Survey
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
-1- Viscosity Determination of a given liquid viscosity Object : to determine of a given liquid viscosity using the falling ball method Theory Viscosity is a measure of the resistance of a fluid which is being deformed by either shear stress or tensile stress. In everyday terms (and for fluids only), viscosity is "thickness" or "internal friction". Thus, water is "thin", having a lower viscosity, while honey is "thick", having a higher viscosity. Put simply, the less viscous the fluid is, the greater its ease of movement (fluidity With the exception of superfluids, all real fluids have some resistance to stress and therefore are viscous. A fluid which has no resistance to shear stress is known as an ideal fluid or inviscid fluid. In common usage, a liquid with the viscosity less than water is known as a mobile liquid, while a substance with a viscosity substantially greater than water is simply called a viscous liquid. The study of flowing matter is known as rheology, which includes viscosity and related concepts. If a ball with density (kg/m3) moves with a constant terminal velocity v (m/s) in a liquid of density ' (kg/m3), as shown in the figure. Then, the sum if the forces acting on the ball equal zero. Therefore total upright forces and downright forces have to be the same. Buoyant force + Friction force Gravitational Force Gravitational force = Buoyant force + Friction force -2- Gravitational Force = mg Buoyant Force = 4 3 Friction Force = fv = 4 3 𝜋 𝑅 3 𝜌𝑔 𝜋 𝑅 3 𝜌′ 𝑔 = 6 RV Where : viscosity coefficient of the liquid in (Pa.s.) R : radius of the ball f : friction coefficient SO, 4 3 𝜋 𝑅 3 𝜌𝑔 4 3 = 𝜋 𝑅 3 𝜌′ 𝑔 + 6 RV By calculation : = 𝟐𝒈𝑹𝟐 ( 𝟗𝒗 ) (𝝆 − 𝝆′ ) Pa.s. Procedure 1- Measure the temperature of the liquid by using of the thermometer fixed in the outer tube 2- Measure the radius of the ball (R) by using of micrometer (repeat three times) 3- Measure the time the ball need to fall a distance 10 cm and then calculate the velocity v (m/s) (repeat three times) 3 4- Givens : density of the boron-glass ball = 2200 kg/m the density of a 40 % sugar solution ' = 1180 kg/m3. Gravity = 9.8 m/s2 5- The tube filled with the liquid is titled by an angle of 10°, therefore the viscosity coefficient will be calculated from the relation : 𝟐 = (𝟐𝒈𝑹 ) (𝝆 − 𝝆′ ) cos 10 𝟗𝒗 -3- Pa.s. Inverse Sqaure Law (ISL) Object : to verify of the inverse square relation between distance and intensity of radiation Theory An inverse-square law is any physical law stating that a specified physical quantity or intensity is inversely proportional to the square of the distance from the source of that physical quantity. In equation form: The divergence of a vector field which is the resultant of radial inverse-square law fields with respect to one or more sources is everywhere proportional to the strength of the local sources, and hence zero outside sources. Newton's law of universal gravitation follows an inverse-square law, as do the effects of electric, magnetic, light, sound, and radiation phenomena. In radiation, as you double the distance between source and detector, intensity of radiation goes down by a factor of four. If you triple the distance, intensity will go down by a factor of nine. If you quadruple the distance, the intensity will go down by a factor of sixteen. -4- Procedure 1234- Set the voltage of the GM tube to the optimal operating voltage (around 900 V) From the Preset menu, set Runs to Zero and set Preset Time to 10 sec. First do a run without radioactive sources to check your background level. Place a radioactive source in the top shelf (2 cm from GM actual detector) and begin taking data 5- Move the source down one shelf each time and take another run. You should see the data accumulating in the data window 1 6- Plot a graph between R (count rate) and the inverse square of the distance (𝑑2 ) 7- Examine the plot shape to verify the fact of inverse square law (straight line) R 𝟏 𝐝𝟐 R1 R2 R3 R (average) -5- d 𝟏 𝒅𝟐 Attenuation Coefficient Object : to investigate the attenuation of x-rays as a function of the absorber thickness Theory Attenuation is the gradual loss in intensity of any kind of flux through a medium. For instance, sunlight is attenuated by dark glasses, X-rays are attenuated by lead, and light and sound are attenuated by water. When x-rays fall