Survey
* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project
* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project
Proceedings of the ASME 2013 Summer Heat Transfer Conference HT2013 July 14-19, 2013, Minneapolis, MN, USA HT2013-17441 Experimental Investigation of Flow Boiling Performance of Open Microchannels with Uniform and Tapered Manifolds (OMM) Ankit Kalani, Satish G. Kandlikar* Department of Mechanical Engineering Rochester Institute of Technology Rochester, NY, USA *[email protected] ABSTRACT Boiling can provide orders of magnitude higher cooling performance than a traditional air cooled system especially related to electronics cooling application. It can dissipate large quantities of heat while maintaining a low surface temperature difference. Flow boiling with microchannels has shown a lot of potential due to its high surface area to volume ratio and latent heat removal. Flow instabilities and early critical heat flux have however prevented its successful implementation. A novel flow boiling design is experimentally investigated to overcome the above mentioned disadvantages while presenting a very low pressure drop. The design uses open microchannels with a tapered manifold (OMM) to provide stable and efficient operation. Distilled, degassed water at atmospheric pressure was used as the fluid medium. Effect of tapered block with varied dimension is investigated. Pressure drop data for uniform and tapered manifold for plain and microchannel chip are presented. A maximum heat flux of 281 W/cm2 at 10 °C wall superheat was recorded with microchannel and tapered manifold without reaching CHF. The maximum pressure drop obtained for the above mentioned configuration was only 3.3 kPa. 1. INTRODUCTION Single phase cooling has been the dominant mode of heat transfer for the last several decades due to its low cost and reliable operation. Miniaturization of electronics and increase in chip power densities has generated a need for high heat flux dissipation. Boiling has the ability to dissipate large quantities of heat due to its latent effects. Cooling with microchannels has shown a lot of potential since the pioneering work of Tuckerman and Pease [1]. Colgan et al. [2] used enhanced microchannels in a single phase study to dissipate heat fluxes of over 1 kW/cm2. They were limited by the high chip temperature and large pumping power requirement. The state of research of flow boiling in microchannels has been well reviewed by many researchers [3–5]. The current literature review was focused on the heat transfer and pressure drop performance of flow boiling in microchannels. For two-phase flow, Kandlikar [6] had pointed out that flow instability, low heat transfer coefficient and flow maldistribution were some of the key issues for the poor performance of the flow boiling system. Various authors have used different techniques to prevent high pressure drop fluctuation in the system. Kandlikar et al. [7] used artificial nucleation sites and inlet restrictors to provide stable flow boiling system. Wang et al. [8] studied flow boiling instability using three different inlet/outlet configurations. Low fluctuations were obtained for the configuration which had inlet restrictions, while no restrictions were placed on the exit side. Wu and Cheng [9] and Lee et al. [10] used trapezoidal cross section, parallel microchannels in their flow boiling study. Lu and Pan [11] achieved stability in their microchannel system using diverging, parallel microchannel as proposed by Mukherjee and Kandlikar [12] with artificial nucleation sites. Design with evenly distributed cavities along the channel 1 Copyright © 2013 by ASME showed the best performance. Hetsroni et al. [13] used parallel triangular microchannels with varied experimental parameters. Zhang et al. [14] extensively studied the Ledinegg instability in microchannels. They concluded that presence of inlet restrictors, increase in the system pressure and the channel diameter, reduction in the number of channels and the channel length lead to a more stable flow in the microchannels. Balasubramanian et al. [15] used straight and expanding microchannel in their flow boiling study. The authors observed lower pressure drop and wall temperature fluctuation with expanding microchannel geometry. Cho et al. [16] and Megahed [17] used cross-linked microchannel in their experimental work. Recently, Sitar et al. [18] used square parallel microchannel of 25 × 25 µm and 50 × 50 µm cross section with FC-72 and water. The authors used a combination of inlet/outlet restrictors, inlet/outlet manifolds and fabricated cavities to limit the instabilities. The authors observed reduction in the onset