Download Using the HP 34401A Digital Multimeter (DMM)

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Pulse-width modulation wikipedia , lookup

Transistor wikipedia , lookup

Mercury-arc valve wikipedia , lookup

Immunity-aware programming wikipedia , lookup

Fuse (electrical) wikipedia , lookup

Power inverter wikipedia , lookup

Ground (electricity) wikipedia , lookup

Variable-frequency drive wikipedia , lookup

Islanding wikipedia , lookup

Electrical ballast wikipedia , lookup

Electrical substation wikipedia , lookup

History of electric power transmission wikipedia , lookup

Schmitt trigger wikipedia , lookup

Ohm's law wikipedia , lookup

Distribution management system wikipedia , lookup

Earthing system wikipedia , lookup

Rectifier wikipedia , lookup

Resistive opto-isolator wikipedia , lookup

Metadyne wikipedia , lookup

Triode wikipedia , lookup

Current source wikipedia , lookup

Power electronics wikipedia , lookup

Three-phase electric power wikipedia , lookup

Voltage regulator wikipedia , lookup

Opto-isolator wikipedia , lookup

Switched-mode power supply wikipedia , lookup

Power MOSFET wikipedia , lookup

Buck converter wikipedia , lookup

Multimeter wikipedia , lookup

Stray voltage wikipedia , lookup

Voltage optimisation wikipedia , lookup

Surge protector wikipedia , lookup

Alternating current wikipedia , lookup

Mains electricity wikipedia , lookup

Transcript
Using the HP 34401A Digital Multimeter (DMM)
We will make frequency use of the HP 34401A multimeter in this lab. It is important to
review the operation of this device, as misuse of the device can blow a fuse, which is
difficult and time-consuming to replace.
The most important thing to remember when using the 34401A is that the probes must be
connected to different terminals depending on whether you are measuring voltage or
current.
When measuring voltages, the probes should be connected as follows:
connect red probe here
HI
connect black probe here
LO
I
When measuring currents, the probes should be connected as follows:
HI
LO
connect black probe here
connect red probe here
I
If you have the probes connected for measuring currents and you try to measure the
voltage of a low-impedance voltage source (such as a power supply), you will blow a
fuse in the instrument. (The ammeter has a very low impedance, so the current through it
will grow too large.) This fuse must be replaced by a lab technician. In order to avoid
this, always remember to check the connection of your probes before making
voltage measurements.
Having the probes connected for voltage measurements is much safer. If you try to
measure a current with the probes connected for a voltage, no damage will occur (but you
will not be able to measure the current). It is always a good idea at the end of your lab
session to connect your probes for voltage measurements so that the group in the next lab
session does not accidentally blow a fuse.