* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Download Rosemount Recommended Calibration Practices for Magnetic
Survey
Document related concepts
Electromagnetism wikipedia , lookup
Hall effect wikipedia , lookup
Neutron magnetic moment wikipedia , lookup
Faraday paradox wikipedia , lookup
Magnetic monopole wikipedia , lookup
Magnetic nanoparticles wikipedia , lookup
Eddy current wikipedia , lookup
Scanning SQUID microscope wikipedia , lookup
Superconductivity wikipedia , lookup
Multiferroics wikipedia , lookup
Magnetoreception wikipedia , lookup
Sound level meter wikipedia , lookup
Force between magnets wikipedia , lookup
Magnetohydrodynamics wikipedia , lookup
Magnetochemistry wikipedia , lookup
Galvanometer wikipedia , lookup
Transcript
Peter Fritsche Director Global Quality Emerson Process Management Rosemount, Inc. 12001 Technology Drive Eden Prairie, MN 55344 T 1 (952) 828-3265 [email protected] March 1, 2010 SUBJECT: Calibration and verification of Electromagnetic Flowmeters This document defines the official Rosemount “manufacturer recommended practice” to assist users in their compliance with EPA regulations governing Green House Gas measurement points. It is suggested that users retain a copy of this document in their compliance files. All Rosemount Magnetic Flowmeters are calibrated prior to shipping in our NIST traceable flow facilities to determine every meters calibration constant (sometimes called k-factor). This establishes how the relationship between the induced voltage on the electrodes and the volume of fluid passing through the pipe. This establishes the calibration factor and accuracy for the meters entire operating range. Our labs are designed to calibrate every meter at three velocities – approximately 1, 3 and 10 feet/second (approx 0.3, 1 and 3 m/s) – to determine the calibration factor, which is then used as the constant over the entire operating range. It is common for some manufacturers to determine the calibration constant by flowing meters at two velocities. Emerson calibrates at three velocities to ensure the meters calibration is linear over the entire flow range. The theory for magnetic flowmeters is based on faradays law, which states: E=C*V (Where C = Calibration constant) C = k * D * B, resulting in D E=k*D*B*V Where: E = Induced Voltage on electrodes k = Unit conversion constant D = Distance between electrodes For a given meter, the calibration helps us take into account small manufacturing variations in D (distance between electrodes) and B (Magnetic field strength delivered by the coils) from meter to meter. The theory shows that the relationship remains constant independent of velocity. Page 1 In-situ Calibration Verification Options: Smart Meter Verification Once in the field, the magnetic flowmeter can be verified using Smart Meter Verification. This internal diagnostic allows users to reconfirm the Magnetic field through key measurements of the magnetic coil inductance and resistance as well as the measuring electrode resistance. In addition, Meter Verification also verifies the calibration of the transmitter. Additional information on the Rosemount Smart Meter Verification option can be found in the attached Technical Note. Accuracy of 5% can be obtained by selecting a test criteria of less than 2.0% when configuring the Smart Meter Verification option. Traditional Meter Verification In addition, to the Smart Meter Verification option, the magnetic flowmeter sensor and transmitter can be verified using the attached Magmeter Verification Document. For new installations, Emerson recommends purchasing 8712E or 8732E transmitters with the Smart Meter Verification option (DA2 option code) and performing a baseline signature during commissioning. For existing installations, Emerson recommends that users with existing 8732E or 8712E utilize the Smart Meter Verification approach. This is available as a licensed option from the factory. For installed transmitters that are not currently licensed, a license code can be purchased by calling Rosemount Customer Central. For users with transmitters of a previous generation, we recommend that they upgrade the transmitter to an 8712E or 8732E with Smart Meter Verification. However, if this is not feasible, the traditional meter verification approach may be utilized. For continuing verification, either approach can also be conducted by one of Emerson’s Instrument and Valve Services Certified Field Service Technicians. It is Rosemount’s recommendation that meters be verified once every three years. When the test criteria is not met in either of the aforementioned approaches or when application requirements warrant, the magnetic flowmeter may be returned to a Rosemount flowlab for verification and/or recalibration through Emerson’s Instrument and Valve Services. Sincerely, Peter Fritsche Director Global Quality Emerson Process Management – Rosemount Flow 1-952-828-3265 Page 2