Survey
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
routine previously used was modified and adapted to ensure high mass accuracy across the full mass range. With the data collected on different types of samples and presented in this study we demonstrate the successful analysis after implementation of the High Mass Range (HMR) mode, successful desolvation and optimized critical hardware operation settings. This makes the instrument an ideal platform to cover the three major workflows in BioPharma: intact mass analysis under denaturing and under native conditions in HMR mode, subunit analysis (reduced mAb and or IdeS digested mAb) in protein mode and peptide mapping in standard mode (see Figure 1). Enabling Mass Spectrometric Analysis Normal Mode of Intact Proteins in Native Conditions on A Hybrid Quadrupole-Orbitrap Mass Spectrometer fo Peptide Mapping Poster Note 64 8 04 Figure 1: Operating modes for the three major BioPharma workflows: Normal Mode, Protein Mode and HMR mode MAbPAC RP column for denatured samples SMART digest MAbPAC SEC column for native samples Acclaim RP C18 column HMR Mode Intact Analysis native & denatured Kai Scheffler1, Eugen Damoc2, Aaron O. Bailey3, and Jonathan Josephs3 Thermo Fisher Scientific, 1Im Steingrund 4-6, Dreieich, Germany, 2Hanna-Kunath-Str. 14, Bremen, Germany, Vanquish UHPLC 3 355 River Oaks Parkway, San Jose, CA, USA B C D Enabling Mass Spectrometric Analysis of Intact Protein ©2016 Genovis AB Kai Scheffler1, Eugen Damoc2, Aaron O. Bailey3, INTRODUCTION Analysis of proteins in native-like conditions free of organic solvents can allow proteins to preserve non-covalent interactions and retain high degrees of folding. This effect has analytical benefits: greater protein folding leads to reduced charge states leading to increased mass separation and increased signal at higher m/z. This strategy has been utilized for analysis of antibodies and antibody drug conjugates present in highly complex mixtures of different antibody/drug combinations [1]. Requirements for performing native MS on antibody samples include scanning towards 8000 m/z and increased transmission optimization for large compounds. Such features are not compatible with current commercially available quadrupole-Orbitrap instruments. Here we show results obtained after successful implementation of modifications aimed at adding the capability to perform native MS analysis without compromising performance of normal operation modes. and FabRICATOR® (IdeS) digest and/or reduction MAbPAC RP column Q Exactive BioPharma • Protein Mode • Enhanced Resolution Mode • High Mass Range Mode ( m/z ≤8000) Subunit Analysis Jonathan Josephs3, Protein Mode F c Thermo Fisher Scientific, w MATERIALS AND METHODS RESULTS Samples: Samples usedmany in this study that are play ammonium hexafluorophosphate Fishersome Scientific, part relate numberto There are factors a key role in the analysis(AHFP, of proteins, of which A0368370), Trastuzumab(buffers, (tradenamesolvents, Herceptin, Roche, UK). sample and preparation additives) while others relate to the mass spectrometer’s source conditions as well as the physical environment inside the instrument Chromatography: A [2,3]. ThermoThe Scientific™ Vanquish™ UHPLC was usedHF for all LC/MS experiments. For(Figure native analysis, 50 Q Exactive Plus and system Q Exactive mass spectrometers 2A) have mM ammonium been acetateintroduced buffer (99.99%, used.option, Reversedwhich phase was chromatography was previously withSigma the Aldrich) Protein was Mode one of many performed with water/0.1% formic acid and acetonitrile/0.1% formic acid on a Thermo Scientific™ MAbPac™ RP advancements for intact protein analysis on the Orbitrap platform. For these two instruments 2.1x50 mm column. an automated HCD gas control was introduced by using an electronically controlled valve for re T a a H e in F to h is d T a o a 1 L Enabling Mass Spectrometric Analysis of Intact Protein The analysis of intact proteins under native conditions is more challenging than under denaturing conditions since the buffers used don’t contain any organic solvents. Performing electrospray from aqueous buffer solutions produces larger solvent droplets size and desolvation is less efficient. Moreover, for large proteins such as intact antibodies the required mass range for analysis under 1 range of more than 6000 m/z due 2 to a smaller number of accepted native conditions requires a mass charges. The increase of the upper mass range on the mass spectrometer was achieved by implementing instrument control software changes. The analysis of molecules across the full mass range including the detection of proteins under native conditions required the use of optimized parameter settings to ensure efficient desolvation in the front region of the instrument, the efficient transfer via multipoles, efficient trapping in the C-trap/HCD region and the sensitive injection and detection in the Orbitrap mass analyzer. Critical parameters are the optimization of in-source Analysis of proteins in native-like conditions free of solventsprocess. can allow proteins preserve fragmentation that strongly influences the support of organic the desolvation Also, for theto transmission non-covalent interactions and retain of folding. This effectto hasensure analytical benefits: efficiency specific voltages have high beendegrees evaluated and optimized robust and greater sensitive protein folding in leads to reduced statesperforming leading toanalyses increased mass separation and experiments increased conditions, performance the higher mass charge range when under native signal at higher m/z. This strategy has been utilized for analysis of antibodies and antibody drug that have not been possible