Download Lab 3: Motor-Pump System Measurements and LabView Interface

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Intelligent maintenance system wikipedia , lookup

Transmission (mechanics) wikipedia , lookup

Sensors for arc welding wikipedia , lookup

Dynamometer wikipedia , lookup

Friction-plate electromagnetic couplings wikipedia , lookup

Variable-frequency drive wikipedia , lookup

Pedelec wikipedia , lookup

Transcript
Lab 3: Motor-Pump System Measurements and LabView Interface
In this lab experience you will create a Virtual Instrument (VI) using LabView.
The VI will be used to capture data for later analysis. A separate instruction
sheet explains the procedure to be followed to create the VI. This document
contains the procedure to be followed after the VI has been created and tested
for successful operation.
Safety Precautions:
 Always remember that water and electricity do not mix. Use extreme
caution when transferring the hose from one position to another. Transfer
the hose with the system powered down.
 Use the rags to dry any water that might have splashed or that might be
leaking from the pump seals or hose connections.
 Do not operate any knobs with wet hands.
Hose in Recirculation Mode
Use this configuration when testing the system and setting up the VI.
Avoid kinks on the hose
To eliminate a kink, turn the coupling as necessary.
Hose in Flow Mode
I) Torque Sensor Calibration
The torque sensor will be calibrated without the use of the VI. This is
because the scale used to measure the force does not have output
connectivity (bummer!).
0)With the power supply OFF, decrease the motor supply voltage to zero
using the large black wheel.
1)Attach the lever arm to the torque sensor’s pump side brass coupling as
shown in the picture. Place the scale underneath the nut at the end of the
lever arm. Do not change the location of the nut. The moment arm
has been carefully measured to be 5.885 inches.
Mount the arm at right angles to the shaft!
2) Turn on the scale, and have the nut barely touch the plate of the
scale, with a zero gram reading. Connect the multimeter terminals to
the brown and white wires coming from the sensor. You can also use
the terminal connections at channel ACH0 on the Elvis board. Turn on
the power supply used to provide 12V to the sensor (the Agilent
supply, not the motor supply).
3) Open a blank spreadsheet and prepare it to calculate the force in N
from the mass in grams (this is what the scale will indicate) and the
torque in N-mm corresponding to each sensor voltage reading.
4) Gradually increase the motor voltage to generate force on the scale.
You will notice that the force keeps increasing even without increasing
the voltage. For this reason, you will have to take simultaneous
readings of the scale and the sensor voltage quite fast. Take 10 to15
data pairs with ascending voltage and another 10 to15 with
descending voltage. Do not exceed 5 amps of current.
5) Return the voltage wheel to zero and turn off the motor supply.
Remove the arm.
6) Plot the sensor voltage against the true torque and extract the
sensitivity in V/N-mm. This number will be entered in the VI next.
II) Data collection
1) Enter the correct value for the torque sensor sensitivity in the VI.
2) Place the hose in flow mode and twist the coupling to eliminate the
kink.
3) Zero out the velocity and torque values using the slider controls of
your VI.
4) You have a limited time to perform the tasks described in step 5, since
water level decreases rapidly and we don’t want to run the pump on
air. Read the instructions carefully.
5) Start with a full tank. Run the VI. Turn on the power supply and set
the voltage to 15 and enter the value in the VI’s front panel. Enter the
current and press the save data button. Wait until the water level
sensor is halfway thru the stroke or so, and press the save data button
again. Stop the VI. Then turn off the motor power. Verify that a file
has been created.
6) Repeat step 3 for the following motor voltages: 14, 13, 12, 11, 10, 9.
Keep track of filenames!!
7) Return the hose to the recirculation position and clean up any water
leaks.
III) Data Analysis and Report
1) Write a very short report summarizing what has been done.
Concentrate on the following steps.
2) Summarize and present the data used to find the torque sensor
sensitivity. Plot the torque-voltage data for the ascending and
descending ranges in a single chart.
3) For the data corresponding to the flow mode, prepare the following
plots:
 Pump flow rate in liters/min vs. speed in rpm.
 Pump torque in N-mm vs. flow rate in liters/min.
 Motor electrical power vs. time and motor mechanical power vs.
time (assume the torque to be equal throughout the shaft) in
one plot.
 Motor electromechanical conversion efficiency vs. time.
 Motor torque in N-mm vs. motor current in A.
 Motor speed in rpm vs. motor voltage.
 Motor speed vs. motor torque.
Wiring Color Codes
DC Motor:
Red:
Black:
Red (shielded):
Black (shielded):
Armature +V
Armature –V
Tachometer +
Tachometer –
Level Sensor:
Brown:
Green:
White:
+5V
Sensor output 0-5V
Common ground
Torque Sensor:
Purple:
Yellow:
Brown:
White:
Excitation (+12V)
Excitation ground
Sensor + output (0-10V)
Sensor – output
Motor Power Supply:
A pair of B/W wires is connected to a parallel bank of capacitors to reduce
voltage ripple.