Download Applicator - Research - Vanderbilt University

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Synaptic gating wikipedia , lookup

Neuropsychopharmacology wikipedia , lookup

Metastability in the brain wikipedia , lookup

Neural engineering wikipedia , lookup

Neuroregeneration wikipedia , lookup

Electrophysiology wikipedia , lookup

Multielectrode array wikipedia , lookup

Single-unit recording wikipedia , lookup

Rheobase wikipedia , lookup

Transcranial direct-current stimulation wikipedia , lookup

Evoked potential wikipedia , lookup

Functional electrical stimulation wikipedia , lookup

Neuroprosthetics wikipedia , lookup

Microneurography wikipedia , lookup

Neurostimulation wikipedia , lookup

Transcript
Non-Transcranial Electroanesthesia Device
Group 2: Ryan Demeter 1, Matt Jackson1, Caroline Schulman1, Matt Whitfield1
Advisors: James Berry, M.D. 2, Paul King, Ph.D.1;
1
Department of Biomedical Engineering, Vanderbilt University, Nashville, TN
2 Vanderbilt University Medical Center Anesthesiology Multi-Specialties Division, Nashville, TN
Transcranial Products
Problem Statement
The goal of this design project is to develop a device to control and administer non-transcranial
electroanesthesia using the vagal nerve. The device will be portable, self sustaining, and rechargeable..
The exact parameters for stimulation will be variable allowing the device to be adaptable to altered
stimulation parameters determined with any future research. The electrical pulses required for the
electroanesthesia will be generated by the computer’s soundcard and then output from the sound jack to
an external amplification circuit. This circuit will amplify the signal to the desired voltage and a set of
electrodes will deliver the signal to the surface of the skin. This project assumes that non-transcranial
electroanesthesia is possible. The device was designed based on the best understanding of both current
electroanesthesia and vagal nerve stimulation. Further research into the actual application of vagal nerve
stimulation to achieve electroanesthesia will be addressed in a later project.
Site of Application
Our Design
2 electrodes on the forehead, 2
electrodes on the back of the head
1 electrode over the right vagus, 1
electrode over the left vagus
Applicator
Multiple pad electrodes
2 pad or 2 needle electrodes
Stimulus Production
Electrical
Soundcard Output
Parameters
35 V, 4 mA, 100 Hz, rectangular pulse
Power
Rechargeable Battery
25 V, 50 uA, 20 Hz, rectangular pulse
(250 us)
Rechargeable Battery
Pricing
$1500-$3000
$1500
Budget
Box and mountings ($214.15)
• Case (Dimensions: 18" x 18" x 10") ($124.12)
• Foam Padding (VU BME Donated)
• Keyboard Hand Rest ($8.75)
• Hardware Mounting ($81.37)
Circuit and Supplies ($155.78)
• 2 Fans ($18.00)
• 3 Amplifiers ($45)
• Supplies from Radio Shack ($92.78)
Applicator ($0)
• Pad Electrode (VU BME Donated)
• Needle Electrode (VUMC borrowed)
Computer ($0, VUMC Loaned)
• 1.6 GHz
• 1 GB Hard drive
Miscellaneous ($130)
Theory Behind Electroanesthesia
Total Budget =$500
•Quicker recovery time and fewer biological effects than gas and intravenous anesthesia (Photiades, 218225).
Patent Search
•Patients heal better and are less affected by the process.
•No patents involving non-transcranial electroanesthesia.
•Less buildup of naturally produced gases in the body.
•Little neural tissue change in mammalian studies.
•A patent was found (4383522, 1983) that describes a transcranial method of delivering
electroanesthesia using electrodes placed on the neck and the forehead.
•Little change in potassium and sodium concentration levels in extracellular and intracellular fluids of
the brain. These electrolytes are important to the function of many systems, homeostasis, and iron
regulation in the body and play a role in nerve stimulation.
•With this method, electric pulses are used in conjunction with muscle relaxants, nitrous oxide and
oxygen to produce anesthesia.
• Another patent (6,393,319, 2002) involves using prerecorded wave forms to generate electric pulses
for therapeutic purposes.
•Decreased gastric acid secretion leading to less of a chance for stomach ulcers seen with other
anesthesia (Sances and Larson, 33-46).
•Our device differs from the above by allowing for dynamic control of the output signal.
Design Specifications
Theory Behind Vagal Nerve Stimulation
LabVIEW virtual instrument
•Pain can be controlled by electrical stimulation of the vagal nerves
(Kirchner et al., Ness et al.).
•The ON/OFF button turns the signal generation on or off but does not
close the program.
•Vagal nerves lead into the brain and allow for the control of pain without
passing current through the brain.
•There are dials to control pulse duration, amplitude, and frequency of the
output signal.
Normal nerve reaction to a stimulus
•Pain control is possible via clips on the earlobes (Cork et al 2004).
Gain Stage
“Relative refractory zone” showing
inhibition of function
•The computer output passes through a gain stage of 23 to amplify the
signal for patient delivery to any desired voltage between 13.8 and
59.8V.
•When inhibitory neurons are tonically excited, other inhibitory neurons are
excited through mutual inhibitory connections (Jefferys et al. 202-208).