so far on this type of mass spectrometer. Also, the standard calibration conjugates presentused in was highly complex mixtures toofensure different antibody/drug combinations [1]. routine previously modified and adapted high mass accuracy across the full mass Requirements for performing native MS on antibody samples include scanning towards 8000 m/z and range. increased transmission optimization for large compounds. Such features are not compatible with With the data collected on different types of samples and presented in this study we demonstrate the current commercially available quadrupole-Orbitrap Here we mode, show results obtained after successful analysis after implementation of the Highinstruments. Mass Range (HMR) successful desolvation successful implementation of modifications aimed atThis adding the the capability to perform MSto and optimized critical hardware operation settings. makes instrument an idealnative platform analysis without compromising performance of normal operation modes. cover the three major workflows in BioPharma: intact mass analysis under denaturing and under native Mass Spectrometry: nitrogen gas in the HCD cell for easier optimization of experimental conditions required for Mass spectrometers in this wished study are Scientific™ Q Exactive™ Plus and Q Exactive™ HF different types ofused analyses tothe runThermo on a single platform. systems with BioPharma Option. The instruments were operated under Tune 2.8 instrument control software in In Normal settings factory-optimized, suitable analyses and ions HMR mode, inMode which pressure the RF applied to theare C-trap was increased from 2,400 Vfor p-pmost to 2,900 Vp-p for better are cooled in the (Fig. trapping gas pressure is 1 which corresponds trapping of the high m/zC-trap ions. Also, to 2B). ensureThe better capture of high-m/z ions setting in the Orbitrap analyzer, the initial to a high vacuum pressure deltafrom (∆HV) ~3.1 e-5mbar. Thethe∆HV is defined as the difference central electrode voltage was adjusted −3.7ofkV to −3.4 kV, while setting during detection remained unchanged The S-lens RFon level was allowed to be increased to a setting of 200 in HMR mode and set to between(−5 HVkV). with HCD gas minus HV with HCD gas off. that for allMode experiments shown trapping here. Inlevel Protein the default gas pressure setting is 0.2 and that corresponds to a ∆HV Kai Scheffler , Eugen Damoc , Aaron O. Bailey3, and Jonathan Josephs3, Thermo Fisher Scientific, The combination of reduced C-trap and HCD cell gas pressures, and trapping ions in the HCD There are2.many factors that role in Plus/HF the analysis of proteins, some ofand which relate to Figure A) Schematic ofplay the aQkey Exactive mass spectrometers differences sample preparation solvents, additives) others relateB) to the Mode, mass in the trapping path (buffers, in the three different operatingwhile modes available: Normal spectrometer’s source conditions as well as the physical environment inside the instrument C) Protein Mode and D) HMR Mode. E) Illustration of improvement in signal intensity [2,3]. The Q Exactive Plus and Q Exactive HF mass spectrometers (Figure 2A) have for +17 charge state of a mAb light chain comparing Protein Mode and Normal Mode. previously been introduced with the Protein Mode option, which was one of many advancements for intact protein analysis on the Orbitrap platform. For these two instruments an automated HCD gas control was introduced by using an electronically controlled valve for nitrogen gas in the HCD cell for easier optimization of experimental conditions required for A of analyses wished to run on a single platform. different types In Normal Mode pressure settings are factory-optimized, suitable for most analyses and ions are cooled in the C-trap (Fig. 2B). The trapping gas pressure setting is 1region whichwhere corresponds source to a high vacuum pressure delta (∆HV) of ~3.1 e-5mbar. The ∆HV is defined difference CIDas is the applied between HV with HCD gas on minus HV with HCD gas off. Exactive Plus: setting is 0.2 and that corresponds to a ∆HV In Protein Mode the default trapping gas Qpressure Standard Orbitrap mass analyzer HF: which is 5x lower than in Normal Mode.Q Exactive Additionally, ions are transferred and cooled in the Ultra High Field Orbitrap mass analyzer HCD cell and thus have a longer flight path (Fig. 2C). The combination of reduced C-trap and HCD cell gas pressures, and trapping ions in the HCD cell prior to mass analysis extends the life time of protein ions resulting in increased signal Trapping in Normal Mode: intensities of isotopically resolved species 2E). B trapping gas(Fig. pressure setting 1 E 100 MAbPAC RP column N2 N2 Protein Normal ModeMode MATERIALS AND METHODS Samples: Samples used in thisSMART study digest are ammonium hexafluorophosphate (AHFP, Fisher Scientific, part number column A0368370), and Trastuzumab (tradename Herceptin,MAbPAC Roche,SEC UK). for native samples Intact Analysis Acclaim RP C18 column Chromatography: native & denatured A Thermo Scientific™ Vanquish™ UHPLC system was used for all LC/MS experiments. For native analysis, 50 mM ammonium acetate buffer (99.99%, Sigma Aldrich) was used. Reversed phase chromatography was performed with water/0.1% formic acid and acetonitrile/0.1% formic acid on a Thermo Scientific™ MAbPac™ RP 2.1x50 mm column. HMR Mode Mass Spectrometry: UHPLC Mass spectrometers used in thisVanquish study are the Thermo Scientific™ Q Exactive™ Plus and Q Exactive™ HF systems with BioPharma Option. The instruments were operated under Tune 2.8 instrument control software in HMR mode,FabRICATOR in which the RF applied to the C-trap was BioPharma increased from 2,400 V p-p to 2,900 Vp-p for better ® Q