Survey
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
A Thesis entitled Evaluation of Phantoms Used in Image Quality Performance Testing of Dental Cone Beam Computed Tomography Systems by Haitham N Alahmad Submitted to the Graduate Faculty as partial fulfillment of the requirements for the Master of Science in Biomedical Science Degree in Medical Physics ________________________________________ E. Ishmael Parsai, Ph.D, Committee Chair ________________________________________ Kerry Krugh, Ph.D, Committee Member ________________________________________ Diana Shvydka, Ph.D, Committee Member ________________________________________ Patricia Komuniecki, Ph.D , Dean College of Graduate Studies The University of Toledo August 2015 Copyright 2015, Haitham N Alahmad This document is copyrighted material. Under copyright law, no parts of this document may be reproduced without the expressed permission of the author. An Abstract of Evaluation of Phantoms Used in Image Quality Performance Testing of Dental Cone Beam Computed Tomography Systems by Haitham N Alahmad Submitted to the Graduate Faculty as partial fulfillment of the requirements for the Master of Science in Biomedical Science Degree in Medical Physics The University of Toledo August 2015 Cone beam computed tomography (CBCT) units dedicated for dental and maxillofacial imaging have gained widespread use in recent years. So far there are no standardized testing methods to evaluate the image quality or the dose on these units. Although the vendors commonly provide an image quality phantom with the machine, the procedures, image quality parameters evaluated, and performance criteria are limited and specific to the vendors. The goal of this study is to evaluate the available phantoms as a testing tool for image quality assessment of dental CBCT units. In addition, the optimal diameter and features of a QA phantom for dental CBCT testing was assessed. Two commercially available phantoms were evaluated to assess the adequacy of each for use in the standardized testing procedures. These included ACR CT phantom (Gammex 464) and CATPHAN (Phantom Laboratory). Additionally, a prototype dental CT phantom (CIRS Inc.) was evaluated. Scans were made on three different machines; iCAT Classic (Imaging Sciences International), ILUMA (IMTEC/3M) and GALILEOS Comfort (Sirona). A CT scan of the three phantoms were also performed with a conventional CT scanner (Toshiba Aquilion 16) to verify the pixel values. The image quality parameters that were evaluated included: image noise, image uniformity, pixel value accuracy, pixel value linearity, contrast scale, and high contrast spatial resolution. There is not one phantom evaluated that provided superior results for all image quality tests. The prototype phantom was adequate for all tests with the exception of the high contrast resolution test whereas the ACR phantom and the CATPHAN due to larger diameter relative to the prototype demonstrated higher noise and non-uniformity, pixel value inaccuracy, and lower contrast scale because of more beam hardening artifact occurring with the large diameter. The optimal diameter for a phantom specialized in dental CBCT testing was found to be between 16 and 17 cm in diameter. To my parents, Nasser and Norah: I hope that you will always be proud of me. Acknowledgements I would like to express my gratitude to Dr. Kerry Krugh. This thesis would not have been possible without your support and mentoring. Thank you for introducing the topic to me and thank you for all the remarks and comments that were very helpful. I would like also to send a special thank you for Dr. Parsai and Dr. Shvydka. You have been nothing but supportive of me since I started the program. I would like to thank Whittaker Family Dental (Defiance, OH), Dr. Matthew Lark dental (Toledo, OH), and Harbor Light Oral & Maxillofacial Surgeons (Toledo, OH) for allowing us to work on their CBCT machines. Also I would like to thank Vladimir Varchena of CIRS Inc. (Norfolk, VA) for lending us the prototype phantom. At last but not the least, this project would also not have been possible without the love and support from my Dad, Mom, my wife Ayesha, and my sister Ghada. Thank you guys for your support and I love you all so much. v Table of Contents Abstractβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦..β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦.iii Acknowledgementsβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦..........v Table of Contentsβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦...β¦β¦β¦β¦.vi List of Tablesβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦...β¦β¦β¦.viii List of Figuresβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦.ix List of Abbreviationsβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦.β¦β¦β¦β¦β¦β¦β¦β¦.β¦xi 1. Introductionβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦..β¦β¦β¦..1 1.1. Radiographic imaging in dentistryβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦..1 1.2. Computed tomographyβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦..β¦β¦β¦.2 1.3. Dental Cone Beam Computed Tomography (CBCT) Technologyβ¦β¦β¦β¦β¦....4 1.4. Quality assurance in dental CBCTβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦..β¦..β¦.9 1.4.1. Radiation dosimetryβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦...β¦β¦β¦β¦..9 1.4.2. Image quality testing in dental CBCTβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦...11 1.5. Existing phantoms designed for dental CBCTβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦.β¦18 1.6. Determination of water-equivalent diameter for human dental and maxillofacial regionβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦..β¦..21 1.7. Objectives of the Study β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦...β¦.23 2. Materials and Methodsβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦.25 vi 2.1. Dental CBCT unitsβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦.β¦..25 2.2. Protocolsβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦.β¦β¦..27 2.3. Set up and Positioningβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦..27 2.4. Phantomsβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦....β¦β¦..28 2.5. Image Quality Analysisβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦....34 2.6. Calculation of water-equivalent diameter (Dw) for human dental and maxillofacial regionβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦..37 3. Resultsβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦..39 4. Discussionβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦52 5. Conclusionβ¦β¦β¦β¦β¦β¦β¦.β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦..55 Referencesβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦..57 vii List of Tables 1.1 Quantitative image quality testing of dental CBCT systems suggested by the SEDENTEXCT projectβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦... 19 2.1 Technical differences between the i-CAT, ILUMA, and GALILEOS β¦.. 25 2.2 Scan protocols used in scanning the phantomsβ¦β¦β¦β¦β¦β¦β¦.β¦β¦β¦... 28 3.1 Noise levels measured in uniformity sections in the phantomsβ¦β¦β¦β¦β¦ 39 3.2 Normalization factors for noiseβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦. 41 3.3 Normalized noise levelsβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦ 3.4 Average noise for each phantom on each machineβ¦β¦β¦β¦β¦β¦β¦β¦β¦... 42 3.5 Average difference between the average mean pixel value of the 41 peripheral ROIs and the central ROIβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦..... 43 3.6 Pixel values of the ACR phantoms materials β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦... 45 3.7 Pixel values of the CATPHAN materials β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦......... 45 3.8 Pixel values of the CIRS prototype materials β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦...... 45 3.9 Values of y-intercept in the three phantomsβ¦β¦β¦β¦β¦..β¦β¦.β¦β¦β¦β¦.. 48 3.10 Contrast scale of the three phantomsβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦ 49 3.11 Calculation of the water-equivalent diameter of the dental and maxillofacial regionβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦.. viii 51 List of Figures 1.1 Principle of computed tomographyβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦.. 3 1.2 Dental cone beam computed tomographyβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦... 4 1.3 Difference between conventional CT and CBCTβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦ 5 1.4 Phantom proposed by SEDENTEXCT project by Leeds Test Objects Inc. 18 1.5 QUART DVT AP phantom designedβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦.. 19 1.6 Phantom suggested by Torgersen et alβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦ 20 2.1 Set up and positioning of the phantoms in dental CBCT unitsβ¦β¦β¦β¦β¦ 27 2.2 Modules of CT ACR accreditation phantomβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦... 30 2.3 Modules of CATPHAN504β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦. 31 2.4 CIRS prototype phantom for dental CBCT unitsβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦ 32 2.5 Sections of the CIRS prototype phantomβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦ 33 2.6 Measuring noise in the uniformity section in each phantom on image acquired by i-CATβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦... 2.7 34 Measuring the difference in pixel values of the periphery from the center in the uniformity sections in each phantom on images acquired by ILUMAβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦.... 2.8 35 Measuring the mean pixel values of different materials in each phantoms on images acquired by i-CATβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦.. 35 ix 2.9 The high contrast spatial resolution section in each phantom on images acquired by i-CATβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦... 36 2.10 Measuring the mean CT number from head CT scanβ¦β¦β¦β¦β¦β¦β¦...... 