* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Download Methods of Intraocular Pressure Measurement
Survey
Document related concepts
Transcript
Clinical Methods of Intraocular Pressure Measurement Dr. Vidya Anandam, MS, Aravind Eye Hospital, Madurai Transpalpebral IOP Measurement 1. Digital Palpation Palpation is the oldest method of rough IOP evaluation. Johann Zacharias Platner was the first scientist to state that the glaucomatous eye was hard. For the examination, the patient is asked to look in downgaze. The redundant skin of the upper eyelid is displaced, and the central meridian of the globe is balloted alternately with the tips of each index finger. By comparing tactile estimations of IOP to formal pressure measurements, the examiner’s sense of touch can be ‘calibrated’ to a limited extent. Bowman’s Grading System Tn = normal tension T+1, T+2, T+3 = indicates degree of increased tension T-1, T-2, T-3 = indicates degree of low tension Although palpation correlates poorly with Goldmann applanation readings, palpation may have a limited role in screening for marked elevations of IOP. Merits • Simplest, least expensive • Instrumentation not required • Useful when external tonometry is not possible, for example, after penetrating keratoplasty or corneal scarring. • Palpation may be the only feasible technique in patients who are unwilling or unable to undergo other methods of IOP measurement. Demerits • Least accurate method of IOP measurement • Palpation is best avoided in eyes with significant trauma or in certain postoperative conditions. 2. Transpalpebral Tonometers Transpalpebral tonometers, such as the TGDc-01 and IGD-02 devices. These portable instruments measure the IOP through the eyelid. The operation of both instruments is based on determining the acceleration of freely falling rod after its interaction with the elastic eye surface. Troost et al demonstrated an increasing underestimation of IOP at elevated pressure levels when compared with Goldmann applanation tonometer. The preview eye-pressure monitor (Bausch & Lomb, Rochester, NY USA) was developed as a psychophysical test for self-tonometry at home. The pencil – shaped instrument is pressed with its probe against the upper eyelid with increasing pressure until visual phenomenon are detected. These phosphenes should appear opposite to where the pressure was applied. The position of probe application can influence the measurement. Application of the probe to the superonasal part of upper lid gives the most accurate results. Merits By not applanating or indentating the cornea scarring, edema of this cornea can be prevented. Variations of the central corneal thickness did not contribute to the difference. Demerits Li et al compared IOP values using the Proview eye pressure monitor with those measured with the Goldmann applanation tonometer and with the TonoPen. The IOPs obtained with the Proview eye pressure monitor were significantly lower. Manometry Manometry is an invasive technique that precisely measures the pressure inside the eye. It is the reference pressure by which all other tonometers should be judged. Manometry is 2 used most commonly as a laboratory technique in performing continuous pressure measurements over time, evaluating the effect of physiologic and pharmacologic manipulations on pressure, and studying aqueous humor dynamics in postmortem eyes. Most widely used tonometers, such as Goldmann’s applanation tonometry, Schiotz’s indentation tonometry, Langham’s pneumotonometry and Kanngiesser’s dynamic contour tonometry have been calibrated and validated on human cadaver eyes against a manometric reference pressure. The ethical use of manometry in living human eyes is restricted to eyes undergoing enucleation or intraocular surgery. Tonometry Tonometers are the instruments for performing tonometry. Their purpose is to obtain an accurate measurement of the IOP with least disturbance to the eye. So far, cornea is the only structure of the eye that is accessible to external tonometry. Each technique has its advantages and disadvantages and none is ideal. The ideal tonometer must record accurate, reproducible measurements, without affecting the pressure or harming the eye. In addition the tonometer should be portable, simple to calibrate easy to maintain and standardize. Indentation Tonometry The shape of deformation is truncated cone. There is no precise shape and these type of tonometers displace a relatively large intraocular volume in response to a standard weight applied to the cornea. As a result conversion tables based on empirical data from in vitro and in vivo studies must be used to estimate IOP. The prototype of this group is the Schiotz tonometer (1905). Because of its simplicity, reliability and relative accuracy, it is the only mechanical indentation tonometer in use today. Other types of indentation tonomers include • Von Graefe (1962) • Donders (1863), Snellen (1968), Monnik (1868), Dor (1869) AECS Illumination • Layerat (1885), Smith (1887), Nicati (1900), Romer (1918) • McLean (1914), Many (1919), Bailliart (1923) • Recording Tonometer by Maurice (1958) • Electronic Tonometer by Mueller (1960) Schiotz Tonometer The body of the tonometer has a • Footplate which rests on cornea • Plunger which moves freely within a shaft in the footplate • Indicator needle • Scale • Additional weights 7.5g, 10g, 15g. A 5.5g weight is permanently fixed to the plunger. Basis concept of indentation tonometry: When plunger indents the cornea • Baseline pressure P0 is raised to Pt. Tonometer measures Pt. The change from P0 to Pt is an expression of resistance an eye offers to the displacement of a volume of fluid (Vc). P0 is estimated from conversion tables. • Friedenwald – developed an empirical formula for linear relationship between logarithm of pressure and volume change in a given eye. Constant used in the formula is (K) coefficient of ocular rigidity Conversion tables use - K = 0.0245 (1948 tables) or - K = 0.0215 (1955 tables). 