of nucleate boiling temperature and an even flow distribution. High heat flux testing has also been undertaken by various researchers. Qu and Mudawar [19] tested a microchannel heat sink with 21 parallel channels and obtained a maximum heat flux of 130 W/cm2. Kuo and Peles [20] used 200 µm × 253 µm parallel microchannel with structured reentrant cavities. Mass flux was varied from 83 kg/m2s to 303 kg/m2s. The authors concluded that lower boiling incipience and increased CHF were observed with structured reentrant cavities. Heat fluxes of up to 643 W/cm2 at 80 °C wall superheat were recorded with 303 kg/m2s mass flux. Liu and Garimella [21] experimentally investigated flow boiling in microchannels with inlet water temperatures of 67 – 95 °C, and mass fluxes of 221 – 1283 kg/m2s. The authors obtained a heat flux of 129 W/cm2 and at exit quality of 0.2. Recently, Kandlikar et al. [22] and Kandlikar [23] used the open microchannel with tapered manifold configuration (OMM) to simultaneously increase the CHF and the heat transfer coefficient. They and obtained a heat flux of 506 W/cm2 at a wall superheat of 26.2°C without reaching CHF. Preliminary work with tapered manifold showed low pressure drop and high heat transfer performance. In the current work, tapered microchannel configuration is further investigated and its effect on pressure drop is studied. Plain and microchannel chips are used with distilled and degassed water as the working fluid at atmospheric pressure. Three tapered manifolds having the same inlet manifold height (0.127 mm) and gradually increasing exit manifold height (200, 400 and 600 µm) are used and their effect on the pressure fluctuations are studied. Results are compared with the uniform manifold for both plain and microchannel chips. 2. NOMENCLATURE kCu q” x Tsat Tc thermal conductivity of copper, W/m K heat flux, W/m2 distance, m saturation temperature, K chip temperature, K Twall ΔTsat G ṁ Ac h wall temperature, K wall superheat, K mass flux, kg/m2s mass flow rate, kg/s cross-sectional area, m2 heat transfer coefficient, W/m2 K 3. EXPERIMENTAL SETUP Figure 1 shows the flow boiling test setup used in the current study. It was similar to that used in an earlier study by Kandlikar et al. [22]. The test setup consisted of a heating block, base plate, intermediate plate and a manifold block. The heater consisted of a copper block with eight 200 W cartridge heaters. The top half of the heater had three equally spaced holes for thermocouple probes and the tip was 10 mm × 10 mm square section contacting the test chips. The manifold block consisted of inlet and outlet openings for the fluid flow. The base plate used was used to support the intermediate plate and the manifold block. The test section was placed in the base plate and a silicone gasket helped in sealing the system. A gasket thickness of 0.127 mm was used for all the test runs. This presents a fixed manifold height in the uniform manifold case, and the inlet manifold height in the case of tapered manifolds. A Micropump© pump was used to provide the required flow rate for the system. A flow rate of 80 mL/min was used for the current work. A supply tank with distilled water was degassed initially before supplying the water to the system. The supply tubing from the tank to the system was wrapped in fiberglass insulation to minimize the heat losses. A subcooling of 10°C was employed at the water inlet into the test section. Figure 1. Schematic of the flow boiling test setup [22]. The intermediate plate consisted of tapered and uniform manifolds. The tapered manifold allowed a tapered gap as seen in Fig. 2 above the microchannel while the uniform manifold had no recess in the intermediate plate. The tapered manifold was designed to allow a larger flow cross-sectional area on the exit side of the fluid path. Three different tapered manifolds were used having the same inlet height (0.127 mm) and gradually increasing manifold heights to 200 µm, 400 µm and 600 µm on the exit side. Polysulfone was used for the manifold block and the intermediate plate. 