37 3.1 Noise levels measured in the three phantomsβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦.. 40 3.2 Average noise levels for each phantom in each machineβ¦β¦β¦β¦β¦β¦β¦ 3.3 Average difference between the periphery and the center in the 42 uniformity section in each phantomβ¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦β¦. 44 3.4 Pixel values of different materials in the phantoms (i-CAT vs. CT)β¦β¦... 46 3.5 Pixel values of different materials in the phantoms (ILUMA vs. CT)β¦.... 3.6 Pixel values of different materials in the phantoms (GALILEOS vs. CT).. 47 3.7 Contrast scale of the three phantoms for each machineβ¦β¦β¦β¦β¦β¦....... 49 3.8 The high contrast resolution of the three phantomsβ¦..β¦β¦β¦β¦β¦β¦.β¦.. 50 x 47 List of Abbreviations 3Dβ¦β¦β¦β¦β¦β¦β¦β¦... AAPMβ¦β¦β¦β¦β¦β¦β¦ ACRβ¦β¦β¦β¦β¦β¦β¦... AECβ¦β¦β¦β¦β¦β¦β¦... CBCTβ¦β¦β¦β¦β¦β¦β¦. CCDβ¦β¦β¦β¦β¦β¦β¦... CT.............................. ... CTDIβ¦β¦β¦β¦β¦β¦...... CSβ¦β¦β¦β¦β¦β¦β¦β¦... DAPβ¦β¦β¦β¦β¦β¦β¦β¦ DLPβ¦β¦β¦β¦β¦β¦β¦β¦ FOVβ¦β¦β¦β¦β¦β¦β¦β¦ FPDβ¦β¦β¦β¦β¦β¦β¦β¦ Gyβ¦β¦β¦β¦β¦β¦β¦β¦... HUβ¦β¦β¦β¦β¦β¦β¦β¦... ICRPβ¦β¦β¦β¦β¦β¦β¦β¦ IIβ¦β¦β¦β¦β¦β¦β¦β¦β¦. kVpβ¦β¦β¦β¦β¦β¦β¦β¦. mAβ¦β¦β¦β¦β¦β¦β¦β¦.. mAsβ¦β¦β¦β¦β¦β¦β¦β¦. MDCTβ¦β¦β¦β¦β¦β¦β¦. MSCTβ¦β¦β¦β¦β¦β¦β¦. NCRPβ¦β¦β¦β¦β¦β¦β¦. QAβ¦β¦β¦β¦β¦β¦β¦β¦.. QCβ¦β¦β¦β¦β¦β¦β¦β¦.. ROIβ¦β¦β¦β¦β¦β¦..β¦β¦ Sv β¦β¦β¦β¦β¦β¦..β¦β¦. TFTβ¦β¦β¦β¦β¦..β¦β¦... TLDβ¦β¦β¦β¦β¦..β¦β¦.. Three Dimensional American Association of Physicists in Medicine American College of Radiology Automatic Exposure Control Cone Beam Computed Tomography Charged Coupled Device Computed Tomography CT dose Index Contrast Scale Dose Area Product Dose Length Product Field of View Flat Panel Detector Gray Hounsfield Unit International Commission on Radiation Protection Image Intensifier killoVoltage Peak milliAmprege milliamprege - second .Multidetector CT Multislice CT National Commission on Radiation Protection & Measurments Quality Assurance Quality Control Region of Interest Sievert Thin Film Transistor Thermolumenence Detector xi Chapter 1 Introduction 1.1. Radiographic Imaging in Dentistry X-rays has been always used in dentistry to help doctors assess patient condition and plan treatments. Intraoral radiography is the technique that has been commonly used where a small dental film or digital sensor is placed in the mouth and an external x-ray source is used to expose the image receptor. However, there was always the need for a more comprehensive imaging technique; a method that enables the imaging of both of the jaws and adjacent structures on one image. Therefore, panoramic imaging was introduced and first made commercially available in the 1960s. Panoramic imaging was considered the most comprehensive test used in dental practice. Panoramic imaging uses the principle of tomography to produce a two-dimensional image of a curved structure. The x-ray tube and the image receptor rotate simultaneously around an imaginary fulcrum trough located in the patient. The structures that are in the focal trough (the jaws and teeth) will appear in the image while the structures in the other layers will be blurred out. 1 Panoramic images, however, suffer from the shortcomings that all plain film radiography techniques suffer from; βmagnification, distortion, superimposition, and misrepresentation of structuresβ (Scarfe & Farman, 2008). Analog tomography was also used in dental applications. There are even some panoramic units that have the capabilities of producing tomograms. This mode of imaging had the advantage of removing the superimposition of structures that is present in the panoramic images. However, imaging by this mode of scanning is limited to only one slice, and there was no capability for three-dimensional (3D) reconstruction or multiplanar viewing. Therefore, the idea of utilizing 3D imaging techniques in the dental practice came up to the surface. 1.2. Computed Tomography In 1972, Hounsfield and Cormack revolutionized medicine when their work lead to the invention of computed tomography. Computed tomography (CT) is an imaging modality in which the x-ray tube produces a thin fan beam of x-rays while rotating around the patient. An array of detectors on the opposite side rotates with the tube in synchrony and collects the beam transmitted through the patient. These transmission measurements can then be reconstructed to give a cross-sectional image of the scanned object. Most of the current commercially available CT scanners use multidetector systems where multiple rows of detectors (e.g. 8, 16, or 64) are used to scan multiple slices per one tube rotation. With the aid of slip ring technology, the tube can be continuously rotated in one direction. The gantry rotations can be progressed through the 2 longitudinal axis of the patient in a step-and-shoot fashion which is known as axial scanning mode, or the gantry can be continuously rotated during the scan while moving the table in the longitudinal axis to cover the area of interest. This simultaneous motion of the gantry and the table results in a spiral or helical trajectory of the tube rotation around the patient. This technique helps to cover more anatomical area in less time (Figure 1-1). A set of reconstructed CT images can be post-processed to result in axial, sagittal and/or coronal images. 3D images of the area of interest can be also formed by stacking the reconstructed 2D images. CT is one of the most common imaging procedures requested by physicians in the United States. About 67 million CT scans were performed in the year of 2006 in the United States (NRCP report No. 160, 2009). One of the reasons for such a higher number of CT exams is the many clinical applications in which CT can be utilized. There is no doubt that CT is a strong diagnostic tool that was made possible by the advancements in computers. In addition to the well-known diagnostic role of CT to detect injuries and abnormalities, it can be also utilized to give information about the stage of cancer, plan radiation treatments and surgeries, guide interventional procedures, and monitor the effectiveness of radiation treatment plans. CT has several advantages over general radiography. It removes the superimposition of structures that plain radiography suffers from. It also produces images with high contrast sensitivity; structures in the same slice that differ in their density by only 0.6% can be 3 Figure 1-1: Principle of Computed Tomography. (Taken from www.fda.gov) easily differentiated. Also the CT technology enables viewing images in different planes (axial, coronal, and sagittal) by using the multiplaner reformation tool (MPR). However, due to radiation dose concerns, the use of conventional CT in dentistry has been limited. Other reasons for this limited role are high acquisition and operating cost and large space requirement. Thus cone beam computed tomography (CBCT) units dedicated for dental imaging were introduced. 1.3. Dental Cone Beam Computed Tomography (CBCT) Technology The dental (maxillofacial) CBCT technology was made commercially available in early 2000s. It produces cross-sectional images and reconstructed 3D images of the area of interest. The CBCT technology was initially developed for angiographic applications. Now this technology is being used in many clinical applications such as radiation therapy, surgery, and fluoroscopy. It is also used in many dental applications such as implant planning, endodontics, maxillofacial surgery, and orthodontics (Pauwels et al., 2015). There are two types of dental CBCT equipment arrangement. The first is the upright CBCT where the patient sits on a chair or stands up while the scanning occurs (Figure 1-2). The second configuration is where the patient lies in the supine position during the scan. The latter looks like a conventional CT scan but smaller Figure 1-2: patient sitting in chair while scanning in dental CBCT unit in size. The upright configuration is more common for dental CBCT units. 4 CBCT is considered a new generation of computed tomography. The major difference between the CBCT and the conventional CT is the geometry of the x-ray beam. Figure 1-3 shows the difference in the beam geometry; in conventional CT, a collimator restricts the x-ray beam into a fan-beam geometry, while CBCT uses conical or pyramidal shaped x-ray beams. Unlike conventional CT, an area detector is used in CBCT (rectangular mostly or circular in shape). In CBCT the gantry rotates around a fulcrum point in the patient only one time. A complete volumetric data set can be collected during this single rotation. The rotation might be a full 360° rotation or a partial rotation (180° plus the fan angle) taking in the range of 10 to 40 seconds. The size of the rotation angle can be fixed or variable depending on the manufacturer. For most manufacturers smaller rotation angles usually are preset with lower mAs which means less dose and nosier images. During the single rotation of the gantry, about 150 to 600 planar projections, also called basis images, are collected. The basis images (projections) look like a series of radiographic images, each captured from a slightly different angle than the next. Figure 1-3: Conventional CT beam geometry (left), CBCT beam geometry (right). (Taken from Journal of Canadian Dental Association) Figure 2-1: sedentexCTIQ phantom by Leeds Test Objects Inc.Figure 1-2: Conventional CT beam geometry (left), CBCT beam geometry (right). (Taken from Journal of Canadian Dental Association) 5 Figure 3-1: sedentexCTIQ phantom by Leeds Test Objects Inc. The number of projections usually depends on the scan time, frame rate, angle of gantry rotation and the speed of rotation. The number of projections can be fixed or variable. When more projections are collected, better image quality can be obtained at the expense of higher radiation dose and longer reconstruction times. Also more projections lead to an increase in the signal-to-noise and reduction of metallic artifacts (Scarfe and Farman, 2008). The tube potential used in dental CBCT typically ranges from 60 to 120 kVp and it is fixed for most vendors. Tubes are usually filtered with aluminum of thickness ranging from 2.5 to 10 mm. The tube current applied typically ranges from 1 to 10 mA which, in some models, can be manually varied by the operator depending on the patient size and desired image quality. Some vendors apply automatic exposure control (AEC) in which a suitable mA value can be obtained from the scout image. Other vendors apply non-patient specific mA modulation in which certain mA values are preset for each scan angle. The x-ray generation in CBCT might be continuous or pulsed. The continuous production of x-ray contributes to a higher dose to the patient. The x-ray beam can be made pulsed to correspond with the sampling of the detector. By this way the dose to the patient is dramatically reduced. In the pulsed mode, the real exposure time is less than the scanning time. For example, the exposure time for a 20-second scan on the i-CAT, a dental CBCT unit by Imaging Sciences International (Hatfield, PA), is only 3.6 seconds. That means that the x-ray turns on for 11.76 ms for each of the 306 projections (Imaging Sciences International Inc., 2006). Pulsed systems may show improved image quality due to the reduced motion effect from the gantry rotation (Pauwels et al., 2015). 