3D Treatment Planning and Intensity-Modulated Radiation
Therapy
Review Article [1] | October 01, 1999
By James A. Purdy, PhD [2]
Three-dimensional (3D) image-based treatment planning and new delivery technologies have
spurred the implementation of external beam radiation therapy techniques, in which the high-dose
region is conformed much
Introduction
hree-dimensional radiation therapy treatment planning (3DRTTP) systems have spurred the
implementation of external beam radiation therapy techniques, in which the high-dose region can be
conformed much more closely to the cancer patient’s target volume than was previously possible,
thus reducing the volume of normal tissues receiving a high dose. This form of external beam
irradiation is referred to as three-dimensional conformal radiation therapy (3DCRT), and its
development and clinical use are discussed in considerable detail elsewhere.[1,2] Several groups
have already reported on their early clinical experience with 3DCRT for the treatment of prostate
cancer, and the results are encouraging.[3-5]
Three-dimensional planning is not just an addition to the current radiation oncology planning
process, but rather represents a radical change in practice, particularly for the radiation oncologist.
The two-dimensional (2D) treatment planning approach emphasizes the use of a conventional
simulator for designing beam portals, based on standardized beam arrangement techniques applied
to whole classes of comparable patients. Three-dimensional treatment planning emphasizes a virtual
simulation, image-based approach for objectively defining tumor and critical structure volumes for
the individual patient.
The use of the terms 2D and 3D as descriptors for the planning process has caused some confusion
in the radiation oncology community. One should recognize that planning the cancer patient’s
treatment is, and always has been, a 3D problem, and 2D planning refers to the process and tools
used. Three-dimensional treatment planning does not require the use of “noncoplanar” beams, a
common misconception. Noncoplanar beams have been used for years in selected sites such as
breast cancer, even though 2D treatment planning systems could not accurately account for the
geometry. Clearly, noncoplanar breast tangential fields represented on a 2DRTTP system by a single,
or at most a few, slices cannot be considered a 3D treatment plan. Radiation oncologists will be able
to transition to 3D planning much more easily if they approach 3DRTTP as a new treatment planning
process, emphasizing image-based target volume design rather than as a reflection of a particular
beam configuration.
One of the important factors contributing to the current 3D process is the standardization of
nomenclature, published in the International Commission on Radiation Units and Measurements
(ICRU) Report 50.[6] This report has provided a language and a way of thinking about the problem
for defining the volume of known tumor, suspected microscopic spread, and marginal volumes
necessary to account for set-up variations and organ and patient motion. Figure 1 illustrates the
ICRU 50 formal definitions for the gross tumor volume, clinical target volume, planning target
volume, and organs at risk. Some brief discussion on the use of the ICRU 50 methodology is
presented in a later section. Also, the reader is referred to the literature for more details regarding
the use of ICRU 50 nomenclature.[7,8]
Table 1 lists the various tasks that make up the 3D planning dose and delivery process in its current
technology state. 3DCRT treatment plans generally use an increased number of radiation beams that
are angled and shaped to conform to the planning target volume using the 3DRTTP system’s
beam’s-eye view and room’s-eye view displays (Figures 2A and 2B). To improve the conformity of
the dose distribution, conventional beam modifiers (eg, wedges or compensating filters) are
sometimes used. This form of 3DCRT must now be referred to as “traditional” or “conventional”
T
Page 1 of 13
3D Treatment Planning and Intensity-Modulated Radiation The
Published on Cancer Network (http://www.cancernetwork.com)
3DCRT, because a more advanced form of 3DCRT, called intensity-modulated radiation therapy, is
already emerging. Intensity-modulated radiation therapy can achieve even greater conformity by
optimally modulating the radiation beam intensity (fluence) throughout each treatment field.[9]
These advances in radiation oncology technology are truly exciting and are occurring at a very rapid
pace. However, as the implementation of 3D planning and dose delivery systems becomes more
widespread, the radiation oncology team must understand that ensuring the safety and accuracy of
this new modality is more difficult (in its current development state) than with the standard 2D
process. Therefore, it is essential that the radiation oncology team stay well informed on the new
technologies and maintain a strong commitment to a rigorous quality assurance program when this
new treatment modality is implemented in the clinic. This article will update a previous review of
advances in 3DRTTP and dose delivery.[10]
Advances In 3D Treatment Planning
Three-dimensional RTTP systems are currently going through a rapid transition period, having
emerged from university-developed systems to Food and Drug Administration-approved
commercially available systems in the early 1990s. These first-generation commercial systems
provide specific functionality, such as virtual simulation or 3D external beam dose planning. Such
limited functionality has resulted in clinics needing multiple treatment planning systems to provide
for various high-tech treatment modalities, like virtual simulation, 3D external beam dose planning,
high-dose-rate brachytherapy, prostate seed implant, and stereotactic radiosurgery. This is clearly
not very satisfactory, resulting in increased costs and inefficiencies.
Over the next few years, the treatment planning system manufacturers will begin to combine these
planning programs into a single integrated image-based 3DRTTP system that provides planning
capability for multiple treatment modalities, resulting in considerable cost savings. In addition, we
should soon see improved efficiencies based on the use of “intranets.” An intranet is a special type
of internet-related technology with the potential to dramatically improve a radiation oncology
clinic’s ability to use 3DRTTP. A radiation oncology clinic’s intranet will make use of client server
(and browser) technology that will provide easy-to-use Windows-like displays, permitting users to
easily navigate through multiple treatment planning and other clinical applications, data, and
graphical displays using point-and-click technology. However, prior to further discussion on
information technology, it is important to describe some of the improved features now readily
available on 3DRTTP systems.
Imaging for 3D Planning
The development of powerful but relatively inexpensive computer systems has allowed the
integration of computed tomography (CT)–based target volume and normal tissue definition with the
process of radiation therapy treatment planning, creating an entirely new device, the CT simulator.
The CT simulator differs in both purpose and function from the conventional simulator.[11,12] A
typical CT simulation uses a helical CT scanner with localization software, a patient registration and
laser marking system, and a dedicated computer workstation for virtual simulation and generation of
digitally reconstructed radiographs.