2 Copyright © 2013 by ASME Figure.3 Schematic of the microchannel copper chip. x Twall = Tc − q"(k 1 ) Figure 2. Schematic of the tapered manifold [22]. Heat flux was calculated using the three thermocouples inserted in the copper block using the one-dimensional heat conduction equation. dT q" = −k Cu (1) dx Three-point backward Taylor’s series approximation was used to calculate the temperature gradient dT/dx. dT dx = 3T1 −4T2 +T3 2∆x (2) A differential Omega® pressure sensor was used for the pressure drop reading. An NI cDaq-9172 data acquisition system with NI-9213 temperature module and NI-9205 pressure module was used to record the temperature and pressure respectively. A LabVIEW® virtual instrument (VI) was used to display and record temperature, pressure and heat flux. 4. TEST SECTION A copper chip with an overall dimension of 20 mm × 20 mm × 3 mm was used as the test section. Only the central 10 mm × 10 mm area of the chip was exposed to the fluid through an opening in the silicone gasket. A square 2 mm × 2 mm groove, as shown in Fig. 3, on the underside of the chip was provided to reduce the heat spreading effect. A thermocouple hole was provided on the side of the test section, to measure the actual chip temperature. A microchannel chip of 450 µm depth and with 181 µm wide channels and 195 µm wide fins was also tested. CNC machines were used to make the microchannels on the copper chip. The wall temperature at the top of the chip surface was calculated using the heat flux obtained from the measured chip temperature Tc and the temperature drop in the copper substrate over the distance x1, which is the distance between the thermocouple location and the top surface. (3) Cu The maximum heat flux tested was limited to 300 W/cm2 due to the heater design. CHF was not reached in any of the testing. 5. RESULTS Both, uniform and tapered manifolds were tested with the plain and microchannel chips and the results for heat transfer and pressure drop are presented in this section. Heat flux was calculated using Eq. 1 for a heater area of 100 mm2. The pressure data was obtained using the differential pressure sensor. For all the test runs, the flow rate was kept constant at 80 mL/min, and a gasket of thickness 0.127 mm was used to provide a fixed height at the inlet manifold. The effect of a uniform height manifold and three tapered manifold is also discussed. Table 1. Manifold configuration and mass fluxes at inlet and outlet. Manifold Taper height (exit) Inlet height (µm) Exit height (µm) Ginlet (kg/m2s) Goutlet (kg/m2s) Uniform 0 127 127 372 372 Taper A 200 127 327 372 238 Taper B 400 127 527 372 175 Taper C 600 127 727 372 138 The configurations for the uniform and the tapered manifolds used in the current system are shown in Table. 1. The inlet height remains constant for all test runs, while the exit height changed depending on the type of manifold used. Both the inlet and exit heights were referenced from the top plane of the microchannel to the manifold cover. The mass flux for inlet and outlet are calculated using the following equation. 3 Copyright © 2013 by ASME ṁ (4) 𝐴𝑐 where the cross-sectional area Ac, is the actual flow area at a given section calculated as the sum of the microchannel and gap height provided by the gasket at that section. The maximum exit quality observed in the current testing is below 0.1. 𝐺= 5.1 Uniform manifold testing: The uniform manifold with plain and microchannel chips were tested first so as to establish the baseline results. The uniform manifold consisted only the constant height provided by the gasket over the chip. Both the inlet and exit manifolds had a height of 0.127 mm. Figure 4 shows the boiling performance of the two chips with uniform manifold. The plain chip showed a slight boiling overshoot at a wall superheat of 17 °C. Linear increase in heat flux was observed with wall superheat. A maximum heat of 227 W/cm2 at 22 °C wall superheat was recorded for plain chip. Microchannel chip showed similar overshoot but performed significantly better than plain chip. A maximum heat flux of 283 W/cm2 at a wall superheat of 13 °C was obtained. Testing was not continued at higher heat fluxes. Figure 5. Pressure drop versus heat flux for plain and microchannel chip with uniform manifold. 5.2 Tapered manifold testing: Tapered manifold was designed to provide an additional flow area in the system along the flow direction to accommodate the vapor flow and reduce the pressure drop. Figure 6 compares the boiling performance for the uniform and tapered manifold C for a microchannel and a plain chip. Expectedly, the microchannel chip performed better than the plain chip. The highest heat flux tested for the microchannel chip was 281 W/cm2 at 10 °C wall superheat. For the plain chip, a maximum heat flux of 208 W/cm2 at a wall superheat of 16 °C was recorded. Boiling overshoot was not observed for both plain and microchannel chips. Figure 4. Boiling performance showing heat flux versus wall superheat for plain and microchannel chip with uniform manifold. Figure 5 shows the heat flux and its corresponding pressure drop for the same. The plain chip shows pressure fluctuations from 100 kPa at low heat fluxes to 160 kPa at high heat fluxes. The introduction of microchannel does show a reduction in pressure drop as the extra