Download Surecell CI2

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts
no text concepts found
Transcript
SureCell - Cl2
Innovation, Quality and Expertise for Gas Detection.
Operating Performance
Operating Principle
3-electrode electrochemical
Gas Detected
Chlorine
Measurement Range
0-20 ppm
Maximum Overload
50 ppm
Expected Operating Life
2 years in air
Output Signal
0.50 +0.10µA per ppm
Temperature Range
Continuous: -20°C to +40°C
2
o13.0+0.1
1.25+0.2
Flat area for
sealing purposes
Intermittent: -40°C to +55°C
Pressure Range
1 atm +10%
Humidity Range (non-condensing)
Continuous: 15 - 90% RH
Intermittent: 0 - 99%RH
<60 seconds
Zero Shift (-40 to +55°C)
<+0.5ppm equivalent
Long Term Output Drift
<2% per month
Repeatability
<+5%
Linearity
Linear <+3%
Recommended Load Resistor
5Ω
Bias Voltage
Not required
Intrinsic Safety Data
Maximum at 50ppm
0.05mA
Maximum o/c Voltage
0.8V
Maximum s/c Current
<1.0A
o20.75+0.25
R2.50
13.9+0.5
<+0.25ppm equivalent
16.60+0.25
90)
Baseline Offset (clean air)
4.00+0.15
Response Time (T5
o20.2+0.1
Physical Specification
Weight
5g (approx)
Housing Material
Noryl 110
Storage Life
6 months in sealed container
Storage Conditions
+10°C to +30°C
Orientation
Any
Warranty Period
18 months from date of despatch
Working
3 Pins o1.6 on
13.5+0.1 PCD
W
Reference
Ordering Details
Part Number
Order From
2112B2035
Sixth Sense
Hatch Pond House, 4 Stinsford Road
Poole, Dorset, England BH17 0RZ
T: +44 (0) 1202 645770
F: +44 (0) 1202 665331
E: [email protected]
Code date: 08/03
W: www.sixth-sense.com
R
C
Counter
Additional Information
Innovation, Quality and Expertise for Gas Detection.
Linearity of CI2 SureCells across Measurement Range
45
40
35
30
25
20
15
10
5
0
0
5
10
15
20
25
30
35
40
45
Applied Concentration (ppm CI2)
Temperature Coefficients of Chlorine SureCells
110
100
Normalised Sensitivity (to 20 C)
90
o
Measured Concentration (ppm CI2)
50
80
70
60
50
40
30
20
10
0
-30
-20
-10
10
20
30
40
Temperature (oC)
Cross Sensitivity Table
Gas
Concentration
Reading
Used (ppm)
(ppm CI2)
Gas
Hydrogen Sulphide
25
-16.3
20000
0
Sulfur Dioxide
50
9.1
Ammonia
50
-1.9
Nitrogen Dioxide
50
1.25
Hydrogen Chloride
9
1.25
Carbon Dioxide
Concentration
Reading
Used (ppm)
(ppm CI2)
In the interest of product improvement, Sixth Sense reserves the right to alter design features and specifications without notice.
Unless otherwise stated, all product specifications are quoted at standard temperature and pressure.
*Note: The figures in this table are typical values and should not be used as a basis for cross calibration. Cross sensitivities may
not be linear and should not be scaled. All data based on a 5 minute gassing. For some cross interferents break through
will occur if gas is applied for a longer time period.