Download doc-file

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Schmitt trigger wikipedia , lookup

Counter-IED equipment wikipedia , lookup

Automatic test equipment wikipedia , lookup

TRIAC wikipedia , lookup

CMOS wikipedia , lookup

Power electronics wikipedia , lookup

Valve RF amplifier wikipedia , lookup

Giant magnetoresistance wikipedia , lookup

Josephson voltage standard wikipedia , lookup

Ohm's law wikipedia , lookup

Switched-mode power supply wikipedia , lookup

Multimeter wikipedia , lookup

Surge protector wikipedia , lookup

Lego Mindstorms wikipedia , lookup

Test probe wikipedia , lookup

Resistive opto-isolator wikipedia , lookup

Current mirror wikipedia , lookup

Rectiverter wikipedia , lookup

Power MOSFET wikipedia , lookup

Opto-isolator wikipedia , lookup

Transcript
DØ-note 4309
July 30, 2003
Characteristics of the Layer 1 Silicon Sensors for
the Run IIb Silicon Detector
M. Demarteau1, R. Demina2, S. Korjenevski2, F. Lehner3, R. Lipton1, H.S. Mao1, R.
McCarthy4, R. Smith1
1) Fermilab, Batavia, USA
2) Kansas State University, Manhattan, USA
3) University of Zurich, Switzerland
4) State University of New York at Stony Brook, USA
Abstract
The characteristics of the prototype Layer 1 Hamamatsu silicon sensors for the Run IIb
DØ silicon detector are described. The results of the electrical and mechanical
characterizations indicate that the overall sensor quality is excellent. The sensors behave
in all aspects very satisfactory and our specifications are well matched.
1
1. Introduction
Extended running of the Tevatron at a center of mass energy approaching 2 TeV provides
an enormous physics potential for the collider detectors. Our current theoretical prejudice
indicates that the ability to identify heavy quarks in the final state is absolutely necessary
to resolve these new physics channels. A silicon strip detector is currently the tool most
suited to identify heavy flavors in proton-antiproton collisions. It is anticipated that the
current silicon detector of the DØ experiment will start performing marginally by an
integrated luminosity of 4 fb-1 due to radiation damage and low signal to noise ratio. It is
for these reasons that the collaboration has designed a new silicon tracker [1]. The silicon
tracking system will be a six layer device, divided into two radial groups. The outer
group is comprised of layers 2 through 5. The basic building block of the outer layers is a
stave. A stave is a two layer structure of silicon sensors. Silicon sensors are mounted on
each side of a rohacell stave core, which has embedded cooling channels. One side of the
stave will have axial readout, the other side stereo readout. The inner two layers, covering
a radius between ~18 mm and ~35 mm, will have axial readout only. These layers have a
significantly reduced radius with respect to the current tracker. The sensors are mounted
on carbon fiber support structures. For Layer 0, the hybrids will be located outside of the
tracking volume and the signals transmitted from the silicon over analogue cables to the
front-end readout chip. This choice is dictated by the cooling requirements, the desire to
minimize the amount of material for the innermost position measurement, and the very
limited amount of space available. For Layer 1 the hybrids are mounted on the silicon
sensors. The innermost layers have a six-fold symmetry. Each layer has two sub-layers,
layers A and B. The sensors for the B sub-layer are mounted on castellations in the
carbon fiber support structure. There will be six sensors in z per azimuthal sector. Our
design thus calls for 144 sensors each for layers 0 and 1. This note describes the results of
the tests done at Fermilab, KSU and Zürich on a set of prototype Layer 1 outer layer
sensors and test structures.
2. Sensor Specifications
All silicon detectors are p+n type single sided sensors, with AC coupling and biased
through poly-silicon resistors. The expected fluence of the inner layer detectors is
expected to be 1.251013 1 MeV eq. n/cm2/fb-1 (Layer 0) and 4.121012 1 MeV eq.
n/cm2/fb-1 (Layer 1) respectively including a safety factor of 1.5. All inner layer sensors
will be provided by Hamamatsu Photonics. The silicon sensors will have a single-guard
ring with peripheral n-well as designed by Hamamatsu in order to improve the high
voltage stability after irradiation. The detailed specifications for the sensors can be found
in [2]. Prototype Layer 1 sensors were only purchased from Hamamatsu for two reasons.
First, the specifications for the layer 0 and layer 1 sensors are the same. Secondly,
Hamamatsu has already delivered sensors for the Layer 00 detector for CDF, which are
very similar to our proposed Layer 0 sensors. A few Layer 00 sensors were obtained from
CDF and were submitted to an extensive irradiation program at the Fermilab booster. The
performance of the sensors with irradiation was well within our specification [3]. Table 1
summarizes some of the main characteristics for the sensors. A detailed mechanical
drawing of the layer 1 sensor, drawing number 3823.210-ME-399436, on which the
2
location of strips, fiducial marks, bonding and testing pads as well as the strip numbering
definition and other features are indicated, is available and was provided to Hamamatsu
(see Fig. 1). Figure 2 shows the equivalent drawing, drawing number 3823.210-ME399678, for the Layer 0 sensors. Sensors are also outfitted with a 20-field scratch pad for
unique identification.
Specifications:
Wafer thickness
Depletion voltage
Leakage current
Junction breakdown
Implant width
Al width
Al strip resistivity
Coupling capacitance
Coupling capacitor
breakdown
Interstrip capacitance
Polysilicon bias resistor
Active Length (mm)
Active Width (mm)
Cut Length (mm)
Cut Width (mm)
Strip Pitch (m)
Readout Pitch (m)
# of Readout strips
Defective strips