Exactive • Protein Mode of high-m/z ions in the Orbitrap analyzer, the initial trapping of the high m/z ions. Also, to ensure better capture (IdeS) digest MAbPAC RP • Enhanced Resolution and/or central electrode voltage was adjusted from −3.7 kV to −3.4 kV, Mode while the setting during detection remained column • High Mass Range Mode ( m/z ≤8000) unchanged (−5reduction kV). The S-lens RF level was allowed to be increased to a setting of 200 in HMR mode and set to that level for all experiments shown here. Subunit Analysis ©2016 Genovis AB Data Analysis: Protein Mode A 0 Trapping in HMR Mode: trapping gas pressure default setting 1 (range 1.0-1.5) 1355.5 1356.0 1356.5 1357.0 m/z For higher gas pressures, high charge states of the same protein decay lower regionfaster wherethan source charge states. This is because center-of-mass collision energy CID is applied Kce=E*m/(M/z) Q Exactive Plus: with M/z: the mass-to-charge ratio for a given charge state Standard Orbitrap mass analyzer • Enhanced Resolution Mode • High Mass Range Mode ( m/z ≤8000) MAbPAC RP column for denatured samples 20 10 D Figure 1: Operating modes for the three major BioPharma workflows: Normal Mode, Protein Mode and HMRSubunit mode Analysis Peptide Mapping Protein Mode Normal Mode 80 2 ©2016 Genovis AB and/or reduction 90 (fixed on setting 1) Figure 2.N A) Schematic of the Q Exactive Plus/HF mass spectrometers and differences 70 x 5.1 60 in the trapping path in the three different operating modes available: B) Normal Mode, Trapping in Protein Mode: 50 C) Protein Mode and D) HMR Mode. E) Illustration of improvement in signal intensity C trapping gas pressure default setting 0.2 40 for +17 charge state of a mAb light chain and Normal Mode. (range 0.2-1) comparing Protein Mode 30 Relative Abundance conditions in HMR mode, subunit analysis (reduced mAb and or IdeS digested mAb) in protein mode The of intact in proteins under conditions andanalysis peptide mapping standard modenative (see Figure 1). is more challenging than under denaturing conditions since the buffers used don’t contain any organic solvents. Performing electrospray from aqueous buffer solutions produces larger solvent size and desolvation is less efficient. Figure 1: Operating modes for the three majordroplets BioPharma workflows: Normal Mode, Moreover, for large proteins such as intact antibodies the required mass range for analysis under Protein Mode and HMR mode native conditions requires a mass range of more than 6000 m/z due to a smaller number of accepted charges. The increase of the upper mass range on the mass spectrometer was achieved by Normal Mode implementing instrument control software changes. The analysis of molecules across the full mass Mapping range including thePeptide detection of proteins under native conditions MAbPAC RP columnrequired the use of optimized for denatured samplesof the instrument, the efficient parameter settings to ensure efficient desolvation in the front region transfer via multipoles, efficient trapping in the C-trap/HCD region and the sensitive injection and detection in the Orbitrap mass analyzer. Critical parameters are the optimization of in-source SMART digest the support of the desolvation process. Also, for the transmission fragmentation that strongly influences HMR Mode column efficiency specific voltages have been evaluatedMAbPAC and SEC optimized to ensure robust and sensitive for native samples Intact Analysis RP mass C18 column conditions, experiments performance in theAcclaim higher range when performing analyses under native native & denatured that have not been possible so far on this type of mass spectrometer. Also, the standard calibration routine previously used was modified and adapted to ensure high mass accuracy across the full mass range. With the data collected on different types of samples and presented in this study we demonstrate the successful analysis after implementation of the High Mass Range (HMR) mode, successful desolvation and optimized critical hardware Vanquish operation settings. This makes the instrument an ideal platform to UHPLC cover the three major workflows in BioPharma: intact mass analysis under denaturing and under native conditions in HMR mode,® subunit analysis (reduced mAb and or IdeS digested mAb) in protein mode Q Exactive BioPharma FabRICATOR and peptide mapping in standard mode (see Figure 1). Mode • Protein (IdeS) digest and cooled in the cell prior to mass analysis extends the life time of protein ions resulting in increased signal RESULTS intensities of isotopically resolved species (Fig. 2E). Q Exactive HF: m: mass of residual gas, nitrogen Ultra High Field Orbitrap mass E: ion energy inside the Orbitrapanalyzer resulting in: Kce is proportional to the charge state z. This explains observations of charge envelope shifts on the m/z scale when comparing data Trapping in Normal Mode: 100 acquired in different modes withtrapping different pressure regimesE in the HCD cell and C-trap region, gas pressure setting 1 90 (fixed on Protein Mode as shown in one example in Figure 6. setting 1) 80 Normal Mode N Here we have investigated and implemented the new High Mass Range (HMR) Mode that is 70 x 5.1 60 especially required for the analysis of proteins under native conditions when samples are kept Trapping in Protein Mode: 50 solvents involved at near pH. C in aqueous buffers with no organic trapping gas pressure default setting 0.2 neutral 40 (range gas 0.2-1) pressure setting is 1 but For HMR mode the default trapping 30 it can be slightly increased up N 20 as protein complexes and to 1.5 for even improved trapping of certain species such 10 heterogeneous large proteins (e.g. antibody drug conjugates). The trapping path in HMR mode 0 in HMR Mode: taking place in the 1355.5 with ion cooling HCD cell. And also, 1356.0 1356.5 mass 1357.0 D is the same as in Protein ModeTrapping trapping gas pressure default setting 1 m/z detection is enabled ranging up(range to m/z 8000 compared to m/z 6000 in the two other modes. 