Download CMR Image acquisition parameters

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Management of acute coronary syndrome wikipedia , lookup

Quantium Medical Cardiac Output wikipedia , lookup

Myocardial infarction wikipedia , lookup

Transcript
CMR Image acquisition parameters
Typical imaging parameters for SSFP cine imaging were: voxel size 2.0x2.0x8.0mm, TR/TE
39.6/1.12 ms, flip angle 55o, matrix 192x192; ShMOLLI (1) T1-maps are based on 5-7 images
with specific TI=100-5000ms, collected using SSFP readouts in a single breath-hold, typically:
TR/TE=201.32/1.07 ms, flip angle=35º, matrix=192x144, 107 phase encoding steps, interpolated
voxel size=0.9x0.9x8mm, cardiac delay time TD=500ms; 206ms acquisition time for single
image; STIR (2): voxel size 1.9x1.5x10.0mm, matrix=256x166, effective echo time TE=61ms,
effective repetition time TR=2 RR intervals during breath-hold, flip angle 180o, echo spacing
6.74ms, TI=170ms, dark blood thickness 200%, dark blood flip angle 180o, turbo factor 25, echo
trains per slice=7; ACUT2E TSE-SSFP (3): voxel size 1.9x1.5x8.0mm, matrix=165x256,
TR/TE=229.70/1.78ms, effective TE=98ms, flip angle 180o, T2 prep duration=24ms, segments=33, 5
shots per slice, bandwidth=781 Hz/Px); phase-sensitive inversion recovery sequence: voxel size 2.0
x 1.5 x 8.0 mm, matrix 144x256, TR/TE=800.20/3.36ms, flip angle 25o).
CMR Image Analysis
Cine images. Analysis of left ventricular ejection fraction was performed using Argus software
(Version 2002B, Siemens Medical Solutions). Semi-quantitative assessment of wall motion
abnormality was performed in MC-ROI. Segments were graded as: normal=1, hypokinetic=2,
akinetic=3, or dyskinetic=4.
Dark-blood STIR images. Quantitative analysis was performed by comparing the LV
myocardium in short axis against adjacent skeletal muscle in the same slice, verified on a
corresponding SSFP image. The T2 signal intensity (SI) ratio was calculated as T2
SImyo:skeletal=SImyocardium/SIskeletal muscle, as previously published (1). Edema is diagnosed when
myocardial T2 SI is ≥ 2.0 compared to that of skeletal muscle (4). Care was taken to exclude
non-suppressed blood pool signal due to slow-flow adjacent to the subendocardium and to avoid
using areas with abnormally low signal for normalization.
Bright-blood ACUT2E images. Quantitative analysis was performed the same way as for STIR
images on coil intensity normalized and distortion corrected images. If no suitable reference
skeletal muscle was available on a slice (i.e. affected by artifacts), then a ROI with a mean SI
and SD similar to that of skeletal muscle of the nearest neighboring slice is placed on the index
slice.
LGE. Images were evaluated qualitatively by two experienced cardiologists for the presence or
absence, pattern (subendocardial, midwall, subepicardial, transmural) and regional distribution of
LGE areas. In addition, focal areas of LGE were defined quantitatively as those with a SI ≥ 2.0
standard deviations above the mean SI of normal myocardium.
T1-maps. After T1-maps were generated, the LV myocardium was contoured and segmented as
described in the main manuscript. Average myocardial T1 values were calculated per case as the
average of the T1 values of all pixels for all slices or segments using the in-house software MC-ROI,
as previously described (5).
References
1.
Piechnik SK, Ferreira VM, Dall'Armellina E, et al. Shortened Modified Look-Locker
Inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 and 3 T within
a 9 heartbeat breathhold. J. Cardiovasc. Magn. Reson. 2010;12:69.
2.
Simonetti OP, Finn JP, White RD, Laub G, Henry DA. "Black blood" T2-weighted
inversion-recovery MR imaging of the heart. Radiology 1996;199:49-57.
3.
Aletras AH, Kellman P, Derbyshire JA, Arai AE. ACUT2E TSE-SSFP: A hybrid method
for T2-weighted imaging of edema in the heart. Magn. Reson. Med. 2008;59:229-235.
4.
Friedrich MG, Sechtem U, Schulz-Menger J, et al. Cardiovascular Magnetic Resonance
in Myocarditis: A JACC White Paper. J. Am. Coll. Cardiol. 2009;53:1475-1487.
5.
Ferreira V, Piechnik S, Dall'Armellina E, et al. Non-contrast T1-mapping detects acute
myocardial edema with high diagnostic accuracy: a comparison to T2-weighted
cardiovascular magnetic resonance. J. Cardiovasc. Magn. Reson. 2012;14:42.