CP-493
Add View Modifiers to CID 4015
Date: 2005/01/12
Status: Letter Ballot
DICOM Correction Item
Correction Number
CP- 493
Log Summary: Add View Modifiers to CID 4015
Type of Modification
Name of Standard
Addition
PS 3.16 - 2004
Rationale for Correction:
The Rolled Lateral and Rolled Medial view modifiers for Mammography in CID 4015 apply to a CC-like view.
When an MLO-like view is rolled, it is either Rolled Inferior or Rolled Superior. These view modifiers are
listed in the Mammography Quality Control Manual (MQCM) and in the recently published BI-RADS® Atlas,
but were omitted from CID 4015. This proposal corrects the oversight.
Sections of documents affected
PS 3.16, Annex B, CID 4015
Correction Wording:
Modify PS 3.16, Annex B, CID 4015
CID 4015
View Modifier for Mammography
Context ID 4015
View Modifier for Mammography
Type: Non-Extensible Version: 20020904ymmdd
Coding Scheme
Designator
(0008,0102)
Code Value
(0008,0100)
Code Meaning
(0008,0104)
Applies only when
view is:
ACR BI-RADS
Equivalent
SNM3
R-102D2
Cleavage
CC
CV
SNM3
R-102D1
Axillary Tail
MLO
AT
SNM3
R-102D3
Rolled Lateral
any
...RL
SNM3
R-102D4
Rolled Medial
any
...RM
SRT
R-102CA
Rolled Inferior
any
…RI
SRT
R-102C9
Rolled Superior
any
…RS
SNM3
R-102D5
Implant Displaced
any
ID
SNM3
R-102D6
Magnification
any
M...
SNM3
R-102D7
Spot Compression
any
S
SNM3
R-102C2
Tangential
any
TAN