6 When CBCT systems were first developed, the detector used, and still for some models, was an image intensifier (II) paired with charged coupled device (CCD). This type has some technical problems. For example, its circular field of view (FOV) suffers from truncation artifacts at the periphery. Also because of the gantry rotation, the sensitivity of the image intensifier can be easily affected by the magnetic field of the earth (Scarfe & Farman, 2008). Now, the flat panel detector (FPD) is the most commonly used in CBCT systems. The FPD is an amorphous silicon panel which consists of a single scintillation phosphor (typically cesium iodide doped with thallium) coupled with a thin film transistor (TFT) array. It detects radiation by indirect capture; the scintillator converts the incident radiation into visible light and the light photons are converted into electric charge by the photoconductor layer (amorphous silicon). The signal is then registered by the TFT panel. The FPD produces image quality superior to the images by the II/CCD detector type. The reconstructed CBCT images obtained using FPD have less noise than the images obtained using II/CCD detector type (Baba, Ueda and Okabe, 2004). In addition, the FPD is less bulky and it has higher resolution, greater dynamic range and shows less peripheral artifacts in the FOV (Scarfe & Farman, 2008). More recently, complementary metal-oxide-semiconductor (CMOS) detectors are being incorporated in dental CBCT (Pauwels et al., 2015). The reconstruction technique used in conventional CT is called the filtered back projection. In this technique each axial slice is reconstructed separately from the 2D projection data. For CBCT, the most common reconstruction technique used is the modified Feldkamp algorithm (or the FDK method) which is a variation of the filtered 7 back projection to keep track of the beam divergence in both cone and fan angles. In this method the 3D images are reconstructed from the 2D projection data that is collected during a single rotation around the patient. Further image processing can be performed to produce 3D images by stacking the reconstructed 2D images. Each element in this 3D volumetric data is called a volume element or a voxel. The minimum size of the voxel that can be reconstructed is determined by the size of the pixelated elements of the detector. Each voxel is assigned a value which represents the attenuation properties of the material in that voxel. Also 2D cross-sectional images can be viewed using the multiplanar reformation tool. CBCT reconstruction time depends on the voxel size chosen, the FOV, and the number of projections images collected. Typical reconstruction time is less than 2 minutes. The FOV size in dental CBCT differs according to the manufacturers. Some units have an option for extended view which is used when trying to scan areas that are larger than the detector size. There are two ways for extended view imaging. The first way is by obtaining two separate scans and then fusing the two data sets to result in volumetric data for the large area of interest. This method has the disadvantage of overexposing the overlapping area. The second way is to shift the position of the detector laterally and collimate the beam asymmetrically (Scarfe and Farman, 2008). 8 1.4. Quality assurance for dental CBCT A typical quality assurance program for evaluating the performance of an imaging system that utilizes ionization radiation for image formation involves two components; image quality and radiation dosimetry. 1.4.1. Radiation dosimetry Radiation dose assessment for any radiation emitting devoice is necessary for the safety of patients. It provides a method of monitoring and estimation of the radiation dose delivered to the patients. In conventional CT, the standardized method of choice to characterize the dose is the CT Dose Index (CTDI). However, because of the nature of the geometry of the beam in CBCT and the amount of scatter, the CTDI cannot be properly adapted in CBCT (Pauwels et al., 2012). The pencil ionization chamber that is used in measuring the dose in conventional CT is 10 cm long which may be useful for CBCT machines with height FOV up to 6 cm. However, it is not ideal for machines with FOV larger than 8 cm because the scatter produced would not be measured which would lead to an underestimation of the dose (Araki et al., 2013). Also, a range of FOV sizes can be chosen which can exceed the size of the pencil chamber. Large FOV can affect the dose distribution (Pauwels et al., 2012). In addition, the iso-center of the scan can be changed from central to peripheral in some scans, which in turns produces an asymmetric dose distribution (Pauwels et al., 2012). Similarly, choosing a full arc rotation of the tube versus a partial arc will affect the dose distribution. (Pauwels et al., 2012). 9 The issues discussed in the prior paragraph lead to an inaccurate estimation of the dose when using the concept of CTDI in dental CBCT. Therefore there is not yet a standardized method for evaluation of the radiation dose in dental CBCT and the most accurate way to characterize the dose is by point-by-point measurement using TLDs in an anthropomorphic phantom (e.g., the RANDO phantom) which is not practical for routine QA radiation dosimetry measurement. Effective dose, measured in Sievert (Sv), allows for the assessment of patient risk and also allows for comparing different types of medical imaging modalities. Dental CBCT offers relatively lower effective dose than conventional CT but higher effective dose when it is compared to panoral radiography. A study done by Dr. Ludlow et al compared the effective dose for 3 commercially available CBCT units: CB Mercury, NewTom 3G, and i-CAT. Twenty-four TLDs were placed throughout the layers of a head and neck RANDO phantom. After taking the measurements, they calculated the total body effective dose by summing the doses for individual organs calculated using both of the 1990 and 2005 ICRP tissue weighting factors. The results showed that the effective doses produced by these CBCT scanners were 4 to 42 times greater than equivalent panoramic images (Ludlow et al., 2006). Based on the results, a 12-inch FOV scan on the i-CAT is equivalent to a dose of 21 single panoramic exposures. Also the results showed lower doses when using smaller FOV and reduced values of kV and mA. They concluded that βCBCT dose varies substantially depending on the device, FOV and selected technique factors. Effective dose detriment is several to many times higher than conventional panoramic imaging and 10 an order of magnitude or more less than reported doses for conventional CTβ (Ludlow et al., 2006). Another study that supports the findings of Dr. Ludlow was done byJ A Roberts et al (2009). They studied the effective dose delivered to patients during a CBCT examination placing TLDs in a RANDO phantom and scanning it using the i-CAT dental CBCT. They calculated the effective dose using both the 1990 and the 2007 ICRP tissueweighting factors. They concluded that βdoses from CBCT are low compared with conventional CT but significantly higher than conventional dental radiography techniques.β The SEDENTEXCT project established by the European Atomic Energy Community (Euratom) under the European Commission on Radiation Protection proposed a new method that enables a more accurate estimation of the dose using a specially designed phantom. The dose indices suggested are dose index 1 (DI1), dose index2 (DI2) and dose area product (DAP). However, this method yet to be accepted amongst the medical physics community as the standard method for assessing the dose in dental CBCT. 1.4.2. Image quality testing in dental CBCT The second component of routine QA evaluation of dental CBCT is an assessment of image quality. The image quality parameters that are important for assessment are image noise and low contrast detectability, image uniformity, pixel value accuracy and linearity, contrast scale, high contrast spatial resolution, and image artifacts. 