area provided by the microchannels reduces the flow resistance. At low heat fluxes, a pressure drop of 40 kPa is observed. For high fluxes, a maximum pressure drop of 60 kPa is seen near a heat flux of 250 W/cm2. The overall pressure fluctuation for the microchannel chip increased with the increase in heat flux. Figure 6. Boiling performance of plain and microchannel chips with the tapered manifold C. Figure 7 compares the pressure drop for the tapered manifold C for the microchannel and plain chip. The effect of the tapered manifold is significant in terms of pressure drop performance for both chips. The plain chip shows a maximum pressure drop of 6 kPa at a heat flux of 208 W/cm2, while the 4 Copyright © 2013 by ASME microchannel chip showed a pressure drop of around 2 kPa for a similar heat flux. The error bars shown are pressure fluctuation for a given heat flux. The solid points shown on the graph are the average values over the pressure range. At intermediate heat fluxes, small negative values of pressure drop are seen indicating a small backflow behavior. However, it is quite infrequent and insignificant and is not affecting the heat transfer performance adversely. The plain chip shows an increasing trend of pressure drop with the heat flux, while the microchannel chip shows only a slight increase with the increasing heat flux. Figure 8. Boiling performance of plain chip with tapered and uniform manifold. Figure 7. Pressure drop performance for plain and microchannel chips with tapered manifold C. 5.3 Heat transfer and pressure drop performance with the plain chip The effect of the two manifolds on plain chip is discussed in this section. Figure 8 shows the boiling performance of tapered and uniform manifolds. The tapered manifold showed an improved performance compared to the uniform manifold. The uniform manifold showed a boiling overshoot and recorded a heat flux of 227 W/cm2 at 22 °C wall superheat. In comparison to the uniform manifold, the three tapered manifolds showed similar performances to one another. Hence, the effect of taper height itself was not significant as seen from the figure. Boiling overshoot was not observed with any of the tapered manifolds. The tapered manifold B recorded a heat flux of 255 W/cm2 at a wall superheat of 17 °C. The pressure drop performance of the plain chip with uniform and the three different tapered manifolds is shown in Fig. 9. The tapered manifolds show a significant pressure drop reduction compared to the uniform manifold. The uniform manifold showed the highest pressure drop values at both low and high heat fluxes. For the tapered manifold, the values were below 20 kPa over the entire range. The highest pressure drop was observed with the tapered manifold B at a heat flux of 255 W/cm2 of 19 kPa. Tapered manifold C showed the lowest pressure drop over the entire heat flux range. A maximum pressure drop of 6.2 kPa at 208 W/cm2 heat flux was recorded for the tapered manifold C. The introduction of a tapered manifold drastically reduced the pressure drop from 150 kPa (uniform) to 6 kPa (tapered manifold C) for similar heat flux values. Figure 9. Pressure drop performance of plain chip with uniform and tapered manifolds. 5 Copyright © 2013 by ASME 5. 4 Heat transfer and pressure drop performance with the microchannel chip Results of the microchannel chip with the uniform and the tapered manifolds are discussed in this section. Figure 10 shows the heat flux versus wall superheat plot for the two manifolds with the microchannels. Unlike the plain chip performance, the effect of varying the taper was observed in the heat transfer performance. The tapered manifold C recorded a heat flux of 281 W/cm2 at a wall superheat of 10 °C. The uniform manifold however showed a better performance than the tapered manifold A, although the testing was not continued to higher heat fluxes. CHF was not reached for any of the tests, hence showing potential for greater heat dissipation. The tapered manifold B dissipated a heat flux 225 W/cm2 at a wall superheat of 9 °C. The slope of the tapered manifold A curve suggests that at higher heat fluxes, it might perform better than the uniform manifold. Further testing of tapered manifold A is suggested and will be undertaken in future work. Figure 11. Pressure drop performance of microchannel chip with uniform and tapered manifolds. DISSCUSSIONS Figure 10. Boiling performance of microchannel chip with tapered and uniform manifold. Three tapered manifolds were tested with the microchannel chip and their pressure drop performance is shown in Fig. 11. The maximum pressure drop observed was 11 kPa for tapered manifold A at a heat flux of approximately 170 W/cm2. A maximum pressure fluctuation of around 6 kPa was observed for all three tapered manifolds. The tapered manifold C showed the lowest pressure drop of 3 kPa at a heat flux of 260 W/cm2 in comparison with the other two tapered manifolds. A maximum pressure drop of 10 kPa was observed with the tapered manifold B, and it showed lower pressure fluctuations compared to the tapered manifold A. In this section, the flow boiling performance of the uniform and the three tapered manifolds for both plain and microchannel chips is presented. The results of heat transfer and pressure drop are further analyzed. 