Download Agilent 6230 Accurate-Mass TOF LC/MS

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts
no text concepts found
Transcript
Agilent 6230 Accurate-Mass
TOF LC/MS
Exceptional MS performance without compromises
Product Note
The Agilent 6230 Accurate-Mass Time-ofFlight (TOF) LC/MS system features Agilent
Jet Stream Thermal Focusing technology for
significantly improved sensitivity, as well as
enhanced MassHunter Workstation software for
superior data mining and analysis capabilities.
These new features, coupled with Agilent’s
True High-Definition TOF (True Hi-Def TOF)
technology, enable the 6230 Accurate-Mass TOF
to deliver exceptional sensitivity, excellent mass
accuracy, fast data acquisition, and streamlined
qualitative and quantitative analyses to meet
your most challenging research needs.
Agilent 6230 Accurate-Mass TOF LC/MS integrates
Agilent True Hi-Def TOF technology, Agilent Jet Stream
Thermal Focusing technology, and MassHunter Workstation
software for sensitive, accurate-mass MS analyses.
Product Note
True Hi-Def TOF Technology for
Exceptional Mass Accuracy,
Sensitivity and Speed
Sensitive, Accurate-Mass
MS Analyses
The Agilent 6230 Accurate-Mass
TOF LC/MS system is designed to
provide superior data quality and
advanced analytical capabilities for
profiling, identifying, characterizing,
and quantifying low molecular-weight
compounds and biomolecules with
confidence. Integrating three core
Agilent technical innovations—True HiDef TOF technology, Agilent Jet Stream
Thermal Focusing technology and
MassHunter Workstation software—
the 6230 TOF platform is ideally suited
for accurate-mass analyses of the
complex samples encountered in
today’s most demanding applications.
Agilent True Hi-Def TOF technology, a
key feature of Agilent’s 6200 Series TOF
LC/MS instruments, delivers outstanding TOF performance characteristics
without any performance compromises.
• Sub 1-ppm mass accuracy improves
confidence and reduces false
positives
• > 20,000 mass resolution
distinguishes target analytes from
interferences
• Up to 5 orders of in-spectrum
dynamic range improves detection of
trace-level targets in the presence of
high abundant compounds
LC sample inlet
Nebulizing gas
Super-heated
sheath gas
MS inlet
Nozzle voltage
Heated
drying gas
Collimated
thermal containment
zone results in enhanced
MS sensitivity
Resistive
sampling
capillary
2
• Low picogram sensitivity enables
identification of very low-abundance
compounds
• Fast data acquisition rates of up to
40 spectra/second ensure maximum
compatibility with fast LC and highthroughput methods
• Wide mass range to 20,000 m/z
Agilent Jet Stream Technology
Enhances LC/MS Sensitivity 5-fold
Agilent Jet Stream Thermal Focusing
technology significantly increases LC/
MS sensitivity by improving the spatial
focusing of electrospray droplets.
The subsequent enhancements in ion
density and desolvation result in higher
MS signal intensities and improved
signal-to-noise ratios. On average, a
5-fold improvement in MS sensitivity
is realized by using Agilent Jet Stream
technology at optimal LC flow rates.
Easy to use and tune, Agilent Jet
Stream technology is available on the
6230 Accurate-Mass TOF LC/MS
to provide maximum sensitivity for multiple applications, including the analysis
of drug candidates and trace levels
of food contaminants, metabolites or
biomarkers.
Agilent Jet Stream Thermal Focusing technology
utilizes super-heated nitrogen sheath gas to
confine the electrospray. Ion density and desolvation are improved, resulting in higher MS signal
intensities and reduced noise.