on a material, rays may suffered from scattering and absorption. The scattering of x-ray quanta at the atoms of the attenuator material causes a apart of the radiation to change direction. This reduces the intensity in the original direction. This scattering can be either elastic or entail an energy loss or shift in wavelength i.e. inelastic scattering. Intensity of x-ray beam passing through matter of constant density and composition decreases exponentially as function of distance traveled. The X-ray intensity transmitted through a dense material is given by : I = Io exp (-x) where I is the transmitted X-ray intensity, I0 is the incident X-ray intensity, μ is the linear attenuation coefficient (in cm-1) and x is the thickness of the material (in cm). This equation shows that the X-ray intensity depends on the − the density of the material (the linear attenuation coefficient μ increases with density) − the thickness of the material I The greater the so-called transmittance I of an attenuator is, the lower is o its attenuating capacity. If we assume that the properties of the incident radiation remain unchanged in spite of attenuation, an increase in the thickness X by the amount dX will cause a decrease in the transmittance T by the amount -6- dT. The relative reduction in transmission is proportional to the absolute increase in thickness. − dT T = 𝜇. 𝑑𝑋 The relation between the linear attenuation coefficient of a material and the attenuator thickness is known as Lambert's law. Procedure 1- Set the tube high voltage to U = 21 kv 2- Set the emission current i = 0.05 mA 3- Press the key TARGET 4- Set the angular step width β = 0° 5- Set the measuring time t = 30 sec. 6- Using the adjust knob, set the angular positions of the absorbers (approximately, 0°, 10°, 20°, 30°, 40°, 50° and 60°) one after another, start the measurement with the scan key and display by pressing REPLAY, write your experiment result -7- Elctro Cardio Gram (ECG) Object : to record and analyze the electrocardiogram (ECG) signals Theory Electrocardiography is a trans-thoracic (across the thorax or chest) interpretation of the electrical activity of the heart over a period of time, as detected by electrodes attached to the outer surface of the skin and recorded by a device external to the body. The recording produced by this noninvasive procedure is termed as electrocardiogram (also ECG or EKG). An ECG test records the electrical activity of the heart. ECG is used to measure the rate and regularity of heartbeats, as well as the size and position of the chambers, the presence of any damage to the heart, and the effects of drugs or devices used to regulate the heart, such as a pacemaker. Most ECGs are performed for diagnostic or research purposes on human hearts, but may also be performed on animals, usually for diagnosis of heart abnormalities or research. An ECG is the best way to measure and diagnose abnormal rhythms of the heart, particularly abnormal rhythms caused by damage to the conductive tissue that carries electrical signals, or abnormal rhythms caused by electrolyte imbalances. In a myocardial infarction (MI), the ECG can identify if the heart muscle has been damaged in specific areas, though not all areas of the heart are covered. The ECG cannot reliably measure the pumping ability of the heart, for which ultrasound-based (echocardiography) or nuclear medicine tests are used. It is possible for a human or other animal to be in cardiac arrest, but still have a normal ECG signal (a condition known as pulseless electrical activity). The ECG device detects and amplifies the tiny electrical changes on the skin that are caused when the heart muscle depolarizes during each heartbeat. At rest, each heart muscle cell has a negative charge, called the membrane potential, across its cell membrane. Decreasing this negative charge towards zero, via the influx of the positive cations, Na + and Ca++, is called depolarization, which activates the mechanisms in the cell that cause it to contract. During each heartbeat, a healthy heart will have an orderly progression of a wave of depolarisation that is triggered by the cells in the sinoatrial node, spreads out through the atrium, passes through the atrioventracular node and then spreads all over the ventricles. This is detected as tiny rises and falls in the voltage between two electrodes placed