Although the current generation of CT simulators provides powerful new 3D capabilities, several
improvements in the hardware are still needed for treatment planning purposes because the CT
scanner currently used was designed for diagnostic radiology use. Specifically, a larger CT gantry
aperture is needed. Current scanners have a 70-cm diameter, which limits some treatment set-ups.
Also, the CT reconstruction size needs to be larger (currently only a 52-cm diameter is provided). CT
simulator couch tops that mimic the geometry of a medical linear accelerator are also needed, as
are improved registration systems for going from the real patient to the virtual patient and then back
to the real patient for treatment. Other issues that continue to be of concern are data storage and
data transfer. Also, a satisfactory process for quantitating the effects of respiratory organ motion or
other organ movements has not yet been developed for CT simulation.
For effective use of CT simulation, the radiation oncologist must become familiar with CT
cross-sectional anatomy and with the appropriate use of CT contrast materials that aid in the
identification of the gross tumor volume and organs at risk. When CT simulation/3DRTTP is first
introduced in the clinic, assistance from a diagnostic radiologist in contouring the organs at risk and
gross tumor volume/clinical target volumes can be invaluable. However, identifying normal tissues
and tumor on a treatment-planning CT can be difficult even for an experienced diagnostic
radiologist, as the planning CT images are typically acquired in a nonstandard diagnostic patient
Page 2 of 13
3D Treatment Planning and Intensity-Modulated Radiation The
Published on Cancer Network (http://www.cancernetwork.com)
position. Since 3D planning requires the definition of target volumes and normal tissue on a CT scan,
strong consideration should be given to incorporating image-based, cross-sectional anatomy training
into radiation oncology residency training programs. In the interim, short courses on image-based
anatomy are needed for practicing radiation oncologists.
While CT is the principal source of image data for 3DRTTP, there is a growing demand to incorporate
the complementary information available from magnetic resonance imaging (MRI). The sharply
demonstrated tumor–soft-tissue interface often seen on an MRI scan in tumors such as in the brain
can be used to better define the gross tumor volume.[13] Several groups have also demonstrated
the value of MRI in distinguishing the prostate gland from surrounding normal structures.[14,15] In
addition, functional imaging modalities such as single photon emission computed tomography
(SPECT) and positron emission tomography (PET) will likely prove to be important in both the target
definition phase of treatment planning and also in the follow-up studies needed to assess efficacy.
For example, Marks et al have reported on the use of SPECT lung perfusion scans to determine
functioning regions of the lung.[16,17] The functional lung volume data are used in calculating
dose-volume histograms rather than the CT-defined anatomy and are referred to as functional
dose-volume histograms. Radiation beams are planned that minimize irradiation of these functioning
areas. Similarly, PET imaging is used to aid in defining the patient’s lung cancer gross tumor
volume/clinical target volumes. However, use of multimodality imaging in the treatment planning
process brings with it the need for better image registration tools.
Image Registration
Spatial registration (fusion) of 3D images from different modalities, most commonly CT, MRI, or
PET/SPECT, is a major area of research for radiology and radiation oncology.[18-20] The approaches
used to register two or more data sets can be divided into manual and automatic, or some
combination. Frequently, approximate alignment is achieved through the use of manually interactive
software followed by final alignment by an automatic algorithm. Currently, only a few 3DRTTP
systems provide an image fusion capability in their software. Of the many automatic methods, those
using statistical criteria to optimize the correlation of anatomic features of the images are especially
promising. They do not require edge detection and need minimal or no data preconditioning. The
approach maximizing the measure of “mutual information”[21,22] and the popular and related
Woods method[23] are examples of promising research in this area. In addition to automated image
registration methods such as these, a general method for formally testing the accuracy of image
registration is also needed.
Image Segmentation
The process of segmenting the cancer patient’s planning CT image data set (ie, contouring the gross
tumor volume, clinical target volume, planning target volume, specified organs at risk, and skin) is
one of the most important but, unfortunately, most time-consuming elements of the 3D treatment
planning process. In our clinic, a prostate cancer patient’s CT data set (typically about 80 slices) can
be segmented in about 30 minutes; the bladder, rectum, and femoral heads are contoured by the
dosimetrist, and the prostate gland is contoured by the radiation oncologist. For other sites, the time
requirement is even greater, and thus, there is still a need to develop software that will automate (or
at least nearly automate) the contouring task.
When contouring the gross tumor volume, it is important to use the appropriate CT window and level
settings to determine the maximum dimension of what is considered to be potential gross disease.
At this time, defining the clinical target volume is more of an art than a science, since current
imaging techniques are not capable of detecting subclinical tumor involvement directly. Thus, the
boundary of the clinical target volume must be defined by the radiation oncologist based on clinical
experience. The planning target volume margin must also be specified by the radiation oncologist, in
consultation with the radiation oncology physicist, based on clinical experience. Unfortunately, there
are limited data for internal organ motion and set-up error for most sites other than prostate.[24]
When defining the planning target volume, the radiation oncologist should recognize that the margin
around the clinical target volume need not be uniform but should account for the asymmetrical
nature of positional uncertainties.
Many 3DRTTP systems still do not possess accurate methods for providing a true 3D margin around
the gross tumor volume (Figure 3). Typically, the margin expansion is drawn or specified in 2D
around the gross tumor volume contour to get the planning target volume. For large contour
differences in neighboring slices, this will yield margins that are too small in the cranial-caudal
direction. Bedford and Shentall describe methods to compute target margins.[25]
When defining the beam aperture, additional margin beyond the planning target volume is needed
to obtain dose coverage because of beam penumbra. Some physicians (and physicists) have
Page 3 of 13
3D Treatment Planning and Intensity-Modulated Radiation The
Published on Cancer Network (http://www.cancernetwork.com)
confused this margin in the past by thinking of the planning target volume boundary as the required
beam edge, which is incorrect. For coaxial beam arrangements, the additional margin is typically 7
to 8 mm, except for the inferior and superior borders. In those directions, 12 to 15 mm margins are
required because of the beam divergence effects to have planning target volume coverage by the
prescription minimum isodose treatment volume.