1948 tables agree more closely with measurements by Goldmann applanation tonometry. Technique • With patient lying supine and fixing on overhead target • Under the effect of topical anesthetic drops, lids are separated by the examiner • Foot plate is placed perpendicular to cornea • Needle shows fine movements in response to ocular pulsations. Vol. XV, No.2, April - June 2015 • If scale reading is less than 4, additional weight should be added to the plunger. • A conversion table is used to derive the IOP in mm Hg from the scale reading and plunger weight. Source of Error 1. Ocular rigidity : conversion tables assume an average coefficient of ocular rigidity (K) In eyes which deviate - False value High K- False high IOP Low K - False low IOP • High ocular rigidity seen in - High hyperopia - Longstanding glaucoma - Vasoconstrictor therapy • Low ocular rigidity seen in - High myopia - Elevated IOP - Osteogenesis imperfect - Miotic therapy - Vasodilators - Retinal detachment surgery - Intravitreal gas 2. Blood volume alteration during indentation tonometry 3. Corneal influences : steeper and thicker corneas – more displacement of fluid, false high IOP 4. Moses effect : cornea moulds into the space between plunger and the hole – false elevation of IOP. Calibration Placing the tonometer on a steel test block results in a scale reading of zero. Applanation Tonometers The shape of deformation with these tonometers is a simple flattening and the shape is constant. Therefore the IOP is derived from mathematical calculations. The force necessary to flatten a small, standard area of cornea is calculated. 3 The applanation tonometers are further differentiated on the basis of the variable that is measured. Variable Force (Fixed Area) This type of tonometer measures the force that is required to applanate a standard area of the corneal surface. The prototype is the Goldmann applanation tonometer (1955). Other tonometers include; • Weber (1867) • Fick (1888) • Schmidt (1955) • Draeger (1966) 49 –51) and Perkins (1965) • Noncontact (air puff ) Tonometers • Reichert ocular response analyser (2005) • Dynamic observing tonometer (1996) Variable Area (Fixed Force) This type of applanation tonometers measure area of cornea that is flattened by a known force (weight). The prototype is the Maklakov-type (1885), the other being posner (1965). Goldmann Applanation Tonometer Goldmann applanation tonometer is based on Imbert – Fick law which states that the pressure inside an ideal dry, thin walled sphere equals the force necessary to flatten its surface divided by the area of the flattening. Where Pt = pressure within a sphere W = external force required for flattening A = total area flattened Applies to a sphere which is • Perfectly spherical • Dry • Perfectly flexible • Infinitely thin But cornea fails as it is • Aspherical • Wet 4 • Not perfectly flexible • Not thin The moisture creates a surface tension(S) and the lack of flexibility requires a force to bend the cornea (B) which is independent of the internal pressure. The outer area of flattening (A) is not the same as the inner area (A1). To account for these characteristics of the cornea, Imbert - Fick law was modified to Where A1 = internal area of applanation When A1 = 7.35mm2, S balances B and W = Pt This happens when the diameter of external area of applanation (A) = 3.06 mm. Description of Tonometer • The instrument is mounted on slit lamp in such a way that the examiner’s view is directed through the centre of the plastic biprism, which is used to applanate the cornea. The biprism is attached by a rod to a housing to adjust the force of the biprism against the cornea. • Two beam splitting prisms convert the circular area of corneal contact into two semicircles. • Prisms are arranged so that inner margins of the semicircles overlap when 3.06 mm of cornea is applanated. Technique • Biprism is placed in the holder - 180° marking aligned with white line on the holder and tension nob is set at 1g. • Cornea is anesthetized with a topical preparation. • Tear film is stained with sodium fluorescein – paper strip or fluorescein solution of 0.25% • The tonometer tip is cleaned with disinfecting solution • Cornea and biprism illuminated by a cobalt blue light from the slit lamp, the angle between the biprism and the illumination is 60 deg. • Low magnification and slit beam is opened maximally. • Patient looks straight. Lids are held against the bony orbit. AECS Illumination • Gentle contact of biprism is made with the corneal apex while observing through the slit lamp – Mono ocular view. • Two semicircles of equal size are seen. • The rings should be approximately 0.25 to 0.30 mm in thickness • The tension knob is rotated until the inner borders of the fluorescein ring touch each other at the midpoint of their pulsations. • Reading on the dial is multiplied by 10 to get the IOP in millimetres of