11 1.4.2.1 Image noise and low contrast detectability Noise is defined as a random or stochastic variation in the signal that, in ideal situations, follows Poisson distribution. According to Poisson statistics the quantum noise per pixel in an image is given by βπ where N is the number of photons. Thus, quantum noise is determined by those factors impacting the amount of photons used per pixel for image formation (kVp, mAs, filtration, voxel size, and number of projections). In addition, there are a variety of other sources of image noise, most notably in CT, the image reconstruction kernel. Noise strongly impacts low contrast detectability which is defined as the ability to evaluate structures that have attenuation properties that are slightly different than that of the background. Due to the low mAs and small voxel size that are typically used in dental CBCT systems, images contain relatively high noise levels. In addition, the nature of the beam in CBCT produces a large amount of scatter radiation. Both the relatively high noise levels and increased scatter radiation result in a reduced contrast detectability of dental CBCT images in comparison with conventional CT. 1.4.2.2 Image uniformity Uniformity across an image is an important parameter to evaluate and quantify. Ideally, tissues that have the same attenuation properties should have same pixel value in any location in the image. Due to the divergence of the x-ray beam used in CBCT, there 12 is a variation in the intensity of the incident x-rays which causes a non-uniform absorption and detection. In addition, dental CBCT images suffer from non-uniformities whenever there is a mass outside the FOV, often called exo-mass. The exo-mass causes beam hardening which reduces the image intensities in parts of the image (Bryant, Drage and Richmond, 2008). 1.4.2.3. Pixel value accuracy and linearity In conventional CT, each voxel in the image is assigned a certain value. This value is calculated using the measured attenuation coefficient of the tissues in that voxel normalized to the attenuation coefficient of water and is called CT number (expressed in Hounsfield units). The following equation used to calculate the CT number of a voxel of tissues: πΆπ# = ππ‘ππ π π’π βππ€ππ‘ππ ππ€ππ‘ππ π₯1000 (1-1) Where: CT#: the CT number of a given material in Hounsfield units µtissue: linear attenuation coefficient of a given material µwater: linear attenuation coefficient of water Any material that possesses absorption properties that are higher than water takes a positive value. Similarly any material that has absorption properties that are lower than 13 water takes a negative value. For example, the CT number of air is -1000, while the CT number of dense bone is approximately +1000. The pixel intensity values measured in dental CBCT images may not accurately represent the true Hounsfield unit values (Scarfe et al., 2012). The reasons of such discrepancy in the pixel values between conventional CT and CBCT are the high amount of scatter, the applied reconstruction algorithm utilized in CBCT, and the reduced technique factors that are usually used. In addition, great variability of the pixel values can be seen in CBCT images from different machines manufacturers (Pauwels et al., 2015). In conventional CT, the CT number values across the image possess a linear relationship with respect to the linear attenuation coefficient of the materials in the image. In dental CBCT, the pixel values should also have a linear relationship with the beam attenuation. One method to assess this relationship is to compare the measured dental CBCT pixel values of a variety of materials with known HU values as determined from conventional CT scan. The linearity of the relationship can be measured by means of the correlation coefficient from linear regression analysis. 14 1.4.2.4. Contrast Scale (CS) The contrast scale (CS) in conventional CT is defined as the change in linear attenuation coefficient relative to the change in CT number and is expressed in the following equation: π (πΈ)βπ (πΈ) πΆπ = πΆπ1 (πΈ)βπΆπ2 (πΈ) 1 2 (1-2) Where, µ1: linear attenuation coefficient of the first material for a specific beam energy µ2: linear attenuation coefficient of the second material for a specific beam energy CT1: the CT number of the first material for a specific beam energy CT2: the CT number of the second material for a specific beam energy In this project, however, the CS is defined as the change in the pixel value between certain materials in the dental CBCT image relative to the change in the HU measured in a conventional CT image for the same materials and is expressed in the following equation: πΆπ = ππ1 (πΈ)βππ2 (πΈ) πΆπ1 (πΈ)βπΆπ2 (πΈ) (1-3) where PV1: the pixel value of the first material in the dental CBCT image PV2: the pixel value of the second material in the dental CBCT image CT1: the HU of the first material in the conventional CT image CT2: the HU of the second material in the conventional CT image 15 Ideally, the contrast scale (i.e., slope) should be equal to unity which means any change in the pixel values measured in the CBCT image corresponds to the same amount of change in the HU measured in the conventional CT image. 1.4.2.5.High contrast spatial resolution The spatial resolution is defined as βthe ability of an image system to distinctly depict two objects as they become smaller and closer togetherβ (Bushberg, 2002). The smallest object that can be resolved is known as the limiting spatial resolution. The spatial resolution in dental CBCT is mostly dependent on the reconstructed voxel size, the physical size of the pixelated detector, and the reconstruction kernel. CBCT is capable of reconstructing isotropic voxels meaning that voxels have equivalent dimensions in the x,y, and z directions. Typical voxel size that can be reconstructed in dental CBCT varies between 0.08 mm to 0.4 mm. The limiting spatial resolution can be evaluated visually by using a line pair pattern which consists of an alternating strip and space arrangement of different widths. In literature, the line pair per distance values that have been reported for dental CBCT images range between 0.6 and 2.8 lp/mm (Brüllmann and Schulze, 2015). In addition, the spatial resolution of an imaging system can be quantified by the modulation transfer function (MTF). The MTF can be calculated by applying a Fourier Transform (FT) on a point spread function or a line spread function obtained by the imaging system. The limiting spatial resolution of the system would be empirically 16 considered at the 10% MTF. The 10% MTF for dental CBCT found in literature is in the range from 0.5 to 2.3 cycles/mm (Brüllmann and Schulze, 2015). 1.4.2.6.Artifacts Dental CBCT images may suffer from cone beam artifacts. The cone beam artifacts result from undersampling the cone angle direction. This affects the image quality in the upper and lower edges of the reconstructed volume. The image quality improves gradually towards the center (Schulze et al., 2011). In addition, the beam divergence causes other artifacts such as aliasing and high amounts of scatter. 17 1.5. Existing phantoms designed for dental CBCT During the past few years some efforts were made to the purpose of designing a comprehensive quality assurance program for dental CBCT. The Safety and Efficacy of a New and Emerging Dental X-ray CT scanners, or the SEDENTEXCT, project established by the European Atomic Energy Community (Euratom) under the European Commission on Radiation Protection was one of the firsts who addressed this issue. The objective of the SEDENTEXCT project was to βdevelop evidence-based guidelines dealing with justification, optimization and referral criteria for users of dental CBCT.β They published provisional guidelines in report 172 in 2008 and published an updated version in 2011. The report included a chapter about quality standards and quality assurance and recommended that a QA program should include six topics; x-ray tube generator performance, quantitative assessment of image quality, display screen performance, patient dose assessment, clinical image quality assessment and clinical audit . They proposed an image quality program using the SEDENTEXCT phantom designed by Leeds Test Objects Inc. This phantom is a PMMA cylinder of 16 cm in diameter and a height of 16.2 cm (Figure 1-4). Nine image quality characteristics can be evaluated using this phantom and these parameters are noise, uniformity, pixel value accuracy contrast resolution, spatial resolution, geometrical accuracy, and metal artifacts. Table (1-1) summarizes the test and their frequency. Since there are no action or tolerance 18 Figure1-4: sedentexCTIQ phantom by Leeds Test Objects Inc. levels so far, they suggested to establish baseline values during the acceptance testing and then monitor these values in the routine testing. Table (1-1 ): testing of image quality of dental CBCT as suggested by the SEDENTEXCT project. (Taken from report 172) Table (2-1 ): testing of image quality of dental CBCT as suggested by the SEDENTEXCT project. (Taken from report 172) Table (2-1 ): testing of image quality of dental CBCT as suggested by the SEDENTEXCT project. (Taken from report 172) Table (2-1 ): testing of image quality of dental CBCT as suggested by the SEDENTEXCT project. (Taken from report 172) Pauwels et al (2011) conducted a study where they evaluated the SEDENTEXCT proposed phantom and stated that it showed promising potential for technical image quality evaluation of CBCT. J Bamba et al (2013) also used this phantom to evaluate three dental CBCT systems and stated that all the essential image quality parameters can be assessed using this phantom. QUART, a German company, also designed a special phantom for testing dental CBCT systems (Figure 1-5). The QUART DVT AP phantom is cylindrical in design with a 16 cm diameter and 15 cm in height. It is designed to evaluate image Figure 1-5: Quart DVT AP phantom (taken from Quart website) uniformity, image noise, contrast resolution and CNR, MTF, limiting spatial resolution and image geometry. 