3D Treatment Planning Displays
Three-dimensional RTTP systems typically display the delineated tumor/target volume and organs at
risk cross-sectional contours as surface-rendered objects using a variety of styles that include
transparency, texturing, and lighting (Figure 2). Some planning systems are beginning to use
volume-rendered objects (Figure 4).[26] Volume rendering provides more soft-tissue structure and
should aid the physician in defining the patient’s clinical target volume(s). However, the
volume-rendered displays will not alleviate the need for contours, as they will still be needed for
calculation of dose-volume histograms.
Near real-time beam’s-eye view and room’s-eye view displays are now standard features for most
commercially available 3DRTTP systems. The beam’s-eye view display provides the planner with an
effective means of choosing beam directions and designing beam apertures to maximize planning
target volume coverage and minimize irradiation of any organs at risk. On most systems, customized
beam apertures are shaped manually around the planning target volume using a computer mouse,
analogous to the manual design of blocks using a wax pencil on a simulation radiograph. More
advanced systems automatically shape treatment apertures to conform to the shape of the planning
target volume projection based on a specified margin. The room's-eye view display, using
transparent renderings of the planning target volume and organs at risk, allows quick graphical
positioning of the treatment isocenter. The room’s-eye view display is also helpful in evaluating the
geometry of multiple beam arrangements. Particularly when noncoplanar beam arrangements are
used, care must be taken to avoid the selection of a beam direction that is untreatable because of
table and gantry collisions or other treatment room restrictions. A needed feature, currently missing
on 3DRTTP systems, would be an automatic collision avoidance feature to prevent lost time and
effort in creating a treatment plan that is undeliverable.
High-quality digitally reconstructed radiographs and rapid computation speed are also essential
features of a modern 3DRTTP system. The digitally reconstructed radiograph serves the same
purpose as a conventional simulator radiograph. However, the target volume and organs-at-risk
contours can be overlaid on the digitally reconstructed radiograph and thus provide a more
quantitative tool to appreciate the geometric relationship of these volumes and the portal aperture.
Advanced 3DRTTP systems provide a full range of digitally reconstructed radiograph image
manipulation features controlling image display and geometric parameters. Particular aspects of the
image presentation can be enhanced by changing the window and level settings, as well as the
contrast and brightness of the pixel display. The CT numbers can be grouped into ranges
corresponding to bone, fat, muscle, or tissues; the CT numbers in each category can be modified by
a weighting factor and redisplayed to provide enhancement or suppression of a structure based on
attenuation values.[27] An example of this type of digitally reconstructed radiograph, referred to as
a digital composite radiograph, is shown in Figure 5.
Dose Calculations
Early generation 3DRTTP systems used dose calculation algorithms similar to those used in 2DRTTP
systems and thus provided improved dose visualization, but not improved accuracy.[8,28] However,
more advanced dose-calculation algorithms that compute the dose from more of a first-principles
approach and thus provide a more accurate means of predicting dose have now become the
standard for 3DRTTP systems. These algorithms use convolution energy deposition kernels that
describe the distribution of dose about a single primary photon interaction site.[29] The convolution
kernels are obtained by using the Monte Carlo technique to interact monoenergetic primary photons
at the origin in a phantom, and to transport the charged particles and scattered and secondary
photons that are set in motion. The energy that gets deposited about the primary photon interaction
site is tabulated and stored for use in the convolution method. In addition to describing how
scattered photons contribute to dose absorbed at some distance away from the interaction site of
primary photons, the convolution kernels take into account charged particle transport.
In the future, the direct use of Monte Carlo simulation for 3D dose planning and evaluation should be
feasible. Groups at the Lawrence Livermore National Laboratory, the National Research Council of
Canada, and several universities are actively pursuing development of such capability.[30-32] This
method is the only one capable of computing the dose distribution accurately for complex
geometries that include interfaces of materials with very dissimilar atomic numbers, such as near
Page 4 of 13
3D Treatment Planning and Intensity-Modulated Radiation The
Published on Cancer Network (http://www.cancernetwork.com)
metal prostheses used for femoral head replacement. The major obstacles preventing this technique
from clinical implementation have been the lengthy computation times required and the need for a
detailed quantitative description of the incident radiation beam.
The first obstacle is rapidly being overcome as the computation power of microprocessors continues
to increase dramatically (doubling about every 18 to 24 months). In addition, researchers are
actively working on reducing the computation time by using clever variance reduction techniques
and the macro Monte Carlo technique.[33,34] Rogers and coworkers are actively working on
overcoming the second obstacle with the development of the Monte Carlo code called “BEAM,”
which simulates the radiation beams from medical accelerators.[30]
A practical point should be made here regarding dose calculation programs as they pertain to dose
prescription and dose distribution evaluation. In the 3D era, because of better visualization tools,
there has been a tendency to emphasize the minimum isodose line or isodose surface covering the
planning target volume for prescription and reporting purposes. This is unfortunate, because dose at
the periphery of the planning target volume can be associated with significant uncertainties due to
the steep dose gradient close to the edge of the beam. Radiation oncologists should understand that
the dose along the central axis of a beam is likely to be more accurate than at the periphery
regardless of the dose calculation algorithm used. In fact, ICRU Report 50 points out that the dose at
the central part of the planning target volume is the most representative single dose value for the
planning target volume and recommends that it, along with the maximum and minimum dose to the
planning target volume, be used for reporting.[6]
Interactive Treatment Plan Optimization
The optimization (more accurately termed plan improvement) of a conventional 3DCRT plan relies on
an iterative, interactive plan optimization approach. Typically, the initial beam arrangement is
selected based primarily on clinical experience using beam’s-eye view and room’s-eye view
displays. The arrangement is then modified based on the evaluation of the dose distribution using
dose-volume histograms, 2D isodose sections, and room’s-eye view 3D isodose volume displays
until the resulting dose distribution is approved by the radiation oncologist. The planned dose
distribution is most often a uniform dose to the target volume (eg, ≤ 7% variation from maximum to
minimum dose at a specified dose level), with doses to critical organs at risk held below some
tolerance level that has been specified by the radiation oncologist.