Layer 0
32020m, sensor warp
less than m
40V<Udep<300V
<100nA/cm2 at RT and
FDV+20V, total current < 4
A at 700 V
>700 V
6 m
2-3 m overhanging metal
< 30 /cm
>10 pF/cm
Layer 1
32020m, sensor warp
less than m
40V<Udep<300V
<100nA/cm2 at RT and
FDV+20V, total current < 4
A at 700 V
>700 V
7 m
2-3 m overhanging metal
< 30 /cm
>10 pF/cm
>100 V
>100 V
<1.2 pF/cm
0.8  0.3 M
77.360
12.800
79.400
14.840
25
50
256
<1%
<1.2 pF/cm
0.8  0.3 M
77.360
22.272
79.400
24.312
29
58
384
<1%
Table 1: Specifications for inner layer sensors
3
Figure 1: Detailed Layer 1 sensor layout.
4
Figure 2: Detailed layout of Layer 0 sensor.
5
3. Prototype Sensors and Test Structures
In April 2002 a set of ten prototype Layer 1 sensors were ordered from Hamamatsu
Photonics, which were shipped to Fermilab on September 12, 2002. In addition we
ordered another three sensors in May 2003, which were shipped to Fermilab on July 13.
These sensors came from an overrun of the previous production, i.e. had the same lot
number. The specifications of the sensors are detailed in reference 1. In Table 1 a
summary of the specifications for the Layer 1 sensors is given. The sensors are produced
on 320 m thick 6” wafers. Figure 3 shows a layout of the wafer. All the 13 sensors
received are from one batch. In addition to the sensors, test structures are an integral part
of the wafer. There are four separate test structures on a wafer of which we receive one of
the two adjacent to the long side of the sensor. They are indicated by the serial numbers
of the four sensors on the wafer. The test structure contains two ‘baby-sensors’. One is
identical to the full-size sensor and is different only in the fact that it has eight readout
strips. The other baby-sensor has no poly-silicon resistors. Adjacent to the baby-sensors
are four implants with aluminization with separate contacts for each. As per our
specifications, there are also silicon diodes on the test
Figure 3: Layout of 6" Hamamatsu wafer
6
structure, with and without guard ring structure. One area on the test structure contains a
field MOS structure for flatband voltage measurements, and monitors for the implant
resistance and coupling capacitance. The remaining features on the test structure are
arrays of mainly poly-silicon resistors.
The characterization of the sensors is divided into three distinct parts. In the next section
the measurements on the test structures will be described. The section following is the
main section, describing the measurements performed on the sensors themselves. All
measurements follow the procedures as outlined in reference [4]. The note concludes
with a summary of the mechanical properties of the sensors. Both the L1 test structures
and sensors were also irradiated and characterized after irradiation. Those results are
described in an accompanying note [5].
4. Electrical Characterization of the Test Structures
Ten test structures were received. Test structures, 9.10.11.12, 13.14.15.16, and
17.18.19.20 were measured at KSU, test structure 5.6.7.8 at Zürich, and test structure
1.2.3.4 at Fermilab. Five parameters are determined from measurements on the test
structures:
i.
Coupling Capacitance and Breakdown Voltage of the coupling capacitor
ii.
Implant resistance
iii.
Aluminum strip resistance
iv.
Poly-silicon resistance
v.
Depletion voltage of silicon diodes
vi.
Total load and interstrip capacitance
vii.
MOS capacitance and flatband shift
4.1. Coupling Capacitance and Coupling Capacitor Breakdown
On the test structure there is a series of four strips consisting of the p+ implant only, the
coupling capacitor and the aluminization. Pads are connected to the implant (DC-pad)
and the aluminum strip (AC-pad) (see Figure 4). The frequency dependence of the
coupling capacitance for each of the four strips was mapped out and is shown in Figure 5.
The low frequency limit gives a coupling capacitance of 115 pF, or 14.8 pF/cm, well
within our specification.
The breakdown voltage was determined for three strips on this structure. A voltage was
applied across the coupling capacitor by placing a probe on the DC-pad and the AC-pad.
Breakdown of the capacitor is defined as the voltage when the current reaches 100nA.
Figure 6 shows the dependence of the current versus voltage for these three channels. A
breakdown well in excess of our specified breakdown value of 100V is observed.
7
The breakdown voltage was also determined on a few strips on a baby sensor on test
structure 1.2.3.4. The coupling capacitors for strips 1, 2, 3, 5, 6 and 7 had a breakdown
value of 180, 150, 160, 200, 240 and 170V, respectively.
Figure 4: Implant and Coupling Capacitor structure on the Test Structure
8
Coupling Capacitor L1 teststructure
120
100
C (pF)
80
60
40
CC1
CC2
20
CC3
CC4
0
0.1
1
10
100
1000
10000
Frequency (kHz)
Figure 5: Frequency dependence of coupling capacitance on Layer 1 test structure.
Coupling Capacitor Breakdown
I (nA)
1000
0
-1000
-2000
CC1
-3000
-4000
-5000
CC2
CC3
0
50
100
150
200
250
300
Voltage (V)
Figure 6: IV-Curve for measurement of Ccc breakdown
4.2. Implant Resistance
9
The same four strips, consisting of the p+ implant, the coupling capacitor and the
aluminization, were used to determine the implant resistance. To facilitate the
measurement, the corresponding pads of two adjacent strips are wirebonded at the far end
of the strip. Probes are connected to the implant (DC) pads. The implant resistance was
measured by mapping the current versus the voltage differential between the pads. Sets of
two strips were measured on test structure 1/2 and give an implant resistance of 130
k/cm, to be compared to 104 k/cm for the outer layer sensors [6], which are specified
to have 15% wider implant strips. The implant resistance however, is not part of our
specifications, but the measured value agrees well with our expectations of a typical p+