19 The center for Evidence-Based Purchasing CEP, a part of the department of health in the United Kingdom, conducted a comparative evaluation of dental CBCT systems in 2010 where they compared the dose and image quality for several CBCT systems from different vendors. For image quality comparison they used CATPHAN 424. They were able to test the high contrast spatial resolution, uniformity, CT number accuracy, and geometric accuracy. However, low contrast detectability module of the CATPHAN was not adequately visualized on any of the systems they evaluated. Another phantom which is not commercially available and was designed for research purposes by Torgersen et al (2014) is shown in Figure 1-6. They tried to design an inexpensive phantom for simplified image quality assurance for image quality assessment of dental CBCT. All of the tests performed by this phantom were objective. They also developed software that can be used to evaluate the images of this phantom. The tests included low and high contrast resolution, uniformity, noise, and geometric linearity. The phantom has a diameter of 16 cm and a height of 70 cm. The SEDENTEXCT phantom and the QUART phantom are the only commercially available phantoms designed for dental CBCT but may not be cost effective for every diagnostic medical physicist to purchase. Figure 1-6: QA phantom designed by Torgersen et al for dental CBCT testing 20 1.6. Determination of water equivalent diameter (Dw) for human dental and maxillofacial region. The phantom that is ideally used in the assessment of image quality parameters of dental CBCT scanners would have the same attenuation properties as the dental and maxillofacial region of the human head. Most phantoms in radiologic testing are made of water-equivalent material. The optimal diameter of a cylindrical phantom can be determined by calculating the diameter of water that has similar photon attenuations as the dental and maxillofacial tissues. The CT number values in CT images expressed in Hounsfield units are calculated by using the linear attenuation coefficients of the materials in the image. The CT number is thus representative of the attenuation properties of a voxel of a tissue and can be used to calculate the diameter of water that would have equivalent attenuation. The methodology and calculation of the water-equivalent diameter (Dw) have been discussed in detail in AAPM task group report 220 report. The following equation was used: 1 π·π€ = 2β[1000 πΆπ(π₯, π¦)π ππΌ + 1] π΄π ππΌ (1-4) (AAPM Task Group 220, 2014) π where, Dw: water-equivalent diameter CT(x,y)ROI: the mean CT number in a chosen ROI AROI: the area of the ROI 21 In addition, by rearranging the previous equation and solving for area (A= π(D/2)2), an attenuation equivalent diameter of materials other than water can be determined as in the following equation: π·πππ‘πππππ = 2β{ 2 π· ( π€) ( 2 πΆππππ‘πππππ +1) 1000 } (1-5) where, Dmaterial: the attenuation equivalent diameter of material other than water CTmaterial: the HU of the material of interest Dw: the water-equivalent diameter calculated from Formula 1-4 22 1.7. Objectives of the Study When dental CBCT was first introduced there was no requirement for medical physics acceptance testing or annual quality control testing of dental CBCT. Since that time only some States have mandated that medical physicists perform testing on these units. Ohio Administrative Code rule 3701:1-66-10 was revised in July 2014 to include CBCT scanners and hybrid imaging systems. Paragraph D of that rule states that a QA testing should be performed annually for CBCT units. The annual testing should be performed by a radiation expert. To date there are no standardized test methods, phantoms, or test criteria for these units The American Association of physicists in medicine (AAPM) has recently formed a task group (TG261) for discussing this issue. Its main objective is to establish a standardized procedures and techniques for the quality control of dental and maxillofacial imaging systems. With this rising interest to comply with the regulations, there is a necessity to develop guidelines and recommendations regarding routine quality control testing of CBCT scanners. The following are the objectives of this study: 1. Evaluate the feasibility of using phantoms common to conventional CT testing for use in dental CBCT quality control testing. These phantoms include the ACR CT phantom and CATPHAN. Both of these phantoms are available and familiar to most diagnostic medical physicists. It would be more economical and convenient if these phantoms were found to be 23 suitable for testing dental CBCT units. Purchasing specially designed QA phantoms for dental CBCT testing may not be cost effective, especially for small dental offices or small medical physics consultancy companies. 2. Evaluate a prototype phantom designed specifically for dental CBCT QC phantom (the CIRS prototype phantom). 3. Evaluate the optimal dimensions of a QA phantom for dental CBCT based on the radiation attention properties of the dental and maxillofacial region of the human head and propose alterations and features for appropriate phantom design based on observations from objectives 1 and 2. 24 Chapter 2 Materials & Methods Evaluation of QA phantoms 2.1 Dental CBCT units Image quality testing was performed on three commercially available CBCT scanners: i-CAT Classic (Imaging sciences International), ILUMA (IMTEC/3M), and GALILEOS Comfort (Sirona). Table 2-1 summarizes specifications for each scanner: Table ( 2-1 ): technical differences between the i-CAT, ILUMA, and GALILEOS. Scanner iCAT classic (Imaging Sciences) ILLUMA (3M/IMTEC) GALILEOS comfort (Sirona) Focal spot (mm) Detector Exposure mode kVp mA FOV DxH (cm) 0.5 FPD Pulsed 120 3-7 16x13 0.3 FPD Continuous 120 1 or 3.8 17x14 0.5 II pulsed 85 5-7 15x15 FPD: Flat panel detector. II: Image intensifier. FOV: Field of view. 25 Figure 3-1: the platform and the tripod used to position the phantoms Figure 3-1: the platform and the tripod used to position the phantoms 2.1.1 i-CAT Classic (Imaging Sciences International) The i-CAT utilizes an x-ray tube with a 0.5 mm focal spot size and 15° anode angle. Technical factors for x-ray generation on this units are 120 kVp (fixed tube voltage) and a tube current that can be varied from 3 to 7 mA. The tube has filtration equivalent to 10 mm of aluminum at 120 kVp. The i-CAT utilizes an amorphous silicon flat panel for detection with a readable area with dimensions of 24 cm by 20 cm. This detector allows for a maximum field of view of 16 cm in image diameter and 13 cm in height which can be collimated down to cover areas of smaller size. 2.1.2 ILUMA (IMTEC/3M) The ILUMA utilizes an x-ray tube with a 0.3 focal spot size and has a filtration of 1.27 mm of copper. The tube voltage is fixed to 120 kV and the tube current has two settings; 1 or 3.8 mA. The ILUMA utilizes a flat panel detector with dimensions of 20 cm by 24 cm which allows a reconstructed FOV of 17 cm in image diameter and 14 cm in height. The exposure mode in the ILUMA is continuous, which means the scan time equals the exposure time. 26 2.1.3 GALILEOS comfort (Sirona) This unit uses an x-ray tube with a 0.5 mm focal spot and has a 2.5 mm of aluminum thickness for filtration. The tube voltage is fixed to 85 kV and the tube current can be varied between 5 and7 mA. The GALILEOS uses an image intensifier for detection with an input window of 21.5 diameter which allow a reconstructed FOV is 15 cm in both diameter and height. The GALILEOS has two programs that can be used for scanning; VO1 and VO2. The difference between the two programs is with the image reconstruction algorithm and thus acquisition parameters are equivalent. 2.2 Protocols A total of eight protocols were chosen to scan the phantoms. The protocols chosen for scanning in this project were those that were commonly used in the clinical setting. Table (2-2) summarizes the scan protocols used. Note that in the following chapters the protocols will simply be referred to by the Reference number in Table 2-2. 2.3 Set up and positioning Figure 2-1: the platform and the tripod used to position the phantoms The phantoms were positioned vertically on a custom- made stand inserted on a tripod. The platform was leveled and lasers were used to center the phantoms FOV. Figure 3-1: in the the platform and the 27 tripod used to position the phantoms Figure 3-1: the platform and the tripod used to position the phantoms Table (2-2 ): protocols used in scanning the phantoms Reconstructed volume Protocol scan (diameter x height) Ref (cm) # time (s) voxel arc (mm) (°) projections 120 6.9 10 0.3 332 160 2 120 6.6 20 0.4 332 306 3 120 3.8 7.8 0.3 190 117 17x14 4 120 3.8 20 0.3 360 302 cylindrical 5 120 1 20 0.3 360 302 6 120 3.8 40 0.3 360 602 16x13 (Imaging Sciences) cylindrical (IMTEC/3M) mA 1 iCAT classic ILUMA kV Galileos Comfort 15 7 85 7 14 0.29 204 200 (Sirona) Spherical 8 85 7 14 0.29 204 200 2.4 Phantoms Three phantoms were evaluated for dental CBCT testing. The phantoms evaluated were the ACR phantom and the CATPHAN which are commonly used in testing conventional CT. The third phantom evaluated is the CIRS prototype which is specially designed for testing dental CBCT phantoms. The details for the each phantom are as follows: 28 2.4.1 CT ACR Phantom Gammex 464 is the only phantom that is authorized by the American College of Radiology to be used in the accreditation process of conventional CT scanners. It can be used for initial QA assessment and routine monthly QA testing for CT machines. Several image quality parameters of CT can be evaluated using this phantom. The CT ACR phantom is 20 cm in diameter and has a height of 16 cm. It consists of 4 modules or sections. Each section has a thickness of 4 cm. The first module is used to evaluate positioning and alignment, CT number accuracy, and slice thickness (Figure 2-2: A). In this module there are 5 cylinders of different materials. These different materials are used to evaluate CT number accuracy. This module also includes two ramps each consists with a series of wires that can be used to assess the slice thickness. The second module is used to evaluate low contrast detectability and contrast-tonoise ratio. This module consists of multiple cylinders of different diameters (Figure 22:B) The third module is made of a uniform tissue-equivalent material used to assess uniformity and noise. This module also has 2 small BBs 10 cm apart which can be measured to assess the geometric accuracy (Figure 2-2:C) The fourth module is used to evaluate high contrast spatial resolution. It has 8 square line pair regions. The limiting spatial resolutions that can be measured are 4, 5, 6, 7, 8, 9, 10 and 12 lp/cm (Figure 2-2: D) 29 B A C D Figure 2-2: the four modules of the ACR CT phantom. Taken from the CT ACR accreditation phantom instructions. Figure 3-2: the four modules of the ACR CT phantom. Taken from the CT accreditation phantom instructions. CATPHAN504 2.4.2 Figure504 3-2: the four modules of thePhantoms ACR CT phantom. Taken from the CT The CATPHAN designed by The Laboratory Inc. and is commonly accreditation phantom instructions. used to evaluate image quality and system performance for both CT and CBCT scanners. It is 20 cm in both diameter and height. It consists of 4 modules (Figure 2-3). Module Figure 3-2: the four modules of the ACR CT phantom. Taken from the CT accreditation phantom instructions. CTP528 (Figure 2-3A) is used to test spatial resolution. It consists of high contrast line pairs ranging from 1 to 21 lp/cm. Module CTP 515 (Figure 2-3B) is used to assess low contrast detectability. 