Automated Computer Optimization
Interest in treatment planning computer optimization and automation has been renewed with the
development of 3DCRT and the corresponding increase in image and graphic data available to the
planner. Investigation of computer optimization has also intensified because of intensity-modulated
radiation therapy’s requirement that optimal nonuniform beam fluences be determined.[35-37] The
task of automated computer optimization can be better understood by considering it as two
components: (1) specification of an optimization criteria, and (2) the optimization (search) algorithm
used. The optimization criteria is expressed mathematically as an objective function whose value
(sometimes referred to as score or cost) denotes the goodness of the treatment plan. The search
algorithm then attempts to determine the plan parameters, which minimizes the objective function
(Figure 6).
Objective functions and associated constraints are typically based on dose and dose-volume
parameters for the target volume and critical normal structures. More recently, objective functions
have been expressed in terms of biological indices, such as some weighted combination of tumor
control probabilities and normal tissue complication probabilities. The development of robust tumor
control probabilities and normal tissue complication probabilities models that accurately predict
clinical outcome will emerge from the ongoing 3DCRT clinical trials as dose-volume outcome data
are gathered and analyzed; but to date, these models should be viewed with some caution.
One of the most promising search algorithms for intensity-modulated radiation therapy is simulated
annealing.[38] This algorithm is derived from statistical mechanics, and attempts to mimic the
behavior of a system of interacting particles that are progressively cooled and allowed to maintain
thermal equilibrium while reaching the ground state. With this approach, there is no restriction on
the order (eg, quadratic) of the objective functions or on the number of constraints. In addition, it
has the potential to escape from the local minimum of the objective function, and thus find the
global minimum (ie, the best plan) (Figure 6). However, simulated annealing is generally slower than
the deterministic optimization algorithms that have been used in the past for computer optimization
and thus, from a practical standpoint, may limit the search parameters (eg, table and gantry
positions). More details of the various computer optimization methods being investigated for
intensity-modulated radiation therapy can be found in the literature.[39,40]
Page 5 of 13
3D Treatment Planning and Intensity-Modulated Radiation The
Published on Cancer Network (http://www.cancernetwork.com)
Plan objective function and optimization search algorithms are an important research area for
intensity-modulated radiation therapy with no clear winner established as yet. As computer power
increases, the optimization search algorithm is unlikely to pose the real problem. Rather, the more
difficult problem will probably be the development of a robust means to specify the optimization
criteria that will be applicable to the diverse population of cancer patients and also satisfy individual
physician preferences. Until such an optimization criteria specification is developed, inverse planning
will continue to resemble the iterative, forward plan-improvement approach, in which a physician still
must evaluate dose distributions (dose-volume histograms and room's-eye view 3D isodose volumes)
and make iterative changes to the objective function input parameters in order to obtain an
acceptable plan.
Electronic Data Exchange
The amount of information that must be processed in this new 3D era is orders of magnitude greater
than in the previous 2D era, and thus requires significant changes in the radiation oncology
department’s computer and communications infrastructure. The planning and delivery of
intensity-modulated radiation therapy can only be made practical, safe, and affordable through the
increased use of computer automation and open architecture, common standards-based medical
information systems. The reader is referred to a previous issue of Seminars in Radiation Oncology, in
which a review of the use of information technology in radiation oncology was presented.[41]
An electronic patient chart and image file is an essential component of the integrated
planning-delivery-verification system that will eventually replace the stand-alone systems used
today. The new system will depend on the development of robust “intranets” and “extranets” to
access data and applications that are stored on the radiation oncology department’s servers, and
perhaps those of other clinics, healthcare organizations, and radiation oncology vendors in various
locations. The electronic records will be shared among the different groups within the department
(eg, radiation oncologists, physicists, dosimetrists, radiation therapists, treatment machine
maintenance technicians, administrators) and also with the other hospital departments as needed.
Thus, the importance of computer networks, servers, distributed databases, and browser technology
for radiation oncology must be stressed.
The Radiation Therapy Oncology Group’s (RTOG) 3D Quality Assurance Center (3DQA)
(http://rtog3dqa.wustl.edu/) is an example of how the World Wide Web can be effectively used in
radiation oncology. The 3D planning and verification digital data for patients entered on the National
Cancer Institute-sponsored 3D dose-escalation trials in prostate cancer, lung cancer, and
supratentorial glioblastoma multiforme that are now in progress are submitted to the 3DQA Center
via the internet.[42] The target volume and designated organs-at-risk contours superimposed on CT
display, verification images, and the 3D dose distribution are reviewed for protocol compliance. Once
reviewed, the data are stored in an open architecture, common standards database that is linked
with the RTOG clinical outcome database in Philadelphia.[43] We believe this new way of submitting
protocol data and performing QA review, as well as the new type of linked databases, will
revolutionize the conduct and analysis of clinical trial outcome data, allowing researchers to better
understand the relationship between doses, volumes, and patient outcomes. This effort, in turn, will
lead to the development of improved tumor control probabilities and normal tissue complication
probabilities models that will make the planning of intensity-modulated radiation therapy more
effective and efficient.
For an institution to participate in RTOG 3DCRT clinical trials, the 3DRTTP system to be used must
have the RTOG Data Exchange Specification implemented.[44] This feature allows the treatment
planning data listed in Table 2 to be submitted to the 3DQA Center for review and inclusion in the
new database. To date, over 700 protocol patient data sets have been submitted to the 3DQA Center
using this exchange format.
The RTOG Data Exchange Specification will ultimately be replaced by the radiation therapy
information objects of Digital Imaging and Communication in Medicine 3.0,[45] which themselves
were motivated in part by the RTOG standard. However, this replacement must await the
implementation of the radiation therapy information objects and their support by both commercial
3DRTTP companies and the RTOG 3DQA Center. The uncertainty of a time frame for broad
acceptance and implementation of the Digital Imaging and Communication in Medicine radiation
therapy objects, coupled with the wider commercial implementation of the RTOG Data Exchange
Specification, suggests that the RTOG data exchange will continue to play an important role in
support of 3DCRT multi-institutional trials for several more years.
The broad experience with 3DRTTP data exchange gained thus far by the 3DQA Center has created
keen awareness of the need for continued work in the area of data exchange by the 3DRTTP
Page 6 of 13
3D Treatment Planning and Intensity-Modulated Radiation The
Published on Cancer Network (http://www.cancernetwork.com)
manufacturers. Users should encourage the manufacturers of their 3DRTTP systems to implement
new features on their systems that will simplify the data submission process. For instance, a utility
could be provided that would allow the user to preselect the data required for submission specific to
the 3DCRT protocol. This utility would in turn create a button on a user interface for submitting data.
