EOP-P61-003-6-001-06E/a
PRODUCT SPECIFICATIONS
for MODEL EG/EL-550
In-line type OZONE MONITOR
Published on August 01, 2011
EBARA JITSUGYO CO., LTD.
MEASURING INSTRUMENT
AND MEDICAL DIVISION
MEASURING INSTRUMENT SALES DEPARTMENT
EAST JAPAN SALES BRANCH
: 3-12, Kurigi 2-Chome, Asao-ku,
Kawasaki-shi, Kanagawa, 215-0033 Japan
Tel. +81-44-981-0560 Fax. +81-44-981-0561
E-mail [email protected]
WEST JAPAN SALES BRANCH
: 2-13, Hiranocho 3-Chome, Chuo-ku,
Osaka-shi, Osaka, 541-0046 Japan
Tel. +81-6-6231-3528 Fax. +81-6-6231-2929
E-mail [email protected]
TECHNICAL ENGINEERING DEPARTMENT
3-12, Kurigi 2-Chome, Asao-ku,
Kawasaki-shi, Kanagawa, 215-0033 Japan
Tel. +81-44-981-0560 Fax. +81-44-981-0561
E-mail [email protected]
E.J. OZONE PRODUCTS
EOP-P61-003-6-001-06E/a
Introduction
The EG/EL-550 series, Ozone monitors designed on the specifications in this document, are USED for
general application only, so that you can not use them in explosion environments.
DANGER
Explosion can result in death or serious injury.
Do not USE this Ozone monitor in the explosive environments.
-
E.J. OZONE PRODUCTS
1
-
EOP-P61-003-6-001-06E/a
1. General Description
The ozone densitometer (hereafter referred as ozone monitor), EG/EL-550 series was actualized as a
small instrument by designing the detector unit and the control unit to one package.
Moreover, by only connecting a pipe on inlet for sample, the monitor can measure ozone
concentration continuously.
Still more, to indicate a light intensity for sensor and also value for span is available, so that can easily
check the monitor and perform the maintenance.
2. Measuring Principle
The monitor is a UV absorption type ozone monitor which detects and measures an absorptive
quantity of UV rays by ozone in the sample gas (water) introduced into the monitor, and measure
ozone concentration.
A low-pressure mercury lamp (emission wavelength 253.7 nm) is used for the light source.
As the light intensity absorbed by ozone existing within the optical path "T" obeys the Lambert-Beer's
Law, ozone concentration can be measured as follows.
C=
A
× Log
αT
Where; C
α
T
Io
Ix
A
=
=
=
=
=
=
Io
Ix
・・・・・
equ. 1
Ozone concentration of sample
Specific absorption coefficient
Path length of cell
Light intensity through the sample free of Ozone
Light intensity passed through sample Ozone
Constant
Mercury lamp
Sensor 1
Sample
T
Sensor 2
Figure 1: Logic Diagram
-
E.J. OZONE PRODUCTS
2
-
EOP-P61-003-6-001-06E/a
3. Specifications
3.1 Dissolved ozone monitor
Model
: EL-550
Measuring principle
: UV absorption method
Object of measurement : Dissolved ozone in water
Measuring Range
: One range to be selected from among the following.
0~10 mg / L
0~20 mg / L
0~30 mg / L
0~50 mg / L
0~100 mg / L
Note
Measuring Interval
Note
: The monitor can measure high ozone concentration, but it can not be
guaranteed that the piping and etc. are not deteriorated by ozone
reaction.
: Continuous measurement
: Before supplying sample water, confirm zero point of the monitor
Sampling Method
: Intake from water pressure itself or pressure sucked with pump
Sample flow rate
: 0.05 ~ 3.0 litter per minute
Note
: When measuring the concentration at low flow late, pipe length should
: be as short as possible so as not to decompose easily ozone.
Continuous service
Pressure
: Less than 0.1 MPa (G)
Span Drift
: Within ± 1% of full scale per month
Zero Drift
: Within ± 3% of full scale per month
Linearity
: Within ± 1 % of full scale
Repeatability
: Less than 1 % of Full Scale
Zero Adjustment
: Adjustment with zero trimmer
-
E.J. OZONE PRODUCTS
3
-
EOP-P61-003-6-001-06E/a
Display function
: Available (selection of ozone concentration, light intensity on the sensor,
or span with rotary switch)
Analog output
: Voltage output ・・・・・ 0 ~ 1 V or 0 ~ 10 V DC
Current output ・・・・・ 4 ~ 20 mA DC (isolated)
Note (1) : Allowable resistance to be connectable, is over than 10 k Ω when
voltage output.
Specify either 1 V or 10 V when ordering.
It is set before shipping.
Note (2) : When current output, a resistance externally connectable is less than 750
Ω.
Power requirement
: 90 ~ 260 V AC, 50/60 Hz
Connection method
of power supply
: Connection through a terminal block
Power consumption
: 50 VA
Dimension
: 220 W × 150 H × 105 D
Note
Piping connection
: Protruding parts are not included
: Flowell
TM
, 20 series, 1/4 inch PTFE
Operating environment : 5 ~ 45 ℃, 90 % RH (non condensation)
Enclosure material
: SUS 304
Weight
: 2.2 kg
Accessory
: Fuse ( 250 V AC, 1 A approved by UL)
Options
: Function for preventing the condensation.
(needs dry air of 0.1 liter per minute)
: Power supply for 24 V DC
Note
: The monitor should be installed in the system cabinet.
