Download Argos TM - Lockheed Martin

Survey
yes no Was this document useful for you?
   Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Document related concepts

Spectrum analyzer wikipedia , lookup

Optical amplifier wikipedia , lookup

Harold Hopkins (physicist) wikipedia , lookup

Vibrational analysis with scanning probe microscopy wikipedia , lookup

Optical coherence tomography wikipedia , lookup

Laser wikipedia , lookup

Spectral density wikipedia , lookup

X-ray fluorescence wikipedia , lookup

Photonic laser thruster wikipedia , lookup

Fiber-optic communication wikipedia , lookup

3D optical data storage wikipedia , lookup

Gaseous detection device wikipedia , lookup

Ultraviolet–visible spectroscopy wikipedia , lookup

Magnetic circular dichroism wikipedia , lookup

Optical rogue waves wikipedia , lookup

Two-dimensional nuclear magnetic resonance spectroscopy wikipedia , lookup

Astronomical spectroscopy wikipedia , lookup

Nonlinear optics wikipedia , lookup

Opto-isolator wikipedia , lookup

Mode-locking wikipedia , lookup

Ultrafast laser spectroscopy wikipedia , lookup

Photoacoustic effect wikipedia , lookup

Transcript
Argos
TM
Application Note:
Photoacoustic Spectroscopy / Trace Gas Sensing
Dr. Anatoliy Kosterev and Dr. Frank Tittel at Rice University (Houston, Texas) are leading
researchers in developing advanced trace gas sensors for applications such as monitoring
pollutant gases and analyzing breath for early and non-invasive detection of diseases.
In these applications, the high power and spectral characteristics of a continuous wave
(CW) optical parametric oscillator (OPO) enable high-sensitivity detection using a photoacoustic technique called Quartz-Enhanced Photo-Acoustic Spectroscopy (QEPAS). In
2005, Dr. Kosterev performed QEPAS feasibility tests using a prototype Aculight® CW OPO.
Figure 1: The QEPAS tuning-fork sensor.
With QEPAS, researchers measure the amplitude of the sound wave that is generated when
gas molecules absorb modulated light. The detection module is a low-cost, compact, quartz
tuning fork, such as those commonly found in digital watches. The QEPAS sensor has a
resonant frequency of ~32 kHz, so the OPO source must be frequency modulated at half
this frequency over a range corresponding to the width of an absorption feature of the gas.
To allow the high frequency modulation rate, the prototype OPO used a 1080-nm distributed
Bragg reflector (DBR) diode laser as the seed for its pump fiber laser. (In Aculight Argos™ CW
OPO systems, a fiber-coupled diode laser can be simply exchanged for the standard fiber
laser seed to enable high modulation frequencies.)
2f Photoacoustic Signal
By applying a slow current sweep to the diode in addition to the modulation signal, Kosterev’s
team recorded 2f wavelength modulation spectra of ammonia and water vapor at 2.8 µm.
75
50
25
0
-25
-50
3367.0
3367.4
3367.8
3368.2
-1
Wavenumber (cm )
Figure 2: The spectrum recorded using the output of an Aculight CW OPO shows
absorption features for water vapor and ammonia.
The Rice group, in cooperation with a team at Nijmegen University in the Netherlands,
has also applied QEPAS to trace ethane detection. They used a CW OPO producing 300 mW at
3µm to detect ethane to a concentration level of 13 parts-per-billion. The Rice researchers now
have an Aculight Argos system that produces 3 W and with it they expect to improve detection
sensitivity by a factor of 10.