mercury. • IOP is measured first in one eye until three successive readings are within 1 mm Hg and then measured in the other eye Methods of disinfecting Goldmann applanation The biprism should be rinsed and dried immediately after use. Soaking the tonometer head for 5 minutes in 3% hydrogen peroxide, 0.5% sodium hypochlorite or 70% isopropyl alcohol meets the guidelines published by the Centers for Disease Control and Prevention (CDC) and the American Academy of Ophthalmology (AAO). However, wiping the tip with a 70% isopropyl alcohol swab is also described to be as effective in virus elimination as disinfectant immersion. Alternatively, disposable tonometer tips or silicone shield over the Goldmann tonometer tip can be used. The goldmann applanation should be caliberated once a month. If the Goldmann tonometer is not within 0.1g of the correct calibration, the instrument must be repaired. Causes of Errors in measurement of IOP with Goldmann Applanation • Inappropriate fluorescein pattern : The fluorescein ring is too wide (overestimation) or too narrow (under estimation). • Inadequate staining causes an underestimation of IOP. • Elevation of eyes more than 15° above the horizontal causes an overestimation of IOP. Vol. XV, No.2, April - June 2015 • Widening of lid fissure excessively causes an overestimation IOP. • Repeated tonometry reduces IOP. • A scarred, irregular cornea distorts the fluorescein rings and makes it difficult to estimate the IOP. • The thickness of cornea affects. If the cornea is thick because of odema IOP is underestimated. If it is thick because of the additional tissue, IOP is overestimated. • Pressure on the globe by the examiner or lid squeezing by the patient, IOP is overestimated. • If corneal astigmatism is greater than 3D, IOP is underestimated for which the rule astigmatism, and overestimated for against the rule astigmatism. The IOP reading is inaccurate by 1mm Hg for every 3D of astigmatism. Perkins Tonometer It uses the Goldmann principle, is handheld, portable, battery – powered, and usable in the supine as well as sitting positions. This brought applanation tonometry to screening clinics, the bedside, and the operating room. The perkins applanation tonometer is particularly useful in measuring the IOP in young children, elderly and in obese patients, permitting measurements without having to position the patient at a slit lamp. Draeger Tonometer It is similar to the Goldmann and Perkins tonometers except that uses a different biprism. The force for applanation is supplied by an electric motor. Like the Perkins instrument, the Draeger tonometer is portable and counterbalanced, so it can be used in a variety of positions and locations. Noncontact Tonometer NCT was introduced by Grolman, its unique advantage being that is does not touch the cornea, other than with a puff of air. A puff of room air creates a constant force which momentarily deforms the cornea. It is postulated that the central cornea is flattened at the moment the pressure measurement is made. 5 Description of the instrument NCT machine is mounted on the table and consists of three subsystems: a. An alignment system: allows the operator to optically align the patient’s cornea in three dimensions (axial, vertical and lateral). b. An optoelectronic applanation monitoring system which consists of 1. A transmitter, which directs a collimated beam at the corneal vertex 2. A receiver and 3. Detector which accepts only parallel, coaxial rays reflected from the cornea c. A pneumatic system which generates a puff of room air, which is directed against the cornea. Schematic Representation of the working principle of NCT A: Light source from transmitter (T) is reflected from undisturbed cornea toward receiver (R), while cornea is aligned by optical system (O). B: Air puff (1) from pneumatic system (P) deforms cornea, which increases the number of light rays (2) received and detected by R. Time (t) from internal reference point to moment of maximum light detection (which presumably corresponds to applanation of the cornea) is converted to intraocular pressure (IOP) (based on calibrations with Goldmann applanation tonometer) and displayed on digital readout. C: Continued air pulse produces momentary concavity of cornea, causing sharp reduction in Right rays received by R. Technique • The patient observes an internal target, the operator aligns the cornea and superimposes a reflection of the target from the patients cornea on a stationary ring. • During this time, light from the transmitter is reflected from the undisturbed cornea which allows only a small number of rays to enter the receiver. When the cornea is properly aligned, the operator depresses a trigger which causes 6 a puff of air to be directed against the cornea, and the IOP is displayed on a digital read out. • At the moment the central cornea is flattened, the greatest number of reflected light rays are received, which is recorded as the peak intensity of light detected. The time from an internal reference point to the moment of maximum light detection is converted to IOP based on prior comparison with readings by Goldmann tonometry. In the subsequent versions air puff is automatically triggered when the alignment criteria are satisfied and the force of air to achieve peak light detection is the measured variable • The time interval for an average NCT measurement is 1 to 3 ms and is random with respect to the phase of the cardiac