either side of -8- the heart which is displayed as a wavy line either on a screen or on paper. This display indicates the overall rhythm of the heart and weaknesses in different parts of the heart muscle. Usually, more than two electrodes are used, and they can be combined into a number of pairs (For example: left arm (LA), right arm (RA) and left leg (LL) electrodes form the three pairs LA+RA, LA+LL, and RA+LL). The output from each pair is known as a lead. Each lead looks at the heart from a different angle. Different types of EKGs can be referred to by the number of leads that are recorded, for example 3-lead, 5-lead or 12-lead ECGs (sometimes simply "a 12-lead"). A 12-lead EKG is one in which 12 different electrical signals are recorded at approximately the same time and will often be used as a one-off recording of an ECG, traditionally printed out as a paper copy. Three- and 5-lead ECGs tend to be monitored continuously and viewed only on the screen of an appropriate monitoring device, for example during an operation or whilst being transported in an ambulance. There may or may not be any permanent record of a 3- or 5-lead ECG, depending on the equipment used. The term "lead" in electrocardiography causes much confusion because it is used to refer to two different things. In accordance with common parlance, the word lead may be used to refer to the electrical cable attaching the electrodes to the ECG recorder. are standard in a "12-lead" ECG. Alternatively, the word lead may refer to the tracing of the voltage difference between two of the electrodes and is what is actually produced by the ECG recorder. Each will have a specific name. For example "lead I" is the voltage between the right arm electrode and the left arm electrode, "Lead II" is the voltage between the right arm and the feet, whereas, Lead II is the voltage between the left arm and the feet. LEAD I LEAD II LEAD III = = = RA-LA RA-LL LA-LL -9- Feature RR interval P wave PR interval PR segment Description The interval between an R wave and the next R wave: Normal resting heart rate is between 60 and 100 bpm. During normal atrial depolarization, the main electrical vector is directed from the SA node towards the AV node, and spreads from the right atrium to the left atrium. This turns into the P wave on the ECG. The PR interval is measured from the beginning of the P wave to the beginning of the QRS complex. The PR interval reflects the time the electrical impulse takes to travel from the sinus node through the AV node and entering the ventricles. The PR interval is, therefore, a good estimate of AV node function. The PR segment connects the P wave and the QRS complex. The impulse vector is from the AV node to the bundle of His to the bundle branches and then to the Purkinje fibers. This electrical activity does not produce a contraction directly and is merely traveling down towards the ventricles, and this shows up flat on the ECG. The PR interval is more clinically relevant. QRS complex The QRS complex reflects the rapid depolarization of the right and left ventricles. They have a large muscle mass compared to the atria, so the QRS complex usually has a much larger amplitude than the P-wave. J-point The point at which the QRS complex finishes and the ST segment begins, it is used to measure the degree of ST elevation or depression present. ST segment The ST segment connects the QRS complex and the T wave. The ST segment represents the period when the ventricles are depolarized. It is -10- Duration 0.6 to 1.2s 80ms 120 to 200ms 50 to 120ms 80 to 120ms N/A 80 to 120ms isoelectric. T wave ST interval QT interval U wave J wave The T wave represents the repolarization (or recovery) of the ventricles. The interval from the beginning of the QRS complex to the apex of the T wave is referred to as the absolute refractory period. The last half of the T wave is referred to as the relative refractory period (or vulnerable period). 