Download 2 Isolated V Actvtd Op Input

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Transistor wikipedia , lookup

Electrical ballast wikipedia , lookup

Flip-flop (electronics) wikipedia , lookup

Electrical substation wikipedia , lookup

Three-phase electric power wikipedia , lookup

Power inverter wikipedia , lookup

History of electric power transmission wikipedia , lookup

Islanding wikipedia , lookup

Ohm's law wikipedia , lookup

Immunity-aware programming wikipedia , lookup

Variable-frequency drive wikipedia , lookup

Current source wikipedia , lookup

Power MOSFET wikipedia , lookup

Resistive opto-isolator wikipedia , lookup

Triode wikipedia , lookup

Analog-to-digital converter wikipedia , lookup

Integrating ADC wikipedia , lookup

Rectifier wikipedia , lookup

Power electronics wikipedia , lookup

Alternating current wikipedia , lookup

Stray voltage wikipedia , lookup

Surge protector wikipedia , lookup

Buck converter wikipedia , lookup

Metadyne wikipedia , lookup

Voltage optimisation wikipedia , lookup

Voltage regulator wikipedia , lookup

Mains electricity wikipedia , lookup

Switched-mode power supply wikipedia , lookup

Schmitt trigger wikipedia , lookup

Opto-isolator wikipedia , lookup

Transcript
Ultrasonics
Application Note 102
• Connector J20 pins 4 and 7 can activate the ultrasound output with an external dc
voltage
• Input voltage range from 4.5Vdc to 30Vdc @ 4mA minimum
• Input current is electronically limited to 14mA maximum
• Isolated input is not polarity sensitive - it cannot be connected backwards
• Optical-isolation rated at 1500VPEAK
• Input is overvoltage protected at 33V with a surgector protection device
• Input is overcurrent protected at 315mA with a metric-style fuse and is replaceable
on the circuit board.
A dc voltage (4.5V Min. to 30V Max.) applied between J20 pin 4 and pin 7 activates the
generator ultrasound output. Removing the dc voltage deactivates the ultrasound output. The
polarity of the dc voltage input signal is not important since either polarity will function
properly (it cannot be connected backwards). This input requires a minimum of 4mAdc to
function at 4.5Vdc and the current is actively limited to 14mA, up to the maximum input
voltage. Noise filtering is provided on this input, which will delay ultrasound activation by
about 20mSec. after voltage is applied to the input. Likewise, the noise filter discharge time
will delay the deactivation of the ultrasound output by approximately the same time when
the input voltage is removed. Input overvoltage protection is provided. The input voltage will
be shorted (crow-barred) if the input voltage exceeds approximately 36Vdc. This also will
open the input protection fuse (F402) if the power source current output capability exceeds
315mA. This input is electrically isolated from all internal generator circuitry.
User
Supplied
Voltage
Source
© Copyright 1998 Dukane Corporation. All rights reserved.
Application Notes
Isolated Voltage Activated Operate Input
Connector
Generator Internal Circuitry
J20
Input is not
Polarity Sensitive
4
315mA
Bridge
Rectifier
dc
Switchable
dc
Voltage
Source
(30Vdc MAX )
Current
Limiter
FUSE
Noise
Filter
33V
O. V.
Limit
OptoCoupler
7
dc Voltage Applied = Ultrasound Activated
dc Voltage Off = Ultrasound Off
Isolated Voltage Activated Operate Input Connection Diagram