Download Anatomy of an IRMS Lecture 2 – Vacuum Systems Reading: see

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Vascular remodelling in the embryo wikipedia , lookup

Transcript
Lecture 2 – Vacuum Systems
Anatomy of an IRMS
Reading: see www.vacuumlab.com/articles
1) Introduction – Start by looking at the molecular basis of pressure and vacuum
a) Units
i) Atmosphere (atm). Very convenient, not commonly used.
ii) SI unit is Pascals (Pa), defined as 1 N/m2 . 1 atm = 101,325 Pa. Very inconvenient for typical
calculations, but still should be used in publications.
iii) Torr – most common in US; 1atm = 760 Torr
iv) Bar (mbar) – most common in Europe. 1 atm = 1.0135 bar.
v) See conversion scale to go between them. To rough approximation, 1mbar = 1 Torr, 1 mTorr
= 10-3 mbar.
b) Vacuum Regimes
Regime
Pressure (Torr)
Pressure
(mbar)
Molecules/mL
Mean free path
Low vacuum
25
33
1018
2 um
13
medium
1e-3
1.33
10
10 cm
High
1e-9
1.33e-6
107
100 km
Ultra high
(UHV)
1e-12
1.3e-9
104
100,000 km
Interplanet
space
1e-15
10
c) Flow regimes
i) Molecular flow can be categorized in two regimes by comparing mean-free path to
characteristic distances of system (ie, tube diameter); Knudsen number
ii) Mean free path a function of density and size: mfp = (sqrt(2)*pi*n*d2)-1, n= molecules/cm3, d
= molecular diameter (cm); thus smaller atoms (He) have a longer mean free path than larger
ones (N2, Ar)
iii) Viscous flow – dominated by interactions between molecules. Pressure-driven flow. Behaves
like water in pipes. Does not fractionate isotopes.
iv) Molecular flow – dominated by collisions with walls. Diffusion-like flow, rate dependent on
mean-free path and probability that a gas molecule flies into pump orifice. Does fractionate
isotopes, because velocity varies with mass.
v) Transition between viscous and molecular flow depends on geometry of physical system.
Smaller diameters will experience viscous flow at much lower pressures. (see graphic).
Becomes very important in talking about dual-inlet systems.
d) Pump-down
i) At low vacuum, flow is dominated by viscous flow, simply need to minimize friction losses.
ii) At high vacuum, system is in molecular flow. Rate of pumping down is governed by the
mean-free path of molecules out of the system. Can be formalized as the conductance of the
system, which is the inverse of resistance.
iii) Math. Can be described in analogous way to electricity; recall V = IR
(1) V is equivalent to pressure (P; units Torr)
(2) I is equivalent to mass flow rates (Q; units Torr-L/sec)
(3) Conductance then has units of L/sec; tubes and orifices have a conductance, but so do
pumps; in pumps this is often called the pumping speed (S), where it reflects a volumetric
1
Lecture 2 – Vacuum Systems
Anatomy of an IRMS
flow rate. Note that pumping speed of a physical pump (turbos) depends on size of
molecule, and is slower for smaller atoms;
(a) The speed of a pump is quoted as the volumetric flow per unit time; will be ~constant
over some pressure range, but realize that this represent decreasing mass flow rate
(4) Combining these we get P = Q/S, usually rearranged to give Q = SP
(a) At fixed mass flow rate (determined by inputs), higher pumping speed or lower
conductance gives lower pressure
(b) At fixed pressure, higher pumping speed gives higher flow rate
(5) When we have multiple components in series (tubes then pump), calculate the effective
conductance just as we would for resistance (Reff = R1 + R2 +…; 1/Seff = 1/S1 + 1/S2+…)
iv) These calculations are often very useful in diagnosing vacuum systems. If mass flow rate and
pressure are known, pumping speed can be calculated. If pumping speed and pressure
differential are known, a limiting conductance can be calculated. In a closed ion source, if
conductance is known then pressure inside ion volume can be calculated.