160ms The ST interval is measured from the J point to the end of the T wave. 320ms The QT interval is measured from the beginning of the QRS complex to the end of the T wave. A prolonged QT interval is a risk factor for ventricular tachyarrhythmias and sudden death. It varies with heart rate and for clinical relevance requires a correction for this, giving the QTc. Up to 420ms in heart rate of 60 bpm The U wave is hypothesized to be caused by the repolarization of the intervenqctricular septum. They normally have a low amplitude, and even more often completely absent. They always follow the T wave and also follow the same direction in amplitude. If they are too prominent, suspect hypokalemia, hypercalcemia or hyperthyroidism usually. The J wave, elevated J-point or Osborn wave appears as a late delta wave following the QRS or as a small secondary R wave. It is considered pathognomonic of hypothermia or hypocalcemia. Procedure : 1- Test subject has to be relaxed and in a resting position, as otherwise the ECG signal can be overlaid by the electrical potentials of the skeletal musculature, falsifying the measurement. 2- To reduce the skin resistance, spread electrode gel on the electrodes and attach them to the appropriate points in the body using the rubber straps. 3- Then, attach the leads to the electrodes as follow i. Red right arm ii. Yellow left arm iii. Green left arm iv. Black right arm 4- Clean the electrodes after each use with a paper towel or similar to prevent a salt layer from forming when the gel dries. Then, spray the electrodes and the corresponding skin sites with disinfectant to ensure maximum hygiene. Examination starting 5- Load settings 6- Start the measurement (F9) 7- The three leads according to Einthoven are recorded simultaneously 8- Stop the measurement with (F9) -11- Lung Volume (Respirometer) Object : to measure the lung parameters to verify the respiration efficiency Theory A respirometer is a device used to measure the rate of respiration of a living organism by measuring its rate of exchange of oxygen and/or carbon dioxide. They allow investigation into how factors such as age, chemicals or the effect of light affect the rate of respiration. Respirometers are designed to measure respiration either on the level of a whole animal (plant) or on the cellular level. These fields are covered by whole animal and cellular (or mitochondrial) respirometry, respectively. The traditional respirometer is used to measure the lung performance parameters. Respiration parameters Tidal Volume TV Volume of air inspired and then expired during breathing in rest Inspiratory Reserve Volume IRV Volume of air a person can inspire above tidal volume Expiratory Reserve Volume ERV Volume of air a person can exhale below resting expiratory volume Inspiratory Capacity IC Volume of a person can inspire above the resting expiratory volume Residual Volume RV Volume of air left in lung after maximum expiratory effort Functional Residual Capacity FRC Air remains in lung after normal expiration Vital Capacity VC Maximum volume of air that can be inspired and then expired Forced Vital Capacity FVC The same VC but under maximum expiratory force -12- Respiratory Volumes and Capacities This is the area of the respiratory system (trachea Anatomic Dead Space and bronchi) in which air is not exposed to blood. (150 cm3) Physiological Dead Space This is the area of alveoli by act of some diseases cannot perfuse the blood to O2. Procedure 1- Connect the spirometer box to Sensor-Cassy input A about 10 seconds before the measurement (warm-up phase). Use a new mouthpiece and a new bacteria filter for every test person and disinfect the sieves regularly with a disinfectant. 2- Load settings i. Compensate the zero point off the displayed volume flux dVA1. Open the window settings volume flux dVA1, select -0and make sure that no air is flowing through the spirometer during this procedure. ii. Start the measurement as soon as possible afterwards with F9 (as long as the thermal error in volume flux dVA1 is still negigble) -13- iii. Breathe evenly through the spirometer three or four times. Then inhale and exhale as much air as possible through the spirometer. Then breathe evenly again. iv. Stop the measurement with F9. Evaluation The tidal volume V1 is the difference between the maximum and minimum for normal breathing. You can determine this e.g. by setting horizontal lines or directly by measuring the difference. You can insert this value at any point in the diagram as text. The vital capacity V2 is calculated as the sum if the inspiratory and expiratory reserve volume and the tidal volume. It can be determined from the minimum and maximum at maximum inspiration and expiration analogous to the tidal volume. -14- Human Eye Model Object : to Theory -15-