•The circuit consists of a precision instrumentation amplifier, resistors,
fans, and a battery power supply contained under the laptop.
•When an excitatory nerve is repeatedly stimulated, it can cease responding
to the stimulus effectively inhibiting the nerve and all of its downstream
connections (See adjacent graphic).
•The power supply to the amplifier is a 9V rechargeable Nickel Metal
Hydride battery.
•At this point, the use of the vagal nerve to administer electroanesthesia is
theoretical, but through the development and further research into vagal
nerve stimulation full anesthesia may be achieved.
Market Analysis
•20 million people undergo surgery using general anesthesia each year in the United States.
•An electroanesthesia device would reduce the cost of anesthesia and the need for a highly trained
anesthesiologist making it especially marketable in less developed countries.
•Electroanesthesia devices in the European market utilize transcranial methods.
Risks
Reduction Technique
Low current and voltage
Insulated housing and wires
Computer, circuit, and fans are grounded
Unplugged during use
Thorough testing of computer code
Backup battery
Electrical Shock
Applicator
•The white foam backing of the pad electrode allows for stretch-ability
and memory over repeated use. The carbon/silver film gives uniform
current distribution and lower impedance levels for more comfort.
•A monopolar 1 in., 28 gauge conical tip needle electrode will be used to
delivery the desired signal for any length of time.
•Substitute for intravenous and gas anesthesia in any medical procedure requiring general anesthesia
below the neck.
Safety Considerations
•A rechargeable battery maintains the portability of the device while also
offering an easy way to recharge the system.
•Pad or needle electrode.
•Nothing on the market in the US similar to the proposed device.
To date, the FDA has not approved any electroanesthesia devices and considers them to be Class III
medical devices under 21CFR868.5400. Approval for such a device needs to go through a Premarket
Approval (PMA) or Product Development Protocol (PDP). Extensive clinical testing, the disclosure of
specifications, intended use, manufacturing methods, and proposed labeling would be required for the
PMA process. In order to collect preliminary data and show proof of concept, animal testing would have
to be conducted first. Before proceeding with animal testing, Institutional Review Board (IRB) approval
would need to be obtained. After these studies, an investigational device exemption (IDE) would need to
be cleared for human trials to begin.
•During waveform generation, a plot showing the current waveform
generation and a plot showing a ten second window of all of the pulses
during the session are displayed.
•Ammons et al. (1983) proposed that there are sites on the brain stem that
can activate descending inhibitory pathways via the stimulation of the vagal
nerves and the surrounding nerves.
•Hammond et al. (1992) reported that “synchronous bursts of potentials
could be dispersed to widespread areas of the brain” through afferent
pathways.
FDA Regulations
Housing
•The laptop and amplification circuit are be housed within an 18" x 18" x
10" compartment constructed from rigid high-density polyethylene.
•Underneath the laptop is space for the circuit, a battery and charger to
run the circuit and fans, a power strip to connect all of the devices, and
extra space for the storage of the recommended vital signs equipment.
Pad Electrode
Needle Electrode
Cessation of Function
Detachment of Applicator
Self-adhesive medical tape to be used
Acknowledgements
Special thanks to Dr. James Berry, Dr. Paul King,, Ray Erlandson, Ray Booker, Patrick Taylor, Dr. Jerry
Collins
References
Ammons et al. "Vagal Afferent Inhibition of Primate Thoracic Spinothalamic Neurons." Journal if Neurophysiology 50.4 (October 1983): 926-940.
Clarke, H.L. "Anaesthesia for Out-Patient Procedures." The West African Medical Journal 11 (1962): 155-160.
Cork et al. "The Effect of Cranial Electrotherapy Stimulation (CES) on Pain Associated with Fibromyalgia." The Internet Journal of Anesthesiology (2004): 1-7.
Hammond et al. "Electrophysiological studies of cervical vagus nerve stimulation in humans: I. EEG effects.." Epilepsia. 33 (1992): 1013-1020.
Jefferys, John GR, Roger D. Traub, and Miles A. Whittington. "Neuronal networks for induced '40 Hz' rhythms." Trends in Neuroscience 19 (1996): 202-208.
Kirchner et al. "Left vagus nerve stimulation suppresses experimentally induced pain." Neurology 55.8 (2000): 1167-1171.
Ness, T.J., R.B. Fillingim, A. Randich, and E.M. Backensto. "Low intensity vagal nerve stimulation lowers human thermal." Pain 86 (2000): 81-85.
Photiades et al.. "Electroaneasthesia in Major Surgery ." The West African Medical Journal (October 1963): 218-225.
Sances Jr., Anthony, and Sanford J. Larson. Electroanesthesia Biomedical and Biophysical Studies. New York: Academic Press, Inc., 1975.
Takakura, Kintomo. “Transcutaneous Electrical Nerve Stimulation for Relieving Pain: Physiological significance of the 1/f frequency fluctuation.” Accessed 11/4/05
<http://www.everbest.com.au/TENSforRelievingPain1ffrequencyfluctuation.html.>