e) Pump-down curves
i) Volume zone (volume of system); only matters during initial pumpdown, usually cleared in a
matter of seconds to minutes; pressure reaches 1 mbar quickly
ii) Below that, flow is dominated by
(1) Desorption from walls, mainly water; in this regime pumpdown time is dominated by
humidity/exposure time and surface area, not volume; called the “drydown zone” of the
pumping curve; this is why we vent instruments with dry N2 instead of to air
(a) Rate of desorption depends on supplying enough energy to drive molecules into the
gas phase; water is adsorbed in multiple layers, with layers closer to the surface more
strongly bound, so rate of desorption slows with time; increasing temperature
increases ambient energy and speeds the rate of desorption
(b) Note that most water molecules desorb, hit another surface, and then many resorb
(sticking coefficient is about 1%); thus the geometry of the system plays important
role in rate of desorption
(c) Common ways to spead up water removal are baking (200-250C) and UV photons
(2) Leaks through seals (o-rings or gaskets); assuming no outright leaks, rate is generally
determined by permeability of the seal; polymer o-rings have much higher rates than
metal gaskets; Viton is better than buna
iii) Turbo pumps can typically reach 10-6 mbar via baking out; below that, pressure is mostly H2
which is not efficiently pumped, thus need to use ion pumps to get better vacuum
f) Designing vacuum systems.
i) Sizing tubing. Generally want wider diameters and shorter length. Diminishing returns as
surface area increases.
ii) Straight flow paths.
iii) Smooth surfaces adsorb less water than rough ones (surface area)
iv) Pump speed. Higher is better, but costs increase exponentially. Need to consider material
being pumped; air is much easier to pump than He
v) Minimize number of seals to reduce permeation; tradeoff between lower leak rates (metal)
versus convenience (polymer)
vi) Virtual leaks. Air trapped at end of a screw has very long free path, will gradually leak out
and take forever to pump down. Special design to provide eliminate trapped air spaces
vii) Cleanliness. Oil and fingerprints degas very slowly (low vapor pressure), and at high pressure
the equivalent volume of a small mass is very large
(1) Calculation: 1mg of a substance with 100g/mol would have a volume of 22.4 E-5L at
STP. At high vacuum (1E-6 mbar, or 1E-9 atm) this becomes 22.4 E4 L or 224,000 liters.
2) Pressure gauges
2
Lecture 2 – Vacuum Systems
Anatomy of an IRMS
a) Review gauge pressure versus absolute pressure. Gauge pressure depends on local atmospheric
pressure, really only useful for high-P measurements.
b) Mechanical – work by measuring displacement or strain under pressure; useful for ~1 Torr to
high pressure
i) Manometer- measure the height of liquid displaced (water or mecury)
(1) McLeod gauge; compresses gas into a liquid manometer; very accurate, down to 10-6 torr,
but inconvenient; often used for calibration of other gauges
ii) Anaeroid – metal pressure plate pushes against a spring and displacement is measured
iii) Bourdon tube – a flattened tube tends to become more circular as pressure differential
increases; when tube is bent into C shape, this causes it to straighten out at pressure drops;
can be surprisingly sensitive
iv) Diaphragm gauge – measures deflection of a flexible membrane. Often implemented as a
‘capacitance manometer”, in which system measures capacitance of the membrane moving
relative to a fixed plate of known capacitance. Baratron gauge is one example. Useful down
to 10-2 Torr (mbar)
c) Thermal conductivity (useful range: 10-3 to 10 Torr)
i) Basic principle is that the rate of heat loss of a wire in vacuum depends on number of
molecular collisions, which depend on density of molecules. Drawback of all these is that
they depend on heat capacity of gas, must be calibrated for specific gases.
ii) Thermocouple gauge; constant current applied to wire, separate thermocouple junction heated
by conduction measures temperature of the wire. This is a 2-wire system.
iii) Pirani gauge. Uses a single platinum wire, applying constant current and measuring change in
voltage as the resistance of the wire changes with temperature.
d) Ionization gauges (useful range: 10-10 to 10-3 Torr)
i) Bombard gas with electrons, which ionizes gas and causes a resulting current.
ii) Hot cathode design – uses a heated filament to boil off electrons (thermionic emission)
iii) Cold cathode design (Penning gauge, magnetron) – uses high voltage (4kV) to induce current
flow; more sensitive and less maintenance, but more difficult to use
3