-
E.J. OZONE PRODUCTS
4
-
EOP-P61-003-6-001-06E/a
3.2 Gas-phase ozone monitor
Model
: EG-550
Measuring principle
: UV absorption method
Object of measurement : Ozone gas
Measuring Range
: One range to be selected from among the following.
0~10 mg / m3
0~20 mg / m3
0~30 mg / m3
0~50 mg / m3
0~100 mg / m3
Note
Measuring Interval
Note
: The monitor can measure high ozone concentration, but it can not be
guaranteed that the piping and etc. are not deteriorated by ozone
reaction.
: Continuous measurement
: Before supplying sample water, confirm zero point of the monitor
Sampling Method
: Intake from gas pressure itself or pressure sucked with pump
Sample flow rate
: 0.05 ~ 3.0 litter per minute
Note
Continuous service
Pressure
Note
: When measuring the concentration at low flow late, pipe length should
be as short as possible so as not to decompose easily ozone.
: Less than 0.1 MPa (G)
: But, outlet of the monitor should be free to atmosphere, when measuring
the ozone.
When the outlet is influenced with a pressure, perform to compensate
the pressure.
Span Drift
: Within ± 1% of full scale per month
Zero Drift
: Within ± 3% of full scale per month
Linearity
: Within ± 1 % of full scale
Repeatability
: Less than 1 % of full scale
Zero Adjustment
: Adjustment with zero trimmer
Display function
: Available (selection of ozone concentration, light intensity on the sensor,
-
E.J. OZONE PRODUCTS
5
-
EOP-P61-003-6-001-06E/a
or span with rotary switch)
: Voltage output ・・・・・ 0 ~ 1 V or 0 ~ 10 V DC
Current output ・・・・・ 4 ~ 20 mA DC (isolated)
Analog output
Note (1) : Allowable resistance to be connectable, is over than 10 k Ω when
voltage output.
Specify either 1 V or 10 V when ordering.
It is set before shipping.
Note (2) : When current output, a resistance externally connectable is less than 750
Ω.
Power requirement
: 90 ~ 260 V AC, 50/60 Hz
Connection method
of power supply
: Connection through a terminal block
:
Power consumption
: 50 VA
Dimension
: 220 W × 150 H × 105 D
Note
Piping connection
: Protruding parts are not included
: Flowell
TM
, 20 series, 1/4 inch PTFE
Operating environment : 5 ~ 45 ℃, 90 % RH (non condensation)
Enclosure material
: SUS 304
Weight
: 2.2 kg
Accessory
: Fuse ( 250 V AC, 1 A approved by UL)
Options
: Function for preventing the condensation.
(needs dry air of 0.1 liter per minute)
Power supply for 24 V DC
Note
: The monitor should be installed in the system cabinet.
-
E.J. OZONE PRODUCTS
6
-
EOP-P61-003-6-001-06E/a
4. Code Number
EG
ozone gas
EL
dissolved ozone
・
- 550
measurig
power
piping
analog
mounting
privation
ranges
rating
connection
output
method
of
[ mg/L ]
0 ~ 10
J
[V]
A
90 ~
※3
2
TM
Flowell
260 AC
20 series
24 DC
inch pipe
V1
0~
※2
condensation
B
wall
1
available
F
Floor
0
none
1V DC
for OD 1/4
B2
0 ~ 20
B3
0 ~ 30
D
※1
V10
joint: material
0~
10V DC
PTFE
U5
0 ~ 50
U10
0 ~ 100
F
TM
Kurabo
Final lock
for OD 1/4 inch
junction material
: PTFE + PFA
※1
Note
※ 1: Mark means an optional good
※ 2: Customer is able to change a position of the fixing metal.
※ 3: The monitor has two kind of outputs for voltages and current. Regarding voltage,
there are two outputs, select it.
<Example of order>
Specifications
1)
2)
3)
4)
5)
6)
0 ~ 50 mg per liter (dissolved ozone) ・・・
100 V AC power supply ・・・・・・・・・
pipe connection with FlowelTM ・・・・・
analog output of 0 1 V DC ・・・・・・・・
Mounting on the floor ・・・・・・・・・
prevention of condensation (none) ・・・
Code number: EL-550 U5A2・V1F0
-
E.J. OZONE PRODUCTS
7
-
U5
A
2
V1
F
0
EOP-P61-003-6-001-06E/a
5. Warranty
WARRANTY:We warrant the ozone monitor as follows.
(1)
・
(2)
・
Warranty period
12 months after customer's acceptance inspection.
Condition
The ozone monitor should be operated under the specified condition deserved in this instruction
manual.
(3) Limitation of warranty
・ Liability and obligations are limited to the supply of parts for replacements.
-
E.J. OZONE PRODUCTS
8
-
EOP-P61-003-6-001-06E/a
-
E.J. OZONE PRODUCTS
9
-
EOP-P61-003-6-001-06E/a
MEMO
-
E.J. OZONE PRODUCTS
10
-
荏 原 実 業 株 式 会 社
計測器・医療本部
計測器営業部
■東日本営業課:〒215-0033
神奈川県川崎市麻生区栗木2丁目3番12号
TEL 044-981-0560 FAX 044-981-0561
E-mail ej-ozone@ejk.co.jp
■西日本営業課:〒541-0046
大阪市中央区平野町3丁目2番13号
平野町中央ビル5階
TEL 06-6231-3528 FAX 06-6231-2929
E-mail ozon-osaka@ejk.co.jp
技
術
部
:〒215-0033
神奈川県川崎市麻生区栗木2丁目3番12号
TEL 044-981-0560 FAX 044-981-0561
E-mail ejozndsn1@ejk.co.jp
取扱店:
Related documents