When clicked, the protocol data would automatically be selected and sent to the 3DQA Center. All
additional edit features shown in our prototype interface would be used only for exceptional
submissions, such as those correcting contour or planning problems in earlier submissions for the
same patient in order to reduce the amount of data requiring resubmission.
Advances in Treatment Delivery
In the 2D paradigm, radiation therapy is delivered with radiation beams of uniform intensity across
the field. Simple beam modifiers such as wedges and compensators are often used to account for
missing tissue and, in some instances, to shape dose distribution. While this approach works well for
simple, regularly shaped target volumes, it is severely limited for complex target shapes or for
geometries in which the planning target volume is in close proximity to several critical organs at risk.
In the 3D paradigm (intensity-modulated radiation therapy), these limitations are overcome by
employing nonuniform beam fluence distributions for a selected number of beams or via arc therapy.
The dose distribution is produced by dividing each beam into tiny subdivisions called “beamlets,”
with each beamlet’s intensity determined by a computer optimization algorithm.
The concomitant development of the multileaf collimator and the medical accelerator computer
control systems, which made possible the use of the multileaf collimator in a dynamic mode, have
spurred the development of intensity-modulated radiation therapy.[9,46,47] Now, several different
methods for intensity-modulated radiation therapy delivery are being used clinically (or are in the
process of being clinically implemented) and are briefly discussed below.
Tomotherapy Intensity-Modulated Radiation Therapy
Mackie et al first proposed an approach called “tomotherapy,” which literally means “slice therapy,”
by which intensity-modulated radiation therapy is delivered using a narrow slit beam.[48] A slit
mini-multileaf collimator system (called MIMiC) of the type proposed by Mackie et al is now
commercially available (Peacock, Nomos Corporation, Sewickley, Pa) and is implemented in a large
number of clinics worldwide. The Peacock system’s MIMiC is mounted to a conventional low-energy
megavoltage medical linear accelerator, and treatment is delivered to a narrow slice of the patient
using arc rotation (Figure 7).[49] The beam is collimated to a narrow slit (approximately 2 cm × 20
cm), and beamlets of varying intensity are created by driving the MIMiC’s leaves in and out of the
radiation beam’s path as the gantry rotates around the patient. A complete treatment is
accomplished by sequential delivery to adjoining axial slices. This system was first implemented
clinically at the Baylor College of Medicine (Houston, Tex).[50] Since then, several institutions have
reported their experience with the Peacock intensity-modulated radiation therapy system.[51-53]
Examples of a conformal dose distribution for a complex geometry that is achievable with this form
of intensity-modulated radiation therapy are shown in Figure 8.
With this approach, a significantly increased number of monitor units is required compared to
conventional 3DCRT without delivery of dose to the patient’s target volume. This results in an
increased workload for radiation shielding requirements, as well as an increased whole-body dose to
the patient, primarily due to leakage radiation. Mutic and Low, using this intensity-modulated
radiation therapy system, reported that the dose 10 cm from the target volume, due to scatter and
leakage radiation, is approximately 2.5% of the total target dose, and is still as great as 0.5% at 30
cm from the target volume.[54] Followill et al concluded that the increased whole-body dose from
intensity-modulated radiation therapy compared to conventional dose delivery poses a slightly
increased risk of secondary fatal malignancies.[55]
Treatment plan verification still depends strongly on dosimetry measurements performed using
ionization chambers, radiographic film, and/or thermoluminescent dosimeters in special
intensity-modulated radiation therapy phantoms.[52,56] At our institution, these measurements are
performed in the evening or on the weekend before each patient’s initial intensity-modulated
radiation therapy treatment session. Intensity-modulated radiation therapy delivery places
significant constraints on the design and operational characteristics of dosimetry equipment due to
the dynamic dose delivery. The dosimetry instrumentation and the entire QA process used for
intensity-modulated radiation therapy are early in their development and will continue to undergo
reevaluation.
A new type of treatment unit designed to deliver tomotherapy is under development at the
Page 7 of 13
3D Treatment Planning and Intensity-Modulated Radiation The
Published on Cancer Network (http://www.cancernetwork.com)
University of Wisconsin (Madison).[57] Figure 9 shows the treatment unit as originally proposed by
Mackie et al, in which intensity-modulated radiation therapy would be delivered as the patient is
moved through a ring-gantry in much the same way as a spiral CT study is performed.[48]
Specifically, the beamlets are created using a mini-multileaf collimator similar to MIMiC and a
low-energy linear accelerator mounted in a modified CT scanner gantry. The original proposal also
included a conventional diagnostic CT system mounted on the same gantry to allow the
simultaneous acquisition of a CT verification scan study.
More recent thinking suggests that megavoltage CT imaging may be all that is needed.[57] This
approach eliminates the field abutment problem inherent in the axial tomotherapy approach and
provides for on-line treatment verification.
Conventional Multileaf Collimator Intensity-Modulated Radiation Therapy
A conventional multileaf collimator, used in a dynamic mode, provides a full-field approach to
intensity-modulated radiation therapy, compared to the slit-field approach of tomotherapy. For a
fixed gantry position, the gap formed by each pair of opposing multileaf collimator leaves is swept
across the target volume under computer control with the radiation beam on to produce the desired
fluence profiles (Figure 10). The setting of the gap opening and its speed for each multileaf
collimator leaf pair are determined by a technique first introduced by Convery and Rosenbloom[58]
and extended by Bortfeld et al[59] and Spirou and Chui.[60] This intensity-modulated radiation
therapy approach, referred to as dynamic multileaf collimator or “sliding window,” was first
implemented for clinical use at the Memorial Sloan-Kettering Cancer Center in New York.[61] The
various steps of treatment planning and delivery for this form of intensity-modulated radiation
therapy and the associated QA programs are discussed in the article by Burman et al.[37]
Another multileaf collimator approach uses a sequence of static multileaf collimator fields (typically
called “segments” or “subfields”) that are set up and delivered at each orientation of the gantry
under computer control (Figure 11). This method is sometimes referred to as “step-and-shoot,” or
“stop-and-shoot,” and the leaf sequences can be determined by the method suggested by Bortfeld
et al.[62] In this approach, the linear accelerator is switched on and off for a predefined number of
monitor units for each subfield automatically without human intervention. This allows a completely
automatic treatment delivery of all subfields of all beams of a treatment plan, including gantry
rotation. Most medical linear accelerator manufacturers now offer this type of computer control
feature. Thus, widespread implementation of this form of intensity-modulated radiation therapy is
anticipated over the next several years.