30 B A C Figure 2-3: modules of Catphan 504 (taken from the phantom laboratory manual) It consists of three groups of supra-slice low contrast targets with nominal contrast levels of 0.3%, 0.5%, and 1.0%. Each group has 9 circles with diameters ranging from 2 to 15 mm. It also consists of three sub-slice groups each having 4 circles of diameters 3,5,7 and 9mm. Module CTP 404 (Figure 2-3C) has 8 targets of different materials (Air, PMP, LDPE, Polystyrene, Acrylic, Derlin, and Teflon) which can be used to assess the accuracy and linearity of the Hounsfield unit scale. This section also contains 4 wire ramps with 23° angulation. The wire ramps can be used to assess alignment and slice thickness. Also the section has 4 3mm holes positioned at 50mm from each other. These can be used to evaluate the measurement accuracy. In the center of this module there are 5 acrylic spheres of different diameters to evaluate the machineβs ability to image volumes. 31 The last module is made of a uniform material (CTP486) module and is used to evaluate image uniformity, noise and artifact. 2.4.3 CIRS QA Phantom prototype This phantom was designed in 2010 by Computerized Imaging Reference systems (CIRS) Inc. which is still under investigation and subject to further development. This prototype is designed to be used in CBCT acceptance and periodic QA testing. It has 5 sections and can be used to measure seven dental CBCT performance parameters. It can also provide a range of bone mineral density references. It has a triangular shape with a height of 17 cm (Figure 2-4). The second layer of the phantom is used to evaluate uniformity and noise. It is made of a water-equivalent material and has a thickness of 20 mm (Figure 2-5). Figure 2-4 : Dental CBCT prototype QA phantom by CIRS Inc. 32 Figure 3-4 : Dental CBCT prototype QA phantom by CIRS Inc. The third section is used to evaluate low contrast detectability. This section has a thickness of 30 mm and its background is made to be water equivalent. There are three sets in this sections. Each set has 8 cylindrical targets and 8 spherical targets. The cylinders are 30 mm long and have diameters of 10, 7, 5, 3.5, 2.5, 1.8, 1.2 and 0.9 mm. The spheres have diameters of 10, 8, 6.5, 5, 4, 3.2, 2.5, and 2 mm. Each set has a different density than the other; the first set is 5 HU, the second is 10 HU and the third is 20 HU above the background (Figure 2-5). The fourth section is used to evaluate CT number linearity, alignment and slice thickness. This layer also is 30 mm in thickness. To assess the CT number linearity, this section has 6 cylinders each with 13 mm diameter and 30 mm in length. Each cylinder has different density to mimic different material; adipose tissue, brain, muscle, bone, dense bone. To assess the slice thickness this section also has 3 stainless steel wires that are angled 30° from the axial plane (Figure 2-5). Figure 2-5: the layers of the dental CBCT prototype (taken from the manual provided by CIRS) 33 2.5 Image Quality analysis All image analysis was done using ImageJ software freely available from NIH (National Institute of Health). Image noise was measured as the standard deviation of a central ROI drawn in the uniformity section for each phantom used. To eliminate slice to slice variation the noise was averaged over twenty consecutive slices. a b c Figure 2-6: measuring noise in the uniformity section in each phantom on i-CAT images. a) ACR phantom. b)CATPHAN. C) CIRS prototype. For measuring the image uniformity, 5 equal circular ROIs were drawn in center, 12, 3, 6, and 9 oβclock positions in the uniformity section in each phantom. The image uniformity was calculated as the mean difference in pixel values of the four periphery positions from that of the center. This again was averaged over twenty consecutive slices 34 a b c Figure 2-7: measuring the difference of the peripheral mean pixel value from the center in the uniformity section in each phantom on the ILUMA images. a) ACR phantom. b) CATPHAN. c)CIRS prototype. The mean pixel value for all materials in each phantom was measured with an ROI. The pixel values were plotted against the βtrueβ HU values obtained from a conventional CT scan. Linear regression was used to fit a line to the data. From the linear equation the Y-intercept was used as an assessment of pixel value accuracy, the slope as an assessment of contrast scale, and the correlation coefficient (R2) as an assessment of pixel value linearity. a b c Figure 2-8: measuring the mean pixel values for different materials on the i-CAT images of each phantom 35 a c b Figure 2-9: high contrast spatial;; resolution section of phantoms acquired on i-CAT scanner. A)ACR phantom. b) CATPHAN. c) CIRS prototype For evaluating the high contrast spatial resolution, the smallest line pair that could be visually resolved in each phantom was recorded. 36 Diameter of dental CBCT QA phantom 2.6 Calculation of water-equivalent diameter (Dw) for human dental and maxillofacial region Ten conventional CT scans of the head area were chosen for this calculation. Since the goal is to measure the equivalent attenuation of the dental and maxillofacial tissues only, the images that had foreign objects, such as metallic implants or bismuth shields, were avoided. Such highly attenuating materials affect the accuracy of the Dw calculation (AAPM Task Group 220, 2014). In each head scan the slices from the inferior of the orbit to the inferior edge of the mandible were selected. An ROI was drawn to include the tissues in each slice. a b c Figure 2-10: measuring the mean CT number value in the CT scans. ROI was drawn to include all tissues in the slice. a) Orbits level. b) mid-level. c) Mandible level The ROI was drawn in such a way to include the entire cross-section. Irrelevant objects such as the patient table was not included in the ROI since it can lead to an overestimation of the attenuation of the ROI (AAPM Task Group 220, 2014). 37 The mean CT number of each ROI was recorded and the average across all the slices was determined. 38 Chapter 3 Results Evaluation of QA phantoms 3.1 Image noise Table 3-1 shows the noise levels measured in each phantom using scan protocols 1 to 8 referenced in Table 2-2. Figure 3-1 depicts the difference in noise between the protocols used. Table 3-1: noise levels measured as the standard deviation of an ROI of uniformity section in each phantom Scan Protocol Machine ACR Phantom CATPHAN 1 iCAT 91.4 91.9 CIRS prototype 47.7 2 iCAT 69.3 72.4 38.8 3 ILUMA 92.5 88.9 51.5 4 ILUMA 62.9 58.8 32.3 5 ILUMA 128.4 115.8 60.5 6 ILUMA 55.5 44.1 24.5 7 GALILEOS 32.4 48.0 10.1 8 GALILEOS N/A 152.0 35.5 39 160.0 140.0 Noise 120.0 100.0 80.0 60.0 40.0 20.0 0.0 iCAT iCAT ILUMA ILUMA ILUMA ILUMA GAL GAL 1 2 3 4 5 6 7 8 Scan Protocols ACR CAT CIRS Figure 3-1: noise levels measured in the phantoms Since image noise is a function of the number of photons used for image formation it is possible to normalize the noise to the image technique. The normalization removes the variability in noise for a given machine for changing technique factors. In particular the noise was normalized to the tube current (mA) and number of projections used for image formation. The mA normalization factor is the square root of the mA in each protocol divided by 7 mA (the highest mA amongst the protocols). The normalization factor of the number of projections is the square root of the number of projections used in each protocol divided by 300. This number was chosen because many protocols used number of projections close to 300. The following equation is for calculating the noise normalization factor (NNF): ππ΄ πππΉ = β( 7 # ππ ππππππ )( 300 40 ) (3-1) The calculation of the noise normalization factor is presented in Table 3-2. The noise normalization factor is multiplied by the noise levels in Table 3-1. The normalized noise levels are presented in Table 3-3. In Tables 3-2 and 3-3 only one protocol was used for the GALIEOS since protocol 7 and 8 differ only in the reconstruction kernel and not in the technical factors. Thus no normalization of the mA or the number of projection can be used for the GALILEOS. Table 3-2: Noise Normalization Factor (NNF) Scan Protocol Machine mA projections mA Normalization factor No. of projections Normalization factor NNF 1 iCAT 6.9 160 0.993 0.730 0.725 2 iCAT 6.6 306 0.971 1.010 0.981 3 ILUMA 3.8 117 0.737 0.624 0.460 4 ILUMA 3.8 302 0.737 1.003 0.739 5 ILUMA 1.0 302 0.378 1.003 0.379 6 ILUMA 3.8 602 0.737 1.417 1.040 7 GAL 7.0 200 1.000 1.000 1.000 Table 3-3: normalized noise levels Scan Protocol Machine ACR Phantom CATPHAN CIRS prototype 1 iCAT 66.3 66.6 34.6 2 iCAT 68.0 71.0 38.1 3 ILUMA 42.6 40.9 23.7 4 ILUMA 46.5 43.5 23.8 5 ILUMA 48.7 43.9 22.9 6 ILUMA 57.9 46.1 25.5 7 GAL 32.4 48.0 10.1 41 An average of the normalized noise of the phantoms was calculated for the i-CAT and the ILUMA. The normalized noise values of the three phantoms from protocols 1 and 2 were averaged and the error bars were determined as the standard deviation of the average. Similarly, the normalized noise values of the phantoms from protocols 3,4,5 and 6 were averaged and the error bars were determined. No averaging was performed on the normalized noise for the GALILEOS as explained above and thus only protocol 7 was used for comparison. Table 3-4 presents the average noise with the error bars and Figure 3-2 shows these values. Table 3-4: Average noise in each phantom for each machine i-CAT ILUMA GALILEOS Averaged Protocolsβ data 1 and 2 3,4,5 and 6 7 only ACR Phantom CATPHAN CIRS Prototype 67.1±1.2 48.9±6.5 32.4 68.8±3.1 43.6±2.1 48.0 36.4±2.5 24.0±1.1 10.1 80.00 Normalized noise 70.00 60.00 50.00 40.00 30.00 20.00 10.00 0.00 i-CAT ILUMA ACR CAT GALILEOS CIRS Figure 3-2: average noise in each phantom for each machine. 