doping of this width.
Alternatively, the implant resistance was measured using the ‘N LW20’ feature on the
test structure. A resistance of 2605  was measured, as shown in figure 7. The feature
having a length of 0.02 cm gives 130 k/cm, in excellent agreement with the direct
measurement on the baby sensor.
p-implant
R=2605 Ohm, length=0.02 cm
R/cm=130kOhm/cm
3.E-04
2.E-04
I (A)
1.E-04
0.E+00
-1.E-04
-2.E-04
-3.E-04
-0.60
-0.40
-0.20
0.00
0.20
0.40
0.60
U (V)
Figure 7: Implant resistance measurement using feature ‘N LW20’
4.3. Aluminum Strip Resistance
The aluminum strip resistance was measured by applying a voltage differential between
two neighboring AC pads and mapping the current versus voltage. Three sets of two
strips were measured on test structure 1 and give a resistance for the aluminization of
22.1 /cm, well within our specification which requires a resistance of less than 30
/cm. Figure 8 shows the measurement for the set consisting of strips 6 and 8. The
measurement on all the other strips gives identical results. The measured aluminum
resistance corresponds to an aluminum thickness of about 1.5 m, if we assume a 2 m
wider metal layer over the actual implant width of 7 m. The introduced ENC noise of a
total series resistance of around 170  for a strip length of 7.74 cm in front of the
10
preamplifier amounts to a noise contribution of less than 200e at our shaping times of 132
ns. It can therefore be neglected.
The aluminum strip resistance was also determined from feature ‘LW6000’ on the test
structure. The feature corresponds to an aluminum trace that meanders around the
implants and MOS structures (see Figure 9). The resistance of this structure was
measured to be 210 for a trace length of 60.6 mm, or 35 /cm as figure 10 shows.
Baby Sensor, Strips 6-8
Voltage (V)
0.6
0.4
0.2
0
-0.002
-0.0015
-0.001
-0.0005
0
0.0005
0.001
0.0015
0.002
-0.2
-0.4
-0.6
I (A)
Figure 8: I-V curve for aluminization on strips 6 and 8 for the baby sensor on test
structure 1.
11
Figure 9: A long aluminum trace surrounds various implant features on the test
structure. The implants in the upper left-hand corner correspond to the p+ and n+
implants.
Aluminium Trace LW6000
3.E-04
2.E-04
I (A)
1.E-04
0.E+00
-1.E-04
-2.E-04
-3.E-04
-0.06
-0.04
-0.02
0.00
0.02
0.04
0.06
U (V)
Figure 10: Measurement of aluminum strip resistivity on feature ‘LW6000’.
4.4. Polysilicon Resistance
4.4.1. Strip Poly-silicon Resistance
The resistance of the poly-silicon resistor was measured on the baby sensor. The baby
sensor has exactly the same layout as the full strip detector. Also on the baby sensor there
12
are intermediate and readout strips. Simply because of space constraints, the poly-silicon
resistors cannot all reside on one end. The resistors alternate between strips.
Consequently, the poly-silicon resistors for the intermediate strips are all at one end, and
the resistors for the readout strip are all at the other end. The poly-silicon resistor value
has been measured on the intermediate strips, by placing the probes on the DC-pad and
the bias line and applying a voltage differential, while the detector was biased. Figure 11
shows the I-V curve as measured for four strips on test structure 5. The measured polysilicon resistor value is 0.7 ± 0.1 M.
Performing the same measurement on the adjacent strip, the readout strip, still provides
valuable information. Placing the probe on the DC pad measures the implant and polysilicon resistance in series. The average value of the resistance was determined to be 1.5
± 0.15 Mslightly lower than the sum of the individual resistances.
Rpoly on baby sensor
1.5E-05
1.0E-05
I (A)
5.0E-06
0.0E+00
strip 1
-5.0E-06
strip 2
-1.0E-05
strip 3
strip 4
-1.5E-05
-5
-4
-3
-2
-1
0
1
2
3
4
5
U (V)
Figure 11: Measurement of the poly-silicon resistance on intermediate strips of the baby
sensor on test structure 5.
4.4.2. Monitor Polysilicon Resistors
The test structure also contains various arrays of poly-silicon resistors for monitoring
purposes. These arrays are labeled ‘PSxx’, with xx a numeric identifier. The resistors
labeled ‘PS-20’, ‘PS-10’ and ‘PS-0’ have all been measured. The resistors show an
“ohmic” behavior in the region between ± 2 V, and have a consistent resistance of 1.05 ±
0.01 Min this region.
4.5. Depletion Voltage of Silicon Diodes
The test structure for the layer 1 sensors has silicon diodes with and without guard ring
structure. The depletion voltage of the silicon diodes was determined using the C-V
method. Figure 12 shows the C-V curve for the diodes for four frequencies. The depletion
13
voltage is lower by about 5-10% with respect to the depletion voltage for the
corresponding sensors on this wafer. In general, the depletion voltage for segmented strip
detectors is higher than the depletion voltage for planar diodes.
1/C2 (pF-2)
C-V L1 diode teststructure
4.50E+03
4.00E+03
3.50E+03
3.00E+03
2.50E+03
2.00E+03
1.50E+03
1.00E+03
10 kHz
100kHz
500kHz
1 MHz
20
40
60
80
100 120 140 160 180
bias (V)
Figure 12: C-V curve for planar diode on test structure 5.
4.6. Total Load and Interstrip Capacitance
The total load and interstrip capacitance was measured on test structure 8. The procedure
as outlined in the Quality Assurance document [4] was followed. The setup with the
equivalent circuit diagram is shown in figure 28. Two measurements are made. The first
measurement involves a three probe setup. Probe 1 is placed on the AC pad of strip N;
Probe 2 and 3 are placed on the AC pads of the adjacent strips, strip N-1 and N+1,
respectively. The probes on strip N-1, N+1 and the connection to the backside of the
detector are tied together at the input of the LCR meter. The capacitance measured this
way is, to first order, the total load capacitance the readout chip sees and is denoted by C L
or C1. In a second measurement one of the probe tips adjacent to the center probe tip, is
