RAD 220 QUALITY ASSURANCE
FOR RADIOGRAPHERS
DIGITAL IMAGING SYSTEMS
DIAGNOSTIC IMAGING: THREE
MAJOR COMPONENTS
• ACQUISITION (CAPTURE)
• PROCESSING
• DISPLAY
FILM-SCREEN IMAGING
Image capture, processing and
display are handled by the
same recording medium.
1
DIGITAL IMAGING
Image capture, processing
and display are handled
separately.
DIGITAL IMAGING: DATA
CONVERSION
• DIGITAL CAMERA
• SCANNING DIGITIZER (LASER)
• SCANNING DIGITIZER (CCD)
COMPUTED RADIOGRAPHY:
IMAGE CAPTURE
• BaFBr:Eu (Barium Flurobromide)
• BaFI:Eu (Barium Fluoroiodide)
• CsBr:Eu (Cesium Bromide)
2
COMPUTED RADIOGRAPHY:
IMAGE PROCESSING & DISPLAY
•
•
•
•
•
•
•
IRD (IMAGE READING DEVICE)
LASER SCANNING
LIGHT DETECTION
ANALOG-TO-DIGITAL CONVERSION
PROCESSING
DIGITAL TO ANALOG CONVERSION
IMAGE DISPLAY
HISTOGRAM
COMPUTED RADIOGRAPHY:
ADVANTAGES OVER FILM-SCREEN
• BETTER CONTRAST RESOLUTION
• GREATER EXPOSURE LATITUDE
• FEWER TECHNIQUE-RELATED
REPEATS
3
CR VERSUS FILM: EXPOSURE LATITUDE
COMPUTED RADIOGRAPHY:
DISADVANTAGES OVER FILM-SCREEN
• LOWER SPATIAL RESOLUTION
• MORE SENSITIVE TO SCATTER
RADIATION
• POTENTIAL FOR CARELESS
TECHNIQUE SELECTION & HIGHER
PATIENT DOSE
MONITORING RADIATION
EXPOSURE (FUJI)
FUJI S-VALUES: THE HIGHER THE S
NUMBER, THE LOWER THE EXPOSURE
RECEIVED BY THE IR.
™0.1 mR: S VALUE = 2000
™1.0 mR: S VALUE = 200
™10.0 mR: S VALUE = 20
THE NORMAL RANGE IS 150 TO 250
4
MONITORING RADIATION
EXPOSURE (KODAK)
KODAK INDEX NUMBERS: THE HIGHER
THE NUMBER, THE HIGHER THE
EXPOSURE RECEIVED BY THE IR.
ƒ 0.1 mR: INDEX = 1000
ƒ 1.0 mR: INDEX = 2000
ƒ 10.0 mR: INDEX = 3000
NORMAL RANGE IS 1800 TO 2200
MONITORING RADIATION
EXPOSURE (AGFA)
AGFA logM: Logarithmic expression
of median exposure. An increase of
0.3 represents a doubling of
exposure. The measured exposure is
compared to a baseline established
by the imaging department.
DIGITAL RADIOGRAPHY
Use of solid-state semiconductor
arrays (called the active matrix) to
capture and convert x-ray signals to
electrical signals. The image
receptor is incorporated into table or
upright grid assembly.
5
DIGITAL RADIOGRAPHY
• INDIRECT CONVERSION
• DIRECT CONVERSION
INDIRECT CONVERSION DR
• FLUORESCENT CRYSTALS (CsI)
• AMORPHOUS SILICON (a-Si)
DETECTORS
• THIN-FILM TRANSISTOR STORAGE
ASSEMBLY
• LOWER SPATIAL RESOLUTION (VS.
DIRECT CONVERSION DR)
• HIGHER QUANTUM DETECTIVE
SENSITIVITY
DIRECT CONVERSION DR
• AMORPHOUS SELENIUM (a-Se)
PHOTOCONDUCTIVE DETECTORS
• THIN-FILM TRANSISTOR (TFT)
STORAGE ASSEMBLY
• BETTER SPATIAL RESOLUTION
• LOWER QUANTUM DETECTIVE
SENSITIVITY
6
WHICH IS BETTER, CR or
DR?
An objective comparison of image
quality is based on a concept
called DETECTIVE QUANTUM
EFFICIENCY (DQE).
DETECTIVE QUANTUM
EFFICIENCY
A measure of detector transfer
efficiency and is defined as:
2
SNRout
SNRin2
DETECTIVE QUANTUM
EFFICIENCY
DQE measures overall image quality as
a function of object detail. This is similar
to the concept of MODULATION
TRANSFER FUNCTION. The higher the
DQE, the better the system is at
displaying small low-contrast objects.
7
BASED ON DQE…
• DIRECT-COVERSION DR IS
GENERALLY BETTER THAN INDIRECT
CONVERSION DR
• INDIRECT CONVERSION DR IS BETTER
THAN CR
• CR IS BETTER THAN FILM-SCREEN
RADIOGRAPHY
WHICH IS MORE
PRACTICAL, CR or DR?
Every imaging department must
decide on the proper mix of
equipment. CR seems offers more
flexibility in terms of use (table vs.
bucky) and portability. DR systems
are incorporated into a table or upright
grid assembly and tend not be very
portable.
WHAT ABOUT QUALITY
CONTROL?
• CR AND DR SYSTEMS LACK UNIFORM
STANDARDS
• RELIANCE ON MANUFACTURERSPECIFIC INSTRUCTIONS ON TESTING
AND MAINTENANCE
• STILL, ALL CR AND DR EQUIPMENT
SHARE COMMON ISSUES
8
QUALITY CONTROL OF DIGITAL
IMAGING EQUIPMENT
•
•
•
•
•
WHO IS RESPONSIBLE
WHAT TO TEST
HOW OFTEN TO TEST
PREVENTIVE MAINTENANCE
SPECIALIZED TESTING
EQUIPMENT
QUALITY CONTROL OF DIGITAL
IMAGING EQUIPMENT
• IMAGING PLATES
• IMAGE READING DEVICE
• IMAGE DISPLAY SCREENS (CRT VS.
LED)
• COMPUTER SYSTEM TESTS
TERMS TO KNOW
•
•
•
•
•
CONTRAST RESOLUTION
SPATIAL RESOLUTION
TEMPORAL RESOLUTION
MODULATION TRANSFER FUNCTION
DETECTIVE QUANTUM EFFICIENCY