6.1 Comparison between the microchannel and the plain chip with uniform and tapered manifold Figure 12 shows the heat transfer performance for tapered manifold C with the microchannel chip and the uniform manifold with both chips. Expectedly, microchannel chip showed significant performance improvement compared to plain chip for both manifolds. The introduction of the tapered manifold yields similar performance at the mid-range heat fluxes, but the heat flux is seen to rise for a given wall superheat at higher heat fluxes. The maximum heat flux obtained with the tapered manifold is greater than uniform manifold with the microchannel chip. Both plain and microchannel chips show a small temperature overshoot with the uniform manifold. No temperature overshoot is observed with the tapered manifold. Figure 12. Boiling performance comparison with tapered manifold with microchannel and uniform manifold with both chips. 6 Copyright © 2013 by ASME Figure 13 shows pressure drop versus the corresponding heat flux with the uniform manifold and the tapered manifold A. The highest pressure drop was observed with a uniform manifold with a plain chip. At high heat fluxes (~225 W/cm2), a pressure drop of 160 kPa was recorded with the plain chip. At a similar heat flux, the microchannel chip with a uniform manifold recorded a pressure drop of 50 kPa. The reduction in the pressure drop was mainly due to the increase in the flow cross-sectional area provided by the microchannels. The tapered manifold C showed the lowest pressure drop of ~2 kPa at a heat flux of 225W/cm2. The combination of tapered manifold with microchannel clearly showed a significant pressure drop reduction over the entire range of heat flux. The expanding cross-sectional area along the flow direction was able to accommodate the increased vapor flow and resulted in an extremely low pressure drop. The overall increase in the pressure fluctuation with increasing heat flux is also limited for the tapered manifold, hence showing a more stable flow with the tapered manifold in comparison to the uniform manifold, while simultaneously offering a better heat transfer performance in terms of higher heat flux at a given wall superheat. °C wall superheat with the microchannel chip, while a heat flux of 250 W/cm2 at 18 °C wall superheat was obtained with the plain chip. Figure 14. Comparison of heat transfer performance for plain and microchannel chip with tapered manifold (B and C). Figure 15 shows the heat transfer coefficient versus heat flux for the microchannel and the plain chips with the tapered manifolds B and C. Heat transfer coefficient is a very important parameter in comparing the thermal performance of different surfaces. Similar to the conclusion from Fig. 14, the microchannel chip performed significantly better than the plain chip for both tapered manifolds. At higher heat fluxes, a maximum heat transfer coefficient 278 kW/m2K for both tapers was recorded. A maximum heat transfer coefficient of 140 kW/m2K at 250 W/cm2 heat flux was observed for the plain chip with taper B. For taper A, similar results were obtained. Figure 13. Pressure drop performance comparison with tapered manifold C with microchannel chip and uniform manifold with both chips. 6.2 Comparison between the microchannel and the plain chip with the tapered manifold Figures 14 and 15 show the boiling performance comparison between the two chips for taper manifolds, B and C. The data for the tapered manifold A was not included in the figures so as to avoid overcrowding of data points. Higher heat dissipation at a given wall superheat is observed with the microchannel chip for both tapered manifolds as seen in Fig. 14. The plain chip shows no effect of taper height on the boiling performance. For the tapered manifold C, a heat dissipation of 282 W/cm2 was achieved at 10 Figure 15. Comparison of heat transfer coefficient for plain and microchannel chip with tapered manifolds B and C. 7 Copyright © 2013 by ASME Table 2 shows the maximum values for heat flux and heat transfer coefficient and their corresponding wall superheat