lifted. The capacitance measured in the second measurement is denoted C2. From the
equivalent circuit diagram one finds that
CL = C1 = 2 Ci + C b
14
C2 = Ci + Cb .
Here, Cb refers to the capacitance of the strip to the backplane and Ci is the interstrip
capacitance. CL is to first order the total load capacitance on the readout chip. The second
measurement can of course be performed for both neighboring strips and the results
should be identical. In the results presented here, unless mentioned explicitly, the
capacitance C2 is the average capacitance measured for the pair (N, N+1) and the pair (N,
N-1):
C2 = ( C2 (N, N-1) + C2 (N, N+1) ) / 2 .
Figure 13 shows the dependence of C1 as function of frequency for two triplets of strips
on the test structure. At a frequency of 1 MHz, the frequency of relevance for operation
with the SVX4 readout chip, the total capacitance is about 8.5 pF, or 1.1 pF/cm. Figure
14 shows the voltage dependence of the total load capacitance. As was also observed for
the outer layer sensors, the capacitance reaches a plateau already at 10 V.
Total strip capacitance (pF)
10
C L (pF)
8
6
4
2
0
1
10
100
1000
Frequency (kHz)
Figure 13: Total load capacitance as function of frequency as measured on the baby
sensor.
The interstrip capacitance to both neighbors and to one neighbor is shown in figure 15.
The high frequency limit is 0.39 pF/cm for the interstrip capacitance to one neighbor and
0.71 pF/cm for the interstrip capacitance to both neighbors. Our specifications call for a
total interstrip capacitance of less than 1.2 pF/cm.
15
Total strip capacitance
20
1kHz
2kHz
CL (pF)
15
5kHz
10kHz
10
20kHz
50kHz
100kHz
5
200kHz
500kHz
0
1MHz
0
30
60
90
120
bias (V)
Figure 14: Total load capacitance as measured on the baby sensor for various
frequencies as function of backside bias voltage
C int (pF)
Interstrip capacitance
7
6
5
4
3
2
1
0
Cint both
Cint both
Cint one neighbor
1
10
100
1000
Frequency (kHz)
Figure 15: Interstrip capacitance as function of frequency, with respect to one and two
neighbors.
16
4.7. MOS capacitance and flatband shift
The delivered test structures from Hamamatsu contain also a p-MOS structure1. This
MOS capacitor is labeled as “FI” in figure 9 and has an area of 0.5 mm2. MOS capacitors
are well suited to determine the fixed oxide charge in the oxide layer. Figure 16 shows
the C-V characteristics of three MOS capacitors on three different test structures. The
measurements were done as outlined in the QA document [4]. We are measuring in the
high frequency regime as opposed to a quasi-static measurement and hence we are
sensitive to the charge in the depletion layer only. The curves in figure 16 look as
expected for high frequency measurements on MOS capacitors. At large positive gate
voltages we are in the accumulation region, since the majority carriers from the nsubstrate are attracted to the silicon-oxide interface. Going to lower gate voltages
depletes the substrate more and more and the capacitance is therefore dropping. The
voltage separating the accumulation from the depletion region is called flatband voltage.
Inversion finally occurs at negative voltages, since an inversion layer of minority carriers
is built directly at the interface.
MOS-FI structure
35
30
capacitance (pF)
25
20
L2-63/64
L2-62
L1-6/7
15
10
5
0
-15
-10
-5
0
5
10
15
20
gate voltage (V)
Figure 16: MOS capacitance measurements on three test structures.
In the case of our MOS capacitors the flatband voltage is at very low values of 1-2 V.
Since the flatband condition corresponds to a flat energy band in the silicon ideally at
zero volt of the gate, the amount of fixed charge in the oxide is less than 110-11 C/cm2
and hence rather small.
1
The substrate is n-type, but the created inversion layer is of p-type.
17
5. Electrical Characterization of the Sensors
Hamamatsu delivered ten prototype sensors to Fermilab in September 2002 and another
three in July 2003. The sensors were probed both at Fermilab and KSU. The following
sets of measurements are performed on the sensors:
i.
I-V curve, to measure total detector leakage current as function of bias voltage up
to 700V.
ii.
Long-term stability of detector total leakage current
iii.
C-V curve, to measure the total detector capacitance as function of bias voltage.
The depletion voltage is extracted from this measurement
iv.
AC-scan, to measure the coupling capacitance and the current of individual strips.
Sensor defects, most commonly shorted strips and pinholes, are identified with
this measurement.
v.
DC-scan, to measure the leakage current of individual strips. Sensor defects, most
commonly leaky strips, are identified with this measurement.
vi.
Interstrip isolation by measuring the interstrip resistance and measurement of
resistance of bias resistor
vii.
Interstrip capacitance and total load capacitance
An IV and CV as well as full strip scans were performed for all sensors. Not all sensors
have undergone the full series of detailed tests like interstrip capacitance measurements,
however. Table 2 shows a summary which measurements were performed on which
sensor. Some of the sensors tested were irradiated and the results are described in
reference [5]. Some other sensors were used to build detector modules. The results of the
module readout noise tests will be described in an accompanying note [7].
Serial
No.
Tested
at
1
3
4
6
7
9
11
12
13
20
21
22
24
FNAL
FNAL
FNAL
KSU
FNAL
FNAL
KSU
KSU
KSU
KSU
FNAL
FNAL
FNAL
IVScan
I(nA)
@
350V
81
67
54
56
482
174
24
28
28
31
54
51
67
CVScan
FDV
(V)
DCscan
strip #
AC-scan
strip #
110
120
110
130
125
105
125
115
120
126
120
130
120
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
13,47,267
320
0
0
0
0
47/48
0
0
0
Comment
broken
HPK
Strip#
267
320
irradiated
irradiated
irradiated
47/48
HPK
I(nA)
@350V
HPK
FDV
(V)
56
56
69
76
228
55
55