1.0-1.5) N The trapping gas pressure in all modes is set and saved in the tune files and since a method allows for segmentation using different tune files different pressure settings can be used within For gas pressures, states of theis same protein decay than lower onehigher LC-MS run. In contrasthigh the charge mass range setting set in the method andfaster the method editor charge This nodes is because center-of-mass collision energy allowsstates. for several with different experiment types using different mass ranges within one LC-MS run. Kce=E*m/(M/z) with M/z: the mass-to-charge ratio for a given charge state B 2 2 2 Relative Abundance INTRODUCTION which is 5x lower than in Normal Mode. Additionally, ions are transferred Data Analysis: TM BioPharma FinderTM 1.0 SP1 software. Data analysis was performed Thermoflight Scientific HCD cell and thus havewith a longer path (Fig. 2C). The combination of reduced C-trap and HCD cell gas pressures, and trapping ions in the HCD A cell prior to mass analysis extends the life time of protein ions resulting in increased signal intensities of isotopically resolved species (Fig. 2E). region where source CID isand applied Figure 2. A) Schematic of the Q Exactive Plus/HF mass spectrometers differences in the trapping path in the three different operating modes available: B) Normal Mode, Q Exactive Plus: Standard mass analyzer of improvement in signal intensity C) Protein Mode and D) HMR Mode. E) Orbitrap Illustration Q Exactive HF: Ultra High Field Orbitrap mass for +17 charge state of a mAb light chain comparing Protein Mode and Normal Mode. m/z analyzer E 100 90 Trapping in Protein Mode: trapping gas pressure default setting 0.2 (range 0.2-1) N2 40 30 20 10 0 Trapping trapping gas pressure default setting 1 (range 1.0-1.5) N2 region where source CID is applied 50 Q Exactive Plus: Standard Orbitrap mass analyzer Q Exactive HF: High Field Orbitrap mass inUltra HMR Mode: analyzer D x 5.1 60 1355.5 1356.0 1356.5 1357.0 m/z Trapping in Normal Mode: trapping gas pressure setting 1 E 100 90 For higher gas pressures, high charge states decay faster thanProtein lower (fixed on setting 1) of the same protein Mode 80 Normal Mode N charge states. This is because center-of-mass collision energy 70 Kce=E*m/(M/z) x 5.1 60 Trapping in Protein Mode: 50 M/z: the mass-to-charge ratio for a given charge state Cwith Relative Abundance 2 trapping gas pressure default setting 0.2 m: mass of residual gas, nitrogen (range 0.2-1) E: ion energy inside the Orbitrap N2 5070.9845 0.89 6048.98036 6048.9855 -0.85 5070.98903 5070.9845 0.89 6048.98036 6048.9855 -0.85 Figure 5. Trastuzumab analyzed under native (A) and denaturing (B) conditions 7026.97170 7026.9619 1.39 resulting in highly similar deconvolution results (C). 60 Da For calibration of the HMR mode enabling mass detection up to m/z 8000 we used ammonium hexafluorophosphate (AHFP) in direct infusion mode. Figure 3 displays the spectra 5923.3 obtained in Normal 5923.3 100 R=4359 versus HMR mode with excellent mass accuracy detected with low 5695.6 (Table 1) even for masses 80 In-source CID 100 abundance at m/z higher than 6000, also shown in the mass accuracy test in Figure 4. A 5 60displays 6169.9 Figure the spectra obtained for intact Trastuzumab analyzed in native and denaturing z=+25 native zoomand pH conditions (aqueous, conditions40 resulting in different charge envelopes 5484.6 inherent to the solvent near neutral pH for native; organic solvent content and low pH for denaturing). Both types of spectra 6438.0 20 result in nearly identical results after spectra deconvolution (Fig. 5C). 70 Relative Abundance C B Protein Mode Normal Mode 80 0 3000 3085.6 4000 5000 m/z 6000 5900 7000 5950 m/z 3085.6 1005. Trastuzumab analyzed under native (A) and denaturing (B) conditions Figure R=3462 2962.2 3291.2 In-source CIDresults 80 80 in2904.1 resulting highly similar deconvolution (C). 60 Da 30 20 resulting in: Kce is proportional to the charge state z. 10 0 This explains observations of charge envelope shifts on the m/z scale when comparing data Trapping in HMR Mode: 1355.5 1356.0 1356.5 1357.0 Dacquired in different modes with different trapping gaspressure pressure default setting 1in the HCD cell and C-trap regimes region, m/z (range 1.0-1.5) as shown in one example in Figure 6. N Here we have investigated and implemented the new High Mass Range (HMR) Mode that is especially required for the analysis of proteins under native conditions are kept For higher gas pressures, high charge states of the same protein when decaysamples faster than lower incharge aqueous buffers with no organic solvents involved at near neutral pH. states. This is because center-of-mass collision energy For HMR mode the default trapping gas pressure setting is 1 but it can be slightly increased up ce=E*m/(M/z) to 1.5 for even improved trapping of K certain species such as protein complexes and with M/z: the mass-to-charge ratio for a given charge state heterogeneous large proteins (e.g. antibody drug conjugates). The trapping path in HMR mode m: mass of residual gas, nitrogen is the same asE:inion Protein Mode ion cooling taking place in the HCD cell. And also, mass energy inside thewith Orbitrap detection ranging up to m/z 8000 compared resultingisin:enabled Kce is proportional to the charge state to z. m/z 6000 in the two other modes. The trapping gas pressure in all modes is set andshifts saved them/z tune files when and since a