Download Slideset () - Journal of Medical Devices

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts
no text concepts found
Transcript
Date of download: 5/2/2017
Copyright © ASME. All rights reserved.
Ocular Blood Flow Response to Applied Force: In Vivo Tests Toward Noninvasive Intracranial Pressure
Monitoring1
J. Med. Devices. 2015;9(2):020911-020911-3. doi:10.1115/1.4030119
Figure Legend:
Blood flow velocity via transcranial Doppler (left) and transducer instrumented with axial load cell (right)
Date of download: 5/2/2017
Copyright © ASME. All rights reserved.
Ocular Blood Flow Response to Applied Force: In Vivo Tests Toward Noninvasive Intracranial Pressure
Monitoring1
J. Med. Devices. 2015;9(2):020911-020911-3. doi:10.1115/1.4030119
Figure Legend:
Inserting the lumbar catheter (left) and position of the ICP monitor in the right frontal lobe (right)
Date of download: 5/2/2017
Copyright © ASME. All rights reserved.
Ocular Blood Flow Response to Applied Force: In Vivo Tests Toward Noninvasive Intracranial Pressure
Monitoring1
J. Med. Devices. 2015;9(2):020911-020911-3. doi:10.1115/1.4030119
Figure Legend:
Sample in vivo test results at 28 mm Hg
Date of download: 5/2/2017
Copyright © ASME. All rights reserved.
Ocular Blood Flow Response to Applied Force: In Vivo Tests Toward Noninvasive Intracranial Pressure
Monitoring1
J. Med. Devices. 2015;9(2):020911-020911-3. doi:10.1115/1.4030119
Figure Legend:
RI and PI before force is applied to the cornea
Date of download: 5/2/2017
Copyright © ASME. All rights reserved.
Ocular Blood Flow Response to Applied Force: In Vivo Tests Toward Noninvasive Intracranial Pressure
Monitoring1
J. Med. Devices. 2015;9(2):020911-020911-3. doi:10.1115/1.4030119
Figure Legend:
RI percent change during applied versus ICP level
Related documents