65
55
67
69
65
85
130
150
140
150
150
130
140
130
140
140
140
150
140
18
Table 2: Summary of measurements on the L1 sensors.
5.1. Characterization at the Manufacturer
The contract with the vendor calls for an electrical characterization of the sensors at the
company. Each sensor should undergo:
o I-V curve to 500 V at a temperature of 25  3C and a relative humidity of <50%
o Optical inspection for defects, opens, shorts and defects, and verification of mask
alignment to better than 2.0 m
o Depletion voltage as determined from the C-V method
o
AC capacitance value measurement and pinhole determination
In addition, the manufacturer is asked to verify the poly-silicon resistor value, implant
resistivity, coupling capacitor value and its breakdown as well as the aluminum resistance
on the test structures. The vendor provides these results as average per delivered batch.
All ten sensors came from a single batch, SWA61589, and the Hamamatsu data
pertaining to the single batch is listed in Table 3. Our measured results on the test
structures, as described in section 4 of this document, are in reasonable agreement with
Hamamatsu’s findings on the aforementioned quantities.
Parameter
Rpoly
Rimplant
RACAL
Cc
Vbreakdown
Average Value
0.75 M
1.6 M
170 
107 pF
155 V
Table 3: Average value of parameters as determined by Hamamatsu using the monitor
patterns pertaining to the single batch of wafers for the Layer 1 sensors.
Figure 17 shows, for completeness, the IV-curves for all 13 sensors as measured by
Hamamatsu. Sensor 7 shows an anomalous current bias voltage behavior, but still meets
our specification of a total detector current less than 4 A at 700V.
19
Total Leakage Current, HPK
500
450
1
3
4
6
7
9
11
12
13
20
21
22
24
400
350
Ileak (nA)
300
250
200
150
100
50
0
0
100
200
300
400
500
600
700
800
Bias Voltage (V)
Figure 17: Total detector leakage current as function of bias voltage as measured by
Hamamatsu for all 13 layer 1 prototype sensors
All sensors have a depletion voltage determined at the vendor between 130-150V, with a
granularity of 10V. HPK measures the C-V dependence in steps of 10 Volts and defines
the full depletion voltage as the lowest voltage where the increase in 1/C2 is found to be
less than 2% [8]. This method tends to overestimate the depletion voltage by about 20V
compared to the DØ measurement, which uses an alternate criterion, discussed in section
5.4.
Table 4 lists all the sensor defects as noted by Hamamatsu. Three sensors out of 13 are
flagged as having a defect. Sensors 6 and 7 have a shorted coupling capacitor. Sensor 20
has an open on two neighbored strips in the coupling capacitor. Unfortunately sensor 6
was irreversibly damaged during the initial probing.
Lot No.
Serial No.
Type
Ch. No.
SWA61589
6
Coupling short
267
SWA61589
7
SWA61589
20
Coupling short
AC-AL open
320
47
AC-AL open
48
Table 4: The defects on the three sensors as noted by Hamamatsu
20
5.2. IV-Scan
All sensors that were tested have good detector leakage current, except one. Sensor 7 has
an anomalous leakage current, but still meets our specification. Figure 18 shows the
difference in total detector leakage current as measured at Hamamatsu and our test sites
and shows that, within the uncertainty of the measurement, we can reproduce the
Hamamatsu measurements. The difference is less than 30 nA, except for sensor L1-0007
and L1-0020 at very high bias voltages. Note also, that the leakage current difference is
almost independent of bias voltage. The small offset in current can be attributed to the
different environmental conditions for the measurements at the two different locations.
Our own measurements were done at a temperature of 19-21C whereas Hamamatsu
measures at 25ºC. The breakdown voltage is well above the specified 700 V for all
sensors.
HPK - Our Measurement
200
150
1
3
100
4
 I (nA)
6
7
50
9
11
12
0
13
20
21
-50
22
24
-100
-150
0
100
200
300
400
500
600
700
800
Bias Voltage (V)
Figure 18: Difference in leakage current between Hamamatsu and our own measurement
for all layer 1 sensors.
5.3. Long Term Stability
A long term test facility is available at Fermilab to monitor sensor behavior under bias for
extended periods. The facility consists of a light and gas tight box, which can
accommodate up to six sensors in aluminum holders and a PC/Labview based readout
system shown in figure 19. Data for sensor currents, box temperature and humidity are
recorded at set intervals, typically every five minutes, during the burn-in period. Figure
20 shows the results of a burn-in of four L1 prototype sensors over a period spanning
approximately 96 hours. Detectors were biased to 300 V. The sensors were purged with
nitrogen at various rates. The lower graph shows the dew point measured inside the dry
box. Note that the minimum dew point meter readout is -40C. Under all conditions, no
21
run-away behavior of the currents is observed. Even sensor 7 shows a stable sensor
leakage current. Room temperatures varied by 3.5C during the test and the currents are
corrected to 20C. Some small residual time dependence is visible. These variations are
most prominent during periods when the temperatures change most rapidly. This residual
variation appears to be due to a time lag between the air temperature recorded by the PC
and the actual sensor temperature. However, no long-term drifts of the leakage currents
have been observed and the sensors turned out to exhibit a rather stable bias behavior.
The measurements were repeated for a 24 hour period at a bias voltage of 500 V and the
currents and dew point measurements are shown in Figure 21. The currents plotted in the
lower figure are again normalized to 20C. Again, no long-term drifts of the leakage
currents are observed, also not for sensor 7. The slight increase for sensor 4 after two
hours precedes the slight increase in temperature that is observed.