method This explains observations of charge envelope oninthe scale comparing data allows for segmentation usingwith different tunepressure files different pressure beC-trap used within acquired in different modes different regimes in the settings HCD cellcan and region, one LC-MS run. In contrast the mass range setting is set in the method and the method editor as shown in one example in Figure 6. allows nodes with and different experiment different mass(HMR) rangesMode oneis Here for we several have investigated implemented thetypes new using High Mass Range 1 LC-MS 2withinthat run. required for the analysis of proteins under native conditions when samples are kept especially B A 2 C 3366.0 3444.2 3526.1 3612.1 In-source 3797.3 2848.3 60 40 100 40 2080 060 Relative Abundance N2 40 native 3000 CID 100 0 100 100 80 80 60 60 40 40 20 z=+48 zoom denatured 5923.3 5923.3 R=4359 5695.6 z=+25 6169.9 4000 5000 m/z 5484.6 20 Relative Intensity Trapping in Normal Mode: trapping gas pressure setting 1 (fixed on setting 1) A Relative Abundance B 5070.98903 7026.97170 7026.9619 Table 1. Theoretical and 1.39 Figure 4. HMR Mode spectral mass accuracy test using measured masses (Figure 3) of For calibration of the HMR mode The enabling mass test detection up to m/z 8000 we used ammonium ammonium hexafluorophosphate. displayed ammonium hexafluorophosphate hexafluorophosphate (AHFP) in direct infusion mode. Figure 3 displays the spectra obtained in Normal results in rms = 0.5 ppm. for calibration of HMR mode versus HMR mode with excellent mass accuracy (Table 1) even for masses detected with low experimental m/z ∆ Mass (ppm) abundance at m/z higher than 6000, also shown in the mass accuracytheoretical test in m/z Figure 4. 670.02805 670.0280 0.07 Figure 5 displays the spectra obtained for intact Trastuzumab analyzed in native and denaturing 1159.02371 1159.0237 0.01 conditions resulting in different charge envelopes inherent to the solvent and pH conditions (aqueous, 2137.01504 2137.0155 -0.21 near neutral pH for native; organic solvent content and low pH for denaturing). Both types of spectra 3604.00204 3604.0029 -0.24 result in nearly identical results after spectra deconvolution (Fig. 5C). 4418.99481 4418.9929 0.43 6000 7000 Deconvolution 3000 4000 5000 4.7ppm m/z 3085.6 148055.503 2962.2 3291.2 In-source CID 2904.1 3366.0 2848.3 3444.2 3526.1 147848.203 147909.303 3612.1 3797.3 147850.188 147908.488 native 3080 6438.0 6000 80 zoom m/z 3100 30 Da 5900 7000 5.0 ppm 148217.603 5950 m/z 3085.6 R=3462 z=+48 1.2 ppm zoom denatured eins in Native Conditions on A Hybrid Quadrupole-Orbitrap B 020 -20 0 -40 -60 -80 -100 3000 4000 denatured 148380.303 148542.203 148540.188 5000 m/z 6000 7000 148218.688 148379.888 3080 4.1 ppm m/z 3100 -2 ppm 3 30 Da Deconvolution c, Im Steingrund 4-6, Dreieich, Germany, Hanna-Kunath-Str. 14, Bremen, Germany, 355 River Oaks Par in aqueous buffers with no organic solvents involved at near neutral pH. For HMR mode the default trapping gas pressure setting is 1 but it can be slightly increased up to 1.5 for even improved trapping of certain species such as protein complexes and heterogeneous large proteins (e.g. antibody drug conjugates). Thedirect trapping path inexperiment HMR mode Figure 3. Full MS spectra of ammonium hexafluorophosphate infusion the same asHMR in Protein Mode with ion cooling taking place in the HCD cell. And also, mass inisNormal and modes. x10 in the two other modes. detection is enabled ranging up to m/z 8000 compared to m/z 6000 The trapping gas pressure in all modes is set and saved in the tune files and since a method Normal Mode 100 x5 within allows for segmentation using different tune files different pressure settings can be used and the method editor 1 80one LC-MS run. In contrast the mass range setting is set in the method 2 6048.9855 100 60allows for several nodes with different experiment types using different mass ranges within one -0.85ppm 80 40LC-MS run. 60 6863.9281 148056.888 -2.9 ppm 4.7ppm 148055.503 100 C 80 5.0 ppm native 148217.603 Relative Abundance Relative Intensity eins in Native Conditions on A Hybrid Quadrupole-Orbitra 60 1.2 ppm Figure 6. 40 Effect of in-source CID supporting desolvation/ declustering in Normal vs. HMR 148380.303 mode 20 147909.303 148542.203 0 -20 147848.203 147850.188 5923.4 100 147908.488 80 -40 Normal Mode -60 In-source CID 100eV denatured -80 single scan (10 uscans) -100 5923.4 NL: 2.06E7 148540.188 NL: 2.06E7 5695.4 zoom 60 148379.888 4.1 ppm 0.24ppm 80 Relative Abundance 60 0.07 ppm 0.01 ppm 100 670.0280 1159.0237 40 80 4418.9929 3604.0029 Normal Mode -0.85ppm 5070.9845 2626.0123 x5 6048.9855 Relative Abundance 100 20 60 0 40 6374.9744 1000 2000 3000 20 0 4000 m/z 5000 6000 0.21ppm 100 Figure 4. HMR Mode spectralHMR massMode accuracy test using 0.43ppm 4418.9929 ammonium hexafluorophosphate. The displayed test 80 0.24ppm 3604.0029 ppm results inppm rms0.01 = 0.5 ppm. 60 0.07 0.89ppm 670.0280 1159.0237 5070.9845 2626.0123 60 40 6048.9855 6863.9281 -0.85ppm 6863.9281 8000 7000 1.39ppm 7026.9619 20 -0.85ppm 6048.9855 theoretical m/z experimental m/z ∆ Mass (ppm) 670.02805 670.0280 6374.9744 1159.023716863.9281 1159.0237 20 0 1000 2000 3000 4000 m/z 7678.9320 Table06000 1. Theoretical and 7500 6500 7000 measured masses m/z (Figure 3) of ammonium hexafluorophosphate for calibration of HMR mode 2137.0155 40 80 5000 6000 0.07 0.01 2137.01504 7000 3604.00204 2137.0155 8000 3604.0029 -0.21 -0.24 4418.99481 4418.9929 0.43 5070.98903 5070.9845 0.89 Table 1. Theoretical 6048.98036 6048.9855 and -0.85 Figure 4. HMR Mode spectral mass accuracy test using measured masses 3) of 7026.97170 7026.9619(Figure 1.39 ammonium hexafluorophosphate. The displayed test ammonium hexafluorophosphate For calibration HMR mode enabling mass detection up to 8000 we ammonium results in rms =of0.5the ppm. form/z calibration of used HMR mode hexafluorophosphate (AHFP) in direct infusion mode. Figure 3 displays the spectra obtained in Normal theoretical m/z experimental m/z ∆ Mass (ppm) versus HMR mode with excellent mass accuracy (Table 1) even 670.02805 for masses670.0280 detected with low 0.07 abundance at m/z higher than 6000, also shown in the mass accuracy test in Figure 4. 