A third multileaf collimator approach, called intensity-modulated arc therapy, has been suggested by
Yu and coworkers.[63-65] In this case, multiple superimposing arcs are used with the beam aperture
changing shape during gantry rotation under computer control, with the radiation beam on to
produce the desired fluence profiles. The cumulative fluence distribution of all arcs generates the
desired dose distribution.
X-ray-Compensating Filter Intensity-Modulated Radiation Therapy
A physical compensation filter approach can also be used to deliver intensity-modulated radiation
therapy. Filters designed using 3DRTTP are capable of compensation for not only the missing tissue
deficit, but also for internal tissue heterogeneities. Filters can be designed by calculating a thickness
along a ray line, using an effective attenuation coefficient for the filter material and dose-ratio
parameters for effective depths that generate the desired intensity-modulated radiation therapy
fluence profile when the filter is placed in the radiation beam. Stein et al are investigating the use of
physical modulators fabricated using low-melting-point alloy poured into foam molds, which are cut
using a computer controlled cutter.[66] In addition, Dubal et al recently described a compensating
filter approach for intensity-modulated radiation therapy for the treatment of breast cancer.[67]
The compensator method for intensity-modulated radiation therapy suffers from the cumbersome
and time-consuming manufacturing process and from the need to enter the treatment room to
change the filters, thus increasing the time allocated for patient treatment. It is likely that this
approach will serve only as an intermediate step to large-scale implementation of multileaf
collimator intensity-modulated radiation therapy treatments.
Conclusion
The use of 3DRTTP and 3DCRT/intensity-modulated radiation therapy dose delivery is advancing
radiation oncology by providing the opportunity for both more conformal dose distributions and more
complete and thorough safety systems. Although much of the current 3DCRT process requires
interactive tasks (some still very laborious), the path is clear toward overcoming the technological
Page 8 of 13
3D Treatment Planning and Intensity-Modulated Radiation The
Published on Cancer Network (http://www.cancernetwork.com)
obstacles so that a nearly automated planning, delivery, and verification system will become a
reality over the next decade. This system will allow radiation oncologists to significantly increase
dose to many tumor sites while concomitantly lowering doses to organs at risk. Most of the tasks will
be automated, thus lowering the overall costs currently needed to provide high-quality external
beam radiation therapy.
These new technologies present a new set of QA challenges. The introduction of 3D conformal
radiation therapy/intensity-modulated radiation therapy techniques must be supported by strong QA
programs. As many of these new technologies are not fully developed, there may be increased
potential for patient treatment error. However, as long as there are rigorous QA procedures in place,
a commitment to adequate staffing and continuing education, and the implementation of
complementary safety and QA devices, such as a record-and-verify system, there should be no
increase in errors. In fact, minimizing the human steps needed to treat a cancer patient with
radiation will likely lead to a reduction in errors.
References:
1. Purdy JA, Emami B (eds): 3-D Radiation Treatment Planning and Conformal Therapy. Madison, Wis,
Medical Physics Publishing, 1995.
2. Meyer JL, Purdy JA (eds): 3-D Conformal Radiotherapy: A New Era in the Irradiation of Cancer.
Basel, Switzerland, Karger. 1996.
3. Hanks GE, Hanlon AL, Schultheiss TE, et al: Dose escalation with 3D conformal treatment:
Five-year outcomes, treatment optimization, and future directions. Int J Radiat Oncol Biol Phys
41:501-510, 1998.
4. Sandler HM, McLaughlin PW, Ten Haken RY, et al: Three-dimensional conformal radiotherapy for
the treatment of prostate cancer: Low risk of chronic rectal morbidity observed in a large series of
patients. Int J Radiat Oncol Biol Phys 33:797-801, 1995.
5. Zelefsky MJ, Leibel SA, Gaudin PB, et al: Dose escalation with three-dimensional conformal
radiation therapy affects the outcome in prostate cancer. Int J Radiat Oncol Biol Phys 41:491-500,
1998.
6. International Commission on Radiation Units and Measurements. Report No. 50: Prescribing,
Recording, and Reporting Photon Beam Therapy. Bethesda, Md, International Commission on
Radiation Units and Measurements, 1993.
7. Purdy JA: Volume and dose specification, treatment evaluation, and reporting for 3D conformal
radiation therapy, in Palta J, Mackie TR (eds): Teletherapy: Present and Future, pp 235-251. College
Park, Md, Advanced Medical Publishing, 1996.
8. Purdy JA: Three-dimensional physics and treatment planning, in Perez CA, Brady LW (eds):
Principles and Practice of Radiation Oncology, pp 343-369. Philadelphia, Lippincott-Raven, 1997.
9. Sternick ES (ed): The Theory and Practice of Intensity-Modulated Radiation Therapy. Madison, WI,
Advanced Medical Publishing, 1997.
10. Purdy JA: Advances in three-dimensional treatment planning and conformal dose delivery. Semin
Oncol 24:655-671, 1997.
11. Coia LR, Schultheiss TE, Hanks G (eds): A Practical Guide to CT Simulation. Madison, Wis,
Advanced Medical Publishing, 1995.
12. Perez CA, Purdy JA, Harms WB, et al: Three-dimensional treatment planning and conformal
radiation therapy: Preliminary evaluation. Radiother Oncol 36:32-43, 1995.
13. Austin-Seymour M, Chen GT, Rosenman J, et al: Tumor and target delineation: Current research
and future challenges. Int J Radiat Oncol Biol Phys 33:1041-1052, 1995.