Agilent 6230 Accurate-Mass TOF LC/MS
2
x10 +ESI TIC Scan Frag=120.0V Pest MX_20 Lettuce_01 ppb in 1_3 water.d
1 1
1
0.9
0.8
MassHunter Workstation Software for
Powerful Data Mining and Analysis
Capabilities
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
x105
2.2
2.1
2
1.9
1.8
1.7
1.6
1.5
1.4
1.3
1.2
1.1
1
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
+ESI Scan (2.501-2.660 min, 13 scans) Frag=120.0V Pest MX_20 Lettuce_01ppb in 1_3 water.d Subtract
390.1763
(M+H)+
202.0431
Resolving power=15,000
Mass Accuracy=1.3 ppm
+
201.1091
204.1217
193.0859
300.2019
234.5509
100
150
200
250
722.4532
300
350
400
450
500
550
600
650
700
750
800
850
900
950
1000
Counts vs. Mass-to-Charge (m/z)
High sensitivity screening (< 1 ppb) of pesticides in lettuce. Thiabendazole and its isotopes are detected at a 0.1 ppb
concentration level in the above chromatogram and extracted mass spectrum.
x10 4
+ESI Scan (11.590-11.755 min, 11 scans) Frag=225.0V CA-2p.d S
x10 6
Agilent’s MassHunter Workstation
Software facilitates faster and easier
processing of information-rich data
generated by the 6230 Accurate-Mass
TOF LC/MS. Compound-centric data
mining and navigation capabilities
enable efficient analyses of complex
MS data. A sophisticated molecular
feature extraction algorithm automatically retrieves all spectral and
chromatographic information for each
component in a sample mixture, including those in overlapping and co-eluting
peaks, thereby saving hours of analysis
time. MassHunter’s integrated mass
profiling tools streamline differential
and statistical analyses between
sample sets. The software can be used
to perform additional processing steps
such as molecular formula generation,
database or library search, deconvolution, isotope pattern matching or
charge-state determination for
confident compound identification.
+ESI Scan (11.590-11.755 min, 11 scans) Frag=225.0V CA-2p.d S
…
6.5
5
6
5.5
5
A
B
4.5
4
4.5
3.5
4
3.5
3
2.5
3
2.5
2
2
1.5
1.5
1
0.5
1
0.5
0
High performance mass accuracy for bovine carbonic anhydrase confirms protein molecular weight and high resolution (data collected in
4 GHz mode) detects protein modifications. Mass spectrum (A) and the maximum entropy deconvolution (B) of bovine carbonic anhydrase.
The calculated average mass of carbonic anhydrase is 29024.85 Da; the mass error is 9 ppm.
3
Designed for the Most Demanding
Applications
Buy online:
www.agilent.com/chem/store
The Agilent 6230 Accurate-Mass TOF
LC/MS system delivers exceptional
MS analyses for applications that
demand the highest accurate-mass
measurements, without compromising
sensitivity and speed. Integration of
Agilent Jet Stream Thermal Focusing
technology ensures the highest
level of sensitivity for your most
challenging samples. Powerful new
data mining tools in MassHunter
Workstation software facilitate
profiling, characterization, identification
and quantification of compounds in
complex mixtures. These performance
characteristics enable the 6230
TOF system to support demanding
applications such as proteomics,
metabolomics, impurity testing, product
degradation studies, forensics, food
safety, and environmental analyses.
Find an Agilent customer center in
your country:
www.agilent.com/chem/contactus
U.S. and Canada
1-800-227-9770
[email protected]
Asia Pacific
[email protected]
Europe
[email protected]
About Agilent Technologies
Agilent Technologies is a leading
supplier of life science research
systems that enable scientists
to understand complex biological
processes, determine disease
mechanisms, and speed drug
discovery. Engineered for sensitivity, reproducibility, and workflow
productivity, Agilent’s life science
solutions include instrumentation,
microfluidics, software, microarrays, consumables, and services
for genomics, proteomics, and
metabolomics applications.
This item is intended for Research Use Only. Not
for use in diagnostic procedures. Information,
descriptions, and specifications in this publication
are subject to change without notice.
Agilent Technologies shall not be liable for errors
contained herein or for incidental or consequential damages in connection with the furnishing,
performance or use of this material.
© Agilent Technologies, Inc. 2008
Printed in the U.S.A. December 17, 2008
5990-3422EN