1159.02371 1159.0237 0.01 Figure 5 displays the spectra obtained for intact Trastuzumab analyzed and denaturing 2137.01504 in native 2137.0155 -0.21 3604.00204 3604.0029 -0.24 conditions resulting in different charge envelopes inherent to the solvent and pH conditions (aqueous, 4418.99481 4418.9929 0.43 near neutral pH for native; organic solvent content and low pH for denaturing). Both types of spectra 5070.9845 0.89 result in nearly identical results after spectra deconvolution (Fig. 5C). 5070.98903 6048.98036 6048.9855 -0.85 7026.97170 7026.9619 1.39 Relative Abundance Figure 5. Trastuzumab analyzed under native (A) and denaturing (B) conditions For calibration of the HMR mode enabling mass detection up to m/z 8000 we used ammonium resulting in highly similar deconvolution results (C). 60 Da hexafluorophosphate (AHFP) in direct infusion mode. Figure 3 displays the spectra obtained in Normal versus HMR mode with excellent mass accuracy (Table 1) even for masses detected with low 5923.3 5923.3 100 at m/z higher than 6000, also shown in the abundance mass accuracy test in FigureR=4359 4. 5695.6 Figure 5 80displays the spectra obtained In-source CID for 100intact Trastuzumab analyzed in native and denaturing A 60resulting conditions in different charge envelopes inherent 6169.9to the solvent and pH conditions z=+25(aqueous, native near neutral pH for native; organic solvent content and low pH for denaturing). Both types of spectra zoom 5484.6 40 result in nearly identical results after spectra deconvolution (Fig. 5C). 6438.0 20 0 5900conditions 5950 Figure 5. Trastuzumab analyzed under and denaturing (B) 3000 4000 5000native (A) 6000 7000 m/z m/z 3085.6 3085.6 resulting similar deconvolution results (C). 100 in highly 60 Da R=3462 2962.2 3291.2 In-source CID 80 80 2904.1 100 60 2848.3 3366.0 3444.2 5923.3 zoom 5695.6 denatured z=+48 5923.3 R=4359 80 3526.1 In-source CID 100 40 Enabling Mass Spectrometric Analysis of Intact Proteins in Native Conditions on A 2B 3612.1 A nce 60 20 40 0 native 3000 4000 z=+25 6169.9 3797.3 5000 5484.6 zoom 6000 7000 3080 3100 148056.888 20 -2.9 ppm 0 100 6170.3 5923.3 NL: 9.83E7 NL: 1.33E8 in-source80CID x10 0.89ppm Trapping gas pressure setting of 1 for both modes Peaks in this box at m/z >6000 can not be detected in Normal Mode on a regular instrument and are shown here for demonstration purposes only to display the full charge envelope. Figure 6. Effect of supporting desolvation/ declustering in Normal vs. HMR 5923.3 6177.1 HMR Mode mode 60 zoom 5923.4 5923.4 In-source CID 200eV 100 40 NL: 2.06E7 NL: 2.06E7 5695.4 single scan (10 uscans) 80 20 Normal Mode zoom 60 In-source CID 100eV 0 6000 7000 5900 5950 6000 40 5000 m/z m/z single scan (10 uscans) Figures 6 and 7 depict the20 influence and importance of source fragmentation supporting Trapping gasand pressure desolvation declustering required to obtain the correct pattern of glycoforms. The data in these 0 6170.3 5923.3 100 setting of 1 for bothacquired modes in nanospray two figures were mode, requiring significantly higher in-source CID settings NL: 1.33E8 NL: 9.83E7 for native analyses. Due to neither gases available in the nanospray source nor a heated probe, the 80 5923.3 HMR Modeprocess needs to be desolvation supported with 6177.1 source fragmentation taking place inside the mass 60 zoom In-source CID 200eV step implemented between the S-lens and the Q00 (Figure 2A). spectrometer, a potential 40 single scan (10 uscans) Peaks in this box at m/z >6000 can not be detected in Normal Mode on a regular instrument and are shown here for demonstration purposes only to display the full charge envelope. 20 Figure 7. Intact Trastuzumab analysis under native conditions with increasing in-source CID settings highlighting the 0importance of this parameter on desolvation/ declustering. 5900 5950 6000 5000 6000 7000 m/z m/z In-source 6187.1 fragmentation supporting 6187.1 Figures 6 and1007 depict the influence and importance of source 100 NL: 1.58E6 CID (eV) 6172.7 desolvation and declustering required to obtain the correct pattern of glycoforms. The data in these 60 50 were60acquired in nanospray mode, requiring two figures significantly higher in-source CID settings 20 20 for native analyses. Due to neither gases available in the nanospray source nor a heated probe, the 6170.4 taking place inside the mass 100 100fragmentation desolvation process needs to be 6170.4 supported with source NL: 3.70E7 spectrometer, a potential step implemented between the60S-lens and the Q00 (Figure 2A). 100 60 20 20 Figure 7. Intact under native with increasing in-source 6170.4 6170.4 100Trastuzumab analysis 100conditions NL: 1.06E8 CID settings highlighting the importance of this parameter on desolvation/ declustering. 150 60 60 In-source 6187.1 6187.1 100 100 NL: 1.58E6 20 CID (eV) 20 6172.7 6170.4 6170.4 10060 10060 NL: 9.21E7 50 200 60 60 20 20 6170.4 100 20 100 60 5000 6000 6170.4 100 20 7000 m/z 6100 60 NL: 3.70E7 6200 m/z 6300 Further parameters20that were found critical in optimizing source conditions for native analysis using 20 6170.4were the capillary 6170.4 size exclusion chromatography (SEC) temperature (also referredNL: to1.06E8 as transfer 100 100 tube) and the probe heater temperature (also referred to as Aux heater temperature). Figure 8 150 60 60 shows one example of two different temperature settings resulting in differences in charge state 20 20 distribution. 