22
GPIB
Temp,
humidity
HP 3456
DVM
Keithley
485 pa
meter
Keithley 706 switch box
Keithley
2410 ps
PC
Labview
Test Box with detector diodes
Figure 19: Setup for long term sensor current stability
23
Long Term Stability Layer 1 sensors at V bias = 300 V
I (nA)
10000
1000
100
L1-09
L1-03
L1-07
10
0.0E+00
L1-04
5.0E+04
1.0E+05
1.5E+05
2.0E+05
2.5E+05
3.0E+05
3.5E+05
Elapsed Time (s)
T ( 0C)
30
20
10
0
-10
-20
-30
Temp
Dew Point
-40
-50
0.0E+00
5.0E+04
1.0E+05
1.5E+05
2.0E+05
2.5E+05
3.0E+05
3.5E+05
Elapsed Time (s)
Figure 20: Four day burn-in results of four outer layer Hamamatsu sensor. The bias
voltage was set to 300V.
24
Temperature and Dew Point, Long Term Test, (4/14-4/15/03)
Temperature ( 0C)
30
20
10
0
-10
-20
-30
Temp
-40
Dew Point
-50
10
100
1000
10000
100000
Time (s)
Long Term Stability, Layer 1 Sensors, Vbias = 500 V
10000
I (nA)
1000
100
L1-09
L1-03
10
L1-07
L1-04
1
10
100
1000
10000
100000
Time (s)
Figure 21: Long term stability of four Layer 1 sensors over a period of 24 hours
at a bias voltage of 500 V. Note the logarithmic scales.
25
5.4. CV-Scan
The CV-scan is used to determine the depletion voltage. The depletion voltage at our
testing centers was extracted as the intercept of the two straight lines in a 1/C 2 - Vbias plot.
Both Fermilab and KSU measured depletion voltages, which were consistently lower
than Hamamatsu. However, the difference can be attributed to Hamamatsu’s alternative
approach to determine the depletion voltage. Figure 22 shows the CV-curves for all 13
Layer 1 sensors. All of them show the clear and distinct characteristics of a typical 1/C 2
versus bias voltage graph. A comparison of the depletion voltages as measured at the test
centers and by Hamamatsu is shown in figure 23.
-2
1/C2 (1/pF)2 * 1 E9
C vs Voltage
3000
L1_001
2500
L1_003
L1_004
L1_006
2000
L1_007
L1_009
1500
L1_011
Vdepl=105-130V
L1_012
1000
L1_013
L1_020
L1_021
500
L1_022
L1_024
0
0
50
100
150
200
250
300
Bias Voltage (V)
Figure 22: CV-curves for all layer 1 sensors; the straight lines indicate the method for
depletion voltage determination.
26
Depletion Voltage HPK (V)
160
150
140
130
120
110
100
90
90
100
110
120
130
Depletion Voltage CV Scan (V)
Figure 23: Comparison of depletion voltage values measured with our CV method and
Hamamatsu measurements; the line indicates one-to-one correspondence.
5.5. AC-Scan
The integrity of each strip is verified with the AC-scan. The AC-scan measurement is
performed probing the AC-pad and the bias ring. A backside voltage of 20V is applied.
The scan consists of two steps. First, a voltage of 80V is applied to the AC pad and the
current through the dielectric layer of a R/O strip is measured. In a second step the
voltage is lowered to 0V and the capacitance is measured with a LCR meter. The
capacitance represents the coupling capacitance of the dielectric layer. Figure 24 shows a
typical AC-scan taken at a frequency of 100 kHz. The coupling capacitance is lower due
to the frequency response of a low-pass filter consisting of coupling capacitor and
implant resistor. At lower frequencies the measured capacitances approach their zerofrequency limit and the values so obtained are consistent with the test structure
measurements in section 4. As a result of our probing we could verify that ten sensors
claimed by Hamamatsu as being flawless, do not have any open, short or pinhole. The
three AC defects reported by the vendor on the three remaining sensors (and listed in
Table 4) were all identified during our probing of the sensors as indicated in Table 2.
However, we also found two more pinholes on sensor 6. Unfortunately these two defects
could not be confirmed since the sensor was irreparably damaged during the initial
probing.
Figure 25 shows the current through the dielectric layer as measured on sensor 13 with
the AC-scan. Good strip capacitors typically have currents smaller than 100 pA. In case
of a pinhole, the dielectric currents are much higher and reach immediately the current
compliance of the SMU, which is set to 100 nA. As can be seen from figure 25 no
pinholes were detected on sensor 13.
27
AC Scan, Layer 1, Sensor 13
80
70
60
Cc (pF)
50
40
30
20
10
0
0
50
100
150
200
250
300
350
Strip Number
Figure 24: AC-scan of coupling capacitance for sensor 13 at a frequency of 100 kHz.
Idielec, Layer 1, Sensor 13
200
180
160
Number of strips
140
120
100
80
60
40
20
0
0.051
0.056
0.061
0.066
0.071
0.076
0.081
0.086
0.091
0.096
0.101
0.106
0.111
0.116
Idiel (nA)
Figure 25: Idielec as measured in AC-scan for sensor 13.
28
5.6. DC-Scan
A full DC-scan was also performed on all L1 sensors, with the backside at full bias. No
anomalies were observed. Figure 26 shows the distribution in strip current for all 3840
R/O strips of the first ten layer 1 prototype sensors that were received in fall 2002. We
require strips to have a leakage current less than 10 nA per strip at the full depletion
voltage (FDV). Only 14 strips have a strip current above 1 nA. None of these strips can
be considered leaky according to above definition.
Ileak for all 10 Layer 1 sensors
Number of Strips
1200
Average leakage per strip is 0.04 nA
1000
800
600
400
14 strips leakage at 1 nA
200
3
1
0.27
0.26
0.25
0.24
0.23
0.22
0.2
0.21
0.19
0.18
0.17
0.16
0.15
0.14
0.13
0.12
0.1
0.11
0.09
0.08
0.07
0.06
0.05
0.04
0.03
0.02
0
0.01
0
I leak/strip (nA)
Figure 26: DC-scan results for the leakage current per strip for all strips of the ten
prototype Layer 1 sensors.
5.7. Polysilicon bias and interstrip resistors
The DC-scans performed at KSU were also used to determine the poly-silicon resistor
values on all intermediate strips by applying a small voltage of 1V across the DC-pad and
the bias rail while the backside was kept at bias potential. Such a scan is sensitive to strip
regions on the sensor with low interstrip resistances or to bias resistor inhomogeneities.