and average pressure drop value for plain and microchannel chip. Taper C with microchannel chip shows the best performance in terms of pressure drop and heat transfer. For the plain chip, similar heat transfer performance is obtained for all three tapers. The tapered manifold show significant pressure drop difference compared with uniform manifold for the plain chip. CONCLUSIONS The current work involves an experimental investigation of flow boiling performance with plain chip and chips with open microchannels with the uniform and the tapered manifolds (OMM). The testing was limited to heat fluxes below about 280 W/cm2 due to heater limitations Distilled water at atmospheric pressure at a flow rate of 80 mL/min was used for all test runs. Table 2. Summary of all test runs for plain and microchannel chip including maximum heat flux, wall superheat heat transfer coefficient and average pressure drop. Chip Plain µchannel q”max ∆Tsat h ∆Pavg W/cm2 °C kW/m2K kPa Uniform Taper A Taper B 227.1 228.6 255.1 22.1 15.8 17.7 102.4 144.4 144.3 158.4 12.6 19.6 Taper C 208.3 15.6 133.5 6.3 Uniform Taper A Taper B 283.2 263.8 239.1 12.9 14.1 8.6 217.9 186.7 277.6 62.1 7.5 6.2 Taper C 281.2 10.1 277.8 3.3 Manifold 1. 2. 3. 4. The tapered manifolds with microchannel chips yield a dramatic enhancement in heat transfer performance, while providing an extremely low pressure drop value. This feature makes it particularly suited to cooling the high performance IC chips. The low pressure drop feature provides a very high coefficient of performance (ratio heat removed to pumping power) for 3D chip cooling architecture. For the tapered manifold C with microchannel chip, a heat flux of 281.2 W/cm2 is dissipated at a wall superheat of 10.1 °C with a heat transfer coefficient of 277.8 kW/m2°C. The corresponding pressure drop is only 3.3 kPa. Further performance enhancements are expected with optimizing the microchannel geometry and the taper configuration. The main mechanism responsible for reducing the pressure drop with the tapered manifolds is the increase in the flow cross-sectional area as the vapor is generated along the flow direction. As seen from Table 2, the cross-sectional area increases for tapered manifolds, and the pressure drop is corresponding lower. The results for plain chips are affected due to the presence of backflow under some heat flux conditions. The liquid flows through the microchannels promoting nucleation and is responsible for delaying the CHF. Further work on establishing the CHF limits for these configurations is suggested by redesigning the test section to deliver higher heat fluxes. This work is continuing in the authors’ lab. 5. 6. 7. 8. 8 Three tapered manifolds with a gradual increase in the gap at the exit (200 µm, 400 µm and 600 µm) and a uniform manifold of a 127 µm gap were tested with the microchannel and the plain chips in the current setup. A heat flux of 227 W/cm2 at a wall superheat of 22 °C was observed for the uniform manifold with a plain chip, while a heat flux of 283 °C at 12 °C wall superheat was recorded for the microchannel chip with same manifold The combination of microchannel chip and the tapered manifold significantly reduced the pressure drop in the system. Taper C with microchannel showed the best performance with the lowest pressure drop of 3.3 kPa compared to the 160 kPa pressure drop with the plain chip and the uniform manifold. A heat flux of 281 W/cm2 at 10 °C wall superheat with taper C was recorded with the microchannel chip. The microchannel chip with the tapered manifolds showed significant performance improvement compared to the plain chip with the uniform manifold. Similar improved performance in heat transfer coefficient for the microchannel chip with tapered manifold was observed in comparison to the plain chip with tapered manifold. A maximum heat transfer coefficient of 278 kW/m2K was recorded with microchannel chip and taper C. The testing was not conducted to the CHF limit, which was reported to be higher that 500 W/cm2 in an earlier publication. The comparison presented here showed that the microchannel chip with a taper C has the best heat transfer performance among the chips and manifolds tested. The main mechanism responsible for the dramatic reduction in pressure drop is due to the increased flow cross-sectional area to accommodate the vapor generated along the flow direction. This combines the inherent benefits of microchannels in providing a superior heat transfer performance. The open microchannels with tapered manifold (OMM) configuration is able to provide significant heat transfer coefficient and CHF enhancements and holds promise in overcoming the limitations posed by