6170.4 100 6170.4 100 NL: 9.21E7 Figure 8. Intact Trastuzumab analysis under native conditions with different probe heater and 200 60 60 capillary temperatures. z=+25 z=+25 Probe heater temp./ transfer capillary temp. 20 5923.3 100 80 5000 20 5695.6 6000 m/z 7000 6169.9 100 6100 80 Relative Abundance 20 0 Figure 3. Full MS0.21ppm spectra of ammonium hexafluorophosphate direct infusion experiment 6000 6500 7000 7500 2137.0155 HMR Mode 0.43ppm 100 m/z in Normal and HMR modes. Relative Abundance 7678.9320 Relative Abundance 7026.9619 Relative Abundance 0 1.39ppm 40 Relative Abundance 20 148218.688 40 5923.3 6200 m/z 6300 60 60 zoom source Further parameters that were found critical in optimizing conditions for native analysis using 5484.6 40 40 275 ° C / 275 ° Cchromatography size exclusion (SEC) were the capillary temperature (also referred to as transfer 6438.0 20 20 z=+22 tube) and the probe heater temperature (also 6730.8 referred to as Aux heater temperature). Figure 8 0 0 5923.3 shows one example100of two different settings 100 resulting in differences in charge state 5923.3 temperature 6169.9 Hybrid Quadrupole-Orbitrap Mass Spectrometer distribution. 80 80 Relative Abundance Relative Abundance 3 355 River Oaks Pa c, Im Steingrund 4-6, Dreieich, Germany, Hanna-Kunath-Str. 14, Bremen, Germany, -2 ppm 6438.0 60 60 zoom ° 175 ° C8./ 175 C Trastuzumab 5695.6 Figure Intact analysis under native conditions with different probe heater and 40 40 6730.8 capillary temperatures. 6170.4 6170.4 Re 100 100 NL: 1.06E8 tube) and the probe heater temperature (also referred to as Aux heater temperature). Figure 8 shows of two different temperature settings resulting in differences in charge state 150 one example 60 60 distribution. 20 20 6170.4 Figure 8. Trastuzumab analysis under capillary temperatures. 200 z=+25 60 Probe heater temp./ 20 transfer capillary temp. 5000 6000 60 z=+25 7000 6169.9 m/z 6100 zoom 80 6200 m/z 60 6300 Relative Abundance Further parameters that were found5484.6 critical in optimizing source conditions for native analysis using 40 40 275 C / 275 ° C ° 6438.0 size exclusion chromatography (SEC) were the capillary temperature (also referred to as transfer 20 20 z=+22 6730.8 tube) and the probe heater0 temperature (also referred to as Aux0 heater temperature). Figure 8 5923.3 5923.3 6169.9 100 100 shows one example of two different temperature settings resulting in differences in charge state 80 80 6438.0 distribution. 60 60 zoom 175 ° C / 175 ° C 5695.6 40 40 with different probe heater and Figure 8. Intact Trastuzumab analysis under native conditions 6730.8 20 capillary temperatures. 20 5484.6 7051.3 z=+20 z=+25 z=+25 7403.8 5923.3 0 100 5000 5695.6 80 6000 6169.9 60 m/z zoom 5923.3 0 100 8000 80 7000 Relative Abundance Probe heater temp./ transfer capillary temp. 5900 5920 5940 5960 m/z 60 Relative Abundance 5484.6 40 40 Figure 9. Intact SEC-MS analysis under native conditions with different resolution 275 C / 275 C Trastuzumab ° ° 6438.0 20 20 z=+22 settings. 6730.8 Base Peak Chromatogram 0 0 100 100 90 5923.3 6169.9 100 buffer salts 100 50 5923.3 R=17.5k ap Mass Spectrometer 0 80 0 100 60 Relative Abundance 175 ° C / 175 ° C 50 0 100 50 0 100 40 20 0 2 4 Time (min) 6438.08 6 5695.6 R=17.5k 80 80 zoom 7051.3 5000 6000 R=70k 60 40 50 20 40 0 30 z=+20zoom/ 7403.8 7000 R=35k 70 60 6730.8 5484.6 R=35k 10 overlay 8000 20 m/z 5900 5920 R=70k 5940 5960 m/z arkway, San Jose, CA, USA Figure 9. Intact Trastuzumab SEC-MS analysis under native conditions with different resolution 50 0 2000 settings. 3000 4000 5000 m/z 6000 10 7000 0 8000 Base Peak Chromatogram 100 0 0 2 4 6 8 5920 5930 m/z 100 buffer salts 50 5910 10 5940 90 R=17.5k 80 R=35k characterization. Mass Relative Abundance Relative Abundance A 100 New Lot 60 80 REFERENCES • This new operating mode extends the instrument’s capabilities to cover all three major workflows for BioPharma characterization. [1] Rosati S, van den Bremer ET, Schuurman J, Parren PW, Kamerling JP, Heck AJ. In-depth • Desolvation/declustering conditions, transfer glycosylation and trapping profiles have been optimized to allow for qualitative and quantitative analysis ions of composite and other micro-heterogeneity improved in HMR modeby forhigh-resolution resolution settings asmass high as 70k. on intactsensitivity monoclonal antibodies native spectrometry using a modified Orbitrap. mAbs. 2013;5(6):917-924. doi:10.4161/mabs.26282. • Critical parameters for online LC-MS analyses native conditions capillary and probe for mass [2] Fenn JB, Mann M, Meng CK, Wong under SF, Whitehouse CM. are Electrospray ionization heater temperatures as biomolecules. key factors in supporting desolvation/declustering. spectrometry of large Science. 1989;246(4926):64-71. [3] Fenn JB, Electrospray wings for molecular elephants (Nobel Lecture). Angew. Chem. Int. Ed. 2003. 42:3871–3894. REFERENCES TRADEMARKS/LICENSING [1] Rosati S, van den Bremer ET, Schuurman J, Parren PW, Kamerling JP, Heck AJ. In-depth qualitative quantitative composite profiles and other © 2016and Thermo Fisher analysis ScientificofInc. All rightsglycosylation reserved. FabRICATOR® is a micro-heterogeneity trademark of Genovis AB. on intact monoclonal antibodies high-resolution native mass spectrometry using a modified All other trademarks are theby property of Thermo Fisher Scientific and its subsidiaries. This information Orbitrap. 2013;5(6):917-924. is notmAbs. intended to encourage usedoi:10.4161/mabs.26282. of these products in any manner that might infringe the intellectual [2] Fenn JB,rights Mannof M, Meng CK, Wong SF, Whitehouse CM. Electrospray ionization for mass property others. spectrometry of large biomolecules. Science. 