42 In general, the noise from the ACR phantom and CATPHAN for a given scanner were comparable. The noise on these two phantoms was within 2.5% in the i-CAT, 12% in the ILUMA and 38% in the GALILEOS. The noise of the CIRS prototype was significantly lower than that in the other two phantoms due to the significant difference in the phantomβs diameter. The noise in the CIRS phantom was lower than that in the the ACR phantom and CATPHAN by 47% for the i-CAT and ILUMA and by 75% on the GALILEOS. 3.2 Image uniformity Table 3-5 shows the calculation of image uniformity in each phantom for the protocols 1 to 8. Figure 3-3 depicts the average difference in pixel value between the center and the periphery in the phantoms. Table 3-5: the average difference between the peripheral ROIs and the central ROI inn the uniformity section in each phantom Scan protocol scanner ACR phantom CATPHAN CIRS prototype 1 iCAT 122.8 127.4 83.2 2 iCAT 144.6 129.2 81.6 3 ILUMA 10.3 31.2 79.7 4 ILUMA 9.0 31.6 2.7 5 ILUMA 9.5 28.8 6.9 6 ILUMA 12.9 31.9 1.9 7 GALILEOS 46.4 106.6 5.2 8 GALILEOS NA 109.8 5.9 43 Uniformity Pixel Value Difference 160 140 120 100 80 60 40 20 0 i-CAT i-CAT ILUMA ILUMA ILUMA 1 2 3 4 5 ACR CAT ILUAM GALILEOS GALILEOS 6 7 8 CIRS Figure 3-3: the average difference between the periphery and center for protocols 1 to 8. For a given scanner, the images of the CIRS prototype show the highest uniformity as compared to the other two phantoms with the exception of protocol 3 which was affected by artifacts. On average the difference in pixel values between the center and periphery in the CIRS phantom, with the exception of protocol 3, was around 60% less than that in the ACR phantom and around 75% less than that in the CATPHAN. 3.3 Pixel value accuracy and linearity Tables 3-6, 3-7 and 3-8 show the pixel values for each material in the ACR phantom, CATPHAN and CIRS prototype, respectively, measured in protocols 1 to 7 referenced in Table 2-2. The tables also show the CT number of each material measured in the CT images by the Toshiba Aquilion 16 conventional CT scanner. 44 Table 3-6: pixel values measured in the ACR phantom ILUMA iCAT GAL Toshiba CT # P1 P2 Avr P3 P4 P5 P6 Avr P7 Air Polyethylene Acrylic -989 -999.9 -1000 -999.9 -1048.8 -1053.9 -1043.8 -1032.9 -1044.8 1430.8 -93 -656.4 -647.5 -651.9 -181.2 -158.6 -169.9 -127.9 -159.4 2130.1 120 -478.5 -479.9 21.8 39.1 28.7 60.2 940 171.3 172.7 504.2 513.2 497.2 545.2 Water 0 -502.6 -523.5 -113.3 -67.6 -78.3 -46.6 37.4 514.9 -76.5 2106.2 Bone -479.2 172.0 -513.1 Material 2525.8 2120.5 Table 3-7: pixel values measured in the CATPHAN ILUMA iCAT GAL Material Toshiba CT # P1 P2 Avr P3 P4 P5 P6 Avr P7 Acrylic 120 -488.6 -486.7 -116.1 -133.4 -175.0 -135.7 319 -375.3 -338.4 -11.7 -18.2 -59.2 -18.9 Teflon 936 13.8 18.5 259.6 296.4 247.8 298.8 -794.8 -847.8 -865.2 -849.6 -292.9 -320.3 -350.6 -319.6 -251.0 -232.8 -265.0 -232.0 -224.3 -207.2 -242.5 -211.1 -140.1 -27.0 275.6 -839.3 -320.8 -245.2 -221.3 2040.31 Delrin -487.6 -356.8 16.2 -994.4 -689.9 -634.5 -599.8 Air -1009 -994.1 -994.6 PMP -190 -689.9 -689.8 LDPE -99 -636.3 -632.7 Polystyrene -41 -598.4 -601.1 2168.06 2404.22 1767.40 2079.51 2027.21 1952.82 Table 3-8: Pixel values measured in the CIRS phantom ILUMA iCAT GAL Material Toshiba CT # P1 P2 Avr P3 P4 P5 P6 Avr P7 Bone 240 280.4 283.2 84.8 151.0 112.5 152.9 1053 1096.3 1102.1 707.1 730.2 665.2 735.7 Adipose -69 -103.4 -107.5 -61.3 -105.4 -131.1 -104.1 Air -1008 -995.8 -999.3 -923.9 -911.4 -901.3 -914.9 Muscle 55 35.6 36.8 8.9 -15.3 -46.4 -15.6 Brain 45 -4.5 -9.2 -22.8 -17.3 -42.8 -14.7 125.3 709.6 -100.4 -912.9 -17.1 -24.4 2389.3 Dense Bone 281.8 1099.2 -105.5 -997.5 36.2 -6.8 45 3016.8 2355.6 1655.4 2367.6 2355.1 Figure 3-4 shows the pixel values as measured in the i-CAT images compared to their values as measured in the CT images. A linear fit was used with the data points. The pixel values in the CIRS prototype were the most linear with an R2 value of 0.99. The pixel values of the ACR phantom and CATPHAN were slightly less linear with R2 values of 0.95 and 0.97, respectively. i-CAT vs CT: Pixel Value Linearity Pixel Value in i-CAT images 1500 y = 1.0219x - 2.592 R² = 0.9968 y = 0.6064x - 491.77 R² = 0.9558 1000 500 0 -1500 -1000 -500 0 500 1000 1500 -500 y = 0.5267x - 537.97 R² = 0.9728 -1000 -1500 CT number by Toshiba Acquilion 16 ACR phanotm CATPHAN CIRS prototype Figure 3-4: pixel values measured in the i-CAT images vs the CT number measured in CT images Figure 3-5 shows the pixel values in the ILUMA images as compared to their values in the CT images for the three phantoms. The data points from the three phantoms were fit to a linear equation. The data points from the ILUMA images of the CIRS prototype phantom and ACR phantom were fairly linear with an R2 value of 0.99. The CATPHAN pixel values, however, showed the least linearity with an R2 value of 0.97. Figure 3-6 shows the pixel values in the GALILEOS images plotted against the real CT numbers of the materials measured in the CT images. The GALILEOS system 46 uses all positive pixel value scale. The pixel values of the CIRS prototype phantom were linear ( R2=0.98). The CATPHAN showed the least linearity of the pixel values (R2=0.90). ILUMA vs CT: Pixel Value Linearity 1000 y = 0.7878x - 78.152 R² = 0.996 y = 0.8137x - 142.08 R² = 0.9722 0 -1500 -1000 -500 0 500 -500 1000 1500 y = 0.5742x - 219.83 R² = 0.9925 -1000 -1500 CT number from Toshiba Aquilion 16 ACR phanotm CATPHAN CIRS Prototype Figure 3-5: pixel values in the ILUMA images vs CT number (Hounsfield units) in CT GALILEOS vs CT: Pixel Value Linearityy = 0.6483x + 2322.5 R² = 0.9845 3500 3000 PV from GALILEOS PV in ILUMA 500 y = 0.562x + 2065.2 R² = 0.9547 2500 2000 y = 0.3175x + 2061.2 R² = 0.9057 1500 1000 500 0 -1500 -1000 -500 0 500 1000 1500 CT number from Toshiba ACR phantom Catphan CIRS Prototype Figure 3-6: pixel values in the GALILEOS images vs CT number (Hounsfield units) in CT 47 The pixel value accuracy of the images can be evaluated by using the y-intercept values from the previous graphs and are summarized in Table 3-9. Ideally the y-intercept would equal zero and the larger the deviation from zero the more inaccurate the pixel values. Table 3-9: the values of the y-intercept in the three phantoms Scanner ACR CAT CIRS i-CAT -491 -537 -2.5 ILUMA -142 -219 -78 GALILEOS 2065 2061 2322 The y-intercept of the CIRS phantom for the i-CAT and ILUMA was the closest to zero indicating a better pixel accuracy in this phantom. The y-intercepts in the ACR phantom and CATPHAN significantly deviates from zero thus indicating poorer pixel value accuracy. The pixel value accuracy cannot be evaluated in the GALILEOS as it does not use a calibrated Hounsfield scale. The better accuracy observed in the CIRS phantom is likely due to the reduced beam hardening effect compared to the ACR phantom and CATPHAN. 3.4 Contrast Scale (CS) The contrast scale measured in each phantom can be taken as the slope from Figures 3-4, 3-5 and 3-6 and are summarized in Table 3-10. Figure 3-7 depicts the difference in contrast scale between the phantoms. 48 Table 3-10: Contrast scale: the slope of the fitted lines in the pixel linearity graphs i-CAT ILUMA GALELEOS ACR phantom 0.61 0.81 0.56 Catphan 0.53 0.57 0.32 CIRS prototype 1.02 0.80 0.65 Contrast scale 1.2 1.02 1 CS (the slope) 0.81 0.8 0.8 0.61 0.6 0.57 0.53 0.65 0.56 0.32 0.4 0.2 0 i-CAT ACR phantom ILUMA Catphan GALELEOS CIRS prototype Figure 3-7: the contrast scale of the three phantoms In the i-CAT images of the CIRS prototype the contrast scale (CS) is 1.02 which Figure 4-10: the contrast scale of the three phantoms indicates contrast sensitivity similar to that found in CT images. On the contrary, the CS of the i-CAT images on the CATPHAN is 0.53 which means a change in the pixel value Figure 4-10: the contrast scale of the three phantoms in the conventional CT image would correspond significantly less change in the pixel value measured in the CBCT image. The reduced contrast scale in the ACR phantom and Figure 4-10: the contrast scale of the three phantoms the CATPHAN is due to difference in phantom diameter from that of the CIRS phantom. The larger diameter causes more beam hardening and increased noise levels. 49 In general, the CS in dental CBCT is less that optimal and the images from the GALILEOS showed the worst CS among the three machines. The low contrast scale would cause reduced low contrast detectability. 3.5 High contrast spatial resolution The high contrast test objects in the CIRS Prototype, made of epoxy resin, did not show enough contrast to be visualized in all of the images from the three scanners. The high contrast test objects in the ACR phantom are located towards the edge of the phantom which made it difficult to be assessed, especially in the GALILEOS images due to the smaller FOV. Figure 3-8 show the smallest line pairs resolved in each phantom for the i-CAT, ILUMA and GALILEOS images. 10 9 line pair/cm 8 7 6 5 4 3 2 1 0 iCAT iCAT ILUMA ILUMA ILUMA 1 2 3 4 5 ACR phantom CATPHAN ILUMA GALILEOS GALILEOS 6 7 CIRS phantom Figure 3-8: the high contrast spatial resolution of the three phantoms. 