the microchannels during flow boiling. Additional testing to cover a higher exit quality and establish the CHF limits of the OMM configuration is proposed. Copyright © 2013 by ASME ACKNOWLEDGEMENTS The work was conducted in the Thermal Analysis, Microfluidics and Fuel Cell Laboratory at the Rochester Institute of Technology in Rochester, NY and supported by the National Science Foundation under Award No. CBET-123602. [14] [15] REFERENCES [1] [2] [3] [4] [5] [6] [7] [8] [9] [10] [11] [12] [13] Tuckerman D. B., and Pease R. F. W., 1981, “Highperformance heat sinking for VLSI,” Electron Device Lett. Ieee, 2(5), pp. 126–129. Colgan E. G., Furman B., Gaynes M., Graham W. S., LaBianca N. C., Magerlein J. H., Polastre R. J., Rothwell M. B., Bezama R. J., Choudhary R., Marston K. C., Toy H., Wakil J., Zitz J. A., and Schmidt R. R., 2007, “A Practical Implementation of Silicon Microchannel Coolers for High Power Chips,” Components Packag. Technol. Ieee Trans., 30(2), pp. 218–225. Thome J. R., 2004, “Boiling in microchannels: a review of experiment and theory,” Int. J. Heat Fluid Flow, 25(2), pp. 128–139. Bertsch S., Groll E., and Garimella S., 2008, “Review and Comparative Analysis of Studies on Saturated Flow Boiling in Small Channels,” Ctrc Res. Publ. Tibiriçá C. B., and Ribatski G., 2013, “Flow boiling in micro-scale channels – Synthesized literature review,” Int. J. Refrig., 36(2), pp. 301–324. Kandlikar S. G., 2002, “Fundamental issues related to flow boiling in minichannels and microchannels,” Exp. Therm. Fluid Sci., 26(2–4), pp. 389–407. Kandlikar S. G., Kuan W. K., Willistein D. A., and Borrelli J., 2006, “Stabilization of Flow Boiling in Microchannels Using Pressure Drop Elements and Fabricated Nucleation Sites,” J. Heat Transf., 128(4), pp. 389–396. Wang G., Cheng P., and Bergles A. E., 2008, “Effects of inlet/outlet configurations on flow boiling instability in parallel microchannels,” Int. J. Heat Mass Transf., 51(910), pp. 2267–2281. Wu H. Y., and Cheng P., 2004, “Boiling instability in parallel silicon microchannels at different heat flux,” Int. J. Heat Mass Transf., 47(17–18), pp. 3631–3641. Lee P. C., Tseng F. G., and Pan C., 2004, “Bubble dynamics in microchannels. Part I: single microchannel,” Int. J. Heat Mass Transf., 47(25), pp. 5575–5589. Lu C. T., and Pan C., 2009, “A highly stable microchannel heat sink for convective boiling,” J. Micromechanics Microengineering, 19(5), p. 055013. Mukherjee A., June 13–15, “Numerical study of the effect of inlet constriction on flow boiling stability in microchannels,” Toronto, Canada. Hetsroni G., Mosyak A., Pogrebnyak E., and Segal Z., 2005, “Explosive boiling of water in parallel microchannels,” Int. J. Multiph. Flow, 31(4), pp. 371–392. [16] [17] [18] [19] [20] [21] [22] [23] 9 Zhang T., Tong T., Chang J.-Y., Peles Y., Prasher R., Jensen M. K., Wen J. T., and Phelan P., 2009, “Ledinegg instability in microchannels,” Int. J. Heat Mass Transf., 52(25–26), pp. 5661–5674. Balasubramanian K., Lee P. S., Jin L. W., Chou S. K., Teo C. J., and Gao S., 2011, “Experimental investigations of flow boiling heat transfer and pressure drop in straight and expanding microchannels – A comparative study,” Int. J. Therm. Sci., 50(12), pp. 2413–2421. Cho E. S., Koo J.-M., Jiang L., Prasher R. S., Kim M. S., Santiago J. G., Kenny T. W., and Goodson K. E., 2003, “Experimental study on two-phase heat transfer in microchannel heat sinks with hotspots,” Ninteenth Annual IEEE Semiconductor Thermal Measurement and Management Symposium, 2003, pp. 242–246. Megahed A., 2011, “Experimental investigation of flow boiling characteristics in a cross-linked microchannel heat sink,” Int. J. Multiph. Flow, 37(4), pp. 380–393. Sitar A., Sedmak I., and Golobic I., 2012, “Boiling of water and FC-72 in microchannels enhanced with novel features,” Int. J. Heat Mass Transf., 55(23–24), pp. 6446– 6457. Qu W., and Mudawar I., 2003, “Measurement and prediction of pressure drop in two-phase micro-channel heat sinks,” Int. J. Heat Mass Transf., 46(15), pp. 2737– 2753. Kuo C.-J., and Peles Y., 2007, “Local measurement of flow boiling in structured surface microchannels,” Int. J. Heat Mass Transf., 50(23–24), pp. 4513–4526. Liu D., and Garimella S. V., 2007, “Flow Boiling Heat Transfer in Microchannels,” J. Heat Transf., 129(10), pp. 1321–1332. Kandlikar S. G., Widger T., Kalani A., and Mejia V., 2013, “Enhanced Flow Boiling over Open Microchannel with Uniform and Tapered Gap Manifold (OMM),” Accepted for publication in the 75th Anniversary issue of Journal of Heat Transfer, July 2013. Kandlikar, S.G., HEAT TRANSFER ENHANCEMENT THROUGH FLOW FIELD INCORPORATING A TAPER, US Patent Application, March 2013. Copyright © 2013 by ASME