Page 9 of 13
3D Treatment Planning and Intensity-Modulated Radiation The
Published on Cancer Network (http://www.cancernetwork.com)
14. Lau HY, Kagawa K, Lee WR, et al: Short communication: CT-MRI image fusion for 3D conformal
prostate radiotherapy: Use in patients with altered pelvic anatomy. Br J Radiol 69:1165-1170, 1996.
15. Roach M III, Faillace-Akazawa P, Malfatti C, et al: Prostate volumes defined by magnetic
resonance imaging and computerized tomographic scans for three-dimensional conformal
radiotherapy. Int J Radiat Oncol Biol Phys 35:1011-1018, 1996.
16. Marks LB, Spencer DP, Sherouse GW, et al: The role of three-dimensional functional lung imaging
in radiation treatment planning: The functional dose-volume histogram. Int J Radiat Oncol Biol Phys
33:65-75, 1995.
17. Marks LB, Munley MT, Spencer DP, et al: Quantification of radiation-induced regional lung injury
with perfusion imaging. Int J Radiat Oncol Biol Phys 38:399-409, 1997.
18. Kessler ML, Pitluck S, Petti P, et al: Integration of multimodality imaging data for radiotherapy
treatment planning. Int J Radiat Oncol Biol Phys 21:1653-1667, 1991.
19. Sailer S, Rosenman JG, Soltys S, et al: Improving treatment planning accuracy through
multimodality imaging. Int J Radiat Oncol Biol Phys 35:117-124, 1996.
20. Rosenman JG, Miller EP, Tracton G, et al: Image registration: An essential part of radiation
therapy treatment planning. Int J Radiat Oncol Biol Phys 40:197-205, 1998.
21. Meyer CR, Boes JL, Kim B, et al: Demonstration of accuracy and clinical versatility of mutual
information for automatic multimodality image fusion using affine and thin-plate spline warped
geometric deformations. Med Image Anal 1:195-206, 1997.
22. Wells WM III, Viola P, Atsumi H, et al: Multi-modal volume registration by maximization of mutual
information. Med Image Anal 1:35-51, 1996.
23. Woods RP, Mazziotta JC, Cherry SR: MRI-PET registration with automated algorithm. J Comput
Assist Tomogr 17:536-546, 1993.
24. Balter J, Ten Haken R, Lam K: Immobilization and set-up verification, in Palta J, Mackie TR (eds):
Teletherapy: Present and Future, pp 471-494. College Park, Md, Advanced Medical Publishing, 1996.
25. Bedford JL, Shentall GS: A digital method for computing target margins in radiotherapy. Med Phys
25:224-231, 1998.
26. Pelizzari CA, Grzeszczuk R, Cheng GTY, et al: Volumetric visualization of anatomy for treatment
planning. Int J Radiat Oncol Biol Phys 34:205-211, 1996.
27. Drebin R, Carpenter L, Hanrahan P: Volume rendering. Comp Graph 22:65-74, 1988.
28. Purdy JA: Photon dose calculations for three-dimensional radiation treatment planning. Semin
Radiat Oncol 2:235-245, 1992.
29. Mackie TR, Reckwerdt P, McNutt T, et al: Photon beam dose computations, in Palta J, Mackie TR
(eds): Teletherapy: Present and Future, pp 103-136. College Park, Md, Advanced Medical Publishing,
1996.
30. Rogers DW, Faddegon BA, Ding GX, et al: BEAM: A Monte Carlo code to simulate radiotherapy
treatment units. Med Phys 22:503-524, 1995.
31. DeMareo JJ, Solberg TD, Smathers JB: A CT-based Monte Carlo simulation tool for dosimetry
planning and analysis. Med Phys 25:1-11, 1998.
32. Hartmann Siantar CL, Bergstrom PM, Chandler WP, et al: Lawrence Livermore National
Page 10 of 13
3D Treatment Planning and Intensity-Modulated Radiation The
Published on Cancer Network (http://www.cancernetwork.com)
Laboratory’s PEREGRINE Project, in Leavitt DD, Starkschall G (eds): XII International Conference on
the Use of Computers in Radiation Therapy. pp 19-22. Salt Lake City, Utah, Medical Physics
Publishing, 1997.
33. Holmes MA, Mackie TR, Sohn W, et al: The application of correlated sampling to the computation
of electron beam dose distributions in heterogeneous phantoms using the Monte Carlo method. Phys
Med Biol 38:675-688, 1993.
34. Neuenschwander H, Mackie TR, Reckwerdt PJ: MMC—A high-performance Monte Carlo code for
electron beam treatment planning. Phys Med Biol 40:543-574, 1995.
35. Gustafsson A, Lind BK, Svensson R, et al: Simultaneous optimization of dynamic multileaf
collimation and scanning patterns or compensation filters using a generalized pencil beam algorithm.
Med Phys 22:1141-1156, 1995.
36. Bortfeld T, Burkelbach J, Boesecke R, et al: Three-dimensional solution of the inverse problem in
conformation radiotherapy, in Breit A (ed): Advanced Radiation Therapy: Tumor Response Monitoring
and Treatment Planning, pp 503-508. Berlin, Springer-Verlag, 1992.
37. Burman C, Chui CS, Kutcher G, et al: Planning, delivery, and quality assurance of
intensity-modulated radiotherapy using dynamic multileaf collimator: A strategy for large-scale
implementation for the treatment of carcinoma of the prostate. Int J Radiat Oncol Biol Phys
39:863-873, 1997.
38. Webb S: Inverse planning for IMRT: The role of simulated annealing, in Sternick ES (ed): The
Theory and Practice of Intensity Modulated Radiation Therapy, pp 51-73. Madison, Wis, Advanced
Medical Publishing, 1997.
39. Mohan R: Intensity modulation in radiotherapy, in Palta J, Mackie TR (eds): Teletherapy: Present
and Future, pp 761-792. College Park, Md, Advanced Medical Publishing, 1996.
40. Rosen II: Treatment planning for IMRT, in Sternick ES (ed): The Theory and Practice of Intensity
Modulated Radiation Therapy, pp 37-49. Madison, Wis, Advanced Medical Publishing, 1997.
41. Purdy JA (ed): Data management in radiation oncology. Semin Radiat Oncol 7(suppl 1):1-94,
1997.