cycle so that the ocular pulse becomes a significant variable, it cannot be averaged as with some tonometer. Glaucomatous eyes have a significantly greater range of momentary fluctuation in IOP. Therefore it is recommended that a minimum of three readings within 3 mm Hg be taken and averaged as the IOP. Advantage of Noncontact Tonometer • The greatest benefit of NCT is the absence of mechanical contact with the cornea. • Elimination of a need for topical anaesthesia thereby avoiding drug sensitivities and toxic reactions. • No potential complications of corneal abrasions, spread of infection (epidemic keratoconjunctivitis, hepatitis B virus). • Can be performed by paramedical personnel with minimal training and in children. Limitations of Noncontact Tonometer • The main limitation is decreasing reliability in higher pressure ranges • Limited use in irregular corneas or poor visual acuity • Less portable than many other tonometers • Noncontact tonometers are rather expensive and require regular calibration. AECS Illumination Ocular response analyser The ocular response analyser (ORA) is a non contact (air puff ) tonometer that does not require topical anaesthesia and provides additional information on the biomechanical properties of the cornea. It uses an air pulse to deform the cornea into a slight concavity. The difference between the pressures at which the cornea flattens inward and outward is measured by the machine and termed corneal hysteresis (CH). The machine uses this value to correct the effects of the cornea on measurement. Applanation – Fixed Force (Variable Area) Tonometers Maklakow Applanation Tonometer IOP is estimated by measuring the area of cornea that is flattened by a known weight. Conversion tables are used as in Schiotz tonometry because of the volume displacement and therefore ocular rigidity is considered in computing the IOP. Combined indentation – Applanation Tonometers • MacKay – Marg and TonoPen • Pneumatic tonometer MacKay – Marg Tonometer The MacKay – Marg tonometer uses a microplunger connected to a sensitive transducer, which converts plunger displacement into an electrical signal that is recorded on a paper chart, much like an electrocardiogram. The MacKay – Marg tonometer may be a relatively more accurate tonometer in scarred or edematous corneas because the IOP reading is less independent of corneal rigidity and elasticity and the endpoint is reached electromechanically not optically. TonoPen The TonoPen is a handheld, battery operated version of the MacKay – Marg tonometer. The tip is covered by a disposable latex cover and applied perpendicularly to gently indent and anesthetized cornea. Each measurement requires Vol. XV, No.2, April - June 2015 several applanations. An acceptable applanation is indicated by an audible click after contact with the cornea. A microprocessor averages several acceptable waveforms and gives a digital readout of IOP on a liquid crystal display, with an estimate of the variability between the component readings. Pneumatic Tonometers (Pneumotonometer) The concept of this tonometer is similar to that of the Mackay-Marg. The sensor in this case, however, is air pressure, rather than an electronically controlled plunger. Rebound Tonometer Rebound tonometers (1997) determine intraocular pressure by bouncing a small tipped metal probe against the cornea. The device uses an induction coil to magnetise the probe and fire it against the cornea. As the probe bounces against the cornea and back into the device it creates an induction current from which the intraocular pressure is calculated. The device is simple, cheap and easy to use. It is portable, does not require the use of eye drops and is particularly suitable for children. Dynamic Contour Tonometry Dynamic contour tonometry (DCT) is a novel method which uses principles of contour matching instead of applanation. This is designed to reduce the influence of biomechanical properties of the cornea on measurement. These include corneal thickness, rigidity, curvature, and elastic 7 properties. It is less influenced by corneal thickness but more influenced by corneal curvature than the Goldmann tonometer. The PASCAL tonometer is currently the only commercial DCT tonometer available. It uses a miniature pressure sensor embedded within a tonometer tip contour matched to the shape of the cornea. The tonometer tip rests on the cornea with a constant appositional force of the one gram. When the sensor is subjected to a change in pressure, the electrical resistance is altered and the PASCAL’s computer calculates a change in pressure in concordance with the change in resistance. F = Forces ‘F’ are equal on both sides of the cornea; P = Intraocular pressure, d = diameter of contour matching area; RC = relaxed corneal radius; RC' = unrelaxed corneal radius. RC' = RC + ∆R. The contour matched tip has a concave surface of radius 10.5mm, which approximates to the shape of a normal cornea when the pressure on both sides is equal. The probe is placed adjacent to the central cornea and the integrated piezoresistive pressure sensor automatically begins to acquire data, measuring IOP 100 times per second. A complete measurement cycle requires about 8 seconds of contact time. During the measurement cycle, audio feedback is generated, which helps the clinician maintain proper contact with the cornea. The device also measures the variation in pressure that occurs with the cardiac cycle.