Survey
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
CP-493 Add View Modifiers to CID 4015 Date: 2005/01/12 Status: Letter Ballot DICOM Correction Item Correction Number CP- 493 Log Summary: Add View Modifiers to CID 4015 Type of Modification Name of Standard Addition PS 3.16 - 2004 Rationale for Correction: The Rolled Lateral and Rolled Medial view modifiers for Mammography in CID 4015 apply to a CC-like view. When an MLO-like view is rolled, it is either Rolled Inferior or Rolled Superior. These view modifiers are listed in the Mammography Quality Control Manual (MQCM) and in the recently published BI-RADS® Atlas, but were omitted from CID 4015. This proposal corrects the oversight. Sections of documents affected PS 3.16, Annex B, CID 4015 Correction Wording: Modify PS 3.16, Annex B, CID 4015 CID 4015 View Modifier for Mammography Context ID 4015 View Modifier for Mammography Type: Non-Extensible Version: 20020904ymmdd Coding Scheme Designator (0008,0102) Code Value (0008,0100) Code Meaning (0008,0104) Applies only when view is: ACR BI-RADS Equivalent SNM3 R-102D2 Cleavage CC CV SNM3 R-102D1 Axillary Tail MLO AT SNM3 R-102D3 Rolled Lateral any ...RL SNM3 R-102D4 Rolled Medial any ...RM SRT R-102CA Rolled Inferior any …RI SRT R-102C9 Rolled Superior any …RS SNM3 R-102D5 Implant Displaced any ID SNM3 R-102D6 Magnification any M... SNM3 R-102D7 Spot Compression any S SNM3 R-102C2 Tangential any TAN