42. Purdy JA, Harms WB, Michalski J, et al: Multi-institutional clinical trials: 3-D conformal
radiotherapy quality assurance, in Meyer JL, Purdy JA (eds): 3-D Conformal Radiotherapy: A New Era
in the Irradiation of Cancer, pp 255-263. Basel, Switzerland, Karger, 1996.
43. Bosch WR, Lakanen TL, Kahn MG, et al: An image/clinical database for multi-institutional clinical
trials in 3D conformal radiation therapy, in Leavitt DD, Starkschall G (eds): XII International
Conference on the Use of Computers in Radiation Therapy, pp 455-457. Salt Lake City, Medical
Physics Publishing, 1997.
44. Harms WB Sr, Bosch WR, Purdy JA: An interim digital data exchange standard for
multi-institutional 3D conforinal radiation therapy trials, in Leavitt DD, Starkschall G (eds): XII
International Conference on the Use of Computers in Radiation Therapy, pp 465-468. Salt Lake City,
Medical Physics Publishing, 1997.
45. National Electrical Manufacturers Association (NEMA): Digital imaging and communication in
medicine (DICOM). Radiotherapy Objects, suppl 11. Washington, DC, NEMA, 1987.
46. Boyer AL: Present and future developments in radiotherapy treatment units. Semin Radiat Oncol
5:146-155, 1995.
47. Fraass BA, MeShan CL, Kessler ML: Computer-controlled treatment delivery. Semin Radiat Oncol
5:77-85, 1995.
Page 11 of 13
3D Treatment Planning and Intensity-Modulated Radiation The
Published on Cancer Network (http://www.cancernetwork.com)
48. Mackie TR, Holmes T, Swerdloff S, et al: Tomotherapy: A new concept for the delivery of dynamic
conformal radiotherapy. Med Phys 20:1709-1719, 1993.
49. Carol MP: Integrated 3D conformal planning/multivane intensity modulating delivery system for
radiotherapy, in Purdy JA, Emami B (eds): 3D Radiation Treatment Planning and Conformal Therapy,
pp 435-445. Madison, Wis, Medical Physics Publishing, 1995.
50. Grant W: Experience with intensity modulated beam delivery, in Palta J, Mackie TR (eds):
Teletherapy: Present and Future, pp 793-804. College Park, Md, Advanced Medical Publishing, 1996.
51. Low DA, Mutic S, Dempsey JF, et al: Quantitative dosimetric verification of an IMRT planning and
delivery system. Radiother Oncol 49:305-316, 1998.
52. Verellen D, Linthout N, van den Berge D, et al: Initial experience with intensity-modulated
conformal radiation therapy for treatment of the head and neck region. Int J Radiat Oncol Biol Phys
39:99-114, 1997.
53. Tsai JS, Wazer DE, Ling MN, et al: Dosimetric verification of the dynamic intensity-modulated
radiation therapy of 92 patients. Int J Radiat Oncol Biol Phys 40:1213-1230, 1998.
54. Mutic S, Low DA: Whole-body dose from tomotherapy delivery. Int J Radiat Oncol Biol Phys
42:229-232, 1998.
55. Followill D, Geis P, Boyer A: Estimates of whole-body dose equivalent produced by beam intensity
modulated conformal therapy. Int J Radiat Oncol Biol Phys 38:667-672, 1997.
56. Low DA, Mutic S, Gerber RL, et al: The quantitative verification of computer-controlled
intensity-modulated radiation therapy dose planning and delivery—Initial experience, in Leavitt DD,
Starkschall G (eds): XII International Conference on the Use of Computers in Radiation Therapy, pp
410-413. Salt Lake City, Medical Physics Publishing, 1997.
57. Mackie TR: Tomotherapy, in Leavitt DD, Starkschall G (eds): XII International Conference on the
Use of Computers in Radiation Therapy, pp 9-12. Salt Lake City, Medical Physics Publishing, 1997.
58. Convery DJ, Rosenbloom M: The generation of intensity-modulated fields for conformal
radiotherapy by dynamic collimation. Phys Med Biol 37:1359-1374, 1992.
59. Bortfeld T, Boyer AL, Schlegel W, et al: Realization and verification of three-dimensional
conformal radiotherapy with modulated fields. Int J Radiat Oncol Biol Phys 30:899-908, 1994.
60. Spirou SV, Chui CS: Generation of arbitrary intensity profiles by dynamic jaws or multileaf
collimators. Med Phys 21:1031-1041, 1994.
61. Ling CC, Burman C, Chui CS, et al: Conformal radiation treatment of prostate cancer using
inversely-planned intensity-modulated photon beams produced with dynamic multileaf collimation.
Int J Radiat Oncol Biol Phys 35:721-730, 1996.
62. Bortfeld T, Kahler DL, Waldron TJ, et al: X-ray field compensation with multileaf collimators. Int J
Radiat Oncol Biol Phys 28:723-730, 1994.
63. Yu CX: Intensity-modulated arc therapy with dynamic multileaf collimation: An alternative to
tomotherapy. Phys Med Biol 40:1435-1449, 1995.
64. Yu C, Symons MJ, Du MN, et al: A method for implementing dynamic photon beam intensity
modulation using independent jaws and a multileaf collimator. Phys Med Biol 40:769-787, 1995.
65. Yu CX. Intensity modulated arc therapy: A new method for delivering conformal radiation
therapy, in Stemick ES (ed): The Theory and Practice of Intensity Modulated Radiation Therapy, pp
Page 12 of 13
3D Treatment Planning and Intensity-Modulated Radiation The
Published on Cancer Network (http://www.cancernetwork.com)
107-120. Madison, Wis, Advanced Medical Publishing, 1997.
66. Stein J, Hartwig K, Levegrün S, et al: Intensity-modulated treatments: Compensators vs. multileaf
modulation, in Leavitt DD, Starkschall G (eds): XII International Conference on the Use of Computers
in Radiation Therapy, pp 338-341. Salt Lake City, Medical Physics Publishing, 1997.
67. Dubal N, Chang S, Cullip T, et al: Intensity modulation for tangential breast treatment (abstract).
Int J Radiat Oncol Biol Phys 42:127, 1998.