50 8 Diameter of dental CBCT QA phantom 3.6 Calculation of water-equivalent diameter (Dw) for human dental and maxillofacial region Table 3-11 summarizes the calculation of Dw using equation 1-1. The minimum Dw calculated was 156 mm while the maximum was 192 mm. The average Dw was 173± 12 mm. Thus a water cylinder with diameter of 173±12 mm will have the same attenuation properties as the dental and maxillofacial scans that were selected. Table 3-11: calculation of water-equivalent diameter of the dental and maxillofacial regions from CT images Scan subject 1 subject 2 subject 3 subject 4 subject 5 subject 6 subject 7 subject 8 subject 9 Subject 10 AROI (mm2) 24056 26973 28890 32510 26028 31815 29938 29908 27894 27985 Mean CT number -200 -126 -215 -171 -268 -91 -171 -156 -223 -123 Dw (mm) 156 173 170 185 156 192 178 179 166 177 173 ±12 The calculation of the diameter of acrylic that would have similar photon attenuation properties as the CT images chosen can be calculated using equation 1-2. Both the Dw value (173 cm) and CT number of acrylic (120 HU) was plugged in equation 1-2 which resulted in an acrylic-equivalent diameter of 164 cm. 51 Chapter 4 Discussion The results demonstrate that the phantom diameter has a significant impact on image quality evaluation. In particular, measurement of the image noise, image uniformity, and pixel value accuracy are all significantly impacted by the phantom diameter. The presence of phantom material outside the image FOV (referred to as βexomassβ) causes additional attenuation and beam hardening such that the evaluation of these image quality parameters is compromised. Additionally, the larger diameter of the ACR and CATPHAN phantoms often caused test objects to be at the very periphery of the image FOV which complicated the assessment (e.g., the line pair objects for the ACR phantom). All of these factors lead to the conclusion that the ACR and CATPHAN phantoms are not ideal for routine assessment of dental CBCT since their diameter is larger than the typical dental CBCT image FOV. On the other hand, the CIRS prototype had its own limitations in that the line pair test objects did not contain enough contrast for adequate evaluation of limiting spatial resolution. Additionally, the relatively small diameter of the phantom did not allow for assessment of image uniformity across a significant portion of the image FOV. Therefore, design modifications are certainly necessary to make it a viable option for routine QA. In comparison to conventional CT images the dental CBCT images possess higher image noise, higher non-uniformities, larger pixel value inaccuracies, and reduced 52 contrast scale. All of these factors contribute to an overall reduced low contrast detectability. Additionally, there is a much larger variation in measured image quality parameters from manufacturer to manufacturer compared to that of conventional CT. with that being said, the ability to define a standardized set of image quality performance criteria, irrespective of dental CBCT scanner manufacturer, may not be possible. Instead, baseline data may need to be established for each manufacturerβs scanner and thus used as performance criteria. The inadequacies of the phantoms that were tested prompted an investigation into the optimal diameter of a phantom to be used for dental CBCT. An optimal phantom would possess attenuation characteristics comparable to the human head as well as fit within the typical FOV for a dental CBCT scanner. Using the methodology established in AAPM Task Group Report 220; the equivalent diameter to achieve attenuation comparable to the human head was found to be 173 mm for water and 164 mm for acrylic. Given a standard FOV for a dental CBCT scanner to be 16 to 17 cm it is recommended that an optimal phantom be approximately 16 cm in diameter and 13 cm to 15 cm in height. This concurs with the dimensions utilized in the commercially available SEDENTEXCT and QUART phantoms. The reduced contrast scale and relatively high noise levels in dental CBCT caused the low contrast detectability objects in each of the tested phantoms to be undetectable. Also, as discussed above, the line pair test objects to assess the spatial resolution in the CIRS phantom were not discernable (they are quoted as 150 HU different from background). Therefore, it is recommended that the test objects for both low contrast 53 detectability and high contrast spatial resolution possess enough attenuation difference to allow for an adequate assessment of these image quality parameters. . 54 Chapter 5 Conclusion 1. The ACR and CATPHAN phantoms have limited capability for use as a routine quality assurance phantom. This is principally due to their relatively larger diameter (20 cm for each phantom) and the presence of a portion of the phantom outside the FOV for the dental CBCT units tested. This caused beam hardening and excessive attenuation that compromised evaluation of several of the image quality parameters. 2. The CIRS prototype was also limited in its functionality as a dental CBCT phantom due to some design flaws. The primary problems involved the inability to evaluate the line pair objects for limiting spatial resolution due to their low contrast. In addition, the low contrast detectability objects were not discerned and the relatively small diameter meant that image uniformity was assessed over a relatively small FOV. 3. Based on the Dw for dental and maxillofacial region of the head and the typical diameter of the FOV for CBCT scanners, an optimal phantom diameter is between 15 to 17 cm and the height is 13 to 15 cm. 4. Due to reduced contrast scale and relatively high levels of image noise the low contrast detectability and high contrast resolution test objects must contain adequate contrast from that of background. In particular, the low 55 contrast test objects used in conventional CT scanners are inadequate for the assessment in dental CBCT and must be on the order of 10% attenuation difference (100HU) from that of background material. 56 References AAPM Task Group 220, (2014). Use of Water Equivalent Diameter for Calculating Patient Size and Size-Specific Dose Estimates (SSDE) in CT. AAPM reports. College Park, MD: AAPM. Araki, K., Patil, S., Endo, A. and Okano, T. (2013). Dose indices in dental cone beam CT and correlation with doseβarea product. Dentomaxillofacial Radiology, 42(5), p.20120362. Baba, R., Ueda, K. and Okabe, M. (2004). Using a flat-panel detector in high resolution cone beam CT for dental imaging. Dentomaxillofacial Radiology, 33(5), pp.285290. Bamba, J., Araki, K., Endo, A. and Okano, T. (2013). Image quality assessment of three cone beam CT machines using the SEDENTEXCT CT phantom. Dentomaxillofacial Radiology, 42(8), p.20120445. Brüllmann, D. and Schulze, R. (2015). Spatial resolution in CBCT machines for dental/maxillofacial applicationsβwhat do we know today?. Dentomaxillofacial Radiology, 44(1), p.20140204. Bryant, J., Drage, N. and Richmond, S. (2008). Study of the scan uniformity from an iCAT cone beam computed tomography dental imaging system. Dentomaxillofacial Radiology, 37(7), pp.365-374. Bushberg, J. (2002). The essential physics of medical imaging. Philadelphia: Lippincott Williams & Wilkins. Fda.gov, (2015). Dental Cone-beam Computed Tomography. [online] Available at: http://www.fda.gov/RadiationEmittingProducts/RadiationEmittingProductsandProcedures/MedicalImaging/Me dicalX-Rays/ucm315011.htm [Accessed 22 Mar. 2015]. Imaging Sciences International Inc., (2006). i-CAT Operator's Manual:Cone Beam Volumetric Tomography and Panoramic Dental Imaging System. Hatfield, PA: Imaging Sciences International Inc. 57 Ludlow, J., Davies-Ludlow, L., Brooks, S. and Howerton, W. (2006). Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CAT. Dentomaxillofacial Radiology, 35(4), pp.219-226. Pauwels, R., Araki, K., Siewerdsen, J. and Thongvigitmanee, S. (2015). Technical aspects of dental CBCT: state of the art. Dentomaxillofacial Radiology, 44(1), p.20140224. Pauwels, R., Jacobs, R., Singer, S. and Mupparapu, M. (2015). CBCT-based bone quality assessment: are Hounsfield units applicable?. Dentomaxillofacial Radiology, 44(1), p.20140238. Pauwels, R., Stamatakis, H., Manousaridis, G., Walker, A., Michielsen, K., Bosmans, H., Bogaerts, R., Jacobs, R., Horner, K. and Tsiklakis, K. (2011). Development and applicability of a quality control phantom for dental cone-beam CT. J Appl Clin Med Phys, 12(4), p.3478. Pauwels, R., Theodorakou, C., Walker, A., Bosmans, H., Jacobs, R., Horner, K. and Bogaerts, R. (2012). Dose distribution for dental cone beam CT and its implication for defining a dose index. Dentomaxillofacial Radiology, 41(7), pp.583-593. Reiser, M. (2009). Multislice CT. Berlin: Springer. Scarfe, W. and Farman, A. (2008). What is Cone-Beam CT and How Does it Work?. Dental Clinics of North America, 52(4), pp.707-730. Scarfe, W. and Farman, A. (2008). What is Cone-Beam CT and How Does it Work?. Dental Clinics of North America, 52(4), pp.707-730. Scarfe, W., Li, Z., Aboelmaaty, W., Scott, S. and Farman, A. (2012). Maxillofacial cone beam computed tomography: essence, elements and steps to interpretation. Australian Dental Journal, 57, pp.46-60. Schulze, R., Heil, U., GroΞ², D., Bruellmann, D., Dranischnikow, E., Schwanecke, U. and Schoemer, E. (2011). Artefacts in CBCT: a review. Dentomaxillofacial Radiology, 40(5), pp.265-273. SEDENTEXCT Guideline Development Panel, (2012). Cone Beam CT for Dental and Maxillofacial Radiology ( Evidence Based Guidelines). Radiation Protection series. Luxembourg: European Commission: Directorate-General for Energy. 58 Sedentexct.eu, (2015). Technical Description of CBCT | SEDENTEXCT. [online] Available at: http://www.sedentexct.eu/content/technical-description-cbct [Accessed 21 Mar. 2015]. Suomalainen, A., Kiljunen, T., Käser, Y., Peltola, J. and Kortesniemi, M. (2009). Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners. Dentomaxillofacial Radiology, 38(6), pp.367-378. Torgersen, G., Hol, C., Møystad, A., Hellén-Halme, K. and Nilsson, M. (2014). A phantom for simplified image quality control of dental cone beam computed tomography units. Oral Surgery, Oral Medicine, Oral Pathology and Oral Radiology, 118(5), pp.603-611. 59