1989;246(4926):64-71. [3] Fenn JB, Electrospray wings for molecular elephants (Nobel Lecture). Angew. Chem. Int. Ed. 2003. 42:3871–3894. TRADEMARKS/LICENSING © 2016 Thermo Fisher Scientific Inc. All rights reserved. FabRICATOR® is a trademark of Genovis AB. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. zoom 60 20 0 100 80 60 Old Lot 40 R=35k 40 20 0 R=35k 40 20 0 100 80 20 5500 m/z 6000 0 6500 5860 5880 5900 5920 m/z 5940 5960 B Relative Abundance 2xA2G0F 100 80 60 40 20 0 -20 -40 -60 -80 -100 148055.632 New Lot 1xA2G0F/1xA2G1F 148218.219 2xA2G1F 148379.910 1xA2G1F/1xA2G2F 148540.955 1xA2G0/1xA2G0F 147909.041 147599.568 148705.783 147762.623 147909.046 148543.700 148056.349 Mass 148218.427 148381.229 Old Lot CONCLUSIONS • We have successfully implemented the High Mass Range Mode on the Q Exactive Plus and Q Exactive HF mass spectrometers allowing for mass detection up to m/z 8000. • This new operating mode extends the instrument’s capabilities to cover all three major workflows for BioPharma characterization. • Desolvation/declustering conditions, ions transfer and trapping have been optimized to allow for improved sensitivity in HMR mode for resolution settings as high as 70k. • Critical parameters for online LC-MS analyses under native conditions are capillary and probe heater temperatures as key factors in supporting desolvation/declustering. REFERENCES www.thermofisher.com [1] Rosati S, van den Bremer ET, Schuurman J, Parren PW, Kamerling JP, Heck AJ. In-depth ©2016 Thermo Fisher Scientificanalysis Inc. All rights reserved. FabRICATOR is a trademark Genovis AB. micro-heterogeneity qualitative and quantitative of composite glycosylation profilesof and other other monoclonal trademarks are antibodies the property ofbyThermo Fisher Scientific and its subsidiaries. This information onAllintact high-resolution native mass spectrometry using isa presented modified as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that Orbitrap. mAbs. 2013;5(6):917-924. doi:10.4161/mabs.26282. might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available [2] Fenn JB, Mann M, Meng CK, Wong SF, Whitehouse CM. Electrospray ionization for mass in all countries. Please consult your local sales representative for details. spectrometry of large biomolecules. Science. 1989;246(4926):64-71. [3] Fenn JB, Electrospray wings for molecular elephants (Nobel Lecture). Angew. Chem. Int. Ed. 2003. Africa +43 1 333 50 34 0 Denmark +45 70 23 62 60 Japan +81 6 6885 1213 Russia/CIS +43 1 333 50 34 0 42:3871–3894. Australia +61 3 9757 4300 Europe-Other +43 1 333 50 34 0 Korea +82 2 3420 8600 Austria +43 810 282 206 Finland +358 10 3292 200 Latin America +1 561 688 8700 TRADEMARKS/LICENSING Belgium +32 53 73 42 41 France +33 1 60 92 48 00 Middle East +43 1 333 50 34 0 Brazil +55 11 2730 3006 Germany +49 6103 408 1014 Netherlands +31 76 579 55 55 © Canada 2016 Thermo Fisher reserved. of Genovis AB. +1 800 530Scientific 8447 Inc. All rights India +91 22FabRICATOR® 6742 9494 is a trademark New Zealand +64 9 980 6700 AllChina other trademarks are (free the property of Thermo Scientific and its subsidiaries. This information call domestic) 800 810 5118 Italy Fisher +39 02 950 591 Norway +46 8 556 468 00 is not intended to encourage use of these products in any manner that might infringe the intellectual 400 650 5118 property rights of others. Old Lot CONCLUSIONS • Critical parameters for online LC-MS analyses under native conditions are capillary and probe 60 40 148381.229 • Desolvation/declustering conditions, ions transfer and trapping have been optimized to allow for improved sensitivity in HMR mode for resolution settings as high as 70k. Figure 10. Comparison of two different lots of Trastuzumab showing significant variation in the pattern of the glycoforms, a known issue in production of biopharmaceuticals. 80 148218.427 5950 Time (min) 100 Figure 9 shows the SEC-LC/MS analysis of intact Trastuzumab under native conditions acquired 70 R=70k 50 R=17.5k on the Q Exactive Plus in HMR mode at different60 resolution settings of 17.5k, 35k and 70k. 0 50 100 Increasing the resolution setting in this casezoom/ is not40 aiming at achieving isotopic resolution but R=35k 50 overlay allowing to resolve possible sodium and potassium 30 adducts [1] arising e.g. from salts and/or 0 100 20 formulation buffer and thus significantly improving mass accuracy after deconvolution. R=70k 50 Figure 10 displays glycoform pattern variability from10 two different lots of Trastuzumab analyzed 0 0 2000 3000 4000 5000 6000 7000 8000 5910 5920 and 5930 reported 5940 5950 with SEC-LC/MS, which ism/z not desired but has been observed previously. The m/z possibility for such a variation is inherent to the production process of biopharmaceuticals and shows the capabilities of mass spectrometry to easily pick up these variations since the analyses of glycoform patterns on the intact antibody level are very reproducible and reliable, considering the use of consistent optimized parameter settings for all experiments in a study. 100 148056.349 heater temperatures as key factors in supporting desolvation/declustering. • We have successfully implemented the High Mass Range Mode on the Q Exactive Plus and Q Exactive HF mass spectrometers allowing for mass detection up to m/z 8000. 5923.3 100 20 5695.6 80 conditions with different probe heater and NL: 9.21E7 60 5923.3 100 6170.4 100 native Relative Abundance 100 Intact -100BioPharma Singapore +65 6289 1190 Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00 Taiwan +886 2 8751 6655 UK/Ireland +44 1442 233555 USA +1 800 532 4752 PN64804-EN 08/16S