Figure 27 shows the result of one scan performed on sensor 13. The obtained values of
0.8 M are in good agreement to the test structure observations reported in section 4 and
meet our specification. If the polysilicon bias resistors would be measured for the readout
strips a combination of the polysilicon resistor and implant resistance is probed. A value
of 2.35 is found for the combination of implant and polysilicon resistance, which is
somewhat higher but still consistent with our expectations and the measurements on the
test structures.
29
Further measurements were carried out to estimate the interstrip resistance between two
strips. Since the interstrip resistance is rather large, very small currents with large
fluctuations and big uncertainties are measured. We can only reliably say that the
interstrip resistance is of the order of several Gand hence well within our
specifications.
Polysilicon Resistor, Layer 1, Sensor 13
400
Number of strips
350
300
250
200
150
100
50
0
600
650
700
750
800
850
900
950
1000
1050
1100
More
Rpoly (kOhm)
Figure 27: Measurement of the polysilicon resistors for sensor 63
30
5.8. Interstrip Capacitance
The interstrip capacitance was measured extensively on a few sensors. The results shown
here are taken from the measurements on sensor 1. The measurements on the other
sensors give consistent results. The procedure followed was already described in the
section on the test structures. The setup with the equivalent circuit diagram is shown in
figure 28. The two measurements made are the measurement of the total load
capacitance, denoted by CL or C1, and the measurement of C2, which is the combination
of the interstrip capacitance and the capacitance to the backplane. From the equivalent
circuit diagram one finds that
CL = C1 = 2 Ci + C b
C2 = Ci + Cb .
Here, Cb refers to the capacitance of the strip to the backplane and Ci is the interstrip
capacitance. CL is to first order the total load capacitance on the readout chip. The
capacitance of the probes and leads was measured to be on average about 1 pF. In case
‘corrected’ results are shown, the measured capacitance of the leads, dependent on
frequency of course, was subtracted from the measurement. Note that in the extraction of
Ci no correction is needed.
probe 2 on
AC pad
(strip N+1)
probe 1
Ext. LCRprobe 3 on adapter meter
AC pad (strip
N-1)
Cintrstr
Cs
Cs
Cs
Test chuck
Figure 28: Setup for interstrip capacitance measurements
Figure 29 shows the dependence of C1 (the load capacitance) as function of frequency for
the triplet of strips (193, 194, 195) on sensor 1. At a frequency of 1 MHz, the frequency
of relevance for operation with the SVX4 readout chip, the total capacitance is about 1.1
pF/cm.
31
Total Load Capacitance, Sensor 1
10.00
Ch.193-194-195
CL (pF)
8.00
6.00
4.00
2.00
0.00
1.0E+02
1.0E+03
1.0E+04
1.0E+05
1.0E+06
Frequency (Hz)
Figure 29: Frequency dependence of Cl (see text for details)
Figure 30 shows the frequency dependence of the total load capacitance, as shown in the
previous figure, for the full range of bias voltages. An initial backside voltage of 5 Volts
was applied; the voltage was then increased to 10 Volts and then incremented in steps of
10 to 130 Volts. The data in Figure 29 was taken at a backside bias voltage of 110 Volts,
which corresponds to full depletion voltage plus 10%.
Total Load Capacitance at various V bias
12
5v
10v
10
20v
CL (pF)
30v
8
40v
50v
60v
6
70v
80v
4
90v
100v
2
110v
120v
0
100
130v
1000
10000
100000
1000000
Frequency (Hz)
32
Figure 30: Frequency dependence of Cl for various backside bias voltage settings
Total Load Capacitance
5.00E+02
12
8.00E+02
1.00E+03
10
2.00E+03
5.00E+03
CL (pF)
8
8.00E+03
1.00E+04
6
2.00E+04
5.00E+04
4
8.00E+04
1.00E+05
2
2.00E+05
5.00E+05
0
8.00E+05
0
50
100
Vbias (V)
150
1.00E+06
Figure 31: Dependence of Cl on backside bias voltage for various frequencies.
Figure 31 shows the dependence on the backside bias voltage for all frequencies. The
dependence on the backside bias voltage decreases for increasing frequency. Moreover,
there is a change in slope of the dependence on backside voltage at high frequency as was
observed for the outer layer sensors.
Using the measurement of C2 and C1 the interstrip capacitance can be determined, using
the two equations given at the beginning of this section. Figure 32 shows the results for
two strips as measured on sensor 1. The ‘wiggle’ in the capacitance, as was observed for
the outer layers at around 100kHz, is not present. The value for the interstrip capacitance
to a single neighbor is 3 pF or 0.39 pF/cm, which is well within our specification and also
in perfect agreement with the measurement on the test structure shown in figure 15.
33
Ci (pF)
Interstrip Capacitance
3.50
3.00
2.50
2.00
Ci-Ch193
Ci-Ch194
1.50
1.00
0.50
0.00
1.0E+02
1.0E+03
1.0E+04
1.0E+05
1.0E+06
Frequency (Hz)
Figure 32: Interstrip capacitance as function of frequency for two strips of sensor 1.
5.9. Summary
Table 2 gives a summary of the measurements. Strip numbers refer to strips that have
been flagged as faulty by our testing procedures. For the layer 1 sensors two strips in
addition to the strips identified by Hamamatsu have been found as defective, but we
could not confirm these defects due to sensor damage. The column labeled HPK strip#
indicates the strip number(s) identified by Hamamatsu as having a defect. The type of
defect that was listed in table 4 and all defects identified by Hamamatsu were confirmed.
Sensor 6 and 7 had pinholes in channels 267 and 320 respectively. The coupling
capacitance for strips 47 and 48 of sensor 20 are lower by about 10-15% compared to the
nominal value.
Note also in table 2 that the KSU I-V measurements have been performed at a lower
temperature, so that in general the recorded leakage currents are somewhat smaller than
at FNAL.
34
