LAMPIRAN A
MEMBEKAL, MENGHANTAR, MENGUJI DAN MENTAULIAH PERALATAN PERUBATAN BAGI EMPAT (4) UNTUK ESU – DIATHERMY
MACHINE, UNTUK JABATAN ORTHOPEDIK, HOSPITAL SIBU, SARAWAK BAGI TAHUN 2014
NO SEBUTHARGA: HS/Q034/2014
BIL.
SPESIFIKASI
1.0
APPLICATION
1.1
The equipment is used for monopolar and bipolar in open and
laparoscopic surgery.
2.0
ESSENTIAL FEATURES
2.1
This system ensures the power delivery remains
virtually constant, by utilizing computer controlled output for
automatic response to changes in tissue impedance - reducing
the need to adjust power settings for different types of tissue
2.2
The system is able to do simultaneous coagulation.
2.3
The system has 3 separate Bipolar modes.
a) Precise mode
b) Standard mode
c) Macro mode
2.4
The system has 3 separate Monopolar Cut modes.
a) Low mode
b) Pure mode
c) Blend mode
2.5
The system has 4 separate Monopolar Coagulation modes.
a) Desiccate mode
b) Fulgurate HCF mode
c) Fulgurate LCF mode
QTY
COMPLIANCE TO
REQUIREMENTS :
YES/NO
COMMENTS / PROPOSED
SPECIFICATION
1
LAMPIRAN A
d) Spray mode
2.6
The system has the ability to change power setting during
activation of Monopolar Cut or Coagulation.
2.7
The system has a low voltage cut and low voltage bipolar
modes.
2.8
The system has instantaneous RMS voltage control to
minimize RF leakage.
2.9
The system has an automatic computer controlled system to
control maximum output voltage which :a) Reduces capacitive coupling and video interference.
b) Minimizes sparking.
c) Reduces potential patient side burns.
2.10
The system also comes with optional panel of :a) RS 232 Port
b) RF Activation Port
c) Expansion Port
2.11
It is also built-in with RECQMS (Return Electrode Contact
Quality Monitoring System, which measures the quality of the
patient to pad interface. It uses a low voltage, low current
interrogation Sentry circuit to perform this.
3.0
TECHNICAL SPECIFICATIONS:
Output Waveforms (Please specify)
3.1
Bipolar
• Precise
• Standard
• Macro
2
LAMPIRAN A
3.2
Monopolar Cut
• Low
• Pure
• Blend
3.3
Monopolar Coag
• Desiccate
• Fulgurate
• Spray
3.4
Low Frequency Leakage (50/60 Hz)
3.5
High Frequency Leakage
4.0
OUTPUT CHARACTERISTICS (Please specify)
4.1 Maximum Peak to Peak Voltage
4.1.1 • Bipolar
Precise
Standard
Macro
4.1.2 • Monopolar Cut
Low
Pure
Blend
4.1.3 • Monopolar Coag
Desiccate
Fulgurate HCF
Fulgurate LCF
Spray
3
LAMPIRAN A
4.2 Maximum Power Setting
4.2.1 • Bipolar
Precise
Standard
Macro
4.2.2 • Monopolar Cut
Low
Pure
Blend
4.2.3 • Monopolar Coag
Desiccate
Fulgurate HCF
Fulgurate LCF
Spray
4.3 Crest Factor
4.3.1 • Bipolar
Precise
Standard
Macro
4.3.2 • Monopolar Cut
Low
Pure
Blend
4.3.3 • Monopolar Coag
Desiccate
Fulgurate HCF
Fulgurate LCF
Spray
4
LAMPIRAN A
4.4
PER
5.0
Dimensions and weight (Please specify)
• Height
• Width
• Length
• Weight
6.0
Input power requirements
• Operating range
• Max. Current.
7.0
Main Connection Electrosurgical System
• Operating range
• Line frequency
• Current
8.0
Power Cord Specification
This unit is equipped from the factory with a 220 VAC CEE
power cord. Should the AC power cord need to be replaced to
match another plug configuration, the replacement plug/ cable/
receptacle configuration must meet or exceed the following
specifications:
- Cable - H05VVF3G1.0 VDE, maximum length 15 ft (5m)
- Plug - minimum 6 A - 250 VAC
- Unit receptacle -IEC female, minimum 6 A - 250 VAC
9.0
Basic Accessories
1. Generator: 1 unit
2. Mono polar Footswitch : 1 unit
3. Bipolar Footswitch : 1 unit
4. Universal Cart: 1 unit
5. Reusable Mono polar Pencil : 5 Pcs
6. Diathermy pad : 50 Pcs
5
LAMPIRAN A
7. Universal Adapter : 1 Unit
8. Clamp/Cord Assembly : 1 Unit
9. Insulated Cushing 6" Forcep : 1 Unit
10. Reusable Bipolar Cord : 1 Unit
10.0
Warranty
One Year against Manufacturing Defect from Date
Commissioning.
Parts availability for at 10 years from date of testing and
commissioning.
Breakdown respond time should be within 48 hours within
warranty period.
Vendor to specify the warranty period for the parts and
accessories.
11.0
Service
Free Preventive Maintenance Service 1 X 6 Monthly During
Warranty Period.
Supplies have to provide at least two (2) copies of the
Operation Manuals in both soft and hard copy during testing
and commissioning.
Supplies has to provide at least two (2) copies of Instructions
for use Manuals with trouble Shoot Guide, Article Numbers
Catalogue and ordering Information.
12.0
Training
Software Training By Application Specialist And Factory
Trained Personnel On Operation Clinical Applications Running
Self-Test And Basic Trouble Shooting.
6