in the first quadrant (source positive voltage and a positive current) or the third quadrant (source
negative voltage and a negative current).
They cannot work in the second quadrant (negative voltage, positive current.. an adjustable load
of negative power) or the fourth quadrant (positive voltage, negative current.. as used for a battery
discharge test).
When the load itself is a source and the power supply is required to work in the second quadrant as
an adjustable load, the control loop may lose control and the power supply will output an
uncontrolled voltage. This could damage or destroy the load.
1/5
When the power supply works in the fourth quadrant (e.g., used in a battery discharge test), the
control loop is also unstable and will quickly drain the battery. This can result in dangerous
conditions, including damage to the battery, power supply, and a very high risk of fire and
explosion.
2.1. Power Quadrants in more detail
The Cartesian coordinate system is a common representation of power supply output capabilities.
The horizontal axis represents voltage, and the vertical axis represents current. The distributions of
the four quadrants of the power supply as shown in Figure 2.
The first quadrant: the power supply provides a positive voltage and a positive current (the
direction of the current flows from the power supply to the load).
The second quadrant: the power supply provides a negative voltage and a positive current (the
direction of the current flows from the power supply to the load).
The third quadrant: the power supply provides a negative voltage and negative current (the
direction of the current flows from the load to the power supply).
The fourth quadrant: the power supply provides a positive voltage and a negative current (the
direction of the current flows from the load to the power supply).
I
The first quadrant:
positive voltage
positive current
The second quadrant:
negative voltage
positive current
V
0
The third quadrant:
negative voltage
negative current
The forth quadrant:
positive voltage
negative current
Figure 2 Distributions of Power Quadrants
2/5
2.2. Principle of DP800 and DP1000
Here is a block diagram of DP800 and DP1000 series:
DP800 and DP1000
Load
Iout
2
3
Floating_GND
1
Floating_GND
OPAMP
+
OUT
-
Control_Signal
Figure 3 Block Diagram of DP800 and DP1000
If a current is forced into the supply (I.E. sinking the current), it will directly affect the working
status of the MOS transistor and result in instability within the control loop of the power supply as
shown in Figure 4.
DP800 and DP1000
Load
Iin
2
3
Floating_GND
1
Floating_GND
OPAMP
+
OUT
-
Control_Signal
Figure 4 Current Anti-irrigation Diagram
In addition, the DP800 and DP1000 series power supply outputs do not have an output relay. When
a specific output channel is disabled (power off) the output voltage is set to 0V and is regulated by
the control loop.
3/5
For charging batteries with the DP800 and DP1000 series power supplies, we recommend using
constant current mode and implement the circuit shown in Figure 5. The external diode can prevent
the flow of current into the supply and prevent damage.
Power Supply
+
-
Battery
Figure 5 Application Program of Battery Charge Test
4/5
Headquarters
RIGOL TECHNOLOGIES, INC.
No.156,Cai He Village,
Sha He Town,
Chang Ping District, Beijing,
102206 P.R.China
Tel:+86-10-80706688
Fax:+86-10-80705070
Email: [email protected]
USA
RIGOL TECHNOLOGIES,USA INC.
7401 First Place, Suite N
Oakwood Village
OH 44164, USA
Toll free: 877-4-RIGOL-1
Office: (440) 232-4488
Fax: (216)-754-8107
Email: [email protected]
EUROPE
RIGOL TECHNOLOGIES GmbH
Lindbergh str. 4
82178 Puchheim
Germany
Tel: 0049- 89/89418950
Email: [email protected]
5/5