6. Mechanical Characterization of the Sensors
The layout of the sensor as specified in drawing number 3823.210-ME-399436, provides
a scratch pad field containing 5x4=20 pads. The vendor was asked to provide in three sets
of 4 scratch pads a unique serial number coding for the sensors, with the rightmost pad
representing the LSB. The remaining pads would be used for QC pass/fail marks. As was
the case for the outer layer sensors, Hamamatsu uses groups of four bits to represent
decimal numbers in binary format, rather than a binary representation as we expected.
All ten sensors from the September delivery were measured on an Optical Gage Products
(OGP) coordinate measuring machine to verify the mechanical dimensions of the sensors.
A reference system was established using the same convention as in the drawing with a
corner fiducial defining the origin. The x-axis runs along the long side of the detector,
parallel to the strip orientation; the y-axis is along the short axis, perpendicular to the
strip orientation. The flatness of the sensors was measured by defining a grid of 11x11
points in x and y and measuring the z position of the top surface of the sensor. The OGP
has an intrinsic z-resolution of a few m. Figure 33 shows the measurements for sensor 1.
The sensor is warped along both the short and long axis. The difference between the
minimum and maximum z-position on the sensor is then determined (see figure 34). The
average of the ‘highest’ and ‘lowest’ point on a sensor is 45 m, with the highest
difference of 48 m. Our specifications call for a sensor warp of less than 50 m, agreed
to by Hamamatsu only on a best effort basis. Although it was on a best effort basis, all
sensors meet the specification.
The flatness data is analyzed further to see if there is a difference in sensor warp along
either the x-, or y-axis. The sensor is divided into eleven slices along the x (y-) axis and
the z-coordinate is plotted as function of the y- (x-) position along the sensor. The graphs
in figure 35 show the measurements for sensor 7. The set of data points for each slice is
fitted to a parabola, which fits the data points very well, and the coefficient of the
quadratic term is extracted. The average values for the coefficient of the quadratic terms
are -0.25 10-4 (mm-1) for the curvature along the x-axis, and -0.28 10-4 (mm-1) for the yaxis. The sensors have a slightly stronger warp along the short axis, perpendicular to the
strips, than along the long axis. This is to be compared to the warp for the outer layer
sensors for which the values are -0.22 10-4 (mm-1) and -0.28 10-4 (mm-1), respectively.
In addition to the flatness measurements, a set of ten measurements each was taken along
the two short edges of the sensor and a set of twenty measurements were taken for each
of the long edges of the sensor. With this data four characteristics were verified: the
sensor cut width and cut length, the accuracy of the cut edges, and the parallelism of the
corresponding cut edges. In the determination of the cut width and length, the average
position along the cut edge was used.
35
3 D G ra p h
L 1 -0 1 s e n s o r
0 .04
0 .0 3
0 .02
0 .0 1
Z D ata
0 .0 0
1 00
80
-0 . 0 1
60
40
-0 . 0 2
-5
20
0
a
-20
at
-15
a ta
Y D
D
-1 0
X
0
-25
-30
C o l 1 vs C o l 2 vs C o l 3
Figure 33: Flatness measurements for inner layer sensor 1. Units along all axes are mm.
(Zmax-Zmin) distribution for 10 sensors
Zmax (  m)
49
48
47
46
45
44
43
42
41
40
39
1
2
3
4
5
6
Sensor #
7
8
9
10
Figure 34: Maximum variance in z for a set of ten Hamamatsu inner layer sensors,
plotted versus sensor serial identifier.
36
Figure 35: Sensor z-position as function of x and y for eleven slices in y and x,
respectively.
The maximum difference between two measurements along each of the four sides for the
ten sensors measured is shown in figure 36. As was the case for the outer layer sensors,
the cut edges are again extremely accurate. The absolute x or y positions of all edge
measurements for the four edges for all ten sensors measured, are plotted in figure 37.
There are 16 measurements for each of the two long edges, and 11 (12) measurements at
the close (far) short edge of the sensor. Combining the measurements in x and y, gives an
average cut width of the sensors of 24.312 mm and an average cut length of 79.404 mm,
to be compared to the nominal values of 24.312 mm and 79.400 mm, respectively. The
angle between the lines fitted to the cut edges averages 90.00 ± 0.004 degrees. All in all,
the sensors are superior in all mechanical aspects, including the sensor warp.
37
Figure 36: Maximum difference between two measurements (in m) along
each of the four sensor cut edges. Each graph corresponds to a different cut
edge.
38
Figure 37: Absolute coordinates of x or y positions along the cut edges for ten sensors
combined. Each graph corresponds to a different cut edge.
39
References
[1]
[2]
[3]
[4]
[5]
[6]
[7]
[8]
DØ Run IIb Silicon Detector Technical Design Report, Chapter 3: Silicon Sensors
Silicon Sensor Specifications for Layers 0 and 1, July 29 03.
A. Bean et al., Studies of the Radiation Damage to Silicon Detectors for Use in
the DØ Run IIb Experiment, DØ note 3958.
A. Bean et al., Silicon Sensor Quality Assurance for the D0 RunIIb silicon
detector: procedures and equipment, DØ note 4120.
T. Bolton et al., Measurements on irradiated L1 sensor prototypes for the DØ Run
IIb silicon detector project, to be submitted as DØ note.
M. Demarteau et al., Characteristics of outer Layer Silicon Sensors for the Run
IIb Silicon detector, to be submitted as DØ note.
Readout of Run IIb Outer Layer Silicon Sensors for the Run IIb Silicon Detector,
A. Nomerotski and E. von Toerne, in preparation
K. Hara, Tsukuba University, private communication
All documentation is available on:
http://www.physik.unizh.ch/~lehnerf/dzero/prr/prr_l1.html
40